Background

As malaria control decreases plasmodial infection and malaria disease, the inadequate sensitivity of current methods for Plasmodium falciparum becomes an increasingly important limitation [1]. Although asymptomatic individuals harbour a substantial fraction of the parasites in the community [2,3,4], those individuals do not seek care because they are not sick and often have too few asexual parasites detectable using microscopy or rapid diagnostic tests (RDTs, ≤ 200 per μl). However, because asymptomatic parasitized individuals can infect mosquitoes [5,6,7], the failure of current methods to identify asymptomatic infected individuals is an obstacle to the interruption of transmission and thus to the improvement of malaria control strategies necessary for its ultimate elimination.

In contrast to conventional diagnostic methods, PCR-based methods permit the exponential amplification of parasite DNA or RNA to detect P. falciparum. For this reason, PCR-based methods have the potential to provide greater sensitivity than microscopy or RDTs [8]. However, the information available about the sensitivity of PCR-based methods for P. falciparum in asymptomatic individuals and the factors affecting their sensitivity is limited. Although the sensitivity of PCR-based methods for P. falciparum is thought to depend on the template copy number [9, 10], the effects of the template copy number on the sensitivity of PCR-based methods for P. falciparum in asymptomatic individuals have not been examined systematically—by using similar PCR-based methods grouped by their template copy numbers.

PCR-based methods used to detect P. falciparum include but are not limited to: (1) conventional PCR, (2) nested PCR and (3) quantitative PCR (qPCR). Of those strategies, only qPCR permits estimation of the initial template DNA concentration from the cycle threshold (Ct) value—the number of amplification cycles required for the fluorescent signal to exceed the threshold level of background fluorescence. For that reason and because DNA is more stable than RNA, these studies emphasize the use of qPCR amplification of parasite DNA to detect P. falciparum in asymptomatic individuals.

To examine the effects of template copy number on the sensitivity of qPCR, we compared qPCR assays with template copy numbers from 1 to 160 for the detection of P. falciparum DNA in laboratory standards and smear-positive filter paper blots from asymptomatic infected human subjects. The hypothesis underlying these studies was that the sensitivity of qPCR should increase with the template copy number of the qPCR assay.

Methods

In silico development and screening of candidate qPCR assays

The qPCR assays examined initially were based on PubMed literature searches for template copy numbers and qPCR assays for P. falciparum. Additional candidate assays were then developed using the AlleleID®7 software from Premier Biosoft (Palo Alto, CA). Each candidate assay was screened using NCBI Primer-BLAST [11] to: (1) estimate its template copy number (number of DNA sequences yielding a positive fluorescent signal) based on the P. falciparum 3D7 genome (ASM2762), (2) examine the AT content of DNA sequences amplified and (3) test for cross-reactivity with host or pathogen DNA. In contrast, because the template copy numbers for mitochondrial DNA sequences cannot be estimated in silico, representative median estimates of mitochondrial genome copy numbers were used for the cytb and coxI assays [12].

Effects of different qPCR reagents and conditions on assay performance

Based on the studies described above, 13 candidate qPCR assays were selected for laboratory testing (Table 1) [13,14,15,16,17,18,19]. Primers and probes were synthesized by Integrated DNA Technologies (Coralville, IA) using additional modifications as noted (Table 2). These pilot studies and the subsequent studies were performed using a BioRad iQ5™ real-time instrument (Hercules, CA) with 96 well microtitre trays and analysed using the BioRad iQ5™ Optical System Software (v 2.1).

