Abstract
One eye each of ten healthy volunteers was tested with the Rodenstock Optic Nerve Head Analyzer on ten separate occasions. A color-coded pallor map was developed for each examination. A color transparency of each pallor map was then obtained from the television monitor of the Optic Nerve Head Analyzer, using a camera mounted on a tripod at a fixed distance from the screen. A color-coded zone within the center of the disc was chosen for each set of ten studies, and the area of this zone was measured by boundary analysis with another computerized image analyzer, the Oasys Image Processor. There was wide intrasubject variation in all ten volunteers, suggesting that the pallor map generated by the Optic Nerve Head Analyzer, when used in the manner described in this paper, is not adequate for following subtle changes in optic nerve head pallor.
Similar content being viewed by others
References
Caprioli J, Klingbeil U, Sears M (1986) Reproducibility of optic measurements with computerized analysis of stereoscopic video images. Arch Ophthalmol 104:1035–1039
Cornsweet TN, Hersh S, Humphries JC (1983) Quantification of the shape and color of the optic nerve head. In: Breinin GM, Siegel IM (ed) Advances in diagnostic visual optics. Springer, New York Berlin Heidelberg, pp 141–149
Mikelberg FS, Douglas GR, Schulzer M (1984) Reliability of optic disc topographic measurements recorded with a video-ophthalmograph. Am J Ophthalmol 98:98–102
Mikelberg FS, Douglas GR, Drance SM, Schulzer M, Wijsman K (1988) Reproducibility of computerized pallor measurements obtained with the Rodenstock Disk Analyzer. Graefe's Arch Clin Exp Ophthalmol 226:269–272
Mikelberg FS, Douglas GR, Drance SM, Schulzer M, Schwartz B, Nagin P, Wijsman K (1988) Measurements of optic nerve head pallor with a video-ophthalmograph and with computerized boundary analysis. Can J Ophthalmol 23:120–123
Miller JM, Caprioli J (1988) Videographic quantification of optic disc pallor. Invest Ophthalmol Vis Sci 29:320–323
Nagin P, Schwartz B, Namba K (1985) The reproducibility of computerized boundary analysis for measuring optic disk pallor in the normal optic disc. Ophthalmology 92:243–251
Shields MB, Martone JF, Shelton AR (1987) Reproducibility of topographic measurements with the optic nerve analyzer. Am J Ophthalmol 104:581–586
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Miller, K.N., Shields, M.B. & Ollie, A.R. Reproducibility of pallor measurements with the optic nerve head analyzer. Graefe's Arch Clin Exp Ophthalmol 227, 562–564 (1989). https://doi.org/10.1007/BF02169452
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02169452