Abstract
Purpose
Surgical intervention with intercorporal stabilisation in spinal infections is increasingly needed. Our aim was to compare titanium and polyetheretherketon (PEEK) cages according to their adhesion characteristics of different bacteria species in vitro.
Methods
Plates made from PEEK, polished titanium (Ti), two-surface-titanium (TiMe) (n = 2–3) and original PEEK and porous trabecular structured titanium (TiLi) interbody cages (n = 4) were inoculated in different bacterial solutions, S.aureus (MSSA, MRSA), S.epidermidis and E.coli. Growth characteristics were analysed. Biofilms and bacteria were visualised using confocal- and electron microscopy.
Results
Quantitative adherence of MSSA, MRSA, S.epidermidis and E.coli to Ti, TiMe and PEEK plates were different, with polished titanium being mainly advantageous over PEEK and TiMe with significantly less counts of colony forming units (CFU) for MRSA after 56 h compared to TiMe and at 72 h compared to PEEK (p = 0.04 and p = 0.005). For MSSA, more adherent bacteria were detected on PEEK than on TiMe at 32 h (p = 0.02). For PEEK and TiLi cages, significant differences were found after 8 and 72 h for S.epidermidis (p = 0.02 and p = 0.008) and after 72 h for MSSA (p = 0.002) with higher bacterial counts on PEEK, whereas E.coli showed more CFU on TiLi than PEEK (p = 0.05). Electron microscopy demonstrated enhanced adhesion in transition areas.
Conclusion
For S.epidermidis, MSSA and MRSA PEEK cages showed a higher adherence in terms of CFU count, whereas for E.coli PEEK seemed to be advantageous. Electron microscopic visualisation shows that bacteria did not adhere at the titanium mesh structure, but at the border zones of polished material to rougher parts.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Infections of the spinal column such as spontaneous osteomyelitis or post-surgical infections are rare but increasing pathologies due to the ageing population and increasing numbers of comorbidities with mortality rates up to 17% [1,2,3]. Mild infections can be treated conservatively whereas patients with sepsis, osseous destruction and/or neurological deficits due to compression of the spinal cord or the nerve roots need decompressive surgery and/or stabilisation. Resection of the affected intervertebral disc(s) and stabilisation with intervertebral fusion is recommended in these cases. The gold-standard is still the implantation of a dorsal titanium screws-rod-system with either a titanium intervertebral cage or a cortical bone autograft [4, 5], although lately studies had shown that cages made of polyetheretherketon (PEEK) can be a safe and feasible alternative implant material in patients with osteomyelitis [6,7,8].
The most common causative bacterial species for spondylodiscitis or osteomyelitis are S.aureus with an incidence between 30 and 80%, S.epidermidis with rates up to over 10% and MRSA (6.1%). Gram-negative bacteria are responsible for up to 25% of spinal infections, with E.coli being reported as the most common (5.6%) [9,10,11].
Even if there are many studies on osteomyelitis, there is still no final guideline for the correct treatment and choice of implant materials [12, 13]. Autologous bone grafts and titanium implants have been shown to have reduced bacterial adhesion characteristics, whereas PEEK cages are radiolucent, allow better follow-up diagnostics with computed tomography (CT) and magnetic resonance imaging (MRI) due to less artefacts and are advantageous according to better load sharing at dissolved endplates in osteomyelitis [14, 15]. Clinical studies concentrate on bacterial counts on explanted stabilisation material of post-surgical infections or re-infection rates after the use of PEEK cages [6, 8]. Experimental studies evaluated the influence of topography and chemistry on bacterial adhesion on different biomaterials [14, 15]. However, to our knowledge, this in vitro study is the first to actually quantitatively evaluate differences in biofilm formation on two intervertebral cages made of PEEK and titanium, which are in clinical use.
