Abstract
Mechanical ventilation (MV) is an important aspect in the intraoperative and early postoperative management of lung transplant (LTx) recipients. There are no randomized-controlled trials of LTx recipient MV strategies; however, there are LTx center experiences and international survey studies reported. The main early complication of LTx is primary graft dysfunction (PGD), which is similar to the adult respiratory distress syndrome (ARDS). We aim to summarize information pertinent to LTx-MV, as well as PGD, ARDS, and intraoperative MV, and to synthesize these available data into recommendations. Based on the available evidence, we recommend lung-protective MV with low tidal volumes (≤6 mL/kg predicted body weight [PBW]) and positive end-expiratory pressure for the LTx recipient. In our opinion, the MV strategy should be based on donor characteristics (donor PBW as a parameter of actual allograft size), rather than based on recipient characteristics; however, this donor characteristics-based protective MV is based on indirect evidence and requires validation in prospective clinical studies.
Similar content being viewed by others
Introduction
Lung transplantation (LTx) is an important treatment option for select patients with end-stage pulmonary disease. Remarkable progress has been made since the modern LTx era began in 1983 [1]. The field of LTx has grown rapidly over the last 30 years with improved surgical techniques and medical management strategies [2, 3]. However, there is little information on mechanical ventilation (MV) strategies after LTx, and no guidelines specific to this setting exist [4•, 5].
Primary graft dysfunction (PGD) represents one of the most common complications observed in the early period following LTx with incidence rates between 10 and 57 % [6••, 7]. PGD is clinically and histologically analogous to the acute respiratory distress syndrome (ARDS) [7, 8] and results from a variety of often simultaneously contributing insults. It is characterized by diffuse pulmonary infiltrates with an abnormal oxygen requirement occurring within 72 h of transplantation [6••, 7]. Histologic examination in PGD shows diffuse alveolar damage [7]. Severe PGD represents both the main risk factor for early mortality after LTx as well as a risk factor for the development of bronchiolitis obliterans syndrome, which is the primary late complication limiting long-term survival of LTx patients [6••, 7, 9]. Therefore, interventions that reduce the rates of PGD could improve both short-term and long-term outcomes for LTx recipients. Management of MV may present an opportunity for such an intervention. Evolving approaches to MV for patients at risk for ARDS and patients with ARDS have resulted in tangible improvements in outcomes [10••, 11–16]. Lung-protective MV strategies incorporating low tidal volumes (V T) limit ventilator-induced lung injury (VILI), reduce morbidity in patients on MV, and improved survival in patients with ARDS [8, 11, 17–19]. Guidelines embrace the use of lower V T in patients with ARDS [17].
The benefits of a lung-protective MV strategy extend to patients at risk for ARDS [13, 20, 21, 22••, 23]. Higher V T were associated with the development of ARDS in patients who came to the intensive care unit without ARDS but had risk factors for it [22••]. Furthermore, in patients with no prior lung injury who received MV during cardiac surgery in the operating room, higher V T settings were associated with higher inflammatory mediator levels [24]. The IMPROVE study provided further evidence that even brief periods of intraoperative lung-protective ventilation result in lower rates of lung injury in surgical patients at intermediate to high risk of pulmonary complications [25••]. While not specifically studied in the context of LTx, the tenets of lung-protective MV are likely generalizable to this conceptually similar setting and, in the absence of direct data, should inform MV strategies.
There are important differences between the LTx recipient and a general intraoperative or postoperative critically ill patient [26, 27]. LTx recipients have mechanical impairments including (1) a fresh thoracotomy wound that creates thoracic cage abnormalities, (2) frequent phrenic nerve dysfunction, and (3) pleural dysfunction [28]. The bronchial anastomoses sites and the allograft airway mucosa are prone to ischemia, poor healing, infection, and subsequent anastomotic airway complications [29]. Another important aspect unique to LTx is that the size of the transplanted lungs can differ significantly from the size of the recipient’s thoracic cavity [30–33, 34•, 35, 36], Fig. 1. In a study of bilateral LTx recipients, V T during MV were substantially higher if the allograft was undersized compared to oversized allografts, when V T were indexed to donor predicted body weight (as an estimate of the actual size of the allograft) [37, 38•].
There are no randomized controlled trials (RCTs) that address MV in the specific context of LTx. We will approach the review of MV of the LTx recipient by first providing a concise summary of potentially generalizable principles derived from key studies in critical care medicine and will then aim to synthesize these principles into strategies that incorporate the unique aspects of LTx [12, 39].
General principles
In the past, MV strategies with V T of 10 to 15 mL/kg were commonly utilized both intraoperatively and in critically ill patients. V T of that size were believed to be necessary to prevent hypoxemia and atelectasis. However, mounting evidence from experimental and clinical studies consistently demonstrates that the application of high V T during MV may aggravate or cause lung injury [40]. MV using large V T can result in overdistention of alveoli and lead to ventilator-induced lung injury (VILI), which can amplify the risk for lung injury [40, 41]. Lung-protective MV refers to the use of low V T and positive end-expiratory pressure (PEEP) [11, 18, 19]. The ARMA study (or tidal volume study), a RCT reported in 2000 by the NHLBI ARDS Network, provided landmark evidence to support a lung-protective MV strategy in the presence of ARDS11. Investigators in that trial examined an approach relating V T to estimated lung sizes expressed as milliliters (mL) per kilogram (kg) predicted body weight (PBW) and compared lung-protective low V T ventilation to conventional V T strategies [11]. V T targets of 6 mL/kg PBW and strategies limiting maximum allowable plateau pressure to 30-cm H2O were compared to V T targets of 12 mL/kg PBW with a maximum allowable plateau pressure of 50-cm H2O. The low V T strategy was associated with reduced 30-day mortality (31 versus 39.8 %, p = 0.007) [11]. The timing of lung-protective ventilation is important for patients who already have ARDS [10••]. ARDS patients who received lung-protective ventilation from the beginning of their lung injury had a lower mortality compared to patients who were initially given larger V T and then were changed to a protective strategy later in their ARDS course [10••]. Each increase of 1 mL/kg PBW in initial V T was associated with a 23 % increase in ICU mortality risk (adjusted hazard ratio 1.23, 95 % confidence interval [CI] 1.06–1.44, p = 0.008) [10••].