Table 1 Template copy number estimates for 13 candidate qPCR assays
Table 2 Design of 13 candidate qPCR assays in ascending order by template copy number

To limit the potential for confounding due to using qPCR reagents from only one source, nine assays (crt, crt76K, ldh(a), 18SrRNA(a), r364(a), coxI, cytb, varATS and r364(b)) were performed using qPCR reagents available from three different sources: (1) Invitrogen Platinum™ Taq with Green PCR buffer (Carlsbad, CA), (2) New England Biolabs (NEB) Luna® Universal Probe qPCR Master Mix (Ipswich, MA) and (3) Quantabio PerfeCTa qPCR ToughMix™ (Beverly, MA). In addition, four factors were examined for their effects on the results obtained with NEB reagents: (1) annealing temperature (Ta), (2) annealing time, (3) magnesium concentration and (4) final primer and probe concentrations.

Evaluation of qPCR assays based on parasite DNA standards

Based on the studies above, ten of the 13 candidate assays were examined after removing the crt76K, crt76T and r364(a) assays. The crt76K and crt76T assays were removed because they were designed to hybridize with polymorphic DNA sequences for amino acids 72–76 of the Plasmodium falciparum chloroquine resistance transporter gene (pfcrt) [13]. In contrast, the r364(a) assay was removed because of false-positive results. The other 10 qPCR assays were then compared for their sensitivity and specificity using the controls listed below:

  1. 1]

    Positive P. falciparum DNA Controls: eight tenfold dilutions from 10+3 to 10−4 picograms (pg) DNA per qPCR for 3D7 (drug-susceptible) and Dd2 (drug-resistant) parasite DNA standards provided by the Biodefense and Emerging Infections Research Resources Repository (Manassas, VA),

  2. 2]

    Negative Human DNA Controls: human DNA extracted from filter paper blots of Plasmodium-negative, human donor blood collected in the United States and.

  3. 3]

    Negative DNA-Free Controls: nuclease-free water to exclude the contamination of assay reagents or mixtures with parasite DNA.

The qPCR conditions used for the NEB reagents in these studies were considered optimal based on the results of pilot studies examining their effects on assay performance. Thermocycler protocols began with 5 min of denaturation at 95° C followed by 45 cycles of two steps: [1] 30 s for denaturation at 95o C and [2] 60 s for annealing and extension at 50o C (crt, crt76T, ldh(b), coxI, varATS and r364(b) assays) or 60° C (crt76K, ldh(a), 18SrRNA(a), 18SrRNA(b), r364(a), cytb and TARE-2 assays). This testing was performed using final reaction volumes of 15 µl with forward and reverse primers (400 nm final primer concentrations) and probes (200 nM final probe concentrations). A double-stranded DNA intercalating dye (Lumiprobe dsGreen—Hunt Valley, MD) was added to the TARE-2 assay as suggested by the manufacturer and melt curve analyses were performed as described previously [19]. Thresholds of detection were set for each assay at the point where amplification became exponential (the curve became linear) for the positive controls and Ct values ≤ 41 cycles were considered positive.

Longitudinal population-based study design, sample collection and smear surveys to identify asymptomatic persons with P. falciparum infection

Children and adults in Dangassa, Mali were enrolled in the population-based, longitudinal cohort studies of the West African International Centers of Excellence for Malarial Research (ICEMRs) at the beginning and end of the rainy season with their informed assent and their parents’ written informed consent after the study protocol had been reviewed and approved by the NIH (NIAID), the Mali and Tulane Institutional Review Boards (FWA000001769, FWA000002055) and the chief and elders of the village. Enrolled subjects then provided 2 drops of blood (~ 50 μl per drop) with a sterile disposable lancet (as recommended by the Mali Ministry of Health) after the tip of the finger had been cleaned with alcohol and dried. Please note because the results of those studies have been described previously [20], the methods described in this section are provided as clinical context for the human blood samples examined in these molecular studies.

