Summary
A previously developed, simple procedure for computer simulation has been applied to non-ideal, linear chromatography. Relations between three different retention measures (the maximum of the peak, the median and the center of gravity) and the skew and the width of the observed elution peak have been suggested from the results of the simulation. These relations are experimentally verified. No one of the retention measures mentioned above is exactly equal to the thermodynamically correct retention volume, but the discrepancy for the median is the smallest.
Similar content being viewed by others
References
J. R. Conder andC. L. Young, Physicochemical Measurements by gas Chromatography, John Wiley & Sons, New York 1979.
M. Goedert andG. Guiochon, Anal. Chem.42, 962 (1970).
T. S. Sørensen, J. Chromatogr.88, 197 (1974).
J. Å. Jönsson, J. Chromatogr.150, 11 (1978).
J. Å. Jönsson, Chromatographia13, 273 (1980).
J.J. Van Deemter, F.J. Zuiderweg andA. Klinkenberg, Chem. Eng. Sci.5, 271 (1956).
J. C. Martin andR. L. M. Symge, Biochem J.35, 1358 (1941).
J. C. Giddings, Dynamics of Chromatography, Part 1, Marcel Dekker Inc, New York 1965.
H. Cramér, Mathematical Methods of Statistics, Princeton University Press, Princeton, 1963, p. 193.
E. Grushka, M. N. Meyers, P. D. Schettler andJ. C. Giddings, Anal. Chem.41, 889 (1969).
J. Å. Jönsson, J. Chromatogr.139, 156 (1977).
J. Å. Jönsson andR. Jönsson, J. Chromatogr.111, 265 (1975).
J. Å. Jönsson, R. Jönsson andK. Malm, J. Chromatogr.115, 57 (1975).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Jönsson, J.Å. Non-ideal effects in linear chromatography. Chromatographia 13, 729–733 (1980). https://doi.org/10.1007/BF02265550
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02265550