Background

The Ebolavirus outbreak dominated international news in 2014 and the World Health Organization reported more than 28,000 cases worldwide and over 11,000 deaths [1]. After the initial outbreak in Guinea in December 2013 [2], there was intense and wide-spread transmission to neighbouring countries including Sierra Leone and Liberia. The virus was also exported to Mali, Nigeria, Senegal, the United States of America, and Europe [3, 4] Transmission was drastically reduced after global intervention, and all the affected countries have been declared “Ebola free” and travel restrictions have been lifted [5]. Nonetheless, 15 countries in Africa are predicted to be at risk of Ebolavirus outbreaks [6], based on the proximity of people to animals that harbour the virus, as well as environmental factors including climate change and urbanization. The Freetown Declaration on the 16th of October 2015 at the end of the Ebola outbreak in Sierra Leone called for action against the re-emergence of Ebolavirus, including the improvement of laboratory facilities for early detection [7]. The most recent reported Ebolavirus disease (EVD) outbreak, in July 2017 in the Democratic Republic of Congo, with 5 laboratory confirmed cases and 4 deaths [8] further stresses the importance of action in the diagnostic space.

Early laboratory diagnosis is pivotal for the prevention of re-emergence of Ebolavirus. Current laboratory diagnosis includes real-time polymerase chain reaction (RT-PCR), antigen-capture enzyme-linked immunosorbent assay (ELISA), immunoglobulin M (IgM) and immunoglobulin G (IgG) ELISA [9,10,11,12,13,14,15,16]. Additionally, there have been recent Ebolavirus assay developments using antibody and antigen detection and isothermal amplification technologies [17,18,19,20]. The assays deployed during the last Ebola outbreak were a combination of antigen-antibody-based and nucleic acid tests: RealStar Zaire Ebolavirus RT-PCR Kit (Altona Diagnostics, Hamburg, Germany); GeneXpert® Ebola Assay (Cepheid, Sunnyvale, California, USA), FilmArray® BioThreat/Ebola Panels (BioFire, Salt Lake City, Utah, USA); ReEBOV Antigen Rapid Test (Corgenix, Broomfield, Colorado, USA); SD Q Line Ebola Zaire Ag test (SD Biosensor, Suwon, Gyeonggi-do, Republic of Korea). The deployment of these assays depended on laboratory capabilities, rapid turn-around and easy use of the Ebolavirus test device. However, these assays did not consider any future detection of the virus; a future Ebolavirus outbreak has been predicted to affect 15 countries in Africa [6]. Additionally, the molecular assay results are not compatible with clinical laboratories that lack PCR expertise and infrastructural requirements, making the assays less relevant to technicians in resource-limited settings. Detection of viral RNA has proven to be most effective for diagnosis of Ebolavirus infections from the early to late stage of illness, however the standard RT-PCR Ebola assay is not readily available in areas where it is most needed. Among the isothermal technologies developed [21,22,23,24,25,26,27,28,29,30], recombinase polymerase amplification (RPA) appears very amenable to resource limited settings [28, 31]. The technology is fast, amplifying nucleic acids in 10 min, and can operate at body temperature or even lower (22 °C to 38 °C) [32]. The sensitivity and specificity of RPA is comparable to PCR, and is amenable to all PCR-derived amplification detection strategies including, real-time fluorescence and lateral flow strips (LFS) [31]. RPA has previously been reported for the detection of several RNA viruses, including Middle East respiratory syndrome coronavirus, Rift Valley fever virus, Ebolavirus and several filoviruses. These viruses were detected using portable fluorescent equipment, a real-time procedure [32,33,34,35,36,37,38,39,40]. Here we describe a rapid, sensitive, and specific assay for Zaire ebolavirus, which includes the current Guinea strain. Our assay uses RPA, but rather than using real-time fluorescent imaging, the assay uses the LFS as a detection format, making results very simple to interpret. A similar approach has also been used in the detection of RNA viruses such as Yellow fever virus [41] and Chlamydia trachomatis diagnosis [42]. The LFS component of the test (Milenia Biotec, Giessen, Germany) is designed to detect an RPA amplicon dual-labelled with FAM and biotin (supplied in the RPA primers and probes and incorporated during the amplification step). The amplicon is captured by gold-nanoparticles labelled with FAM-specific antibodies in the sample application area, before traversing to immobilized anti-biotin antibodies bound at a test line; precipitation of the gold nanoparticles at the test line results in the appearance of a red-blue band. Excess gold nanoparticles are also captured by species-specific antibodies bound to a control line; appearance of a red-blue band in the control line confirms correct operation of the strips in the absence of dual-labelled amplicon.

