Purpose of Review
To review the clinical use, advantages, and potential limitations of intraoperative aberrometry (IA) and its application in cataract surgery in the normal, astigmatic, and atypical eye.
Recent Findings
The use of IA is similar to traditional preoperative biometry methods for intraocular lens (IOL) determination in short and normal eyes. The use of IA resulted in equivalent outcomes to modern formulas with optimized axial lengths in long eyes and exceeded traditional formulas. Improved outcomes compared to traditional methods were noted with the use of IA when used for both astigmatism correction and IOL determination in patients with a history of corneal refractive surgery.
Summary
The greatest value in using IA is with atypical eyes. There is a surgeon learning curve with this technology to allow for optimal, consistent, and reproducible results. Despite the increasing popularity of IA, the number of published studies on this technology is limited.
Similar content being viewed by others
References
Papers of particular interest, published recently, have been highlighted as: • Of importance •• Of major importance
Hatch KH. Intraoperative aberrometry. In: Narang P, Trattler WB, editors. Optimizing suboptimal results following cataract surgery. New York: Thieme; 2019.
Donaldson K, Fernández-Vega-Cueto L, Davidson R, Dhaliwal D, Hamilton R, Jackson M, et al. Perioperative assessment for refractive cataract surgery. J Cataract Refract Surg. 2018;44(5):642–53. https://doi.org/10.1016/j.jcrs.2018.02.022.
Mackool RJ, Ko W, Mackool R. Intraocular lens power calculation after laser in situ keratomileusis: aphakic refraction technique. J Cataract Refract Surg. 2006;32(3):435–7. https://doi.org/10.1016/j.jcrs.2005.11.045.
Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS. A three-part system for refining lens power calculations. J Cataract Refract Surg. 1988;14(1):17–24.
Hoffer KJ. Clinical results using the Holladay 2 intraocular lens power formula. J Cataract Refract Surg. 2000;26(8):1233–7.
Gavin EA, Hammond CJ. Intraocular lens power calculation in short eyes. Eye (Lond). 2008;22(7):935–8.
Gokce SE, Zeiter JH, Weikert MP, Koch DD, Hill W, Wang L. Intraocular lens power calculations in short eyes using 7 formulas. J Cataract Refract Surg. 2017;43(7):892–7. https://doi.org/10.1016/j.jcrs.2017.07.004.
•• Sudhakar S, Hill DC, King TS, Scott IU, Mishra G, Ernst BB, et al. Intraoperative aberrometry versus preoperative biometry for intraocular lens power selection in short eyes. J Cataract Refract Surg. 2019;45(6):719–24. https://doi.org/10.1016/j.jcrs.2018.12.016. This retrospective consecutive case series highlights that IA was no better than conventional biometry for IOL power determination.
Zaldivar R, Shultz MC, Davidorf JM, Holladay JT. Intraocular lens power calculations in patients with extreme myopia. J Cataract Refract Surg. 2000;26(5):668–74.
Bang S, Edell E, Yu Q, Pratzer K, Stark W. Accuracy of intraocular lens calculations using the IOLMaster in eyes with long axial length and a comparison of various formulas. Ophthalmology. 2011;118(3):503–6. https://doi.org/10.1016/j.ophtha.2010.07.008.
Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD. Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg. 2011;37(11):2018–27. https://doi.org/10.1016/j.jcrs.2011.05.042.
•• Hill DC, Sudhakar S, Hill CS, King TS, Scott IU, Ernst BB, et al. Intraoperative aberrometry versus preoperative biometry for intraocular lens power selection in axial myopia. J Cataract Refract Surg. 2017;43(4):505–10. https://doi.org/10.1016/j.jcrs.2017.01.014. This retrospective consecutive case series suggests that cases with IA had better refractive outcomes than conventional biometry–based formulas in myopic patients.
Jin H, Limberger IJ, Ehmer A, Guo H, Auffarth GU. Impact of axis misalignment of toric intraocular lenses on refractive outcomes after cataract surgery. J Cataract Refract Surg. 2010;36(12):2061–72. https://doi.org/10.1016/j.jcrs.2010.06.066.
U.S. Food and Drug Administration (2005). Summary of safety and effectiveness data P930014/S15, AcrySof® Toric IOL.
• Hatch KM, Woodcock EC, Talamo JH. Intraocular lens power selection and positioning with and without intraoperative aberrometry. J. Refract Surg. 2015;31(4):237–42. https://doi.org/10.3928/1081597X-20150319-03. This study discusses that patients with toric IOLs placed with IA were 2.4 times more likely to have less than 0.5 D of residual refractive astigmatism than conventional methods.
Waisbren E, Ritterband D, Wang L, Trief D, Koplin R, Seedor J. Intraoperative biometry versus conventional methods for predicting intraocular lens power: a closer look at patients undergoing toric lens implantation for astigmatism correction. J Eye Cataract Surg. 2017;3:27. https://doi.org/10.21767/2471-8300.100027.