Open questions remain regarding the importance of limiting plateau pressure to <30 cm H2O, limiting V T to 6 mL/kg PBW, the optimal setting of PEEP and the role for recruitment maneuvers within the lung-protective ventilation strategies for patients with ARDS [18, 19, 42–47]. However, the benefits of a lung-protective MV strategy appear to extend even to patients without lung injury but who are at risk for the development of ARDS [13, 20, 21, 22••, 23]. Greater V T were associated with the development of ARDS in patients who came to the intensive care unit without ARDS but had risk factors for it [20, 21, 22••]. In the context of donor management for transplant, a RCT compared low V T (6 mL/kg PBW) against a standard donor ventilation strategy (VT 10–12 mL/kg PBW) and showed that a significantly higher proportion of donor lungs from the low V T group could be utilized for LTx (54 versus 27 %, P = 0.004) [13]. Based on the above evidence, lung-protective ventilation strategies should remain the preferred method of MV for most critically ill patients (with or without the presence of ARDS) [17, 22••, 23].
The principles of lung-protective low V T MV have recently been extended to even brief periods of MV, as required for general anesthesia during surgical procedures. Increasing evidence shows that in anesthetized patients without ARDS, lung-protective MV can lower the risk of pulmonary complications and ARDS [24, 25••, 48]. The IMPROVE study, a RCT of lung-protective intraoperative MV, provided compelling evidence that lung-protective ventilation benefits surgical patients at intermediate to high risk of pulmonary complications [25••]. The study demonstrated lower rates of pulmonary and extrapulmonary complications in the 7 days following surgery (27.5 versus 10.5 %, p = 0.001), when individuals received lung-protective ventilation (V T = 6–8 mL/kg predicted body weight [PBW], PEEP = 6–8 cm H2O, and 30-s recruitment maneuvers of 30-cm H2O every 30 min) intraoperatively rather than conventional ventilation (V T = 10–12 mL/kg PBW, no PEEP, and no recruitment maneuvers) [25••]. A recent meta-analysis of RCTs evaluated the effect of intraoperative lung-protective ventilation with lower V T on clinical outcomes in patients undergoing surgery [48]. This meta-analysis of 19 RCTs showed that anesthetized patients who received ventilation with lower V T during surgery had lower risks of lung injury and pulmonary infection than those who received conventional ventilation with higher V T [48].
Lung transplant-specific issues in mechanical ventilation of the recipient
Intraoperative considerations
There are several unique aspects regarding the intraoperative period during LTx [49–53]. Adult LTx can be performed with or without the use of cardiopulmonary bypass in the absence of severe pulmonary hypertension. An off bypass procedure is the preferred approach in many programs when feasible. Cardiopulmonary bypass is an independent predictor for the development of severe PGD in several studies [6••, 38•]. To reduce the likelihood of requiring cardiopulmonary bypass, the least functional lung, as determined by preoperative quantitative ventilation and perfusion imaging, is usually resected and replaced first during a bilateral sequential LTx. Occasionally, patients with cystic fibrosis will have such voluminous purulent secretions that single lung ventilation, as required for an off-bypass LTx, can be difficult. Careful bronchoscopic airway clearance should be routinely done in the operating room before the start of the LTx in such patients with significant airway secretions. For a single LTx, a lateral/anterior thoracotomy is performed. For a bilateral sequential LTx, a clamshell incision or bilateral anterior thoracotomies are commonly used [54]. Alternatively, a median sternotomy can also be performed for bilateral lung transplantation on cardiopulmonary bypass. After implantation of the allograft, it can be important to control the rate of reperfusion of the allograft by gradually releasing the clamp from the pulmonary artery to minimize reperfusion injury. During the period of single lung ventilation, the entire cardiac output passes through the first implanted allograft, while the pulmonary artery on the contralateral side is clamped. Increased pulmonary blood flow results in greater sensitivity to develop VILI [55]. Consequently, careful attention to the size of the V T can be especially important during this vulnerable period. We recommend V T of 6 mL/kg donor PBW. The V T should be further adjusted for single lung ventilation by reducing V T approximately 50 %. PEEP of +5 cm H2O should be used and in case of difficulties with oxygenation, PEEP of up to +10 cm H2O can be considered. After rewarming of the allograft and following deflation episodes, careful recruitment maneuvers to allow complete initial inflation are used by manual bag inflation, while trying to avoid peak inspiratory pressure above 30-cm H2O. Since the lungs are visible in the operating field, the anesthesiologist should be in close communication with the LTx surgeon to assure that all atelectatic lungs areas are visibly seen as recruited. An association between increased FiO2 at reperfusion and a higher risk of severe PGD has been reported in several studies [6••, 38•]. This suggests that the lowest FiO2 should be used to maintain appropriate partial pressure of oxygen in the arterial blood [(PaO2) >70 mmHg] and hemoglobin oxygen saturations [(SpO2) >92 %]. Many LTx recipients have significant pre-transplant chronic hypercarbia from their end-stage lung disease. Intraoperative permissive hypercapnia with pCO2 in pre-transplant range can be helpful to allow for optimal cerebral perfusion and to facilitate the use of low V T. However, the allograft vasculature is often sensitive to elevated pCO2, which can cause vasoconstriction and elevated pulmonary arterial pressure, and these factors need to be considered in the setting of permissive hypercapnia. Inhaled nitric oxide (iNO) or inhaled prostacyclin can be considered in case of pulmonary hypertension or to facilitate protective MV settings in case of significant PGD by improving oxygenation. However, the routine use of iNO has no beneficial impact on outcomes [56–58].