Our assay using the RPA-LFS method is highly amenable for low-resource laboratory, and has the potential to be deployed in future Ebolavirus outbreaks. The aim of this study was to develop an assay that is comparable and faster than RT-PCR with easy to interpret results. Additionally, we sought an assay that could be easily deployed in an outbreak situation, where limited resources preclude PCR laboratory facilities.

Methods

Sample preparation

Ebolavirus strains and related viruses (Table 1) were grown on Vero E6 cells and harvested from infected cell culture supernatant after centrifugation at 12,000 g. Stocks determined to have > 106 PFU/mL by standard plaque assay were used for RNA extraction (140 μL) using an RNeasy Mini Kit (QIAGEN, Australia) according to the manufacturer’s instructions. All the procedures and manipulation of Ebolavirus-infected cultures were performed in a Biosafety Level 4 Laboratory at the CSIRO Australian Animal Health Laboratory. All samples were eluted in 50 μL nuclease free water, and 5 μL was used in a 20 μL reverse transcription reaction. Reverse transcription was performed using SuperScript IV Reverse Transcriptase (Thermo Fisher Scientific, MA, USA) and random hexamers (Thermo Fisher Scientific, MA, USA) according to manufacturer’s instructions, and 5 μL resultant cDNA was used as template for RPA-LFS. Dengue 1–4 and West Nile virus (Kunjin NSW 2011 strain) RNA extractions were a gift from Professor Roy Hall (University of Queensland, Brisbane, QLD, Australia) and Plasmodium falciparum (malaria) genomic DNA was a gift from Prof James McCarthy (QIMR-Berghofer, Brisbane, QLD, Australia).

Table 1 Ebolavirus strains and related viruses used for RT-RPA-LFS assay

Ebola RPA assay

Primers and probes were synthesized by Integrated DNA Technology (Iowa, USA), and purified by standard desalting. Assays were first optimised using synthetic gene fragments of one of the isolates (Accession number; KJ660348.1), from 470 to 2210 bp (1794 bp) as a template (data not shown). Optimal primers EBZ#3F (5’ TCT CGT CCT CAG AAA GTC TGG ATG ACG CCG) and EBZ#3R (5’ Biotin-TAC TTG ATA CAC TGG GAT GAC TCT TTG CCG) and probe EBZ#3P (5’ FAM-CTY ACT GAA TCT GAC ATG GAT TAC CAC AAG ATC /idSp/TR ACA GCA GGT CTG TCC /3SpC3/) amplified a 132 base pair fragment of the Ebolavirus Makona (Guinea) strain (Accession number KJ660348.1, amplification occurred between nucleotides 476 and 608). The Ebola RPA assay was performed using the TwistAmp™ nfo kit in pellet format (TwistDx, Cambridge, United Kingdom) according to manufacturer’s instructions, but with modification to the primers and probe concentrations and final volume. Briefly, 29.5 μL of rehydration buffer was mixed with 2.1 μL of each forward and reverse primer (10 μM) and 0.6 μL of the target specific probe (10 μM). Then 34.3 μL of this master mix was added to the dry reagent pellet, followed by 5 μL of the template, and the pellet resuspended by aspirating and dispensing several times. The reaction was activated by addition of 2.5 μL magnesium acetate (280 mM) to the reaction mix, followed by incubation at 37 °C for 30 min. After amplification, 1 μL of the amplified product was diluted with 9 μL of running buffer (Milenia Biotec, Giessen, Germany) and added to the sample pad of the HybriDetect lateral flow strip (Milenia Biotec, Giessen, Germany). Strips were placed into tubes containing 100 μL of running buffer for 5 min, and photographed using a digital camera. Grey-scale converted images were analysed using ImageJ software (National Institutes of Health, MD, USA) to determine band-intensity, by measuring the mean grey value (limit to threshold), using a fixed area measurement, and subtracting from the maximum grey value (255). For each test band, the average of two neighbouring white spaces was subtracted from the band intensity to normalize the results. A sample was defined as positive if the normalised band intensity was 1.3 times higher than the standard deviation of the two neighbouring white space values.