Woodcock MG, Lehmann R, Cionni RJ, Breen M, Scott MC. Intraoperative aberrometry versus standard preoperative biometry and a toric IOL calculator for bilateral toric IOL implantation with a femtosecond laser: one-month results. J Cataract Refract Surg. 2016;42(6):817–25. https://doi.org/10.1016/j.jcrs.2016.02.048.
Solomon JD, Ladas J. Toric outcomes: computer-assisted registration versus intraoperative aberrometry. J Cataract Refract Surg. 2017;43(4):498–504. https://doi.org/10.1016/j.jcrs.2017.01.012.
Solomon KD, Sandoval HP, Potvin R. Correcting astigmatism at the time of cataract surgery: toric IOLs and corneal relaxing incisions planned with an image-guidance system and intraoperative aberrometer versus manual planning and surgery. J Cataract Refract Surg. 2019;45(5):569–75. https://doi.org/10.1016/j.jcrs.2018.12.002.
Kessel L, Andresen J, Tendal B, Erngaard D, Flesner P, Hjortdal J. Toric intraocular lenses in the correction of astigmatism during cataract surgery: a systematic review and meta-analysis. Ophthalmology. 2016;123(2):275–86. https://doi.org/10.1016/j.ophtha.2015.10.002.
• Packer M. Effect of intraoperative aberrometry on the rate of postoperative enhancement: retrospective study. J Cataract Refract Surg. 2010;36(5):747–55. https://doi.org/10.1016/j.jcrs.2009.11.029. This is a retrospective case-control chart review that shows that when IA was used for LRIs, there was a significant reduction in the odds ratio of subsequent laser vision enhancement.
• Canto AP, Chhadva P, Cabot F, Galor A, Yoo SH, Vaddavalli PK, et al. Comparison of IOL power calculation methods and intraoperative wavefront aberrometer in eyes after refractive surgery. J Refract Surg. 2013;29(7):484–9. https://doi.org/10.3928/1081597X-20130617-07. This retrospective study indicates that IA was the most accurate method to predict IOL power in patients with history of keratorefractive surgery. It compares refractive outcomes for RK, PRK, and LASIK.
Demill DL, Moshirfar M, Neuffer MC, Hsu M, Sikder S. A comparison of the American Society of Cataract and Refractive Surgery post-myopic LASIK/PRK intraocular lens (IOL) calculator and the ocular MD IOL calculator. Clin Ophthalmol. 2011;5:1409–14. https://doi.org/10.2147/OPTH.S24131.
Yesilirmak N, Palioura S, Culbertson W, Yoo SH, Donaldson K. Intraoperative wavefront aberrometry for toric intraocular lens placement in eyes with a history of refractive surgery. J Refract Surg. 2016;32(1):69–70. https://doi.org/10.3928/1081597X-20151210-02.
Fram NR, Masket S, Wang L. Comparison of intraoperative aberrometry, OCT-based IOL formula, Haigis-L, and Masket formulae for IOL power calculation after laser vision correction. Ophthalmology. 2015;122(6):1096–101. https://doi.org/10.1016/j.ophtha.2015.01.027.
Cionni RJ, Dimalanta R, Breen M, Hamilton C. A large retrospective database analysis comparing outcomes of intraoperative aberrometry with conventional preoperative planning. J Cataract Refract Surg. 2018;44(10):1230–5. https://doi.org/10.1016/j.jcrs.2018.07.016.
Runde MR. Modern preoperative intraocular lens calculation is better than intraoperative aberrometry for normal eyes. J Cataract Refract Surg. 2019;45(2):253–4. https://doi.org/10.1016/j.jcrs.2018.10.036.
Zhang Z, Thomas LW, Leu SY, Carter S, Garg S. Refractive outcomes of intraoperative wavefront aberrometry versus optical biometry alone for intraocular lens power calculation. Indian J Ophthalmol. 2017;65(9):813–7. https://doi.org/10.4103/ijo.IJO_163_17.
Davison JA, Potvin R. Preoperative measurement vs intraoperative aberrometry for the selection of intraocular lens sphere power in normal eyes. Clin Ophthalmol. 2017;11:923–9. https://doi.org/10.2147/OPTH.S135659.
Stringham J, Pettey J, Olson RJ. Evaluation of variables affecting intraoperative aberrometry. J Cataract Refract Surg. 2012;38(3):470–4. https://doi.org/10.1016/j.jcrs.2011.09.039.
Masket S, Fram NR, Holladay JT. Influence of ophthalmic viscosurgical devices on intraoperative aberrometry. J Cataract Refract Surg. 2016;42(7):990–4. https://doi.org/10.1016/j.jcrs.2016.04.022.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of Interest
All authors declare that they have no conflicts of interest.
Human and Animal Rights and Informed Consent
This article does not contain any studies with human or animal subjects performed by any of the authors.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is part of the Topical Collection on Cataract
Rights and permissions
About this article
Cite this article
Yong, J.J., Hatch, K.M. Intraoperative Aberrometry for IOL Selection. Curr Ophthalmol Rep 7, 213–218 (2019). https://doi.org/10.1007/s40135-019-00222-z
Published:
Issue Date:
DOI: https://doi.org/10.1007/s40135-019-00222-z