Several situations frequently necessitate the use of cardiopulmonary bypass during LTx. Patients with severe pulmonary hypertension, for example, are most safely transplanted on bypass. After allograft implantation while on bypass, protective resting ventilator settings should be used with V T 4–6 mL/kg donor PBW (further reduced for single lung ventilation) and PEEP of +5 cm H2O. Before coming off cardiopulmonary bypass, it can be helpful to bronchoscopically remove blood clots and secretions from the allograft airways to maximize allograft function and facilitate successful weaning from bypass [59]. More recently, veno-arterial extracorporeal membrane oxygenation (ECMO) has emerged as a valid alternative method of support and was associated with decreased rates of pulmonary and renal complications, as compared with cardiopulmonary bypass [60]. Occasionally, the chest remains open following the LTx [61]. If pressure-assist-control MV modes are used in that setting, the pressure control should be carefully adjusted to assure lung-protective low V T, as increased respiratory system compliance with an open chest is possible. Table 1 summarizes recommendations for the intraoperative MV of the LTX recipient.
Postoperative considerations
The goals of controlled MV immediately following LTx are to protect the allografts from injury while improving function and facilitating early weaning and extubation.
Bilateral lung transplant
A bilateral LTx is the most common LTx in the modern era [3]. There are limited data on MV after a LTx; however, a murine model of LTx demonstrated that the mode of mechanical ventilation applied during the early phase of reperfusion influenced the severity of PGD [62]. A protective ventilatory strategy that minimized pulmonary mechanical stress by low V T was associated with less PGD and improved lung function after LTx. The study concluded that VILI might be an under-recognized phenomenon that contributes significantly to PGD after LTx and that protective ventilatory strategies with low V T could potentially lead to improved outcomes after LTx [62]. In a single-center observational cohort study, the implementation of a management guideline for respiratory and hemodynamic status within the first 72 h after LTx resulted in less severe PGD [63••]. The respiratory portion of the protocol was based on a lung-protective low V T ventilation strategy [63••]. The study also gave parameters for hemodynamic support that emphasized the use of vasoactive drugs over fluid administration to maintain a lower central venous pressure [63••, 64].
In an international survey of the LTx community, the majority of respondents indicated a preference for using lung-protective approaches to mechanical ventilation after LTx [4•]. Low V T based on recipient characteristics were frequently chosen [4•]. Donor characteristics often were not considered and frequently were not known by the team managing mechanical ventilation after LTx [4•]. In a single-center study, the relationship between donor-recipient lung size mismatch and postoperative MV V T in a cohort of bilateral LTx patients was evaluated, Fig. 1. V T settings were expressed as absolute values (in mL) and also as fractions of recipient and donor PBW [37]. Postoperative absolute V T settings were comparable between subsets of patients with undersized, matched, and oversized allografts, and V T settings according to recipient PBW was also similar. V T settings according to donor PBW, however, revealed significant differences between undersized, matched, and oversized subsets (11.4 ± 3.1 versus 9.4 ± 1.2 versus 8.1 ± 2.1, respectively; P < 0.05) [37]. Thus, during mechanical ventilation after bilateral LTx, patients with undersized allografts received relatively greater V T compared to those with oversized allografts when VT was related to donor PBW (as an estimate of the actual allograft size). Postoperatively, a single-center report linked hyperinflation of undersized allografts (i.e., donor lungs smaller than recipient thorax) to an increased risk of early allograft failure [65]. The results of other studies have demonstrated that patients with undersized allografts had worse outcomes, specifically increased rates of PGD, tracheostomy, and resource utilization [30, 38•]. In an ancillary study to the LTx outcome group, an undersized allograft was associated with a significantly increased risk of ISHLT grade 3 PGD after bilateral LTx [38•]. Furthermore, a series of studies revealed an association between undersized allografts and risk of first-year mortality [30–33, 34•, 36, 38•, 66–69]. The mechanisms associating an undersized allograft with a higher risk of PGD and a higher risk of first-year mortality are unclear. Hyperinflation of significantly undersized allografts by V T set according to recipient characteristics could increase the risk of VILI. A hypothesis generated from these investigations of lung size mismatch and clinical outcomes after LTx is that a lung-protective mechanical ventilation strategy based on estimates of the allograft size (i.e., donor PBW) could be protective for patients with undersized allografts. A clinical trial of allograft protective mechanical ventilation with V T settings of 6 mL/kg donor PBW compared with routine mechanical ventilation after LTx could test this hypothesis [70]. Although a majority of respondents to a survey did not consider donor characteristics, they indicated that they might modify MV settings if they knew the donor characteristics [4•]; thus, we recommend that donor characteristics should be communicated to and known by the team managing the MV [4•, 30, 38•, 66]. This could be especially important in case of size reduced and lobar transplants [71, 72].
