Introduction

The underlying goal of treating CNS malignancies is to maximize tumor eradication while preserving parenchymal brain function. Tilting the therapeutic index towards eradicating tumor cells while protecting normal tissue may be improved by reducing setup uncertainty. Image guided radiation therapy (IGRT) has the potential to improve accuracy through patient localization.

The term clinical target volume (CTV) is defined in the International Commission on Radiation Units and Measurements (ICRU) Reports as “a tissue volume that contains a gross tumor volume (GTV) which is the gross palpable or visible/demonstrable extent + and location of the malignant growth, and/or subclinical microscopic malignant disease, which has to be eliminated. This volume has to be treated adequately in order to reach the aim of therapy: cure or palliation” [1]. Planning target volume (PTV) is defined as “a geometrical concept, and it is defined to select appropriate beam sizes and beam arrangements, taking into consideration the net effect of all the possible geometrical variations and inaccuracies in order to ensure that the prescribed dose is actually absorbed in the CTV” [1]. The PTV is composed of two factors: (1) the internal margin (IM) which relies on temporal changes in position, volume, and shape of the CTV and the (2) setup margin (SM) which accounts for uncertainties in patient position and beam delivery that is inherent with fractionated irradiation [24]. If the margin between the CTV and PTV is too large, there is a higher likelihood of excessive radiation to normal tissue [5]. Inversely, if the margin is too small, an undesirable outcome may occur due to inadequate radiation of the target tissue. The present study highlights patient features and technical interventions which may play a role in influencing the setup margin.

We tested the following hypotheses: 1) that the use of days 1, 2, and 3 cone-beam CT scan or orthogonal kV imaging predicted patient position during the treatment course, 2) non technology related patient factors, specifically a) prescribed medications b) weight gain, c) and tumor location predicted setup uncertainty, and 3) application of custom CTV to PTV expansion from the first three fractions reduces dose to organs at risk (OAR).

Methods

Under an IRB-approved protocol and in compliance with the Helsinki Declaration, we reviewed the treatment records of 29 patients with brain tumors immobilized with an aquaplast mask and standard base plate or an S-frame and aquaplast mask for simulation. The treatment planning CT with 1 mm neutral gantry axial slices was fused with a gadolinium-enhanced MRI with 3 mm zero tilt axial, coronal, and sagittal slices for target delineation. The CTV was uniformly expanded by 3 mm to create the PTV for all cases to minimize fusion errors. This approach is similar to a phantom study shown to have an accuracy of autofusion less than 0.5 mm [6]. Anatomic verification of deep brain electrode placement based on CT/MRI fusion has been shown to have accuracy of a similar magnitude [7]. OAR were contoured (temporal lobes, brainstem, bilateral hippocampus, cochlea, hypothalamus, and pituitary gland) and were used for inverse-planned static gantry IMRT. The equipment utilized was a Varian linear accelerator with orthogonal kilovoltage imagery.

All patients were treated with daily fractionated radiation with the dose dependent on tumor histology: 21 patients with high grade glioma received 60 Gy in 30 fractions; 4 patients with low grade glioma, 2 patients with meningioma, and 2 patients with pituitary adenoma received 50.4 Gy in 28 fractions. The number of beams and angles was chosen based on tumor location and proximity of OAR.

A total of 517 treatments were delivered with online corrections made and recorded in the vertical, lateral, and longitudinal axis. The three dimensional vector of uncertainty was calculated as the square root of the sum of the squares of each axis:

3 D shift = Vert 2 + Lat 2 + Long 2

Patients found to have setup uncertainty less than 3 mm were planned with the measured setup uncertainty, with no change in optimization parameters. The setup uncertainty was dependent on the localization. A bony match as opposed to a soft tissue match was utilized since the base of the skull served as a surrogate for intracranial targets. We determined the effect of patient specific factors (prescribed medications, weight gain, and tumor location) and technical interventions on setup uncertainty, and plotted the mean +/- standard deviation for each factor. Student’s t-test was used to compare between groups.

Results

Image Guidance Modality and Setup Uncertainty

We first determined the influence of Cone Beam Computed Tomography (CBCT) versus On Board Imaging (OBI) imaging on patient set up variability. For each of the 29 patients, shift data was recorded and analyzed for both techniques. There was no difference between techniques for shifts in the vertical (2+/-4 mm CBCT vs 2+/-2 mm OBI, p = 0.45), longitudinal (1.3+/-1.4 mm CBCT vs 1.4+/- 1.4 mm OBI, p = 0.07), or three dimensional (3.5+/-4.2 mm CBCT vs 3.6+/- 1.9 mm OBI, p = 0.52) vector averages. There was a statistical difference in the lateral vector (1.4+/-1.6 mm CBCT vs 1.7+/- 1.5 mm OBI, p < 0.001) of small magnitude. We found no clinically meaningful difference between the two imaging modalities for evaluating translational setup uncertainty.