Methods
Biomaterials
For baseline experiments round plates (diameter: 12 mm, height: 2 mm) made from PEEK, smoothly polished titanium (Ti), and titanium with 2 surfaces, one side roughly polished and the other with a porous trabecular structure (TiMe) were provided by a medical device company (Spineart, Plan-les-Ouates, Switzerland)(Fig. 1a). Additionally, original interbody oblique cages made of PEEK (L32 mm, H12 mm; Spineart, Plan-les-Ouates, Switzerland) and 3D printed titanium cages with a porous trabecular structured centre (TiLi; L32 mm, H12 mm, W10.5 mm, 6°angle; Spineart, Plan-les-Ouates, Switzerland) were used (Fig. 1b). Titanium and PEEK plates were machine produced followed by a bead-blasting process to remove burrs. The TiMe plates, however, were 3D printed in one piece, the same way as the manufacturing process of the TiLi cages, without any post-treatment. Only the contour of the TiLi cages is polished, the trabecular structure remaining intact. All samples were supplied sterilised as the standard cages for human implantation.
Microorganisms and growth conditions
Bacteria strains were thawed from a glycerol cryo stock and plated over night at 37 °C on trypticase soy broth (TSB) agar plates or lysogeny broth (LB) agar plates, respectively. For inoculation of biomaterials single colonies of methicillin-susceptible and methicillin-resistant S.aureus (MSSA: ATCC 25,923, MRSA: ATCC 43,300), and S.epidermidis (ATCC 12,228) strains were cultured in 2 ml TSB and incubated over night at 37 °C with 200 rpm shaking. Gram-negative E.coli (ATCC 25,922) was grown in LB.
In order to assess bacterial survival in phosphate-buffered saline (PBS), bacterial overnight cultures were diluted 1:200 in PBS, and 200 µl placed in a microtiter plate. Growth was monitored after 0, 8, 24, 28, 32, 48, 52, 56, 72, 76 and 80 h in a microplate reader measuring the optical density OD600 (Tecan, Männedorf, Switzerland).
Quantification of bacterial adherence
Bacteria from an overnight culture were diluted by measuring the OD600, and stock solutions containing 105 CFU were prepared in PBS. The three different biomaterials were inoculated in bacterial stock solutions for 8, 24, 32, 48, 56, 72 and 80 h, and the cages for 8 and 72 h, respectively. For each time point, sample and bacteria control was incubated only in PBS. Afterwards biomaterials were rinsed with PBS, covered with 2 ml PBS, and surface adherent bacteria were removed into solution by 3 min 100% ultrasound in an ultrasound bath (Bandelin BactoSonic, Berlin, Germany). Serial dilutions of bacteria suspensions were plated in at least duplicates on Mueller–Hinton agar plates, and growth was monitored after 24 h by counting CFU macroscopically.
Confocal laser scanning microscopy analysis
Biofilms were formed in vitro in µ-Angiogenese slides (ibidi, Martinsried/Planegg, Germany) and at appropriate time points fixed in 3.7% paraformaldehyde and stained with 4′,6–Diamidine–2′–phenylindole dihydrochloride (DAPI) nucleic acid stain (Sigma-Aldrich, MO, USA) for visualisation of cells. Stained biofilms were visualised under a TCS SP8 X confocal laser scanning microscope (Leica, Wetzlar, Germany). Biofilms were observed using a 63 × oil-immersion objective, and images of different regions of each surface were acquired with 1024 × 1024 resolution. Two independent biological experiments of in vitro biofilm formation of the different strains were performed, and representative images were selected. Images were reconstructed from average intensity projection through confocal image Z-stacks series using ImageJ [16].
Electron microscopy analysis
Scanning electron microscopy (SEM) was performed to localise MSSA, MRSA, S.epidermidis and E.coli on the cage surfaces. The cages were inoculated with bacterial stock solution (105 CFU), incubated for 72 h, fixed in 4% paraformaldehyde, and PEEK cages were additionally sputter coated with carbon to enable conductive behaviour and prevent static electricity. High-resolution images of the cage surfaces were generated with a secondary electron detector attached to a Crossbeam 340 (Carl Zeiss Microscope GmbH, Jena, Germany) system, which was operated at an electron beam voltage of 5 kV and a working distance of 5 mm. Three different observers (TK, SW and SvK) evaluated the electron microscopic images and chose representative sections.
