Introduction

For many years attempts have been made to design a system in which a mass spectrometer would have an independent external source of sample ionization working under atmospheric pressure. In the commercial mass spectrometers the most popular technique of sample ionization is the direct analysis in real time (DART). In this technique ions are produced electronically or vibronically from excited-state species in gases such as helium, argon, or nitrogen that ionize atmospheric molecules or dopant molecules [1,2,3,4,5,6]. The ions generated from atmospheric or dopant molecules undergo ion–molecule reactions with the sample molecules to produce analyte ions. DART has been successfully used for analysis of chemicals, including chemical agents, pharmaceutics, metabolites, pesticides and environmentally significant compounds.

Flowing atmospheric pressure afterglow mass spectrometry (FAPA) is an analytical chemistry technique for sensitive detection of low molecular weight organic compounds [7,8,9,10,11,12,13]. The method of analysis of organic compounds by mass spectrometry with the flowing atmospheric pressure afterglow ion source in the ambient atmosphere and has been proven to be a promising tool for direct and rapid determination of numerous compounds. The construction of the argon plasma torch and its operating parameters have a direct impact on the quality of the spectra obtained in the mass spectrometer. Argon plasma is generated between the two electrodes inside the quartz tube. The first electrode is inside quartz tube and the second electrode is outside at the end of the quartz tube. The analytes were thermally desorbed and ionized by FAPA and identified in the mass spectrometer analyzer. Thermal desorption of analytes is a specific method of introducing a sample into a plasma stream. The order of release of the compounds depends on their volatility. The method offers fast and reliable structural information, with no need of pre-separation, and can be an alternative to the Electron Impact, GC/MS, and ESI for fast analysis of organic compounds.

Zeiri et al. [14] have presented a direct-current microplasma-based flowing atmospheric pressure afterglow for the use in ambient desorption-ionization mass spectrometry. The annular-shaped microplasma is formed in helium between two concentric stainless-steel capillaries that are separated by an alumina tube. Their results suggested that ionization occurred mainly through protonation, with only a small contribution of fragmentation and adduct formation. Limits of detection for several small organic molecules were in the sub-picomole range. Dai et al. [15] have demonstrated the application of a microwave induced plasma ionization (MIPI) source to ion mobility spectrometry (IMS). The MIPI source is composed of a copper cavity and a hollow quartz discharge tube. MIPI-IMS was used to detect some volatile organic compounds in monitoring of air pollutants.

The compact microwave discharge, termed the microstrip plasma (MSP) that was previously used for optical emission spectrometry has been used as an afterglow ion source for mass spectrometry by Pfeuffer et al. [16]. The primary background ions present in the afterglow were the ionized and protonated water clusters. An exponential dilution chamber was used to introduce volatile organic compounds into the MSP and yielded limits of detection in the 40–7 ppm range (v/v). The MSP afterglow was found capable of desorption and ionization of analyte species directly from a solid substrate, which suggested its possible use as an ion source for ambient desorption/ionization mass spectrometry.

We present the construction and application of a new microwave argon plasma torch which was designed to be used in the ion trap mass spectrometry. The application of this technique is presented on the example of analysis of pure compounds, analytes included in molecularly imprinted polymers (MIPs) and commercial drugs in the form of tablets.

Experimental

Instruments

MS and MSn spectra were recorded using an amaZon SL ion trap (Bruker, Bremen, Germany) equipped with an electrospray ion source in infusion mode or microwave argon plasma. The apparatus was operated using the so-called “enhanced resolution mode” (mass range: 50–2,200 m/z, scanning rate: 8100 m/z per s). The typical ESI–MS source settings were found to be optimal also for the microwave ionization source, with the exception of the mass spectrometer entrance glass capillary voltage, which was lower l (1 kV) than the standard ESI setting (4.5 kV). The temperature of the glass capillary was set to 200 °C, the drying gas flow was maintained at 3 L min−1. The mass spectrometer was operating in the positive ion mode. For MS2 experiments, the isolation width was set to 2 m/z and the fragmentation amplitude was in the range from 0.5 to 0.8 unit. The microwave argon plasma ionization source was positioned on the axis of the inlet of the mass spectrometer with the tip located ca. 50 mm from the MS inlet. The vapors generated in such a way were directly introduced into the argon plasma jet stream, which resulted in their ionization. The overall average time of analysis was ca. 5 min. The mass spectrometer was equipped optionally in an outside microwave argon plasma source (Fig. 1). The microwaves were used to generate argon plasma allowing the ionization of analyte particles thermally released from the heated crucible whose temperature could be changed from 100 to 250 °C, with a temperature increase rate of 3 °C s−1. The mini crucible allowing the temperature-controlled desorption was placed ca. 10 mm below the ion stream. The temperature of the compounds desorption was 200 ± 10 °C.

