Background

Uveitis, a significant cause of blindness worldwide with a global prevalence of 38–114.5 per 100,000 [1, 2], describes a group of conditions characterised by intraocular inflammation [3, 4]. Although uveitis can occur at any age [3], it commonly affects the working-age group and therefore has a substantial socioeconomic impact [5].

Anterior uveitis is characterised by inflammation of the anterior uveal tract. The acute symptoms are primarily pain, redness and photophobia, but vision can be affected. With adequate treatment, these symptoms are reversible; however, permanent and vision-threatening complications (such as posterior synechiae formation, cataract and secondary glaucoma) can develop in the presence of prolonged inflammatory activity.

Inflammation in the anterior chamber (AC) can be observed as changes of the aqueous humour. The aqueous humour is a clear fluid in the anterior chamber of the eye that provides nutrition, removes excretory products of metabolism, transports neurotransmitters, stabilises the ocular structure and contributes to the regulation of the homeostasis of ocular structures [6]. In health, the aqueous humour is well isolated from the blood by blood-aqueous barriers including the endothelial cells of the iris capillaries and retinal vessels and non-pigmented ciliary body epithelium. The aqueous is mostly water, but also contains electrolytes, carbohydrates, glutathione, urea, oxygen, carbon dioxide, and proteins [6, 7]. In the absence of inflammation, only small proteins are present in low concentrations in the aqueous. Inflammation causes breakdown of the blood-aqueous barriers and results in the release of inflammatory cells and proteins of higher molecular weight into the eye [8, 9]. The accumulation of proteins in the aqueous humour can be observed as flare or Tyndall effect, which is an optical phenomenon of cloudiness in the aqueous humour due to increased protein content.

The ability to accurately measure the concentration of proteins in the aqueous is clinically important for detecting acute inflammatory episodes and assessing response to treatment [10]. There are several established methods for quantifying aqueous protein levels, including clinical examination and laser flare photometry. However, with the exception of aqueous paracentesis (invasive sampling of the aqueous humour), existing methods are surrogate measures relating to the change in optical properties of the aqueous humour when protein levels are increased. Each method has its limitations, and it is unclear whether measurements from different methods agree with each other. The following section discusses the pros and cons of existing measures for aqueous protein concentration.

Invasive quantification of aqueous protein sampling: aqueous paracentesis

Currently, the only means of directly measuring the protein content of the aqueous is by aqueous paracentesis and analysis in a laboratory setting [11, 12]. While this method provides the most accurate quantification of aqueous constituents, it is not used routinely in clinical practice as it is invasive and carries risk of sight-threatening complications [13]. Consequently, its use is mainly limited to the research setting, and non-invasive methods are used preferably in clinical practice.

Non-invasive quantification of aqueous protein level

Clinical examination

The current clinical standard for AC flare grading was defined at the Standardisation of Uveitis Nomenclature (SUN) working group consensus meeting in 2005 [4]. Prior to this, several systems existed for grading AC flare [14,15,16,17]. Slit-lamp biomicroscopy enables a clinician to grade increasing protein concentrations in the aqueous humour due to back-scattering of light emitted from a slit-lamp. The SUN system grades flare according to the observer’s ability to visualise details of the iris and lens behind the aqueous. Grades range from 0, which corresponds to no visible flare, to 4+, which corresponds to intense flare (Table 1).

Table 1 The Standardisation of Uveitis Nomenclature (SUN) working group grading scheme for anterior chamber flare [4]

Grading of flare based on clinical examination is subjective and subject to high intra- and inter-observer variability [18]. In addition, it is a non-continuous scale, leading to large steps in disease activity between categories. Detection and monitoring of inflammatory changes in anterior uveitis is critically important for clinical management, decision making and clinical trials investigating therapeutic agents [19]. The need for a more robust and reproducible method of measurement is well-recognised within the uveitis community. Newer imaging techniques have become available in recent years; therefore, it is timely to carry out a review of whether they can be purposed for this unmet need and evaluate how they compare to the current standard.

