Abstract
We investigated the volume averaging effect for air-filled cylindrical ionization chambers to determine the correction factors in a small photon field for a given chamber. We measured output factors with several cylindrical ionization chambers, and by using a mathematical method similar to deconvolution, we modeled the non-constant and inhomogeneous exposure function in the cavity of the chamber. The parameters in the exposure function and the correction factors were determined by solving a system of equations that we had developed by using the measured data and the geometry of the given chamber. The correction factors (CFs) were very similar to those obtained from Monte Carlo (MC) simulations. For example, the CFs in this study were found to be 1.116 for PTW31010 and 1.0225 for PTW31016 while the CFs obtained from MC simulations were reported as being approximately between 1.17 and 1.20 for PTW31010 and between 1.02 and 1.06 for PTW31016 in a 6-MV photon beam of 1 × 1 cm2. Furthermore, the method of deconvolution combined with the MC result for the chamber’s response function showed a similar CF for PTW 30013, which was reported as 2.29 and 1.54 for a 1 × 1 cm2 and a 1.5 × 1.5 cm2 field size, respectively. The CFs from our method were similar, 2.42 and 1.54. In addition, we report CFs for PTW30013, PTW31010, PTW31016, IBA FC23-C, and IBA CC13. As a consequence, we suggest the use of our method to measure the correct output factor by using the fact that an inhomogeneous exposure causes a volume averaging effect in the cavity of air-filled cylindrical ionization chamber. The result obtained by using our method is very similar to that obtained from MC simulations. The method we developed can easily be applied in clinics.
Similar content being viewed by others
References
M. Stasi, B. Baiotto, G. Barboni and G. Scielzo, Med. Phys. 31, 2792 (2004).
P. Francescon, W. Kilby, N. Satariano and S. Cora, Phys. Med. Biol. 57, 3741 (2012).
D. M. Duggan and C. W. Coffey, 2nd, Med. Dosim. 23, 153 (1998).
I. J. Das, G. X. Ding and A. Ahnesjo, Med. Phys. 35, 206 (2008).
F. Crop, N. Reynaert, G. Pittomvils, L. Paelinck, C. De Wagter, L. Vakaet and H. Thierens, Phys. Med. Biol. 54, 2951 (2009).
W. Ulmer and W. Kaissl, Phys. Med. Biol. 48, 707 (2003).
C. H. Sibata, H. C. Mota, A. S. Beddar, P. D. Higgins and K. H. Shin, Phys. Med. Biol. 36, 621 (1991).
P. D. Higgins, C. H. Sibata, L. Siskind and J. W. Sohn, Med. Phys. 22, 1663 (1995).
G. Bednarz, M. Saiful Huq and U. F. Rosenow, Phys. Med. Biol. 47, 3643 (2002).
D. Gonzalez-Castano, J. Pena, F. Sanchez-Doblado, G. H. Hartmann, F. Gomez and A. Leal, Med. Biol. Eng. Comput. 46, 373 (2008).
R. Alfonso et al., Med. Phys. 35, 5179 (2008).
P. Francescon, S. Cora and N. Satariano, Med. Phys. 38, 6513 (2011).
C. Bassinet et al., Med. Phys. 40, 071725 (2013).
T. S. Underwood, H. C. Winter, M. A. Hill and J. D. Fenwick, Med. Phys. 40, 082102 (2013).
C. M. Ma and A. E. Nahum, Phys. Med. Biol. 36, 413 (1991).
F. Sanchez-Doblado et al., Phys. Med. Biol. 48, 2081 (2003).
C. Fox, T. Simon, B. Simon, J. F. Dempsey, D. Kahler, J. R. Palta, C. Liu and G. Yan, Med. Phys. 37, 477 (2010).
D. Czarnecki and K. Zink, Phys. Med. Biol. 58, 2431 (2013).
P. H. Charles, S. B. Crowe, T. Kairn, R. T. Knight, B. Hill, J. Kenny, C. M. Langton and J. V. Trapp, Phys. Med. Biol. 58, 4501 (2013).
A. Ralston, P. Liu, K. Warrener, D. McKenzie and N. Suchowerska, Phys. Med. Biol. 57, 2587 (2012).
G. Cranmer-Sargison, S. Weston, J. A. Evans, N. P. Sidhu and D. I. Thwaites, Med. Phys. 38, 6592 (2011).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Park, K., Choi, W., Park, S. et al. Determinations of the correction factors for small fields in cylindrical ionization chambers based on measurement and numerical calculation. Journal of the Korean Physical Society 67, 57–62 (2015). https://doi.org/10.3938/jkps.67.57
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.3938/jkps.67.57