Determination of analytical sensitivity and specificity

Following similar viral RNA assay development studies that employed RPA technology in establishing limit of detection using cloned gene [34, 35], we also performed standard ten-fold dilutions of an in-vitro synthesis (Genscript, USA) of the Zaire ebolavirus NP gene (Kikwit isolate) cloned into EcoRI/XhoI sites of pCAGGS. These pCAGGS were tested in two replicates using real-time TaqMan PCR with previously described Ebolavirus species specific primers and probe [43]. The assay was performed using SuperScript III Platinum Taq One-Step quantitative PCR Kit (Thermo Fisher Scientific, MA, USA) and the ABI7500 in Fast Mode. The plasmid dilutions were subsequently tested with RPA-LFS for determination of analytical sensitivity. Related Ebolaviruses (Table 1) and Dengue 1–4, Plasmodium falciparum, West Nile virus were also tested with RPA-LFS for determination of the analytical specificity of the assay.

Results

Establishing the RPA-LFS assay

Zaire ebolavirus species specific RPA primers and probes were designed based on the published sequence of the Guinea strain (Accession number; KJ660348.1) with the use of PRIMER-BLAST [44], which generated 20 nucleotides that were extended manually to a longer 30 nucleotides to be more amenable for RPA reactions. The sequences of the selected primers and probes were compared with an alignment of the NP gene sequences of seven isolates of the virus (Accession number; AF086833.2, KC242785.1, KC242796.1, KC242800.1, KC242792.1, KJ660346.1, KJ660347.1) to confirm homology with thee strains, which represented the various human outbreaks between 1976 and 1995 and the 2014 outbreak [2, 45,46,47]. Labelled primers and probes were tested for RPA followed by LFS detection (Zaire RPA-LFS) using a synthetic Ebolavirus template, and tested for optimal reaction times between 5 to 40 min. The optimal reaction time at 37 °C was 30 min, plus 5 min incubation of the LFS in the running buffer. A visible red colour band at the control and test lines were observed for positive controls while the no template control (negative control) had only one band at the control region of the strip. The control band on both strips indicated a valid RPA-LFS assay.

Analytical sensitivity and specificity of the RPA-LFS assay

The detection threshold of the Zaire RPA-LFS was determined using a dilution series of plasmid containing the NP gene (1.34 × 1010 to 1.34 × 101) copies / μL), and viral RNA from cultured Ebolavirus and comparing results to quantification using the gold standard real-time PCR assay. Detection using Zaire RPA-LFS showed the same sensitivity to RT-PCR, with a detection limit of 1.34 × 102 copies / μL (Fig. 1) and was notably faster (30 min versus > 2 h). We note that a faint band was sometimes observed in no template negative controls, however, this faint band was clearly distinguishable from true positives, as demonstrated by ImageJ analysis of black pixel density. Such image analysis could be performed in field situations through the use of a lateral flow reader. We note that the faint test bands in the no template negative controls and Fig. 2 have been shown to disappear using higher dilutions of RPA product in the lateral flow device (e.g. 1:100–1:200) [48] compared to the 1:10 dilution that we used, however, this may affect sensitivity of the assay. We also observed that in the case of very high concentrations of hybridization product, the intensity of the control band was affected, but this did not affect interpretation of results.

Fig. 1
figure 1

Analytical sensitivity of Ebolavirus RPA-LFS test. Sensitivity testing using both plasmid containing the Ebolavirus gene (top) and Zaire Ebolavirus RNA (bottom). a RT-PCR cycle threshold (Ct) values for a single sample, along with corresponding photograph of LFS with control bands (all samples) and test bands (positive samples) compared to copy number of serially diluted template DNA or RNA (copies/μL) and no template control. b Normalised pixel density (black values) from the assay displayed in (a). c Positive results compared to number of runs tested at that dilution. d Analytical sensitivity displayed as percentage of correct results from all runs. The test line appeared at every dilution down to 102 copies/μL, which was comparable with Real-Time PCR

Fig. 2
figure 2

Specificity of the Ebolavirus RPA-LFS test. a Representative photograph of LFS with control bands and test bands of different viral isolates, Plasmodium falciparum, positive control genomic template DNA, or no template control. b Normalised pixel density (black values) from the test displayed in (a). c Positive results of all test runs compared to number of individual runs. d Specificity displayed as percentage of correct results from all runs

In order to determine cross reactivity or analytical specificity of our assay, we tested cultured viral RNA from Reston, Sudan and Marburg viruses, as well as different strains of Zaire ebolavirus species (Table 1). Marburg viruses show the same haemorrhagic symptoms as Ebolaviruses and Ebola cases were mistaken for Marburg virus when it was first discovered in 1976 [46]. We also tested Plasmodium falciparum, Dengue virus 1–4 and West Nile virus (Kunjin NSW 2011 strain), as these pathogens are likely to be isolated from clinical samples in the same geographic region. Our Zaire ebolavirus RPA-LFS did not show a positive test result when exposed to the Marburg RNA, any of the other ebolavirus species, or other pathogens (Fig. 2), indicating the assay was specific to Zaire ebolavirus species. We confirmed this specificity by demonstrating a test line consistently appeared when exposed to viral RNA from cultures of different Zaire ebolavirus strains (Fig. 3).