Laser flare and cell photometry

In 1988, Sawa et al. first described laser flare photometry (LFP) as a new method to precisely determine protein concentration in the aqueous humour [20]. LFP is a rapid and non-invasive technique which quantifies AC flare by projecting a laser beam through the anterior chamber and measuring the amount of back-scattered light. Particles present in the aqueous reflect photons of light, which are then measured by an inbuilt photomultiplier. The intensity of back-scattered light is proportional to the concentration and size of proteins in the aqueous chamber. Most models of LFP are capable of measuring flare, but the technology can also be used to count AC cells. However, assessment of AC cells using LFP is less accurate as inflammatory cells cannot be differentiated from other large particles such as pigment particles, debris or red blood cells [9].

Optical coherence tomography (OCT)

OCT is an imaging technique which has become available more recently. It uses coherent light to capture high resolution 2- and 3-dimensional images of structures within the eye. It is a fast and non-invasive method of obtaining high precision images and measurements of ocular structure [21]. Anterior segment OCT (AS-OCT) can provide cross-sectional imaging of the AC. Recently, OCT has demonstrated the ability to detect higher light reflectivity signals on AS-OCT during active inflammation [21].

Aim

The primary aim of this review is to identify all available non-invasive, instrument-based techniques with the potential to quantify protein levels in the aqueous as a measure of AC inflammation. We will evaluate the level of correlation of these measurements with clinical grading using slit lamp examination and/or actual aqueous protein concentration. Where reported, we will also evaluate the level of reliability of each test.

A secondary aim of the review is to evaluate the correlation between clinical grading and protein concentration measured by laboratory analysis of aqueous samples.

The following questions are proposed:

  • Which index tests have the potential to quantify AC inflammation in uveitis?

    • What is the level of correlation between the index test and clinical grading?

    • What is the level of correlation between the index test and aqueous protein concentration?

    • What is the level of repeatability of each test?

  • What is the level of correlation between clinical grading using the slit lamp and aqueous protein concentration?

Methods

Protocol

This protocol has been designed in accordance with the guidelines of Preferred Reporting Items for Systematic Review and Meta-Analysis Protocol (PRISMA-P) [22]. The systematic review will be reported in accordance to the PRISMA guidelines [23].

Systematic review registration

This systematic review has been registered on PROSPERO (CRD42017084167) [24].

Searches

Databases

Our search strategy will include the following areas of interest: anatomical location (‘anterior chamber’), the pathological finding (‘flare’) and the disease context (‘uveitis’). To achieve optimal sensitivity, no search terms will be applied for the ‘technologies/tests’. For bibliographic databases, free text and index terms will be combined for each search element where possible. A sample search strategy for MEDLINE is included in the Appendix 1. We will search:

  • Bibliographic databases of published studies.

    • MEDLINE (Ovid), 1946 to present

    • Embase (Ovid), 1947 to present

    • Cochrane Central Register of Controlled Trials (CENTRAL), database inception to present

    • Centre for Reviews and Dissemination database (Health Technology Assessments and the Database of Abstracts and Reviews of Effects), database inception to present

  • Registers of clinical trials

  • Abstract and conference proceedings

    • British Library’s ZETOC.

    • Conference proceedings Citation Index (Web of Science).

  • Dissertations, theses

  • British library Ethos

  • ProQuest. www.proquest.com

  • Grey literature

No restrictions will be placed on year or language of publication. The literature search results will be entered onto EndNote (Clarivate Analytics, Philadelphia, PA) to facilitate removal of duplicates, study selection, recording decisions and references. References to other works will be considered for inclusion.

Selection criteria

Participants/population

Those with evidence of anterior chamber flare and/or a diagnosis of uveitis, irrespective of active or inactive inflammation, will be included. There will be no restrictions on age, gender, ethnicity, or underlying aetiology or anatomical subtype. Studies with only healthy participants and studies on animals will not be included.

Index test

Any study reporting one or more non-invasive, instrument-based technology with the potential for quantifying AC flare or aqueous protein measurements will be included. To address the secondary outcome (clinical grading versus laboratory measurements of aqueous proteins), any clinical grading system can be considered the index test.

Reference test

The reference test should be either the current clinical standard—clinician-based AC flare grading, or the laboratory standard—measurements of aqueous protein from aqueous paracentesis.

For the clinical grading versus laboratory analysis, the reference test must be a laboratory-based technique for measuring aqueous protein concentration of extracted aqueous samples.

Primary outcomes

The level of correlation (correlation coefficient) of index test measurements with clinical grading using slit lamp examination and/or actual aqueous protein concentration.