Statistics
Statistical testing was performed using Graphpad Prism 8. A Shapiro–Wilk and D’Agostino-Pearson omnibus normality test was performed, and Kruskal–Wallis test was used for statistical analysis and data comparison. Results were deemed significant at p < 0.05.
Results
Quantification of bacterial adherence to implant materials
Plates: Two plates of each material were inoculated for time points 24, 32, 48, 56, 80 h (n = 10), three plates for time points 8 and 72 h (n = 6) and each bacterial species (in total 192 samples). The solution of surface adherent bacteria for each plate was transferred to three agar plates (in total 576 agar plates). A moderate or sharp increase in cell attachment of the bacterial solutions could be detected for MSSA, MRSA and E.coli after 72 or 80 h, respectively. For S.epidermidis only a mild increase of adherence on titanium (Ti) could be detected (Supplemental Fig. S1a and Fig. 2a). Overall, bacterial load of the Ti incubated solution was lower after 72 and 80 h compared to PEEK and TiMe. Detailed bacterial counts in mean CFU with standard deviation (SD) are shown in supplemental Fig. S1a for all plate materials. Comparable to the following results of the cages, the results of cell attachment of the bacterial solution of PEEK vs. TiMe plates are of particular interest. The only significant difference in those two materials was at 32 h for MSSA with significantly less bacterial adherence on TiMe plates compared to PEEK plates (p = 0.02). No significant difference could be found for the other time points or for MRSA, S.epidermidis and E.coli (Fig. 2b).
Cages: PEEK cages and TiLi cages were inoculated for 8 and 72 h, respectively (4 cages each), with all three gram-positive bacteria and E.coli (in total 64 cages). The bacterial solution of each cage was then transferred to three agar plates (in total 192 agar plates) (Fig. 3a and Supplemental Fig. S1b). No significant difference could be detected after 8 h for MSSA, MRSA and E.coli. S.epidermidis showed a significantly lower amount of CFU at both time points with p = 0.02 and p = 0.008 after inoculation of the isolated solution of the bacterial adherence on the TiLi cage (Fig. 3). After 72 h, a significant difference with less bacterial attachment for MSSA and S.epidermidis could be detected for the TiLi cages with p = 0.002 and p = 0.008, respectively. The gram-negative E.coli species showed a different behaviour with a significantly higher CFU for the TiLi cages after 72 h (p = 0.05). No significant difference could be detected at both time points for MRSA (Fig. 3b).
All controls showed no bacterial growth after the respective time points of inoculation in PBS only.
Confocal laser scanning microscopy analysis
When analysing biofilm formation in an in vitro assay different types of multicellular layers can be observed (Fig. 4). After 8 h of growth Staphylococci accumulate in small cell clusters forming a dense and thick biofilm after 80 h. As depicted in Fig. 4 in the XZ view for lateral presentation these bacteria form a compact cell mass with no space in between. The rod-shaped E.coli on the other hand show an even cell distribution without clustering after 8 and 80 h of growth.
Electron microscopy analysis
Representative images of the plates and cage surfaces generated by SEM are shown in Fig. 5. No bacteria were found at the porous structured side of the plates or in the trabecular mesh centre of the titanium cages (Fig. 5a “TiMe mesh” and Fig. 5c). Smoothly polished titanium plates also revealed small grooves of the manufacturing process where bacteria could adhere, but less than on the PEEK plates or cages with a more inhomogeneous surface (Fig. 5a). In both, PEEK and TiLi cages all bacteria were only adhering at the border zones of different surface structures. In detail, they were localised in the transition area from polished titanium to the mesh centre in the TiLi cage. The transition zone is illustrated in Fig. 5c. Bacteria were located on the polished part of the titanium cage which is shown in the right panel of Fig. 5b representing an image section of Fig. 5c. On PEEK cages bacteria were adhering primarily on the transition area from smooth to serrated parts.
