Dear Editor,

We would like to clarify some points about the DIFFUSE (Defining fluticasone propionate/formoterol particulate size) study (December 2013, p 39) [1] in light of the recent review by De Maria and colleagues [2]. Our in vitro study assessed the effects of inhalation flow rate on the aerodynamic particle size distribution (APSD) of fluticasone propionate/formoterol pMDI (FP/FORM; flutiform®, Jagotec AG, Muttenz, Switzerland) compared with three other ICS/LABA combinations: fluticasone propionate/salmeterol DPI (FP/SAL; Seretide® Accuhaler®, GlaxoOperations UK Ltd, Herts, UK), budesonide/formoterol DPI (BUD/FORM; Symbicort® Turbohaler®, AstraZeneca AB, Södertälje, Sweden) and beclometasone dipropionate/formoterol pMDI (BDP/FORM; Foster™, Chiesi Farmaceutici S.p.A, Parma, Italy) [1].

Contrary to the suggestions in the review, the patient information leaflets for FP/FORM and BDP/FORM do not specify a flow rate or ‘a long breath’ [3, 4]. Inhalation flow rates of 30–60 L/min are recognised as ‘clinically effective’ for pMDIs and DPIs, according to the In-Check DIAL [5]. pMDIs are active aerosols with relatively low resistance, and patients often use them at flow rates >30 L/min [68]. Hence, the flow rate dependency of inhalers is of interest to physicians as well as regulators. Including a faster flow rate of 60 L/min in our study was therefore appropriate, and is relevant to the real-life use of these inhalers.

All inhalers were tested across pre-defined parameters, using the same apparatus and the same ambient conditions, to ensure a fair and appropriate comparison. The review questions the use of an Andersen Cascade Impactor (ACI): an ACI is arguably the most widely used tool for testing inhalers, and is routinely used at flow rates of 30, 60 L/min, and above [914]. Indeed, the authors of the review have themselves used an ACI to assess pMDIs at flow rates of 30 and 60 L/min [9, 15].

Importantly, we adapted and calibrated the ACI for use at 60 L/min (as recommended by the European Pharmacopoeia [16]), although this was unclear in our paper. Following standard practice, stages 0 and 7 were removed, and new stages (−0 and −1) were added, providing ‘target cut-points’ for 60 L/min with the same particle size cut-off diameters as for 30 L/min [17].

The review suggests that a ‘low-dose number’ and uncoated ACI plates may have led to an underestimation of BDP/FORM particle size information. Bouncing and consequent re-entrainment of particles might affect assessment of particle size for solid particles impacting onto ACI stages, and the effect is particularly pronounced for high-payload powder presentations and can lead to an overestimation of fine particles. Hence, the plates were coated with surfactant when testing the two DPIs [1], as recommended by the European Pharmacopoeia. Coating may be not necessary for testing pMDIs [16, 17] and is not required for FP/FORM pMDI, but we accept the possibility that uncoated plates could lead to an overestimation of extra-fine particles for BDP/FORM solution and therefore an underestimation of the Mass Median Aerodynamic Diameter (MMAD). This may explain the high Geometric Standard Deviation (GSD) for BDP/FORM at 60L/min; further work is needed to understand this. However and importantly, uncoated plates would not lead to an underestimation of the fine particle fraction (FPF) nor explain the flow rate dependent effects of the FPF for BDP/FORM.

Internal qualification of the methods confirmed that the number of actuations (i.e. discharges) did not affect the results. The number of discharges was limited in accordance with the Pharmacopoeias, to avoid build-up effects and to align the number of discharges with the dose received by the patient, as closely as was acceptable for analysis. Moreover, the total amount of drug recovered from all stages, including from the ACI induction port, was within USP limits.

Contrary to the suggestion in the review, comparing the FPFs of different ICSs is entirely acceptable, in fact the authors of the review have also done such a comparison [9]. There is no reason why the APSDs of FP/FORM and BDP/FORM cannot be compared accurately and reliably; the amount of each drug component deposited at each ACI stage was analysed using established methods specific to each of the individual chemical entities in each presentation. For each product, the FPF of the LABA showed the same pattern as for the ICS; as both products contain similar doses of the same LABA (formoterol), it seems unlikely that the assay would underestimate the dose of formoterol emitted by the BDP/FORM device, but not by FP/FORM.

The review authors also query whether it is appropriate to express the FPF as a percentage of the labelled (or total) dose. For regulatory purposes, the FPF is often reported as a percentage of the delivered (ex-actuator) dose, but it is equally valid (and common practice) to describe FPFs as a percentage of the labelled dose. When comparing different ICS/LABA products, the labelled dose (e.g. BDP/FORM: 100/6 μg) may be more relevant to prescribers, who may find it more difficult to relate to FPF data expressed for the delivered dose (e.g. BDP/FORM 84.6/5.0 μg). Moreover, using the delivered dose could bias results in favour of inefficient delivery systems which have lower delivered doses, or mask flow rate dependent changes to the FPF if the delivered dose is similarly affected by flow rate.

Unpublished in-house data generated by Chiesi via a next generation impactor, show that BDP/FORM produces a “consistent fine particle distribution” as reflected by a consistent MMAD across flow rates, whereas our study suggested a decrease in MMAD at the higher flow rate. It is possible, as noted earlier, that methodological differences may account for this discrepancy. However, it is important to note that, in their study, the fine particle dose (FPD) for BDP varied from 34 μg at 30 L/min to 47 μg at 60 L/min, a 38% relative increase [2]. These data clearly support the findings of the DIFFUSE study that the FPD (and therefore also the FPF) of BDP/FORM is flow rate-dependent. By contrast, both our and Chiesi’s data show that FP/FORM produces a consistent FPD across both flow rates [1, 2].

The review authors question our statement that particles <1 µm are prone to being exhaled. Several papers state that very small particles can be prone to exhalation [18, 19]. Submicron particles diffuse by Brownian motion [18] and deposit upon collision with the airway wall. A failure to breath-hold is one of the most common inhalation errors made by patients [8] and the ERS/ISAM taskforce notes that a longer particle residence time (and thus a long breath hold) will result in greater deposition [20]. Hence, our paper notes that very small extra-fine particles (0.1–1.0 μm) “are prone to exhalation, especially if the breath hold is insufficient”, and we still consider this statement valid and fair and supported by the literature. However, we also advocate that patients hold their breath for as long as is comfortable to maximise deposition of all particle sizes.

In conclusion, in our study FP/FORM produced a high and consistent FPF of approximately 40% at flow rates of 28.3 and 60.0 L/min, whereas the FPF was flow rate-dependent for other products tested [1]. Our study was fair, robust and appropriate: all inhalers were tested under identical conditions using industry standard methods, apparatus and analyses, consistent with European Pharmacopoeia and specific to the molecules analysed [16]. In their review, the authors question our findings and note that in-house studies show BDP/FORM has a consistent particle size distribution (i.e. MMAD) at flow rates of 28–30 and 60 L/min [2]. However, their own data clearly demonstrate that the FPD, and therefore FPF, of BDP/FORM is flow rate-dependent. This is in line with, and is supportive of, our findings.