The first drop of blood was applied to a freshly cleaned glass slide to produce a thick smear and the second to a GE Healthcare Life Sciences Whatman FTA filter paper (Pittsburgh, PA) for qPCR. After drying, the thick smear was rinsed in hypotonic phosphate buffer to lyse intact red blood cells before staining with Giemsa to permit the identification of asexual P. falciparum parasites by microscopy using oil immersion magnification. Based on 7500 white blood cells per μl as the average white cell count, the number of asexual P. falciparum parasites in oil-immersion fields containing 300 white blood cells was multiplied by 25 to estimate the number of asexual parasites per μl of blood [21]. After filter paper blots had been obtained and dried in a covered container, they were stored at ambient temperature in Mali, transported to the U.S., stored at − 20 °C and processed for DNA extraction and laboratory testing in New Orleans.

Afebrile individuals with thick smears positive for asexual P. falciparum parasites who had no symptoms or signs of malaria (i.e., no chills, fever, headache, myalgias or arthralgias) were classified as smear-positive asymptomatic persons with P. falciparum infection. In contrast, subjects with negative thick smears (no asexual P. falciparum parasites in microscopic fields containing 300 white blood cells) and no symptoms or signs of malaria were classified as healthy and uninfected. Members of the cohort who had symptoms or signs consistent with malaria and positive thick smears were examined and treated for their symptoms and signs on the same day without charge.

Evaluation of qPCR assays based on smear-positive asymptomatic subjects

Based on the filter paper blots for subjects 5–14 years of age collected during thick smear surveys in 2015, 96 filter paper blots from smear-positive asymptomatic subjects were selected randomly to estimate the clinical sensitivities of the qPCR assays. After those specimens had been sorted into groups based on their asexual parasite densities: (1) ≤ 200, (2) 201–999, (3) 1,000–1999 and (4) 2000–5000 per µl, Thermo Fisher Scientific ChargeSwitch® Forensic DNA Purification Kits (Waltham, MA) were used to extract DNA from 3 mm diameter punches of 50 µl blood spots on the filter papers. Six microliters of the 150 µl DNA eluate from each specimen were tested using qPCR. Thermocycler conditions were the same as those for the parasite DNA standards.

Testing parameters and statistical analyses

The effects of template copy number were examined after grouping the qPCR assays by their mean template copy numbers: [a] 1 copy (n = 3 assays: crt, ldh(a) and ldh(b)), [b] 3 copies (n = 2 assays: 18SrRNA(a) and 18SrRNA(b)), [c] 32 copies (n = 4 assays: cytb, coxI, r364(b) and varATS) and [d] 160 copies (n = 1 assay: TARE-2). Based on the parasite DNA standards and filter paper blots which yielded positive results (Ct values ≤ 41) for each of the 10 assays examined, 2-way ANOVA testing was performed in R Studio [22] to examine the effects of: [a] template copy number, [b] the initial template DNA concentration and [c] interactions between those factors and the final Ct values. Analytical sensitivities for parasite DNA were based on the Limit of Detection (LoD): the lowest DNA concentration detected consistently in three independent replicates. Clinical sensitivities for filter paper blots were based on the fraction of consistently positive results obtained for each sample using two to three independent replicates compared by χ2.

To compare Ct values for filter paper blots to those for parasite DNA standards, the units for the blood samples used to prepare the filter paper blots (asexual parasites per µl) were converted to pg DNA per qPCR based on 38.9 parasite genomes per pg parasite DNA, a 3 mm punch with 50 µl blood in each 10 mm diameter filter paper blot and the 6 µl of the 150 µl DNA eluate used for qPCR. Next, these results (pg DNA per qPCR) were used to solve for the y-intercepts (interpolated Ct values) of the regression lines for the P. falciparum 3D7 and Dd2 DNA standards. Finally, the interpolated Ct values were compared to the mean Ct values for filter paper blots using unpaired t-tests. Similar conversions were used to compare the sensitivity of qPCR for parasite DNA standards to its sensitivity for the DNA in smear-positive filter paper blots. Because blood was applied directly to filter paper blots and volumes (~ 50 µl) were estimated based on spot diameters (~ 1 cm), these comparisons aimed to understand how the qPCR results for parasite DNA standards relate to those for filter paper blots rather than provide accurate differences.