Fig. 3
figure 3

RPA-LFS detection of Zaire ebolavirus strains. a Photograph of LFS with control bands and test bands of different Zaire ebolavirus strains, positive control (1.34 × 106 copies/μL synthetic template DNA) or no template control. b Normalised pixel density (black values) from the assay shown in (a). c Positive results compared to number of individual runs. d Specificity displayed as percentage of correct results

Discussion

Reliable and rapid diagnosis of patients suspected of having EVD is critical to prevention, limiting the spread and management of treatment. This need was clearly made apparent in the Freetown declaration, which was a collective response to the Ebolavirus outbreak that killed more than 11,000 people, made by delegates from the African Society for Laboratory Medicine and the World Health Organisation in Freetown, Sierra Leone, in October 2015. The declaration included a strong call to build and strengthen laboratory networks to prevent, detect and respond to new and future diseases threats [7]. Importantly, innovative diagnostic tools that are comparable to current gold standards for resource-limited settings have been identified as critical to EVD diagnosis. In particular, a major challenge with RT-PCR is its inappropriateness for settings where there is lack of expertise and infrastructure to support its implementation. Typically, PCR facilities are many miles away from district hospitals where they are critically needed. Hence, suspected EVD samples sent to central facilities can take two or more days before the patients are aware of their status. This delay has major impacts on patient isolation, care and treatment.

In the very active space of Ebola diagnostics post the 2014 outbreak, the challenge still remains to develop a rapid, sensitive, selective Ebolavirus test, and to undergo full regulatory assessment and field testing. Tests for IgM/IgG [49, 50] cannot detect early-stages of the disease before patients have mounted an immune response. Antibody-based diagnostics detecting Ebolavirus antigens like ReEBOV Antigen Rapid Test (Corgenix, Broomfield, Colorado, USA); SD Q Line Ebola Zaire Ag test; (SD Biosensor, Suwon, Gyeonggi-do, Republic of Korea); Ebola (Senova Immunoassay Systems, Weimar, Germany); OraQuick Ebola Rapid Antigen Test (OraSure Technologies, Bethlehem, Pennsylvania, USA) are reasonably easy to operate and offer rapid on-site detection incorporating small quantities of blood or serum as material. However, these antibody based diagnostic methods are not as sensitive as their RT-PCR test counterparts. Thus, currently used reference assays solely detect viral RNA, such as the RealStar Filovirus Screen RT-PCR Kit 1.0 and its improved RealStar Zaire Ebolavirus RT-PCR Kit version, Altona Diagnostics. These RT-PCRs were originally designed for research only and hence mostly include labour-intensive RNA extraction steps, temperature-sensitive reagents, expensive laboratory equipment and the need for manual read-out. Therefore, highly trained staff are making multiple precision steps in sophisticated high containment laboratories to deliver an EVD result with a turnaround time from four to 6 h. In contrast, an automated system integrates sample purification, nucleic acid amplification and detection of the target sequence using RT-PCR, additionally lowering the risk for technical personnel. A verified example of an automated system is the GeneXpert® System combined with the Xpert® Ebola Assay (Cepheid, Sunnyvale, California, USA), able to deliver results in ~ 2.5 h. Other companies have also designed automated or semi-automated Ebola assays (Idylla™ EBOV test (Biocartis, Mechelen, Belgium); BioThreat/Ebola Panels (BioFire, Salt Lake City, Utah, USA) for point-of-care; these systems tend to be easily portable (7.8 to 25.2 kg) but come with a high price tag.