When there is severe PGD, mechanical ventilation may not be able to safely meet the LTx recipients’ needs in terms of oxygenation and minute ventilation, and the ventilator settings needed may be harmful to the allograft. Many LTx centers use veno-venous (VV)-ECMO as rescue strategy for severe PGD [73, 74•, 75]. The advantages of using VV-ECMO are that it allows using protective ventilator settings and allows minimizing sedation [73, 74•, 75]. Ventilator rest settings on VV-ECMO commonly use very low V T of approximately 4 mL/kg (donor PBW) with PEEP 5–8 cm H2O [76, 77]. There is a prospective trial in progress testing whether ultra-protective ventilation using a tidal volume of 3 mL/kg combined with extracorporeal carbon dioxide removal will improve outcomes in severe ARDS compared with conventional low-V T ventilation [78]. Furthermore, if a single dual-lumen bicaval cannula can be utilized for VV-ECMO, physical therapy and mobilization can occasionally be resumed.
Some patients fail extubation or have complications that require longer duration of mechanical ventilation or VV-ECMO. In these cases, early tracheostomy is often performed [79–81]. This allows for safe weaning trials that lessen the risk of airway complications from repeated intubations and constant high pressure on the bronchial anastomoses [79–81]. Patients also have better comfort, oral hygiene, clearance of pulmonary secretions, and a lower risk of vocal cord injury.
Single-lung transplants
Single LTx represent a minority of procedures done in the modern era [3]. When managing these patients, it is important to consider that the native lung has end-stage disease from different etiologies and should not be relied upon to share the volumes and pressures during mechanical ventilation equally with the allograft. In idiopathic pulmonary fibrosis (IPF), the native lung is less compliant than the allograft, and most of the V T will likely go to the more compliant allograft. Lung-protective ventilator V T should be reduced, and we prefer an initial V T of 4–6 mL/kg of the donor’s PBW. Liberalization of V T may be necessary to minimize patient sedation and to allow for early extubation. Recipients of a single LTx for IPF can also have an IPF flare in the native lung triggered by the LTx surgery. This can lead to more severe hypoxemia from shunt physiology through a very noncompliant IPF lung. Recipients of a single LTx for COPD on the other hand have a very compliant native lung, which has severe expiratory airflow obstruction. This can lead to overdistention of the recipient’s native lung from dynamic hyperinflation and auto-PEEP. Here an approach to mechanical ventilation that maximizes expiratory time, by using a short inspiratory time, a low respiratory rate and a VT that allows for full expiration are important. If these difficulties cannot be managed with conventional mechanical ventilation, patients may require independent lung ventilation with a double-lumen endotracheal tube and different ventilator settings for each lung [5]. However, independent lung ventilation generally requires heavy sedation and a preferable approach can be to utilize VV-ECMO, or extracorporeal CO2 elimination as a rescue strategy, as discussed above.
Bronchial anastomoses
A key aspect unique to LTx is the presence of the bronchial anastomoses. Anastomotic airway complications occur in approximately 10–20 % of LTx recipients and often present in both acute and long-term problems [29, 82–87]. Anastomotic airway complications include infection, stenosis, and dehiscence [29, 82–87]. In general, the bronchial circulation is not restored during transplant, and ischemia of the transplanted airway and airway mucosa frequently occur after LTx [29, 88]. Thus, the bronchial anastomoses sites are prone to poor healing, infection, and anastomotic airway complications. There may be collateral flow from the pulmonary circulation, but the pulmonary circulation has relatively low vascular pressure and thus the magnitude of collateral flow is probably small. Therefore, positive pressure mechanical ventilation could potentially impair perfusion to transplanted airways, especially when high inflation pressures are required. In addition, any allograft parenchymal pathology such as PGD, infection, or rejection will reduce the pulmonary flow to the major bronchi and thereby impair anastomotic healing. Alternatively, it is possible that PEEP may increase perfusion through microscopic collateral vessels by redistributing blood flow from the pulmonary vessels which in this setting could be acting as a vascular capacitance bed. This theory is supported by a dog model of LTx without restoration of the bronchial arterial circulation, where increasing the PEEP from 5- to 10-cm H2O was associated with increased retrograde bronchial mucosal blood flow to the bronchial anastomoses [89]. However, positive pressure ventilation can also contribute to bronchial wall and anastomotic stress. High airway pressures and prolonged ventilation times have been linked to the risk for anastomotic airway complications in some studies, however not in others [82, 85, 90]. The concern regarding high airway pressures and anastomotic airway complications are likely reflected in the responses on approaches to peak inspiratory pressure (PIP) and PEEP during MV after LTx in an international survey [4•]. Almost all respondents (91 %) reported routinely assessing airway pressures and most had a PIP limit [4•]. The median limit was 30-cm H2O (interquartile range (IQR) 30–35-cm H2O). The PIP limit differed significantly between volume assist/control (VAC) users and pressure assist/control (PAC) users (median 35 [IQR 35–40] versus median 30 [IQR 20–35], p = 0.002). In that survey, the maximum acceptable PEEP level after LTx averaged 11-cm H2O (IQR 10–12.5-cm H2O) [4•]. However, there is little evidence guiding optimal setting of PEEP and PIP for the LTx recipient and regarding how much pressure is too much for the anastomoses.