Results

In silico development and screening of candidate qPCR assays

Based on these studies, 13 candidate qPCR assays were selected for laboratory testing: nine developed previously and four developed during these studies (Tables 1 and 2). The template copy numbers for these assays ranged from 1 to 160, the AT content of DNA sequences amplified was < 80% and no cross-reactivity with host or pathogen DNA was detected using the NCBI Primer-Blast algorithm [11].

Effects of different qPCR reagents and conditions on assay performance

Based on six tenfold dilutions of 3D7 P. falciparum DNA (from 10+2 to 10−3 pg DNA per qPCR), Ct values for qPCR assays were generally similar with reagents from three different manufacturers (Additional file 1). However, false-negative results for DNA standards were obtained using Quantabio reagents for the crt assay and confirmed by independent replicates in which neither fluorescent signals nor new amplicons were detected using qPCR and agarose gel electrophoresis. Because the qPCR assays performed well with NEB reagents, which were supplied as a master mix, subsequent studies were performed using the NEB reagent master mix.

Although most of the qPCR conditions examined with NEB reagents did not affect assay performance, increasing the annealing temperature (Ta) from 50 to 60 ℃ did affect several TaqMan assays. At the higher Ta (60 ℃), Ct values increased for five assays (crt, ldh(b), coxI, varATS and r364(b)) and one assay (crt76T) yielded false-negative results. In contrast, the Ct value for one assay (cytb) decreased. For the six remaining assays, there were no consistent differences in mean Ct values at 50 ℃ vs. 60 ℃. The optimal Ta for each assay was used to perform these studies.

As mentioned above, the crt76T assay yielded false-negative results using a Ta of 60° C, which may have been caused by the five mismatched nucleotides in the crt76T probe (TGAAT) vs. the Dd2 DNA template (ATTGA). Because the crt76T and crt76K assays were designed to hybridize with the polymorphic DNA sequences of pfcrt, those assays were excluded from further study. In contrast, the r364(a) assay yielded false-positive results with negative controls (nuclease-free water, human DNA). Based on gel electrophoresis, amplicons from those false-positive assays were smaller than those from true-positive assays (~ 100 vs. ~ 125 bp, Additional file 2). Because the false-positive results obtained with negative controls did not permit evaluation of the sensitivity of the r364(a) assay, the r364(a) assay was excluded from further study.

Evaluation of qPCR assays based on Ct values for parasite DNA standards and filter paper blots from smear-positive asymptomatic individuals

Of the two sets of 8 serially diluted parasite DNA standards and the 96 smear-positive filter paper blots examined, positive results were obtained for the qPCR assays with the four highest 3D7 DNA concentrations (10+3 to 100 pg DNA per qPCR), the five highest Dd2 DNA concentrations (10+3 to 10−1 pg DNA per qPCR) and 51 of the 96 smear-positive filter paper blots (4 of 24 with ≤ 200 parasites per µl, 14 of 24 with 201–999 parasites per µl, 15 of 24 with 1000–1999 parasites per µl and 18 of 24 with 2000–5000 parasites per µl). Based on 2-way ANOVA testing, Ct values for the 3D7 and Dd2 parasite DNA standards (Figs. 1 and 2) and filter paper blots (Fig. 3) decreased as parasite DNA (i.e., amounts of DNA and parasite densities) and template copy numbers increased (p < 0.001).

