To the Editor,
We present a unique case of the use of a Nexfin™ HD monitor (BMEYE, Amsterdam, NL) for continuous real-time noninvasive blood pressure (BP) monitoring in a patient with previous bilateral radial artery ligation.
The Nexfin device uses photoplethysmography via a finger cuff to provide noninvasive continuous monitoring of BP, stroke volume, heart rate (HR), and cardiac output.1 The device has been validated as a feasible method for continuous hemodynamic monitoring in the majority of patients undergoing surgery and in the intensive care unit,1,2 and it has been found comparable with a traditional BP cuff2 and thermodilution monitoring.3 While this device cannot completely replace the gold standard of invasive intra-arterial monitoring,1,4 it may still be a suitable option for patients with increased risk factors with radial artery cannulation or where invasive techniques are not warranted or practical.1,3 The pulse pressure method used by the Nexfin device relies on an optimal arterial pressure tracing, and any physiologic or artifact interference may cause inaccuracies in Nexfin readings.3 The Nexfin device has limited applicability in critically ill patients with poor hand perfusion due to systemic hypotension, use of systemic vasopressors, hypothermia, or severe peripheral edema.1,4 Nevertheless, reports are lacking that describe the performance of the Nexfin device in patients with absent radial arteries.
We report the case of a 69-yr-old male, American Society of Anesthesiologists physical status III, undergoing right femoral-femoral bypass graft and extended profundoplasty for lower limb ischemia due to peripheral vascular disease. The patient’s medical history inlcuded coronary artery disease with coronary artery bypass graft (CABG) surgery 13 years prior as well as multiple previous peripheral revascularization procedures. He had both radial arteries harvested for use as vascular conduits for his CABG surgery. He also had a history of hypertension, type 2 diabetes mellitus, and a 7.5 pack-year smoking history. During the preoperative evaluation, both of the patient’s hands appeared well perfused and ulnar pulses were present bilaterally. An Allen’s test was not performed. We elected to proceed with femoral-femoral artery bypass under spinal anesthesia. In addition to standard noninvasive monitors, including a brachial BP cuff on the left arm, a Nexfin cuff was placed on the patient’s right index finger. Baseline brachial noninvasive BP was measured on both of his arms to ensure there was no pre-existing difference in BP. After a successful subarachnoid block, the surgery proceeded uneventfully.
During the procedure, the mean systolic and diastolic BP readings by traditional BP cuff monitoring were found to be approximately 165 mmHg and 72 mmHg, respectively, compared with 111 mmHg and 46 mmHg, respectively, using the Nexfin monitor (Figure). This discrepancy in systolic and diastolic BP (approximately 54 mmHg and 26 mmHg, respectively) was consistent throughout monitoring. The patient maintained a HR of 50-60 beats·min−1 as measured by both standard and Nexfin monitoring, and he remained hemodynamically stable throughout the procedure and postoperatively until discharge two days later.
Radial artery harvesting has been shown to produce minimal differences in hand circulation up to two years post-procedure due to ulnar compensation. Longer term, however, there can be markedly reduced perfusion due to atherosclerotic changes and intimal thickening.5 Our case showed a significant and sustained discrepancy in BP readings between the traditional BP monitor and Nexfin, suggesting that Nexfin is likely not a reliable assessment of BP in patients with limited or absent radial perfusion. These findings are consistent with previous discrepancies in Nexfin readings found in patients with poor hand perfusion of various etiologies.1,3,4
Further research with the Nexfin monitor in patients with limited or absent radial perfusion will help determine whether absent radial arteries should be considered an exclusion criterion for the use of Nexfin for BP monitoring.
References
Ameloot K, Palmers PJ, Malbrain ML. The accuracy of noninvasive cardiac output and pressure measurements with finger cuff: a concise review. Curr Opin Crit Care 2015; 21: 232-9.
Vos JJ, Poterman M, Mooyaart EA, et al. Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia. Br J Anaesth 2014; 113: 67-74.
Thiele RH, Bartels K, Gan TJ. Inter-device differences in monitoring for goal-directed fluid therapy. Can J Anesth 2015; 62: 169-81.
Hohn A, Defosse JM, Becker S, Steffen C, Wappler F, Sakka SG. Non-invasive continuous arterial pressure monitoring with Nexfin does not sufficiently replace invasive measurements in critically ill patients. Br J Anaesth 2013; 111: 178-84.
Chim H, Bakri K, Moran SL. Complications related to radial artery occlusion, radial artery harvest, and arterial lines. Hand Clin 2015; 31: 93-100.
Conflicts of interest
None declared.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Simula, N., Earle, R., Vaghadia, H. et al. Effect of absent bilateral radial arteries on the accuracy of continuous real-time noninvasive blood pressure monitoring with the Nexfin™ system. Can J Anesth/J Can Anesth 62, 1128–1129 (2015). https://doi.org/10.1007/s12630-015-0413-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12630-015-0413-0