Modes of ventilation
Immediately after surgery, there are many different providers and support staff involved in the management of the MV of the LTx recipient. An international survey indicated that the ventilator settings were determined by intensivists in 50 % of centers, pulmonologists in 42 %, surgeons in 28 %, anesthesiologists in 26 %, and respiratory therapists in several instances (multiple answers were allowed) [4•]. Approximately equal percentages of respondents reported using PAC ventilation (37 %) and VAC ventilation (35 %) [4•]. This requires careful attention to the ventilator inputs and outputs as different providers have different preferences and levels of experience with specific ventilator modes. VAC modes are most likely to have consistent tidal volumes but require attention to peak and plateau airway pressures. PAC modes can avoid high peak but not transpulmonary pressures, sometimes providing larger V T than intended. We emphasize that limiting peak inspiratory pressures does not assure that transpulmonary pressure remains in a lung-protective range, except during general anesthesia or deep sedation. Therefore, we prefer the VAC or pressure-regulated volume control (PRVC) modes, rather than PAC, during the period of controlled mechanical ventilation in the ICU. Management guidelines have been successfully implemented at individual LTx centers and can help to facilitate a consistent approach to mechanical ventilation of the LTx recipient [4•, 63••]. Table 2 summarizes recommendations for the postoperative MV of the LTX recipient.
Summary
Lung transplantation is a very specialized field with unique surgical and medical aspects. The principles of lung-protective ventilation have a strong evidence base in patients at risk for or with ARDS. Much of the recommendations presented in this review of lung transplant recipient mechanical ventilation are extrapolated from data in the general patient populations because of the close relationship between PGD and ARDS, as well as the general influence of anesthesia on the respiratory system. All LTx recipients are at risk for PGD, which is similar to ARDS, and should receive mechanical ventilation according to the principles of lung-protective ventilation with low tidal volumes. In our opinion, the low tidal volume strategy should be based on donor characteristics (i.e., donor predicted body weight as a parameter reflecting the actual allograft size), rather than based on LTx recipient characteristics.
References
Papers of particular interest, published recently, have been highlighted as: • Of importance •• Of major importance
Grossman RF, Frost A, Zamel N, et al. Results of single-lung transplantation for bilateral pulmonary fibrosis. The Toronto Lung Transplant Group. N Engl J Med. 1990;322:727–33.
Eberlein M, Garrity ER, Orens JB. Lung allocation in the United States. Clin Chest Med. 2011;32:213–22.
Yusen RD, Edwards LB, Kucheryavaya AY, et al. The registry of the International Society for Heart and Lung Transplantation: thirty-first adult lung and heart-lung transplant report—2014; focus theme: retransplantation. J Heart Lung Transplant. 2014;33:1009–24.
Beer A, Reed RM, Bolukbas S, et al. Mechanical ventilation after lung transplantation. An international survey of practices and preferences. Ann Am Thorac Soc. 2014;11:546–53. The study is an international survey of the lung transplant community to determine ventilator management strategies in lung transplant recipients.
Lucangelo U, Del Sorbo L, Boffini M, et al. Protective ventilation for lung transplantation. Curr Opin Anaesthesiol. 2012;25:170–4.
Diamond JM, Lee JC, Kawut SM, et al. Clinical risk factors for primary graft dysfunction after lung transplantation. Am J Respir Crit Care Med. 2013;187:527–34. This 10-center prospective cohort study identified risk factors for grade 3 PGD. History of donor smoking, higher FiO2 during allograft reperfusion, use of cardiopulmonary bypass and pulmonary arterial hypertension were independent predictors of severe PGD.
Lee JC, Christie JD. Primary graft dysfunction. Clin Chest Med. 2011;32:279–93.
Ranieri VM, Rubenfeld GD, Thompson BT, et al. Acute respiratory distress syndrome: the Berlin Definition. Jama. 2012;307:2526–33.
Daud SA, Yusen RD, Meyers BF, et al. Impact of immediate primary lung allograft dysfunction on bronchiolitis obliterans syndrome. Am J Respir Crit Care Med. 2007;175:507–13.
Needham DM, Yang T, Dinglas VD, et al. Timing of low tidal volume ventilation and ICU mortality in ARDS: a prospective cohort study. Am J Respir Crit Care Med 2014. This is a prospective study of 482 patients with ARDS and 11,558 twice-daily tidal volume assessments. Higher tidal volumes shortly after ARDS onset were associated with a greater risk of ICU mortality. Per each 1 ml/kg PBW increase in initial tidal volume, there was a 23 % increase in ICU mortality risk (adjusted hazard ratio, 1.23; 95 % confidence interval [CI], 1.06–1.44; P = 0.008) .
Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The Acute Respiratory Distress Syndrome Network. N Engl J Med 2000; 342:1301–8.
Klesney-Tait JA, Eberlein M. Lung protective ventilation in donors: an ounce of prevention. Chest. 2014;146:4–6.
Mascia L, Pasero D, Slutsky AS, et al. Effect of a lung protective strategy for organ donors on eligibility and availability of lungs for transplantation: a randomized controlled trial. Jama. 2010;304:2620–7.
Brower RG, Rubenfeld GD. Lung-protective ventilation strategies in acute lung injury. Crit Care Med. 2003;31:S312–6.