Secondary outcome

Intra/inter-observer reliability and repeatability of an index test (kappa statistic) and the level of correlation between clinical grading and actual protein concentration (clinical grading as the index test and protein concentration as the reference test).

Type of study

There will be no restrictions on study design; however, evaluation of correlation between index test and clinical grading using slit-lamp examination/aqueous protein measurements require both tests to be done in a cross-sectional manner. If multiple time points are included, we will include results from all time points, but comparisons between tests will be cross-sectionally. Only those studies where measurements are taken within a reasonable time point (within 24 h of each other) will be included. Case reports involving only one subject, commentaries, opinion articles and pictorial articles will not be included.

Selection process

Titles and abstracts of studies will be screened for relevance to the review, to remove obviously irrelevant studies Two independent reviewers will carry out study selection and reach consensus by discussion or referral to a third reviewer.

Full text of potentially relevant articles will be retrieved and assessed for inclusion in the review against the full selection criteria.

Data extraction

A standardised data extraction form in Microsoft Excel (Microsoft, Washington, US) will be used to extract data from included studies. The extraction process will be carried out by two independent reviewers with referral to a third reviewer if necessary. Information extracted from all studies will include:

  • Study characteristics

    • Title, authors, publication year, journal and language

    • Sample size

    • Study design

    • Index test used

      1. 1.

        Manufacturer and model (including resolution, default settings)

      2. 2.

        Measurements acquisition protocol

      3. 3.

        Measurement analysis protocol

    • Clinical grading system measurements

      1. 1.

        Clinical grading system used and whether any modifications were made

      2. 2.

        Number of observers

    • Aqueous protein concentration

      1. 1.

        Context and reason for aqueous extraction

      2. 2.

        Aqueous extraction protocol

      3. 3.

        Aqueous analysis protocol

  • Patients’ characteristics

    • Age, gender and ethnicity

    • Underlying aetiologies (anatomical subtype, aetiological classification)

    • Active or inactive disease

    • If the study involves a therapeutic intervention: treatment details (indication, drug, dosage, route, subject pre/post treatment status, length of follow-up and number of time points suitable for review question)

  • Outcomes and findings

    • Data will be extracted in preparation for four separate analyses:

      1. 1.

        Evaluation of correlation between index tests and actual aqueous protein concentration: The correlation coefficient will be directly extracted.

      2. 2.

        Evaluation of correlation between index tests and clinical grading systems: The correlation coefficient reported will be directly extracted.

      3. 3.

        Evaluation of correlation between the clinical grading system and aqueous protein concentration: The correlation coefficient will be directly extracted.

      If no correlation coefficient is reported, Index and reference test measurements will be extracted and, provided they are matched, used to calculate the correlation coefficient. If the two measurements are not matched, we will contact the authors for matched measurements.

      1. 4.

        Evaluation of reliability and repeatability of an index test: Studies reporting intra- and inter-observer reliability will be analysed separately for assessment of repeatability. The reported kappa values will be extracted for intra-observer reliability, inter-observer reliability or both.

Cross-sectional measurements may be nested within longitudinal studies whose aims are not primarily to compare performance or correlation between two tests. In this situation, measurements at each time point will be extracted and analysed as individual cross-sectional comparisons.

Quality assessment

All included studies will be assessed for quality using elements of the Quality Assessment of Diagnostic Accuracy Studies tool (QUADAS2) [25]. Although QUADAS2 is designed for comparison of diagnostic accuracy, rather than agreement of test measurements, it is the most suitable existing tool for evaluating risk of bias in diagnostic tests. We have modified elements of the existing QUADAS2 signalling questions to suit the aims of this systematic review (see Appendix 2). For example, the original QUADAS2 framework includes a question regarding whether thresholds were used and pre-specified. We have removed this item as we are not aiming to assess diagnostic accuracy. Similarly, we have supported the QUADAS2 with additional items such as the addition of the question ‘were index test acquisition and analysis parameters determined a priori and consistent for all study participants?’ Existing questions regarding blinding during test interpretation, applicability to the review question, and flow and timing of tests are kept as they remain important for this review.

Assessment will be carried out at the study level. Two independent reviewers will carry out quality assessment and refer to a third reviewer if needed. For studies assessing correlation between two tests, the four risks of bias domains below will be rated as low, high or unclear.