Fig. 1
figure 1

Effects of the amounts of 3D7 DNA and template copy number on Ct values. 3D7 P. falciparum DNA standards were tested by 2-way ANOVA for the effect of template copy number on the Ct value, which decreased as the amounts of parasite DNA and template copy numbers increased (mean Ct values and standard deviations are provided for 3D7 P. falciparum DNA from 10° to 10+3 pg per qPCR, p < 0.001)

Fig. 2
figure 2

Effects of the amounts of Dd2 DNA and template copy number on Ct values. Dd2 P. falciparum DNA standards were tested by 2-way ANOVA for the effect of template copy number on the Ct value, which decreased as the amounts of parasite DNA and template copy numbers increased (mean Ct values and standard deviations are provided for Dd2 P. falciparum DNA from 10−1 to 10+3 pg per qPCR, p < 0.001)

Fig. 3
figure 3

Effects of asexual parasite density and template copy number on Ct values. Ct values for smear-positive samples decreased as asexual parasite densities and template copy numbers increased (p < 0.001). Mean Ct values and standard deviations are grouped by their mean parasite densities and ranges of: 100 (≤ 200), 600 (201–999), 1500 (1000–1999) and 3500 (2000–5000) per μl. Note that the regression lines for the parasite DNA standards (Figs. 1, 2) were steeper than those for the filter paper blots (Fig. 3), consistent with DNA concentrations from 1000-fold (10+3) to 10,000- fold (10+4) for the DNA standards vs. 200-fold (10+2.3) for the filter paper blots

Note that these results do not suggest confounding—because p values for the effects of parasite DNA concentrations and template copy numbers on the Ct values varied from 0.05 to 0.5 and the slopes of their regression lines were fundamentally similar (Additional file 3). In addition, the slopes of the regression lines for the parasite DNA standards (Figs. 1 and 2) were steeper than those for the filter paper blots (Fig. 3), consistent with DNA concentration ranges from 1000-fold (10+3) to 10,000- fold (10+4) for the DNA standards vs. 200-fold (10+2.3) for the filter paper blots.

Evaluation of qPCR assays based on sensitivities for parasite DNA standards and filter paper blots of smear-positive asymptomatic subjects

Based on the drug-susceptible (3D7) and drug-resistant (Dd2) parasite DNA standards, the limits of detection (LoDs) for P. falciparum DNA decreased (sensitivities increased) as the template copy numbers increased (Fig. 4). Note that the TARE-2 assay yielded consistently positive results for the lowest concentration of Dd2 DNA tested initially (10−4 pg DNA per qPCR) but did not yield positive results for an additional dilution of Dd2 DNA (10−5 pg DNA per qPCR).

Fig. 4
figure 4

Limits of detection for 3D7 and Dd2 P. falciparum DNA by template copy number. The LoDs of qPCR assays for drug-susceptible (3D7) and –resistant (Dd2) P. falciparum DNA decreased (analytical sensitivities increased) as the template copy numbers of the qPCR assays increased (p < 0.001). Note that the TARE-2 assay (n = 160) yielded consistently positive results for the lowest concentration of Dd2 DNA tested initially (10−4 pg DNA per qPCR) but did not yield consistently positive results for an additional dilution of Dd2 DNA (10−5 pg DNA per qPCR)

Based on filter paper blots stratified by parasite density, the sensitivities of qPCR assays for P. falciparum increased as parasite densities and template copy numbers increased (Fig. 5 and Additional file 4). Despite the gains in sensitivity due to increased template copy number, qPCR assays failed to detect more than half of the smear-positive filter paper blots at low parasite densities (≤ 200 asexual parasites per µl).

Fig. 5
figure 5

Detection of smear-positive samples by parasite count and template copy number. The clinical sensitivities of the qPCR assays for P. falciparum DNA reported here increased with parasite density and the template copy number in smear-positive, asymptomatic human subjects (n = 24 subjects per group, p < 0.001)

Based on χ2 testing, qPCR assays with template copy numbers from 32 (n = 4) to 160 (n = 1) were as expected more sensitive than assays with copy numbers of 1 (n = 3) (273/384 and 74/96 vs. 165/288, p < 0.001). Likewise, the qPCR assay with a template copy number of 160 (TARE-2) was more sensitive than the assays with template copy numbers of 3 (74/96 vs. 124/192, p < 0.05). In contrast, differences between the sensitivities of other qPCR assays (with template copy numbers of 1 vs. 3, 3 vs. 32 and 32 vs. 160) were not significant (p ≥ 0.1). These differences in significance were consistent with the differences between the template copy numbers of the assays: i.e., assays with greater differences between their template copy numbers (e.g., 1 vs. 160) were more likely to have differences between their outcomes than assays with smaller differences in template copy numbers (e.g., 1 vs. 3).