Diaz JV, Brower R, Calfee CS, et al. Therapeutic strategies for severe acute lung injury. Crit Care Med. 2010;38:1644–50.
Fessler HE, Brower RG. Protocols for lung protective ventilation. Crit Care Med. 2005;33:S223–7.
Dellinger RP, Levy MM, Rhodes A, et al. Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock: 2012. Crit Care Med. 2013;41:580–637.
Hager DN, Krishnan JA, Hayden DL, et al. Tidal volume reduction in patients with acute lung injury when plateau pressures are not high. Am J Respir Crit Care Med. 2005;172:1241–5.
Janz DR, Ware LB. Approach to the patient with the acute respiratory distress syndrome. Clin Chest Med. 2014;35:685–96.
Determann RM, Royakkers A, Wolthuis EK, et al. Ventilation with lower tidal volumes as compared with conventional tidal volumes for patients without acute lung injury: a preventive randomized controlled trial. Crit Care. 2010;14:R1.
Gajic O, Frutos-Vivar F, Esteban A, et al. Ventilator settings as a risk factor for acute respiratory distress syndrome in mechanically ventilated patients. Intensive Care Med. 2005;31:922–6.
Serpa Neto A, Cardoso SO, Manetta JA, et al. Association between use of lung-protective ventilation with lower tidal volumes and clinical outcomes among patients without acute respiratory distress syndrome: a meta-analysis. Jama. 2012;308:1651–9. A meta-analysis of 20 studies that evaluated lower vs higher tidal volumes in surgical and ICU patients without ARDS concluded that lower volumes were associated with reduced rates of lung injury, mortality, pneumonia, and atelectasis.
Serpa Neto A, Simonis FD, Barbas CS, et al. Association between tidal volume size, duration of ventilation, and sedation needs in patients without acute respiratory distress syndrome: an individual patient data meta-analysis. Intensive Care Med. 2014;40:950–7.
Zupancich E, Paparella D, Turani F, et al. Mechanical ventilation affects inflammatory mediators in patients undergoing cardiopulmonary bypass for cardiac surgery: a randomized clinical trial. J Thorac Cardiovasc Surg. 2005;130:378–83.
Futier E, Constantin JM, Paugam-Burtz C, et al. A trial of intraoperative low-tidal-volume ventilation in abdominal surgery. N Engl J Med. 2013;369:428–37. This was a multicenter, randomized, parallel group trial of intraoperative ventilation strategies in patients undergoing abdominal surgery with elevated risk of pulmonary complications. It showed lower rates of major complications, pneumonia, and length of stay in the lung protective ventilation group as compared to non-protective ventilation.
Lau CL, Patterson GA, Palmer SM. Critical care aspects of lung transplantation. J Intensive Care Med. 2004;19:83–104.
Schuurmans MM, Benden C, Inci I. Practical approach to early postoperative management of lung transplant recipients. Swiss Med Wkly. 2013;143:w13773.
Arndt A, Boffa DJ. Pleural space complications associated with lung transplantation. Thorac Surg Clin. 2015;25:87–95.
Machuzak M, Santacruz JF, Gildea T, et al. Airway complications after lung transplantation. Thorac Surg Clin. 2015;25:55–75.
Eberlein M, Arnaoutakis GJ, Yarmus L, et al. The effect of lung size mismatch on complications and resource utilization after bilateral lung transplantation. J Heart Lung Transplant. 2012;31:492–500.
Eberlein M, Bolukbas S, Reed RM. eComment. Gender mismatching in lung transplantation: lung size mismatch is the issue! Interact Cardiovasc Thorac Surg. 2013;16:435–6.
Eberlein M, Bolukbas S, Reed RM. Bilateral lobar lung transplantation and size mismatch by pTLC-ratio. Eur J Cardiothorac Surg. 2013;44:394–5.
Eberlein M, Reed RM, Bolukbas S, et al. Lung size mismatch and survival after single and bilateral lung transplantation. Ann Thorac Surg. 2013;96:457–63.
Eberlein M, Reed RM, Maidaa M, et al. Donor-recipient size matching and survival after lung transplantation. A cohort study. Ann Am Thorac Soc. 2013;10:418–25. This is an analysis of the SRTR lung transplant database in post lung allocation score era showed that donor to recipient lung size matching (pTLC ratio) is an independent predictor of death in the first year after LTx. Specifically, the more undersized the allograft is, the higher the risk of 1-year mortality.
Eberlein M, Reed RM, Permutt S, et al. Parameters of donor-recipient size mismatch and survival after bilateral lung transplantation. J Heart Lung Transplant. 2012;31:1207–1213.e1207.
Reed RM, Eberlein M. Sizing strategies in heart and lung transplantation: you cannot manage what you do not measure. Futur Cardiol. 2014;10:303–6.
Dezube R, Arnaoutakis GJ, Reed RM, et al. The effect of lung-size mismatch on mechanical ventilation tidal volumes after bilateral lung transplantation. Interact Cardiovasc Thorac Surg. 2013;16:275–81.
Eberlein M, Reed RM, Bolukbas S, et al. Lung size mismatch and primary graft dysfunction after bilateral lung transplantation. J Heart Lung Transplant. 2015;34(2)233–40. Ancillary study to the Lung Transplant Outcomes Group Study analyzing 812 patients who were bilateral lung transplant recipients. Oversized lungs with a higher pTLC ratio were independently associated with decreased odds of grade 3 primary graft dysfunction .