In contrast, here we report a molecular assay for Zaire ebolavirus species detection using a LFS, and only requiring simple and inexpensive equipment (such as a 37 °C heating block and lateral flow reader). The assay is simple, fast, comparable with the current gold standard, and reduces the infrastructural requirements and technical needs commonly associated with the RT-PCR-based assays. Our assay has the same detection limit as PCR (Fig. 1) and was highly specific for the Zaire ebolavirus strains, which have been responsible for different outbreaks since 1976 as well as the recent outbreak (Figs. 2 and 3). The assay employs a novel isothermal amplification technology, RPA, which has previously been applied for the detection of DNA and RNA organisms [31]. Our assay sensitivity and specificity are similar to previously published studies using RPA for the detection of RNA viruses [32,33,34,35,36,37,38,39,40].

RPA is yet to demonstrate its potential in the detection of nucleic acids directly from clinical samples, apart from urine [42]. For blood testing, it is anticipated that RNA extraction would need to be performed first, possibly using rapid methods that do not require centrifuges, such as the SpeedXtract Nucleic Acid Kit, with a 15 min extraction protocol and requires only 20 μL of blood serum (Qiagen Hilden, Germany) [51, 52]. This nucleic acid kit was deployed in combination with a prototype diagnostic that employs real time version of RPA during the recent Ebolavirus outbreak and is based on a magnetic particle technology that removes inhibitors [52]. Similarly, although we used a two-step reverse transcription approach in sample preparation, for field detection a one-step method could be used as demonstrated previously by RPA-detection of Rift Valley fever virus [35], where MuLV reverse transcriptase was included in the RPA reaction pellets, making the cDNA synthesis step easier and amenable to resource-limited settings. In addition, a disposable detection device that allows the assay to be performed in a closed system could be used to prevent cross-contamination of samples, contain the spread of amplified products, and protect the operator during the procedure [53]. We note that previous RPA assay publications have indicated a constant shaking speed is required for a more stable signal on the strip, and that agitation also improves sensitivity and amplification time, particularly when the detection limit is low [42]. However, our RPA was performed without agitation and yet achieved similar sensitivity to RT-PCR. We chose to ignore the shaking event, because in a typical routine diagnostic laboratory this step might be forgotten or omitted as a result of scheduled turnaround time with other routine diagnostic assays or work load. It is possible that with shaking our sensitivity may have been even more improved and could reduce the reaction time below 30 min.

Our RPA-LFS for Zaire ebolavirus species, as a qualitative assay, when combined with the technologies for rapid low-resource detection described above, could provide a rapid response to Ebolavirus outbreaks in the future. The cost of components is approximately USD $10 per test, without the sample preparation step. In the 2014/2015 Ebolavirus outbreak over 28,000 cases were diagnosed. If another epidemic were to build with a similar scale, it is critical to take space limitations in high-containment facilities into consideration. Therefore, complete inactivation of Ebolavirus samples is essential to allow specimens to be manipulated outside of biological contained facilities. Currently, our test system requires RNA extraction with a commercially available kit. However, this step may be easily replaced by the use of TRIzol® LS reagent (Invitrogen Corp., Carlsbad, CA), which successfully inactivates members of the Filovirus species [54]. This would enable the assay to be performed outside of a high containment facility, accelerating turnaround time to diagnose an EVD case. However, replacing the labour-intensive low-throughput RNA isolation with either a safe, low-cost, high-throughput version, or eliminating RNA extraction completely, is still highly desirable. Such an optimised Ebolavirus test version could have high potential to be compatible with unprocessed clinical material, such as whole blood obtained by finger-stick. Indeed, Clancy and colleagues (2015) have successfully shown the detection of Streptococcus pneumoniae with whole blood as material using an isothermal amplification approach by RPA [55]. A field trial of our Ebolavirus test or an optimized version during an Ebolavirus outbreak is essential. Yet, it is impossible to predict the strain that future outbreaks may be caused by. Our primers and probe may also detect these future strains because the genetic diversity within the Ebolavirus (particularly Zaire ebolavirus species) has been very low to date, with a maximum of 2.7% nucleotide difference between sequences [56]. This minimal diversity is also evident with the Guinea strain which showed 97% identity with previous strains [2]. Additionally, the designed oligonucleotides that utilized the Guinea strain, detected previous strains, demonstrating a versatility that may help detect future strains. Similar assays could be designed to detect other Filovirus species responsible for outbreaks.

Conclusion

We have developed an Ebolavirus assay that could be implemented in low-resource laboratories that do not have the capacity for RT-PCR, and this assay could potentially be helpful in the next outbreak. However, field evaluation of the assay in a typical clinical setting will help to determine clinical sensitivity and specificity respectively, and areas of improvement.