Discussion
Our study evaluates and compares PEEK and 3D printed trabecular metal (titanium) cages according to their bacterial adhesion characteristics aiming to gain recommendations for the use in cases of osteomyelitis or spondylodiscitis where there is still no final guideline what kind of material for intercorporal cages is best to prevent re-infections and also guarantee bony fusion [17]. Both is provided by autologous bone from the iliac crest, but as it is still associated with a high risk of donor site morbidity and complications cages are more often used in recent years [11]. As there are numerous cages for spinal implantation on the market which vary in their materials and surfaces, we decided to compare a rather new 3D-printed titanium cage with a common PEEK cage. For better comparability of the size and surface design we used special manufactured plates of the same materials in the first stage of the study to assess the differences in bacterial adhesion characteristics. We focused on the comparison between PEEK and titanium cages with a trabecular surface and centre as it was suspicious that bacteria would adhere especially at parts with larger surface [18, 19].
Bacteria were chosen according to their frequency, reported in literature as well as the incidence of the patient cohort in our institution, in causing either spontaneous spondylodiscitis or osteomyelitis and post-surgical infections of the spine. MSSA was chosen as one of the gram-positive bacteria as it is the most common bacteria causing spontaneous spondylodiscitis with an incidence up to 80% and is also predominant in iatrogenic spondylodiscitis. S.epidermidis and MRSA are representatives for perioperative contamination and a multiresistant hospital germ. E.coli was chosen as it is the most often detected gram-negative bacteria which is mainly located in the lower intestine from where especially postoperatively a smear infection of the wound can occur [10, 11, 20,21,22].
In our study, an overall increase of bacterial growth could be detected at 72 h, which is 24–48 h later than shown by Bal and Gorth et al. who used titanium, PEEK and silicon nitride inoculated with S.aureus, S.epidermidis, Pseudomonas aeruginosa, E.coli and Enterococcus in a similar setting [15]. Due to the high cost of the original cages, we chose representative inoculation times of 8 and 72 h as an increase of bacterial adherence on the plates could be shown after 72 h for most bacteria strains. The quantitative analysis showed significantly higher CFU counts on PEEK compared to TiLi cages for MSSA and S.epidermidis after 72 h, for S.epidermidis also after 8 h. Potential reasons might be the rougher machined PEEK surface which provides a better anchorage for the bacteria with micron scale crevices, overlaps and ridges which was previously shown in literature [14, 23, 24]. Against our expectations, SEM images showed no bacteria in the titanium mesh structure cages, which harbour the overall largest surface of the plates or cages. This might be owed to the blast finishing process making the surface features of the titanium mesh more rounded as opposed to the rather sharp edges of the PEEK finish. Both findings are in line with the study of Rochford et al. who also showed a significantly higher number of bacteria on rougher machined PEEK samples with more crevices and folds compared to injection moulded PEEK with a relatively smooth surface and their titanium samples [14]. It is also known that bacterial adhesion is favoured on recessed portions of micro patterned surfaces, and adhesion is greatly impaired when patterning is smaller than the size of the bacteria [25]. Micro-and nanoscale patterning on even the blast finishing of the trabecular structured centre of the titanium cages is still smaller than the irregularities of the PEEK surface which can also be a reason for favoured bacterial adhesion on PEEK rather than the trabecular centre of the titanium cages. SEM also revealed that only the border zones from smoothly polished parts to the mesh structure of the titanium cages seemed suitable for bacterial attachment. Rochford et al. demonstrated similar findings with adherence of bacteria on modified PEEK surfaces on the plateaus between pits, but not in the pits [14].