For studies investigating the reliability of a single index test, additional considerations will be made for:

  1. 1.

    Intra-observer reliability studies: the conditions under which the index test was performed should be reported and standardised.

  2. 2.

    Inter-observer reliability studies: any differences between observer characteristics, such as seniority and experience, should be reported.

Data synthesis

Studies will be included in four groups for data synthesis, one for each outcome: correlation between index test and the clinical grading, correlation between index test and aqueous protein concentration, correlation between the SUN grading system and laboratory measurement of aqueous protein concentration, and reliability of index test.

For each outcome, a narrative synthesis of tabulated evidence will be conducted and, where possible, supported with a meta-analysis.

  1. 1.

    Evaluation of correlation between index test and the SUN grading system

    These studies will be grouped by the type of technology used for the index test (i.e. LFP, OCT). If the data permits, correlation coefficients between each index test versus clinical grade will be compared and pooled for meta-analysis using a random effects model.

  2. 2.

    Evaluation of correlation between index test and laboratory measurement of aqueous protein concentration

  3. 3.

    Evaluation of correlation between the SUN grading system and laboratory measurement of aqueous protein concentration

  4. 4.

    Evaluation of reliability of an index test

    These studies will be grouped by type of technology as above. We will analyse intra- and inter-observer reliability separately for each technology. If the data permits, reliability values (such as kappa values) will be compared and pooled for meta-analysis.

Studies will be assessed for clinical and methodological homogeneity to determine whether it is possible appropriately pool data for meta-analysis. The I2 statistic will be used to quantify heterogeneity across studies in each meta-analysis. Correlation coefficients will be normalised using Fisher’s Z transformation for meta-analysis and back transformed for inference. Meta-analysis of kappa statistics will be performed using an inverse-variance weighted random effects model with standard errors estimated from reported 95% confidence intervals. All data will be reported narratively, regardless of whether data pooling and meta-analysis are possible or not.

Depending on the study data, appropriate sensitivity analyses will be carried out if there is significant heterogeneity between studies. Subgroup analyses for anatomical and aetiological subtypes of uveitis, index test technology and experience of graders (for clinical AC flare grading) will be considered to explore sources of heterogeneity. It is expected that only a small number of studies will meet inclusion criteria, which will limit our capacity to carry out meaningful subgroup analyses.

Discussion and potential impact

The assessment of inflammation in anterior uveitis currently relies on imperfect clinical methods. An increase in aqueous protein concentration is a detectable change which occurs in the presence of inflammatory breakdown of the blood ocular barrier. Laboratory measurement of aqueous protein concentration in aqueous samples obtained through paracentesis is a good gold standard test but carries too many risks to make it practical for monitoring patients. Thus, clinicians must resort to surrogate markers for aqueous protein levels such as ‘flare’.

Whilst the SUN grading system marked a significant effort towards unifying the method for assessing uveitic inflammation, it continues to rely on clinical examination and the subjective appearance of aqueous ‘clarity’. Many factors can also affect an observer’s ability to observe flare including the slit-lamp optics, degree of illumination, ambient conditions and the observer’s level of expertise [26].

The potential for instrument-based techniques for assessing AC inflammation carries significant advantages over clinical grading systems because they are objective, less operator dependent and produce a continuous numerical value that is precise at even low levels of inflammation. Objective measures of inflammation for monitoring AC cells and vitreous haze have been proposed, and OCT measurement of central macular thickness is already established in routine clinical practice [19, 21, 27, 28]. With the use of automated analysis software, these techniques could allow comprehensive objective disease monitoring, potentially even in a virtual care model.

To date, the use of non-invasive instrument-based techniques for assessing AC flare has not been assessed in a systematic way. The need for more objective measures of disease activity is well-recognised within the uveitis community, and given the rise in new imaging technologies over recent years which have the potential to meet this demand, it is timely to carry out a review of how they compare to the current standard. This systematic review will identify all technologies available for quantifying ‘flare’ and their level of reliability and correlation with laboratory measurement of aqueous protein concentration and/or clinical grading. The findings of this review could contribute to the validation process of instrument-based methods for monitoring inflammatory activity and guiding treatment decisions in uveitis.