Abstract
The accuracy and repeatability of a new ambulatory radionuclide detector (VEST) for left ventricular systolic (ejection fraction) and diastolic (peak filling rate) measurements were assessed. Seventeen patients underwent equilibrium radionuclide angiography immediately before and immediately after a VEST study. The accuracy was evaluated at the beginning and at the end of the VEST studies. Limits of agreement for the ejection fraction were −1%: 2% at the beginning of the VEST study and −4% at the end. Limits of agreement for the peak filling rate were −0.6: 0.6 at the beginning of the VEST study and −0.7: 0.5 at the end. For both measurements the limits of agreement were well within the clinical range. Repeatability was evaluated in a second group of 11 patients who underwent VEST studies in 2 separate days. The coefficient of repeatability (twice the standard deviation of the differences between the 2 studies) was 13 for the ejection fraction and 0.4 for the peak filling rate. Thus, the VEST is an accurate and repeatable method to measure both the ejection fraction and peak filling rate.
Similar content being viewed by others
References
Altman DG, Bland JM (1983) Measurement in medicine: the analysis of method comparison studies. Statistician 32:307–317
Bairey CN, Yang L de, Berman DS, Rozanski A (1990) Comparison of physiologic ejection fraction responses to activities of daily living: implications for clinical testing. J Am Coll Cardiol 16:847–854
Betocchi S, Cuocolo A, Pace L, et al. (1987) Effects of intravenous verapamil administration on left ventricular diastolic function in systemic hypertension. Am J Cardiol 59:624–629
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurements. Lancet l:307–310
Bonow RO, Bacharach SL, Green MV, et al. (1981a) Impaired left ventricular diastolic filling in patients with coronary artery disease: assessment with radionuclide angiography. Circulation 64:315–323
Bonow RO, Rosing DR, Bacharach SL, et al. (1981b) Effects of verapamil on left ventricular systolic function and diastolic filling in patients with hypertrophic cardiomyopathy. Circulation 64:787–796
Breisblatt WM, Weiland FL, McLain JR, Tomlinson GC, Burns MJ, Spaccavento LJ (1988) Usefulness of ambulatory radionuclide monitoring of left ventricular function early after acute myocardial infarction for predicting residual myocardial ischemia. Am J Cardiol 62:1005–1010
British Standard Institution (1979) Precision of tests methods I: guide for the determination and reproducibility for a standard test method (BS 5497). BSI, London
Cuocolo A, Sax FL, Brush JE, Maron BJ, Bacharach SI, Bonow RO (1990) Left ventricular hypertrophy and impaired diastolic filling in essential hypertension. Diastolic mechanisms for systolic dysfunction during exercise. Circulation 81:978–986
Kayden DS, Wackers FJT, Zaret BL (1990) Silent left ventricular dysfunction during routine activity after thrombolytic therapy for acute myocardial infarction. J Am Coll Cardiol 15:1500–1507
Kiess MC, Dimsdale JE, Moore RH, Liu P, Newell J, Barlai-Kovach M, Boucher CA (1988) The effects of stress on left ventricular ejection fraction. Eur J Nucl Med 14:12–16
Inouye IK, Massie BM, Loge D, Simpson P, Tubau JF (1984) Failure of antihypertensive therapy with diuretics, beta-blocking and calcium channel-blocking drugs to consistently reverse left ventricular diastolic filling abnormalities. Am J Cardiol 53:1583–1587
Miller TR, Fountos A, Biello DL, Ludbrook PA (1987) Detection of coronary artery disease by analysis of ventricular filling. J Nucl Med 28:837–843
Pace L, Bacharach SI, Bonow RO, Cannon RO III, Green MV, Larson SM (1989) Diagnosis of coronary artery disease by radionuclide angiography: effect of combining indices of left ventricular function. J Nucl Med 30:1966–1971
Soufer R, Wohlgelernter D, De Vita NA, et al. (1985) Intact systolic left ventricular function in clinical congestive heart failure. Am J Cardiol 55:1032–1037
Tamaki N, Gill JB, Moore RH, Yasuda T, Boucher CA, Strauss HW (1981) Cardiac responses to daily activities and exercise in normal subjects assessed by an ambulatory ventricular function monitoring. Am J Cardiol 59:1164–1169
Tamaki N, Yasuda T, Morris RH, Gill JB, Boucher CA, Hutter AM, Gold HK, Strauss HW (1988) Continuous monitoring of left ventricular function by an ambulatory radionuclide detector in patients with coronary artery disease. J Am Coll Cardiol 12:669–679
Wilson RA, Sullivan PJ, Moore RH, Zielonka JS, Alpert NM, Boucher CA, McKusick KA, Strauss HW (1983) An ambulatory ventricular function monitor: validation and preliminary results. Am J Cardiol 52:601–606
Yang L de, Bairey CN, Berman DS, Nichols KJ, Odom-Mayron T, Rozanski A (1991) Accuracy and reproducibility of left ventricular ejection fraction measurements using an ambulatory radionuclide left ventricular function monitoring. J Nucl Med 32:796–802
Author information
Authors and Affiliations
Additional information
Offprint requests to: L. Pace, Sanseverino 5/A, 1-80128 Napoli, Italy
Rights and permissions
About this article
Cite this article
Pace, L., Cuocolo, A., Nappi, A. et al. Accuracy and repeatability of left ventricular systolic and diastolic function measurements using an ambulatory radionuclide monitor. Eur J Nucl Med 19, 800–806 (1992). https://doi.org/10.1007/BF00182823
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF00182823