For colonisation of E.coli on the cages, quantitative analysis and SEM imaging revealed diverse results with a significantly higher count of CFU on TiLi cages than on the PEEK cages after 72 h, but many typically rod-shaped E.coli bacteria on the PEEK cages on SEM. Usually, microorganisms attach more likely to hydrophobic and non-polar surfaces like plastic than to hydrophilic and polar surfaces like metal [26]. Several gram-negative bacteria have, however, other than gram-positive, flagella and pili which help to adhere on rougher surfaces of implant devices [21, 25, 26]. The titanium cages have a smoother surface on the polished parts, but the surface of the trabecular centre might improve the adhesion of gram-negative bacteria. The trabecular centre is also a striking difference to the titanium, smoothly machined, which is used by Bal et al. where they did not find any difference of E.coli adhesion on titanium or PEEK [15]. E.coli can also form long polysaccharide chains being involved in the interaction with material surfaces at the initial adhesion which is a distinct difference between Staphylococci and E.coli and can possibly be responsible for the higher CFU on titanium cages for E.coli compared to PEEK in our study [25].
A major limitation of our study is the low number of diluted samples. However, although significant statistical results could be obtained, further studies with larger numbers might consolidate our results which show that the newer titanium cages are not inferior to and might be even advantageous over PEEK in osteomyelitis due to their individual bacterial adhesion characteristics. As Rochford et al. have shown the ability of coagulase positive S.aureus to cause clotting of blood plasma, which itself can cause bacterial adhesion [14], a study protocol with and without the influence of human blood proteins could be interesting as well and interactions of bacteria with biomaterial surfaces are of course more complex in vivo than in vitro models can simulate [22].
Overall, our study could confirm a non-inferiority of titanium cages even with a trabecular structure compared to PEEK cages according to bacterial adherence. We could show that titanium might be advantageous, at least in gram-positive bacteria.
For clinical practice, our findings imply that despite the enlarged surface of the 3D-printed titanium cage Staphylococci adhere less than on PEEK cages, and therefore, it should be taken into consideration to give the advantage in infectious processes when stabilisation of the spine is needed. For cases, where E.coli is isolated or suspected, it might be better to use PEEK instead of titanium cages. In the future, additional studies are needed to further elucidate potential in vivo factors affecting the adhesion characteristics on different implant materials.
Abbreviations
- °C:
-
Degree Celsius
- CFU:
-
Colony forming units
- CT:
-
Computed tomography
- E.coli :
-
Escherichia coli
- h:
-
Hours
- H:
-
Height
- kV:
-
Kilovolt
- L:
-
Length
- LB:
-
Lysogeny broth
- min:
-
Minutes
- ml:
-
Millilitre
- mm:
-
Millimetre
- MRI:
-
Magnetic resonance imaging
- MRSA:
-
Methicillin-resistant Staphylococcus aureus
- MSSA:
-
Methicillin-sensitive Staphylococcus aureus
- OD:
-
Optical density
- PBS:
-
Phosphate-buffered saline
- PEEK:
-
Polyetheretherketon
- rpm:
-
Rounds per minute
- S.aureus:
-
Staphylococcus aureus
- SD:
-
Standard deviation
- SEM:
-
Scanning electron microscopy
- S.epidermidis:
-
Staphylococcus epidermidis
- Ti:
-
Polished titanium plate
- TiLi:
-
Porous trabecular cantered titanium cage
- TiMe:
-