Klesney-Tait JA, Parekh K. Diamonds in the rough: identification of usable donor lungs. Am J Respir Crit Care Med. 2013;188:410–2.
Dreyfuss D, Soler P, Basset G, et al. High inflation pressure pulmonary edema: respective effects of high airway pressure, high tidal volume, and positive end-expiratory pressure. Am Rev Respir Dis. 1988;137:1159–64.
Slutsky AS, Ranieri VM. Ventilator-induced lung injury. N Engl J Med. 2013;369:2126–36.
Jaswal DS, Leung JM, Sun J, et al. Tidal volume and plateau pressure use for acute lung injury from 2000 to present: a systematic literature review. Crit Care Med. 2014;42:2278–89.
Fan E, Wilcox ME, Brower RG, et al. Recruitment maneuvers for acute lung injury: a systematic review. Am J Respir Crit Care Med. 2008;178:1156–63.
Briel M, Meade M, Mercat A, et al. Higher vs lower positive end-expiratory pressure in patients with acute lung injury and acute respiratory distress syndrome: systematic review and meta-analysis. Jama. 2010;303:865–73.
Brower RG, Lanken PN, MacIntyre N, et al. Higher versus lower positive end-expiratory pressures in patients with the acute respiratory distress syndrome. N Engl J Med. 2004;351:327–36.
Miller 3rd RR, Macintyre NR, Hite RD, et al. Point: should positive end-expiratory pressure in patients with ARDS be set on oxygenation? Yes. Chest. 2012;141:1379–82.
Schmidt GA. Counterpoint: should positive end-expiratory pressure in patients with ARDS be set based on oxygenation? No. Chest. 2012;141:1382–4.
Gu WJ, Wang F, Liu JC. Effect of lung-protective ventilation with lower tidal volumes on clinical outcomes among patients undergoing surgery: a meta-analysis of randomized controlled trials. Cmaj. 2014.
Baez B, Castillo M. Anesthetic considerations for lung transplantation. Semin Cardiothorac Vasc Anesth. 2008;12:122–7.
Castillo M. Anesthetic management for lung transplantation. Curr Opin Anaesthesiol. 2011;24:32–6.
Miranda A, Zink R, McSweeney M. Anesthesia for lung transplantation. Semin Cardiothorac Vasc Anesth. 2005;9:205–12.
Rosenberg AL, Rao M, Benedict PE. Anesthetic implications for lung transplantation. Anesthesiol Clin N Am. 2004;22:767–88.
Singh H, Bossard RF. Perioperative anaesthetic considerations for patients undergoing lung transplantation. Can J Anaesth. 1997;44:284–99.
Parekh K, Patterson GA. Technical considerations in adult lung transplantation. Semin Thorac Cardiovasc Surg. 2004;16:322–32.
Broccard AF, Hotchkiss JR, Kuwayama N, et al. Consequences of vascular flow on lung injury induced by mechanical ventilation. Am J Respir Crit Care Med. 1998;157:1935–42.
Meade MO, Granton JT, Matte-Martyn A, et al. A randomized trial of inhaled nitric oxide to prevent ischemia-reperfusion injury after lung transplantation. Am J Respir Crit Care Med. 2003;167:1483–9.
Khan TA, Schnickel G, Ross D, et al. A prospective, randomized, crossover pilot study of inhaled nitric oxide versus inhaled prostacyclin in heart transplant and lung transplant recipients. J Thorac Cardiovasc Surg. 2009;138:1417–24.
Yerebakan C, Ugurlucan M, Bayraktar S, et al. Effects of inhaled nitric oxide following lung transplantation. J Card Surg. 2009;24:269–74.
Eberlein M, Parekh K, Hansdottir S, et al. Plastic bronchitis complicating primary graft dysfunction after lung transplantation. Ann Thorac Surg. 2014;98:1849.
Bermudez CA, Shiose A, Esper SA, et al. Outcomes of intraoperative venoarterial extracorporeal membrane oxygenation versus cardiopulmonary bypass during lung transplantation. Ann Thorac Surg. 2014;98:1936–43.
Shigemura N, Orhan Y, Bhama JK, et al. Delayed chest closure after lung transplantation: techniques, outcomes, and strategies. J Heart Lung Transplant. 2014;33:741–8.
de Perrot M, Imai Y, Volgyesi GA, et al. Effect of ventilator-induced lung injury on the development of reperfusion injury in a rat lung transplant model. J Thorac Cardiovasc Surg. 2002;124:1137–44.
Currey J, Pilcher DV, Davies A, et al. Implementation of a management guideline aimed at minimizing the severity of primary graft dysfunction after lung transplant. J Thorac Cardiovasc Surg. 2010;139:154–61. This study showed that the implementation of an evidence-based guideline for managing respiratory and hemodynamic status of lung transplant recipients led to a reduction in severity of primary graft dysfunction.
McIlroy DR, Pilcher DV, Snell GI. Does anaesthetic management affect early outcomes after lung transplant? An exploratory analysis. Br J Anaesth. 2009;102:506–14.
Kozower BD, Meyers BF, Ciccone AM, et al. Potential for detrimental hyperinflation after lung transplantation with application of negative pleural pressure to undersized lung grafts. J Thorac Cardiovasc Surg. 2003;125:430–2.
Eberlein M, Diehl E, Bolukbas S, et al. An oversized allograft is associated with improved survival after lung transplantation for idiopathic pulmonary arterial hypertension. J Heart Lung Transplant. 2013;32:1172–8.
