Abstract
An injection moulding method was used for fabricating solid-state reference electrodes (Ag/AgCl type) based on a polymer/inorganic salt composite. In this method, a silver/silver chloride wire was placed inside a mould into which the mixture of polymer and inorganic salt was injected. The obtained solid-state composite reference electrodes were extensively tested to study the influence of different parameters such as solution composition, the concentrations and mobility of ions and pH on the potential stability of the electrodes. These experiments revealed that the composite reference electrodes are insensitive to the matrix effect, have excellent potential readings stability and considerably reduced leakage of inorganic salt. The composite reference electrodes were compared favourably to high-quality commercial reference electrodes. It was concluded that the composite reference electrodes obtained by injection moulding are of analytical quality allowing for continuous, prolonged and intensive usage.
Similar content being viewed by others
References
Inzelt G, Lewenstam A, Scholz F (2013) Handbook of reference electrodes. Springer, Berlin
Blaz T, Migdalski J, Lewenstam A (2005) Analyst 130:637–643
Anastasova-Ivanova S, Mattinen U, Radu A, Bobacka J, Lewenstam A, Migdalski J, Danielewski M, Diamond D (2010) Sens Actuators B 146:199–205
Diamond D, McEnroe E, McCarrick M, Lewenstam A (1994) Electroanalysis 6:962–971
Bakker E (1999) Electroanalysis 11:788–792
Cicmil D, Anastasova S, Kavanagh A, Diamond D, Mattinen U, Bobacka J, Lewenstam A, Radu A (2011) Electroanalysis 23:1881–1890
Ciobanu M, Wilburn JP, Buss NI, Ditavong P, Lowy DA (2002) Electroanalysis 14:989–997
Ciobanu M, Wilburn JP, Lowy DA (2004) Electroanalysis 16:1351–1358
Guth U, Gerlach F, Decker M, Oelssner W, Vonau W (2009) J Solid State Electrochem 13:27–39
Kisiel A, Marcisz H, Michalska A, Maksymiuk K (2005) Analyst 130:1655–1662
Mattinen U, Bobacka J, Lewenstam A (2009) Electroanalysis 21:1955–1960
Michalska A, Kisiel A, Maksymiuk K (2012) The 14th international meeting on chemical sensors, pp 683–686
Rius-Ruiz FX, Kisiel A, Michalska A, Maksymiuk K, Riu J, Rius FX (2011) Anal Bioanal Chem 399:3613–3622
Vonau W, Oelssner W, Guth U, Henze J (2010) Sens Actuators B 144:368–373
Mousavi Z, Granholm K, Sokalski T, Lewenstam A (2013) Analyst 138:5216–5220
Granholm K, Ek P, Sokalski T, Harju L, Bobacka J, Ivaska A (2009) Electroanalysis 21:2014–2021
Acknowledgments
The study was supported by the SalWe Research Program for IMO (Tekes—the Finnish Funding Agency for Technology and Innovation grant 648/10). The authors would like to thank Carl-Johan Wikman for his technical assistance with the injection moulding.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Granholm, K., Mousavi, Z., Sokalski, T. et al. Analytical quality solid-state composite reference electrode manufactured by injection moulding. J Solid State Electrochem 18, 607–612 (2014). https://doi.org/10.1007/s10008-013-2294-x
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10008-013-2294-x