Two surface titanium mesh plate
- TSB:
-
Trypticase soy broth
- W:
-
Width
- µl:
-
Microliter
References
Kehrer M, Pedersen C, Jensen TG, Lassen AT (2014) Increasing incidence of pyogenic spondylodiscitis: a 14-year population-based study. J Infect 68:313–320. https://doi.org/10.1016/j.jinf.2013.11.011
Murillo O, Grau I, Lora-Tamayo J et al.(2015) The changing epidemiology of bacteraemic osteoarticular infections in the early 21st century. Clin Microbiol Infect 21:254.e1-254.e8. https://doi.org/10.1016/j.cmi.2014.09.007
Kurtz SM, Lau E, Ong KL et al.(2012) Infection risk for primary and revision instrumented lumbar spine fusion in the Medicare population. J Neurosurg Spine 17:342–347. https://doi.org/10.3171/2012.7.SPINE12203
Ruf M, Stoltze D, Merk HR et al.(2007) Treatment of vertebral osteomyelitis by radical debridement and stabilization using titanium mesh cages. Spine 32:275–280. https://doi.org/10.1097/01.brs.0000261034.83395.7f
Robinson Y, Tschoeke SK, Finke T et al.(2008) Successful treatment of spondylodiscitis using titanium cages: a 3-year follow-up of 22 consecutive patients. Acta Orthop 79:660–664. https://doi.org/10.1080/17453670810016687
Schomacher M, Finger T, Koeppen D et al.(2014) Application of titanium and polyetheretherketone cages in the treatment of pyogenic spondylodiscitis. Clin Neurol Neurosurg 127:65–70. https://doi.org/10.1016/j.clineuro.2014.09.027
Tschöke SK, Fuchs H, Schmidt O et al.(2015) Single-stage debridement and spinal fusion using PEEK cages through a posterior approach for eradication of lumbar pyogenic spondylodiscitis: a safe treatment strategy for a detrimental condition. Patient Saf Surg 9:35. https://doi.org/10.1186/s13037-015-0083-4
Shiban E, Janssen I, da Cunha PR et al.(2016) Safety and efficacy of polyetheretherketone (PEEK) cages in combination with posterior pedicel screw fixation in pyogenic spinal infection. Acta Neurochir (Wien) 158:1851–1857. https://doi.org/10.1007/s00701-016-2924-z
Nolla JM, Ariza J, Gómez-Vaquero C et al.(2002) Spontaneous pyogenic vertebral osteomyelitis in nondrug users. Semin Arthritis Rheum 31:271–278. https://doi.org/10.1053/sarh.2002.29492
Sobottke R, Seifert H, Fätkenheuer G et al.(2008) Current diagnosis and treatment of spondylodiscitis. Dtsch Aerzteblatt Online. https://doi.org/10.3238/arztebl.2008.0181
Korovessis P, Vardakastanis K, Fennema P, Syrimbeis V (2016) Mesh cage for treatment of hematogenous spondylitis and spondylodiskitis. How safe and successful is its use in acute and chronic complicated cases? a systematic review of literature over a decade. Eur J Orthop Surg Traumatol 26:753–761. https://doi.org/10.1007/s00590-016-1803-x
Berbari EF, Kanj SS, Kowalski TJ et al.(2015) 2015 Infectious diseases society of america (IDSA) clinical practice guidelines for the diagnosis and treatment of native vertebral osteomyelitis in adults. Clin Infect Dis 61:e26–e46. https://doi.org/10.1093/cid/civ482
Stüer C, Stoffel M, Hecker J et al.(2013) A staged treatment algorithm for spinal infections. J Neurol Surg Part A Cent Eur Neurosurg 74:087–095. https://doi.org/10.1055/s-0032-1320022
Rochford ETJ, Poulsson AHC, Salavarrieta Varela J et al.(2014) Bacterial adhesion to orthopaedic implant materials and a novel oxygen plasma modified PEEK surface. Coll Surf B Biointerfaces 113:213–222. https://doi.org/10.1016/j.colsurfb.2013.09.012
Bal S, Gorth D, Puckett S et al.(2012) Decreased bacteria activity on Si3N4 surfaces compared with PEEK or titanium. Int J Nanomed 7:4829–4840. https://doi.org/10.2147/IJN.S35190