Fig. 1
figure 1

The measuring system: microwave argon plasma ionization source, heated crucible and amaZon SL mass spectrometer

Argon plasma torch, made of a 4/2 mm diameter (DIA) quartz tube through which argon was flown at the rate of ca. 100 ml min−1, powered by a 100 W/2450 MHz microwave power generator. The impedance tuning is accomplished using a dual dielectric tuner that tunes to the maximum discharge volume and stability. The argon plasma cavity (quarter wavelength, ca. 3 cm) is made of stainless steel. The diameter of the external screen is 10 mm, and the total length of the resonator section is 4 cm. The construction of the argon plasma source is shown in Fig. 2.

Fig. 2
figure 2

The construction of the argon plasma source

Materials and Chemicals

The compounds used in the studies are presented in Table 1. The compounds were purchased from Sigma-Aldrich (St Louis, MO, USA).

Table 1 The compounds studied

Molecularly imprinted polymers (MIPs) are those that have been processed using the molecular imprinting technique, which leaves cavities in the polymer matrix strictly corresponding to a particular template molecule. All reagents used for the synthesis of MIPs were commercial products. Atrazine, pendimethalin, acrylic acid (AA), methacrylic acid (MAA), ethylene glycol dimethacrylate (EGDMA), 2,2′-azobisisobutyronitrile solution 0.2 M in toluene (AIBN), Aluminum Oxide Activated, basic, 4 Brockman I, and chloroform solvent were of the p.a. grade were obtained from Sigma-Aldrich (St. Louis, MO, USA).

Synthesis of Atrazine-MIP, Quercetin-MIP and Pendimethalin-MIP

A portion of 0.4 mmol of the template molecule-atrazine and 1.6 mmol functional monomer-MAA were dissolved in 25 ml of methanol, while 0.6 mmol of the template molecule-pendimethalin and 2.4 mmol the same functional monomer were dissolved in 10 ml chloroform. A portion of 1 mmol of the template molecule-quercetin and 4 mmol functional monomer-AA were dissolved in a mixture of 5 ml chloroform and 5 ml DMF. Subsequently, previously purified with Aluminum Oxide Activated, basic, Brockman I, 8 mmol or 12 mmol cross-linking agent-EGDMA was added. The ratio of the template:monomer:cross-linking agent was 1:4:20, respectively. The pre-polymerization solution was sonicated and purged with nitrogen for 20 min. Afterwards, 0.5 ml or 1 ml initiator-AIBN was added. The tube was sonicated and purged with nitrogen for additional 20 min, sealed and placed in an oven for 24 h at 60 °C. After polymerization, MIPs were dried under reduced pressure, and ground using mortar and pestle. The particles of used templates were extracted with a mixture of ethanol and acetic acid (9:1, v:v) via Soxhlet extraction for 90 h. Then, the final product MIPs was dried at 60 °C under vacuum and ground. The structure of MIP is presented in Fig. 3.

Fig. 3
figure 3

The structure of analyte/template-MIP

In order to confirm the binding of analytes in the polymer structures, FT-IR and TG analysis were performed. The concentration of the analytes in the obtained MIPs was 48 mg/1 g of atrazine, 101 mg/1 g of quercetin and 71 mg/1 g of pendimethalin.

Synthesis of NIPs

As a control experiment, non-imprinted polymers (NIPs), with no templates during the polymerization, were also prepared and treated in the identical manner.

LOD Determination Procedure

The results of microwave argon plasma ionization source-MS experiments, obtained with the use of a solution of pure analytes, analyte-MIPs, and bioanalytes in medical tablets, were used for evaluation of the work of the new measurement system. All experiments were repeated three times. The relative standard deviation (RSD) obtained for all data did not exceed 20%. In order to determine the background signals, the spectra of particular matrices (NIPs) without the analytes were recorded. The limit of detection (LOD) is the concentration of a substance below which the identity of the substance cannot be distinguished from analytical artefacts. The limit of detection (LOD) was calculated according to the definition: LOD = mean blank value + 3 × standard deviation. The signal m/z analyte was three times higher than the noise level. In the experiments in which the LOD of pure compounds was determined, 20 μL of the solution of the analytes at different concentrations in methanol was introduced into the crucible and heated to about 200 °C, whereas in the case of composite matrices, 100 mg of the mixture of MIP/NIP or tablets/NIP with different concentrations of the analytes were introduced into the crucible and also heated to ca. 200 °C.

The limit of quantification (LOQ) is the concentration of a substance below which the concentration cannot be determined with an acceptable accuracy. The LOQ value was calculated as LOQ = 3 × LOD.

Results and Discussion

To assess the application of the microwave argon plasma source in mass spectrometry analysis we used the following pure compounds: theophylline, caffeine and quinine. Theophylline is a methylxanthine drug used in the therapy of respiratory problems and asthma under a variety of brand names. Theophylline bears structural and pharmacological similarity to theobromine and caffeine, and it is readily found in in tea and cocoa. Caffeine stimulates nervous system and reversibly blocks the action of adenosine on its receptor and consequently prevents the onset of drowsiness induced by adenosine. Quinine is a medication used to treat malaria and babesiosis. Quinine is also an ingredient of tonic water that gives it its bitter taste.