Comparison of qPCR outcomes for DNA standards vs. filter paper blots

Based on conversions from parasites per μl to pg DNA per qPCR, the Ct values for filter paper blots were higher than those interpolated from the 3D7 and Dd2 DNA regression lines (mean Ct = 33.47 ± 3.17 vs. 28.73 ± 3.22 and 27.86 ± 3.09, respectively; p < 0.001). In addition, comparing the mean LoDs (analytical sensitivities) for 3D7 and Dd2 DNA standards using converted units (32 and 6 parasites per μl) to the clinical sensitivities for filter paper blots (fewer than half detected at ≤ 200 parasites per μl blood) suggest qPCR was more sensitive for the parasite DNA standards than the filter paper blots. Likewise, comparing the clinical sensitivities for filter paper blots using converted units (fewer than half detected at ≤ 0.9 pg DNA per qPCR) to the analytical sensitivities for 3D7 and Dd2 DNA standards (mean LoDs of 0.15 ± 0.30 and 0.03 ± 0.04 pg DNA per qPCR) suggests qPCR was more sensitive for the parasite DNA standards than the filter paper blots.

Discussion

In silico development and screening of qPCR candidate assays

The range of template copy numbers reported here for P. falciparum-specific qPCR assays is limited: from 1 to 160. In addition, because more assays were available for low template copy numbers (≤ 3 copies) than for higher template copy numbers, not all of the assays available for lower template copy numbers were examined in these studies and only one assay was available for ≥ 100 copies (TARE-2 for 160 copies). However, despite those limitations, the template copy numbers in the literature and those reported here permit an examination of the effects of template copy number on the sensitivity of qPCR for P. falciparum DNA based on DNA standards and smear-positive filter paper blots from asymptomatic subjects.

Effects of different qPCR reagents and conditions on assay performance

Because pilot studies showed consistent qPCR assay performance with reagents from most manufacturers, the Ct values and sensitivities reported here are unlikely to have been confounded by the use of one set of reagents for qPCR testing (i.e., NEB reagents). In addition, because the annealing temperature was the most important factor for assay performance and was optimized for each assay, the results reported here are unlikely to have been confounded by the conditions used to perform the qPCR assays.

Based on the pilot studies, three assays were excluded from further examination. Two (crt76K and crt76T) were excluded because they were designed to hybridize with highly polymorphic DNA sequences (amino acids 72–76 in pfcrt) [13]. In addition, the r364(a) assay was excluded because they produced false-positive results with negative controls (water, human DNA). Those results were unexpected because false-positive results had not been reported previously with the r364(a) assay [17, 23, 24]. Although agarose gel electrophoresis (Additional file 2) suggested that false-positive results with the r364(a) assay could have resulted from hybridization of the probe to unintended smaller qPCR products, DNA contamination below the level of detection by agarose gel electrophoresis could not be excluded.

Evaluation of qPCR assays based on Ct values for parasite DNA standards and filter paper blots from smear-positive asymptomatic subjects

Two-way ANOVA testing of Ct values for parasite DNA standards and smear-positive filter paper blots revealed that Ct values decreased as the amounts of parasite DNA and template copy numbers increased. In addition, because the slopes of the regression lines for the DNA standards were steeper than those for the filter paper blots (consistent with their greater range of DNA concentrations), those data are internally consistent.