Setup Uncertainty First Three Treatment Days vs. Subsequent Treatments via CBCT Image Guidance

We next determined whether the first three days of IGRT would predict patient setup for subsequent fractions. Patients underwent CBCT for the first three fractions and remarked before the fourth fraction. As shown in Figure 1, there was no statistical difference (p = NS) in setup uncertainty between the first three days of treatment and all subsequent treatments via CBCT image guidance. We also analyzed post treatment images to quantify intrafraction motion. For 100 fractions, the mean difference from pre-treatment to post-treatment images was 0.7 mm.

Figure 1
figure 1

Initial three IGRT fractions predict future patient position with CBCT. The vertical, lateral, longitudinal, and total three dimensional setup uncertainty averages were plotted with +/- standard deviations for the first three days of treatment in black and all subsequent treatments in white. Student’s t-test was used to determine significance between groups.

Immobilization Device and Setup Uncertainty

The use of an S-frame table extender allows the use of additional beam angles to improve isodose conformality and OAR sparing. However, there may be setup error introduced due to flexion or insertion variability. We, therefore, compared the vertical, lateral, longitudinal, and three dimensional setup uncertainty for patients immobilized via an aquaplast mask and base plate (N = 6) vs. an aquaplast mask and S-Frame (N = 23) in Figure 2. S-frame immobilization demonstrated a clinically significant increase in setup uncertainty in the vertical axis (2.5 +/- 3.7 mm S Frame vs. 1.3 +/- 0.9 mm for base plate, p < 0.001) which contributed to the observed significant difference in 3D setup variability.

Figure 2
figure 2

S-frame extension contributes setup uncertainty compared with base plate immobilization. Setup uncertainties with +/- standard deviations for patients immobilized via aquaplast mask and base plate are demonstrated in black and those immobilized via aquaplast mask and S-Frame in white. * Indicates p < 0.05 by Student’s t-test.

Medications and Setup Uncertainty

The time from radiation CT simulation and planning to the completion of radiation therapy can be up to eight weeks for a 30 fraction course, and patients may change during this time frame which may impact setup in a thermoplastic mask. We measured the impact of corticosteroid administration (14 yes vs. 15 no), sedative administration (11 yes vs. 18 no), anti-seizure administration (14 yes vs. 15 no), and weight change (23 stable vs. 6 with at least a 10 lb gain) on setup uncertainty. Patient medication history data were reviewed, and the status for three prescribed medications was recorded. There was no statistical difference in the proportion of patients prescribed medications and those not prescribed medications (p = NS). Neither steroid administration nor weight change was associated with differences in setup variability. However, as shown in Figure 3, patients administered sedatives had increased setup uncertainty (3D vector of 3.3+/-1.6 mm for non-sedated vs. 4.1+/-3.8 mm for sedated patients). Those receiving anti-seizure medications had statistically decreased setup uncertainty in the vertical, lateral, and longitudinal axes of minimal clinical magnitude of less than 0.5 mm in each vector.

Figure 3
figure 3

Influence of sedatives on setup uncertainty. Setup uncertainty data in the vertical, lateral, longitudinal, and total three dimensional setup uncertainty were plotted with +/- standard deviation. Setup uncertainty was statistically significant in patients who were prescribed sedatives, shown in white (* p < 0.05).

Tumor Location and Setup Uncertainty

We hypothesized that central tumors would have less setup uncertainty since rotational errors would not influence location as opposed to lateral tumors. We classified targets (n = 29) according to location: Pituitary (n = 3), frontal lobe (n = 8), temporal lobe (n = 7), occipital/posterior fossa (n = 9), and other (n = 4). There was no statistical difference in the proportion of tumor location (p = NS). Figure 4 shows the 3D vector uncertainty average for the five groups with pituitary targets having the least variability (2.8+/-2.0 mm) and temporal lobe tumors demonstrating the highest (4.4+/-4.4 mm, p <0.05 across all groups by ANOVA). In this analysis, the distance from the center of the skull correlated with setup uncertainty.

Figure 4
figure 4

Tumor location influences setup uncertainty. The 3D vector +/- the standard deviation for each of the tumor locations are shown. P < 0.05 by ANOVA across all groups.

Effect of Custom PTV on Dose to OAR

Minimizing CTV to PTV expansion can lead to decreased dose to normal tissues in close proximity to radiation targets. Three of the 29 patients had each fraction setup variation less than 3 mm, all three of them with central targets. We applied the measured setup uncertainty and applied their individual variation to generate a new smaller PTV for each case. Their IMRT plans were reoptimized, and as shown in Figure 5, application of patient-specific CTV expansions results shift of DVH curves to the left (arrows), decreasing dose to the brainstem, hippocampus, and temporal lobe. The decrease in mean hippocampus dose would in this case be predicted to preserve neurocognitive function based on dose-volume relationships reported previously [8].