In order to determine the suitability of the method for analysis of functional polymers, MIPs containing in their structure the analytes atrazine, pendimethalin and quercetin trapped in the cavities were obtained. Atrazine is a herbicide of the triazine class used in preemergence and postemergence applications. Pendimethalin is an herbicide used to control annual grasses and certain broadleaf plants. Quercetin is a plant flavonol found in many fruit, vegetables, leaves, and grains.

The third group of materials used for testing was that of active ingredients contained in commercial drugs in the form of tablets. Ketoprofen is one of the propionic acid class of nonsteroidal anti-inflammatory drugs with analgesic and antipyretic effects. It acts by inhibiting the body’s production of prostaglandin. Propyphenazone is a derivative of phenazone with similar analgesic and antipyretic effects, while bisoprolol is used specifically in the treatment of hypertension and heart failure.

The first groups of analyzed systems were pure compounds. The microwave argon plasma ionization source-MS spectra as well as MS2 spectra were analyzed for this group of compounds (Figs. 4, 5 and 6). For all analyzed low molecular weight compounds in the positive ion spectrum, the m/z signals [M + H]+ were recorded, whereas in the MS2 spectra the main decay product of this ion was recorded.

Fig. 4
figure 4

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of theophylline (b)

Fig. 5
figure 5

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of caffeine (b)

Fig. 6
figure 6

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of quinine (b). The fragmentation spectrum of ions [M + H]+ confirms the paths of fragmentation obtained in the ESI-MS2 fragmentation of protonated quinine as reported by Kudlacek et al. [17]

The second group of analyzed compounds included MIPs containing the analytes in their structure. Molecularly imprinted polymers are the polymeric matrices obtained using the imprinting technology for defined analytes, useful to separate and analyze complicated samples such as biological and environmental samples. The analytes from the polymer structures thermally released into the argon plasma stream were ionized. The recorded microwave argon plasma ionization source-MS spectra (Figs. 7, 8 and 9) for these compounds (positive ions) showed an intense ion signal [M + H]+ while the fragmentation ions in the MS2 spectrum were characteristic of these compounds and additionally confirmed the correct [M + H]+ signal assignment.

Fig. 7
figure 7

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of atrazine direct from the atrazine-MIP (b). The fragmentation spectrum of ions [M + H]+ confirms the paths of fragmentation obtained by Dong et al. [18]

Fig. 8
figure 8

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of pendimethalin direct from the pendimethalin-MIP (b)

Fig. 9
figure 9

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of quercetin direct from the quercetin-MIP (b)

The microwave argon plasma source-MS technique makes it possible to analyze compounds from complex matrices. The analyzed compounds are released from these matrices by thermal desorption and directed to the argon plasma stream. The active ingredients from commercial medical tablets were analyzed. Very good results were obtained for all studied systems (Figs. 10, 11 and 12). In the MS spectra (positive ions), intense signals were observed m/z [M + H]+, whereas the fragmentation spectra (MS2) of ions [M + H]+ confirmed the structure of the analyzed analytes.

Fig. 10
figure 10

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of ketoprofen (b) direct from the medicine tablet

Fig. 11
figure 11

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of propyphenazone (b) direct from the medicine tablet

Fig. 12
figure 12

The microwave argon plasma source-MS (positive ions) spectrum (a) and fragmentation (MS2) spectrum of bisoprolol (b) direct from the medicine tablet. The fragmentation spectrum of ions [M + H]+ confirms the paths of fragmentation obtained by M.F.Tutunji et al. [19]

For all tested systems: pure compounds, MIPs with absorbed analyte and medical tablets with active ingredients, the linear range of determinations and the limit of analyte detection were determined (Table 2).

Table 2 Comparison of analytical parameters for studied materials

The LOD value depended on the structure of the analyte and ranged from 0.002 μg to 0.005 μg for the analyzed compounds, and the LOQ value ranged from 0.006 μg to 0.015 μg.

Conclusion

A new microwave argon plasma ionization source was designed and constructed for the use in ion trap mass spectrometry for analysis of solid samples and analytes on solid materials. The microwave plasma ionization source was used to generate argon plasma allowing the ionization of analyte particles thermally released from the heated crucible whose temperature could be regulated from 100 to 250 °C. The system with a microwave argon plasma torch was used to determine low molecular weight compounds from different matrices. The obtained results indicate that the proposed microwave burner with argon as the plasma gas, can be used in MS analysis in which the ionization source of samples is exchangeable. In the proposed method, analytes are thermally released from solid matrices without any sample preparation steps prior to qualitative and quantitative analysis. This method significantly reduces the time of analysis without analyte loss, as in the classical sample isolation by filtration or sludge spinning phases.