Evaluation of qPCR assays based on sensitivities for parasite DNA standards and filter paper blots from smear-positive asymptomatic subjects

The LoD results (Fig. 4) show that the sensitivities of qPCR for 3D7 (drug-susceptible) and Dd2 (-resistant) P. falciparum DNA increased with template copy number. However, there was an exception to this pattern for template copy numbers from 32 to 160. Within that range, the LoD decreased (sensitivity increased) for Dd2 DNA, but not for 3D7 DNA. Although this result raises the possibility of strain-specific differences in TARE-2 between 3D7 and Dd2 parasites, that hypothesis cannot be confirmed (or rejected) until comparable sequence data become available for the 3D7 and Dd2 parasite genomes.

The LoDs reported here for 3D7 DNA agree with those reported previously (ldh(a) [25], 18SrRNA(a) [15, 26], 18SrRNA(b) [16, 26], varATS and TARE-2 [19]). However, because the LoDs for drug-resistant Dd2 DNA were lower than those for drug-susceptible 3D7 DNA, additional DNA was obtained to exclude technical variation in preparation of the DNA standards as a cause of those differences. The Ct values for the new Dd2 DNA standards were similar to those for the initial Dd2 DNA standards. Please note that the small number of replicates used to estimate the LoDs (n = 3) likely contributed to the differences observed [27]. In addition, because these results were obtained using DNA from only two parasite strains and the genetic diversity of P. falciparum varies across the globe, additional studies will be necessary to establish the generalizability (validity) of these observations.

The studies of smear-positive filter paper blots indicate that the clinical sensitivity of qPCR for P. falciparum in asymptomatic subjects increased with both parasite density and template copy number, although it was inadequate at lower parasite densities (≤ 200 per μl). However, it is difficult to compare these results to other studies of asymptomatic subjects because data for infected, asymptomatic individuals are rarely stratified by asexual parasite density. Because these studies were based on smear-positive asymptomatic children in Dangassa, additional studies will be necessary to determine whether they are generalizable to other malaria-endemic areas. Although the increased clinical sensitivity of qPCR assays with higher template copy numbers was most apparent with low-density infections (≤ 200 asexual parasites per µl), the most sensitive qPCR assays failed to identify more than half of the smear-positive filter paper blots at asexual parasite densities ≤ 200 per μl and may likewise fail to identify epidemiologically important reservoirs of infection in malaria-endemic areas.

Comparison of qPCR outcomes for parasite DNA Standards and filter paper blots based on unit conversions

These results are consistent with other reports suggesting that parasite DNA from filter paper blots may have higher Ct values and lower PCR sensitivity than DNA extracted from whole blood—likely due in part to the smaller volumes of blood processed [28,29,30]. Furthermore, qPCR has been shown to have greater sensitivity for P. falciparum DNA in asymptomatic individuals than microscopy and RDTs based on the isolation of red blood cells (and Plasmodium) from larger volumes of whole blood (≥ 500 µl) to concentrate the DNA used for testing [31]. Because matching whole blood samples were not available for the filter paper blots examined in these studies, those potential limitations on the qPCR clinical sensitivities reported cannot be excluded.

Other potential limitations on the clinical sensitivity of qPCR for filter paper blots from smear-positive asymptomatic subjects

As noted above, it is difficult to compare the clinical sensitivities reported in these studies to those in other studies because qPCR results for P. falciparum are rarely stratified by parasite densities based on microscopy. For that reason and because the clinical sensitivities reported here were based on the microscopy results and molecular methods used to examine filter paper blots, potential limitations related to those factors require further consideration: (1) false-positive blood smears, (2) DNA extraction method and (3) additional molecular approaches for detecting P. falciparum using filter paper blots.