Rueden CT, Schindelin J, Hiner MC et al.(2017) Image J2: imageJ for the next generation of scientific image data. BMC Bioinformatics. https://doi.org/10.1186/s12859-017-1934-z
Grados F, Lescure FX, Senneville E et al.(2007) Suggestions for managing pyogenic (non-tuberculous) discitis in adults. Jt bone spine 74:133–139. https://doi.org/10.1016/j.jbspin.2006.11.002
Ren Y, Wang C, Chen Z et al.(2018) Emergent heterogeneous microenvironments in biofilms: substratum surface heterogeneity and bacterial adhesion force-sensing. FEMS Microbiol Rev 42:259–272. https://doi.org/10.1093/femsre/fuy001
Palmer J, Flint S, Brooks J (2007) Bacterial cell attachment, the beginning of a biofilm. J Ind Microbiol Biotechnol 34:577–588. https://doi.org/10.1007/s10295-007-0234-4
Duarte RM, Vaccaro AR (2013) Spinal infection: state of the art and management algorithm. Eur Spine J 22:2787–2799. https://doi.org/10.1007/s00586-013-2850-1
Costerton JW, Stewart PS, Greenberg EP (1999) Bacterial biofilms: a common cause of persistent infections. Science 284:1318–1322. https://doi.org/10.1126/science.284.5418.1318
Arciola CR, Campoccia D, Montanaro L (2018) Implant infections: adhesion, biofilm formation and immune evasion. Nat Rev Microbiol 16:397–409. https://doi.org/10.1038/s41579-018-0019-y
Harris LG, Meredith DO, Eschbach L, Richards RG (2007) Staphylococcus aureus adhesion to standard micro-rough and electropolished implant materials. J Mater Sci Mater Med 18:1151–1156. https://doi.org/10.1007/s10856-007-0143-0
Wu Y, Zitelli JP, TenHuisen KS et al.(2011) Differential response of Staphylococci and osteoblasts to varying titanium surface roughness. Biomaterials 32:951–960. https://doi.org/10.1016/j.biomaterials.2010.10.001
Berne C, Ellison CK, Ducret A, Brun YV (2018) Bacterial adhesion at the single-cell level. Nat Rev Microbiol 16:616–627. https://doi.org/10.1038/s41579-018-0057-5
Jamal M, Ahmad W, Andleeb S et al.(2018) Bacterial biofilm and associated infections. J Chin Med Assoc 81:7–11. https://doi.org/10.1016/j.jcma.2017.07.012
Acknowledgments
We thank Spineart, Switzerland, for providing the samples and cages for this study and Tim Rolvien, MD, Department of Osteology and Biomechanics and Department of Orthopaedics, University Medical Centre Hamburg-Eppendorf, Hamburg, Germany, for establishing contact to SvK.
Funding
Open Access funding enabled and organized by Projekt DEAL..
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors have any potential conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Fig. S1 a In vitro incubation and growth comparison of different bacterial strains on diverse materials (plates). Quantitative analysis of time-dependent bacterial growth of MSSA, MRSA, S.epidermidis and E.coli from 8 to 80 h (h) on PEEK-, titanium-(Ti) or titanium mesh (TiMe) plates (mean and standard deviation (SD) in colony forming units (CFU) x 104) (n=2, 3 cultures each for 24, 32, 48, 56 and 80 h; n=3, 3 cultures each for 8 and 72 h). b In vitro incubation and growth comparison of different bacterial strains on diverse materials (cages). Quantitative analysis with individual values in CFU x 104 after 8 and 72 h for PEEK (upper panel) and 3D-printed titanium (lower panel) oblique cages (mean and SD in CFU x 104) (n=4, 3 cultures each)
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Krätzig, T., Mende, K.C., Mohme, M. et al. Bacterial adhesion characteristics on implant materials for intervertebral cages: titanium or PEEK for spinal infections?. Eur Spine J 30, 1774–1782 (2021). https://doi.org/10.1007/s00586-020-06705-z
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00586-020-06705-z