Figure 5
figure 5

Planning with customized PTV allows decreased dose to critical normal structures. A. A patient with a centrally-located malignant glioma underwent replanning with a CTV to PTV expansion of 1.5 mm as measured by IGRT in the first three days of treatment. This decreased dose to the brainstem (blue, max from 5935 to 5550 cGy and mean from 4153 to 3596 cGy), hippocampus (orange, max from 6052 to 5761 cGy and mean from 3892 to 3006 cGy), and temporal lobe (yellow, max from 6550 to 6311 cGy and mean from 2623 to 2205 cGy). B. Axial representative isodose curves from original and reduced PTV, respectively.

Discussion

The treatment of intracranial malignancies has been greatly enhanced by the use of radiation [8, 9]. The ultimate objective is to administer a dose localized at the target volume coupled with the desired effect of sparing normal tissue, thus, minimizing the likelihood of long-term neurocognitive deficits. Reducing the clinical target volume has been shown to be adequate for tumor control with the intention of diminishing the potential cognitive side effects inherent with radiation therapy [10]. The use of three-dimensional conformal radiation therapy has been effective in increasing tumor control and in reducing acute side effects [11].

The determination of the setup uncertainty prior to the initiation of radiation plays an important role in the treatment of CNS malignancies. Daily setup variation may be underrecognized and may have an adverse impact, including target underdose and normal structures receiving a higher dose than anticipated [12, 13]. Daily localization based on cone beam CT imaging may reduce the required setup margin lessening normal tissue exposure to radiation [14]. Patient motion during treatment may affect the dose to critical structures and, therefore, establishing risk volumes are recommended to accurately ascertain the dose administered to normal tissues [3].

Closely monitoring the target volume is vital during radiation therapy to ensure that the doses to the target and normal tissues are not altered which may have deleterious repercussions [15]. Beltran et al. have demonstrated that the target volume during radiation therapy in craniopharyngioma may increase and decrease (-20.7% to 82%), stressing that surveillance imaging is necessary to determine the appropriate dose to the target volume while at the same time closely observing the amount to the normal tissue [15]. Beltran et al. have also recommended daily localization which decreases the PTV margin and minimizes insufficient tumor coverage due to setup uncertainties [2]. In addition, a decrease in the CTV margin lessens the dose directed to normal cerebral tissues [2].

Several studies have addressed daily setup variability in the treatment of head and neck cancer patients using conventional mask immobilization. Hong et al. showed a 3.33 mm absolute average daily setup error in any single dimension while other studies have reported a range of 3–5 mm in head and neck setup variation using weekly portal film measurements [12, 16, 17]. The patients in the current study were either immobilized with an aquaplast mask and either a base plate or S-frame. In our study, immobilization with an S-frame significantly increased setup uncertainty (3D p < 0.001) primarily in the vertical direction, consistent with flexion of the insert as it protrudes beyond the head of the table. Beltran et al. reported that the setup margin was smaller for those patients who were treated under anesthesia [4]. In the present study, setup uncertainty was statistically significant in patients who received sedatives (3D p < 0.001). We hypothesize that patients at risk for claustrophobia requiring oral sedation in fact would have demonstrated larger variation without sedation, and in our study the sedation was unable to completely ameliorate their anxiety. Patients prescribed either steroids or anti-seizure medications demonstrated no statistical difference in setup uncertainty.

We have shown that there was no statistical difference in setup uncertainty between CBCT and OBI imaging modalities. In addition, setup uncertainty was not significant for the first three days of treatment versus all subsequent treatment days via CBCT guidance. The first three days of CBCT predicted setup uncertainty for subsequent treatments and permitted customized CTV to PTV expansion. Patient weight gain of greater than ten pounds was not associated with increased setup uncertainty. The data demonstrated that patients with a tumor located in the temporal lobe experienced greater setup uncertainty whereas those with pituitary tumors showed the least setup uncertainty.

Conclusion

Setup uncertainty in the course of radiation treatment may be influenced by several factors, including the immobilization device utilized in therapy, specific medications, and certain tumor locale. To minimize the setup uncertainty, various institutions have instituted a policy of daily and continuous evaluation of interfraction and intrafraction motion for all head and neck cancer patients undergoing intensity-modulated radiation therapy [12, 18].

Image-guidance potentially maximizes the therapeutic index of brain irradiation by minimizing setup uncertainty. Custom CTV to PTV expansion results in a reduced dose administered to the OAR which diminishes the possibility of developing neurocognitive, learning, and neuroendocrine deficits.