Background

Candida auris is an emerging, often multi-resistant, yeast that causes invasive infections in healthcare settings [1, 2]. C. auris has the ability to cause large healthcare outbreaks [3, 4]. Patients may be colonized for months [3, 5] and C. auris has been shown to remain viable on surfaces for at least 14 days [6, 7]. It is widely considered that the environment may be a reservoir for transmission of C. auris. The Centre for Disease Control and prevention (CDC) recommends post discharge terminal cleaning and disinfection of patients’ rooms and cleaning and disinfection of areas outside the rooms where patients received care [8]. It has been demonstrated that conventional cleaning and disinfection often lacks consistency and that additional disinfection, using non-touch technologies such as hydrogen peroxide vaporisation or germicidal ultraviolet (UV-C) light can further reduce the surface bioburden and transmission of microorganisms [9]. However, a recent publication by Cadnum and colleagues showed a relative resistance of C. auris and other candida species to UV-C and that extended exposure time (20 to 30 min) might be needed [10]. In their experiment, Cadnum and colleagues (2018) used a mobile device that emits 254-nm continuous UV-C light. We tested the hypothesis that the pulsed-xenon ultraviolet (PX-UV) system that emits broad spectrum UV-C (200–280-nm) in short pulses could reduce the exposure time needed to decrease C. auris. As a comparison, we also tested the PX-UV system on C. parapsilosis, another candida known to be responsible for outbreaks in healthcare settings and survive for at least 28 days in the environment [7].

Methods

Prior cultures and growth of clinical strains of C. parapsilosis and C. auris was carried out in a broth liquid culture medium at 37 °C until concentration ranges 105–106 colony-forming units (CFUs) per millilitre were obtained. Glass slides were inoculated with 10 μl of C. auris stock culture and allowed to dry.

Slides were positioned perpendicular to the floor at a distance of 1.25 m from the floor. Exposure time were run uninterrupted for 5-, 10- and 15-min cycles. The experiment was conducted at 1-m distance from the centre of the robot and repeated at a 2-m distance. Tests were conducted in triplicate. Three controls treated in identical conditions were prepared with each organisms. The controls remained covered during the experiment. After exposure, the control and PX-UV exposed slides were incubated in 10 ml phosphate-buffered saline (PBS) and agitated for 30 s to extract CFUs. Serial dilutions were plated onto Chromagar Candida and incubated at 36*for 48 h. The total viable count was quantified into number of colony forming unit. The percent reduction in the pathogens were calculated in comparison with unexposed control slides. The experiment was repeated for C. parapsilosis.

Results

The laboratory testing results for C. auris are detailed in Table 1, stratified according to cycle time (5, 10 and 15 min) and distance from the centre of the UV light bulb (1 and 2 m). At 1-m distance from the PX-UV light bulb, there was a reduction of 99.4% of C. auris CFU observed after a 5-min cycle, while no growth was observed at 10 and 15 min. At 2-m distance from the light bulb, the C. auris CFU reduction was 90.2% after a five minutes cycle, and 99.6% after a 10-min cycle. There was no growth at 15 min.

Table 1 Laboratory testing results for C. auris and C. parapsilosis

At 1-m distance, the average reduction in C. parapsilosis CFU was 98.5% after a 5-min cycle. There was no growth at 10 and 15 min. At 2-m distance, the average efficacy was only 15.7% after a 5-min cycle while it was 95.2% after a 10-min cycle. There was no growth after a 15-min cycle.

Discussion

We found that at 1-m distance from the PX-UV light bulb, there was a 99.4% reduction in C. auris after a 5-min cycle and no growth after 10- and 15-min cycles. Similarly, for C. parapsilosis there was a 98.5% reduction after a 5-min cycle and no growth after 10- and 15- min cycles. We found that at 2-m distance, there was a 15.7% reduction in C. auris after a 5-min cycle, a 95.2% reduction after a 10-min cycle, and no growth after a 15-min cycle. For C. parapsilosis, there was a 90.2% reduction after a 5-min cycle, a 99.6% reduction after a 10-min cycle, and no growth after a 15-min cycle.

The limitations of this study include a relative small study numbers and only one strain of C. auris. The strengths of this study include the inclusion of two common Candida species with varying distances from the PX-UV bulb. Previous studies have not tested the efficacy of PX-UV for Candida spp. Further studies are needed in hospital environment, to assess the cumulative impact of repeated sessions.

Candida auris and Candida parapsilosis are two pathogens commonly involved in outbreaks in healthcare settings outbreaks. They are often recovered from the hospital environment, where they can survive for a long time. Additional disinfection with hydrogen peroxide vapour is effective on C. auris and C. parapsilosis. However, this method is labour intensive as it requires sealing of the bed space, or doors and vents in case of a single-bed room, and monitoring of the concentration of H2O2 inside the enclosure before permitting patients or staff to enter into the room. The cycle time can last 2 to 8 h according to the H2O2 vapour system used. By comparison, the PX-UV technology, is easy to use and has significantly shorter cycle times that makes it easier to disinfect all areas outside the room where the patient received care, as recommended by the CDC.

Conclusions

We found a 99.6% reduction after a 10-min cycle at 2-m distance, and a 95.2% reduction after a 10-min cycle at 2-m distance. The PX-UV mobile device is easy to use and has short cycle times compared to other no-touch room disinfection technologies. Further studies are needed in hospital environment, to assess the cumulative impact of repeated sessions.