Eberlein M, Permutt S, Brown RH, et al. Supranormal expiratory airflow after bilateral lung transplantation is associated with improved survival. Am J Respir Crit Care Med. 2011;183:79–87.
Eberlein M, Permutt S, Chahla MF, et al. Lung size mismatch in bilateral lung transplantation is associated with allograft function and bronchiolitis obliterans syndrome. Chest. 2012;141:451–60.
Eberlein M, Reed RM. Letter by Eberlein and Reed regarding article, "transplantation for idiopathic pulmonary arterial hypertension: improvement in the lung allocation score era". Circulation. 2014;129:e457.
Diamond JM, Ahya VN. Mechanical ventilation after lung transplantation. It's time for a trial. Ann Am Thorac Soc. 2014;11:598–9.
Shigemura N, Bhama J, Bermudez C, et al. Lobar lung transplantation: emerging evidence for a viable option. Semin Thorac Cardiovasc Surg. 2013;25:95–6.
Shigemura N, D'Cunha J, Bhama JK, et al. Lobar lung transplantation: a relevant surgical option in the current era of lung allocation score. Ann Thorac Surg. 2013;96:451–6.
Castleberry AW, Hartwig MG, Whitson BA. Extracorporeal membrane oxygenation post lung transplantation. Curr Opin Organ Transplant 2013.
Hartwig MG, Walczak R, Lin SS, et al. Improved survival but marginal allograft function in patients treated with extracorporeal membrane oxygenation after lung transplantation. Ann Thorac Surg. 2012;93:366–71. This was a retrospective, single-center study of patients requiring VV ECMO after lung transplantation. ECMO provided a reasonable support option with improved outcomes compared to conventional management.
Rehder KJ, Turner DA, Hartwig MG, et al. Active rehabilitation during extracorporeal membrane oxygenation as a bridge to lung transplantation. Respir Care. 2013;58:1291–8.
Del Sorbo L, Goffi A, Goligher E, et al. Setting mechanical ventilation in ARDS patients during VV-ECMO: where are we? Minerva Anestesiol. 2014.
Marhong JD, Telesnicki T, Munshi L, et al. Mechanical ventilation during extracorporeal membrane oxygenation. An international survey. Ann Am Thorac Soc. 2014;11:956–61.
Bein T, Weber-Carstens S, Goldmann A, et al. Lower tidal volume strategy (approximately 3 mL/kg) combined with extracorporeal CO2 removal versus 'conventional' protective ventilation (6 mL/kg) in severe ARDS: the prospective randomized Xtravent-study. Intensive Care Med. 2013;39:847–56.
Feltracco P, Milevoj M, Alberti V, et al. Early tracheostomy following lung transplantation. Transplant Proc. 2011;43:1151–5.
Padia SA, Borja MC, Orens JB, et al. Tracheostomy following lung transplantation predictors and outcomes. Am J Transplant. 2003;3:891–5.
Waller EA, Aduen JF, Kramer DJ, et al. Safety of percutaneous dilatational tracheostomy with direct bronchoscopic guidance for solid organ allograft recipients. Mayo Clin Proc. 2007;82:1502–8.
82 Murthy SC, Blackstone EH, Gildea TR, et al. Impact of anastomotic airway complications after lung transplantation. Ann Thorac Surg. 2007; 84:401–409, 409.e401–404
Porhownik NR. Airway complications post lung transplantation. Curr Opin Pulm Med. 2013;19:174–80.
Puchalski J, Lee HJ, Sterman DH. Airway complications following lung transplantation. Clin Chest Med. 2011;32:357–66.
Santacruz JF, Mehta AC. Airway complications and management after lung transplantation: ischemia, dehiscence, and stenosis. Proc Am Thorac Soc. 2009;6:79–93.
Thistlethwaite PA, Yung G, Kemp A, et al. Airway stenoses after lung transplantation: incidence, management, and outcome. J Thorac Cardiovasc Surg. 2008;136:1569–75.
Weder W, Inci I, Korom S, et al. Airway complications after lung transplantation: risk factors, prevention and outcome. Eur J Cardiothorac Surg. 2009;35:293–8.
Widdicombe J. New perspectives on basic mechanisms in lung disease. 4. Why are the airways so vascular? Thorax. 1993;48:290–5.
Yokomise H, Cardoso PF, Kato H, et al. The effect of pulmonary arterial flow and positive end-expiratory pressure on retrograde bronchial mucosal blood flow. J Thorac Cardiovasc Surg. 1991;101:201–8.
Date H, Trulock EP, Arcidi JM, et al. Improved airway healing after lung transplantation. An analysis of 348 bronchial anastomoses. J Thorac Cardiovasc Surg. 1995;110:1424–32.
Compliance with Ethics Guidelines
Conflicts of Interest
The authors declare that they have no conflicts of interests to declare
Human and Animal Rights and Informed Consent
This article contains no studies with human or animal subjects performed by the author.
Author information
Authors and Affiliations
Corresponding author
Additional information
This article is part of the Topical Collection on Lung Transplant
Rights and permissions
About this article
Cite this article
Barnes, L., Reed, R.M., Parekh, K.R. et al. Mechanical ventilation for the lung transplant recipient. Curr Pulmonol Rep 4, 88–96 (2015). https://doi.org/10.1007/s13665-015-0114-8
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13665-015-0114-8