Background

The Zenith Alpha thoracic endovascular graft (Cook Medical, Bloomington, IN, USA) is a low-profile system that facilitates the deployment of its proximal portion (Illig et al. 2015; Torsello et al. 2012) and could be a viable alternative for treating patients with an aortic arch proximal landing zone, to facilitate precise deployment (Pane et al. 2018).

This is the first report of the successful use of a snare to troubleshoot the deployment failure of the proximal portion of a Zenith Alpha thoracic endovascular graft.

Case presentation

An 80-year-old woman who underwent thoracic endovascular aortic repair (TEVAR) for subacute phase Stanford type B ulcer-like projection aortic dissection revealed type Ia and Ib endoleaks (ELs) on a 1-year follow-up computed tomography angiogram (CTA). We determined that additional stent grafting was indicated.

Preoperative CTA showed that the proximal neck diameter was between 40 and 43 mm; therefore, during the initial TEVAR, two stent grafts (SGs) were utilized. Under general anesthesia, 8-F, 5-F, and 6-F sheaths were inserted via the left common femoral artery, right femoral artery, and left brachial artery, respectively.

The type Ib EL, located in the distal descending thoracic aorta, was identified using digital subtraction angiography (DSA). A 0.035-in. Lunderquist (Cook Medical, Bloomington, IN, USA) guidewire (GW) was inserted via the left common femoral artery. Thereafter, a Zenith Alpha proximal component (diameter/length: 46/233 mm) with an introduction system diameter of 20-F was inserted. It was placed approximately 15 mm from the SG inferior component (Th12 level), implanted previously.

The type Ia EL was identified in the aortic arch, and a Zenith Alpha proximal component (diameter/length: 46/179 mm) with an introduction system diameter of 20-F was inserted. An attempt was made to deploy the SG directly posterior to the bifurcation of the left subclavian artery in Zone 3 (approximately 15 mm from the previously placed SG’s upper end); however, the inner curved side of the proximal portion stopped perpendicular to the vascular wall and failed to deploy properly (Fig. 1). Various unsuccessful attempts at repositioning were performed. Finally, the sheath in the left brachial artery was replaced with a 6-F guiding sheath (length: 53 cm), which was inserted into the aortic arch. The right femoral artery sheath was replaced with a 6-F sheath, and an Amplatz Goose Neck snare (Medtronic, Minneapolis, MN, USA) (wire diameter: 25 mm) was inserted through it and used to grasp the distal side of the abovementioned guiding sheath (Fig. 2a). The end of the left brachial artery sheath was rotated toward the inner curved side of the SG proximal component. Using a 4-F shepherd hook catheter, we passed a 0.035-in. Radifocus (Terumo Medical Corporation, Tokyo, Japan) GW under the inner curved side of the SG proximal component.

Fig. 1
figure 1

Fluoroscopic image immediately after additional stent grafting. The inner curved side of the proximal portion of the additional SG stopped perpendicular to the vascular wall (red arrow) and failed to deploy properly

Fig. 2
figure 2

Schema of the double-snare technique. a Snare grasping of the distal side of the guiding sheath via left brachial artery. b Pulling the snare wire and advancing the guiding sheath toward the inner curved side of the proximal portion of the SG. c Passing the Radifocus GW through the temporarily opened snare loop. d The hard loop structure (guiding sheath end/IMA catheter/GW) under the inner curved side of the proximal portion of the SG. e, f Pulling the guiding sheath toward the left subclavian artery and repositioning (yellow arrow) the inner curved side of the proximal portion of the SG

We then pulled the snare wire forward, to the greatest extent possible, and by placing tension on it, faced the left brachial artery sheath toward the lesser curved side of the proximal portion of the SG (Fig. 2b); we used the 0.035-in. Radifocus GW, to insert a 6-F internal mammary artery (IMA) catheter as deep as possible into the insufficiently deployed portion. After passing through and rotating it, we passed the Radifocus GW through the temporarily opened snare loop, and by grasping it (Fig. 2c), we could pass the hard loop structure (guiding sheath end/IMA catheter/GW) under the inner curved side of the SG proximal portion (Fig. 2d). We then carefully and slowly pulled the guiding sheath to its original position, which repositioned the inner curved side of the SG proximal portion to a position parallel to the vascular wall (Fig. 2e, f), where it was successfully deployed (Fig. 3).

Fig. 3
figure 3

Fluoroscopic image immediately after repair of the incomplete deployment of the SG using the double-snare technique. A catheter and GW are inserted through a guiding sheath and rotated under the stent-graft. The GW is passed through the loop (yellow arrow) of a snare, permitting deployment of the SG (red arrow)

We opened the snare loop and used a Tri-Lobe Balloon Catheter (W. L. GORE & Associates, Flagstaff, AZ, USA) to apply pressure from the lumen side of the SG, which allowed us to remove the 6-F IMA catheter. Once again, we applied pressure on the SG proximal side using the Tri-Lobe Balloon Catheter, reinforced the proximal sealing (Fig. 4), and performed DSA. Type Ia and Ib ELs remained but were less than those before additional TEVAR. One week later, she was discharged. One year later, CT showed no interval change in size of aortic aneurysm with dissection, and she has been followed on an outpatient basis.

Fig. 4
figure 4

Fluoroscopic image under reinforcement of proximal sealing using ballooning. Pressure applied on the proximal side of the SG using Tri-Lobe Balloon Catheter (red arrow), which reinforces the proximal sealing

Conclusions

The proximal portion of the Zenith Alpha thoracic endovascular graft proximal component has a mechanism that reduces bird beak formations proximal to the aortic arch to a great extent. However, based on our experience, if the diameter of the proximal blood vessel is 40 mm or larger, incomplete deployment is a possibility when placing the SG in a region from the proximal aortic arch to the isthmus.

In the present case, the incomplete deployment might have been caused by interference with the previously placed proximal bare stent portion of SG. When placing additional SGs following TEVAR, caution must be exercised to prevent the bare stents from interfering with each other, particularly when the Zenith Alpha thoracic endovascular graft is used.

Non-deployment of the proximal portion of a Zenith Alpha thoracic endovascular graft could be rectified using the double-snare technique described here, in which a snare wire is passed through the left subclavian and femoral artery for Zone 3 placement.