  1. 1)

    False-positive blood smears: Because the limited clinical sensitivities reported could potentially have been explained by higher frequencies of false-positive smears based on the microscopy results for the Dangassa study site, the 18SrRNA(a) assay [15] was used to cross-examine filter paper blots from smear-positive symptomatic children (n = 62) at a different study site (Dioro, Mali) where microscopy slides were read by a different team and infection was confirmed using rapid diagnostic tests for P. falciparum histidine-rich protein 2. Because the 18SrRNA(a) assay results were similar for the filter paper blots from asymptomatic children at Dangassa and those from symptomatic children at Dioro with infections confirmed based on parasite antigen (25/48 vs. 17/24 with ≤ 999 parasites per µl and 38/48 vs. 34/38 with ≥ 1000 parasites per µl, p ≥ 0.1), the results of those studies suggest the clinical sensitivities reported are reproducible and unlikely limited to the microscopy results for the Dangassa study site.

    To test for false-positive smears caused by non-falciparum Plasmodium, filter paper blots negative for P. falciparum by qPCR were examined using a qPCR for Plasmodium malariae [32] (the most common non-falciparum plasmodial infection reported in this region of Mali [33, 34]). Because those results were negative, false-positive smears for P. falciparum from non-falciparum Plasmodium such as P. malariae are unlikely to explain the limited clinical sensitivities of the qPCR assays for P. falciparum reported here.

  2. 2)

    DNA extraction method: Because a number of methods have been used to extract DNA from filter paper blots, pilot studies were performed to examine the sensitivity of the ChargeSwitch® method. The Chelex®-based method was selected for this comparison because it is thought to have greater sensitivity at low parasite densities [35,36,37,38]. However, based on smear-positive filter paper blots from symptomatic subjects (n = 66) in Dioro, Mali, the frequencies of positive PCR results were similar with Chelex® and ChargeSwitch®: (1) 26/66 vs. 22/66, p = 0.5 for 18S rRNA genes [15] and (2) 24/66 vs. 25/66, p = 0.8 for the repetitive element 364 (r364) DNA sequences [10]. In addition, the potential for negative results based on failed ChargeSwitch® extractions was excluded by PCR amplifying human IFNy DNA for those extractions. Although these results suggest the sensitivity of the ChargeSwitch® is similar to that of other methods, alternative methods that process larger volumes of blood from filter paper blots may warrant further study [39].

  3. 3)

    Additional molecular approaches for detecting P. falciparum using filter paper blots: Because pre-amplification of parasite DNA may improve the sensitivity of qPCR for P. falciparum in asymptomatic individuals [40], 12 filter paper blots were examined with and without pre-amplification using four qPCR assays with varying template copy numbers: ldh(b), 18SrRNA(a), coxI and r364(b). Compared to specimens negative in every assay examined, results for these 12 specimens were expected to improve with pre-amplification because they were positive in the assay with the highest template copy number (TARE-2, n = 160), although they were negative in qPCR assays based on single copy DNA templates (crt, ldh(a), ldh(b)). Based on the Ct values for specimens positive without and with pre-amplification (n = 14 per group), pre-amplification increased the amount of DNA template (mean Ct values fell from 36.8 ± 1.2 to 5.2 ± 1.5, p < 0.001). However, the fraction of positive samples was similar without and with pre-amplification (16/48 vs. 18/48, p = 0.83). These results indicate that pre-amplification of filter paper blots from field studies can increase the amount of parasite DNA template, but may not improve the fraction of the qPCR assays (samples or subjects) positive for P. falciparum.

    Because the generalizability of these findings is limited by the strategies used to amplify parasite DNA from filter paper blots, future studies should include alternative strategies such as DNA sequences from novel search algorithms [41], RNA sequences with higher template copy numbers [42, 43] and PCR-based methods linked to immunoassays [44].

Conclusions

Based on parasite DNA standards and smear-positive filter paper blots from subjects with asymptomatic P. falciparum infection, these studies indicate the sensitivity of qPCR assays increases with their template copy number. However, despite the increased sensitivity of qPCR assays with higher template copy numbers, qPCR assays with still greater sensitivity will be necessary to identify asymptomatic P. falciparum-infected individuals with low parasite densities (≤ 200 per µl) in order to interrupt transmission.