Artifacts caused by insufficient contrast medium filling during C-arm cone-beam CT scans: a phantom study

  • Published:
Radiological Physics and Technology Aims and scope Submit manuscript

Abstract

We investigated artifacts due to late-arriving contrast medium (CM) during C-arm cone-beam computed tomography. We scanned a phantom filled with water or with 100, 50, or 5 % v/v concentrations of CM and then virtually produced CM-delayed projection data by partially replacing the projection images. Artifacts as a function of concentration, percentage of filling time, and size and position of the filling area were assessed. In addition, we used an automatic power injector with different injection delays to inject CM during the scans. A decrease in filling times caused by a lag in CM arrival during the scan resulted in a decrease in pixel values, distortion of the filling area, and appearance of streak artifacts. Even a delay of approximately 20 % in CM arrival in the total scan time resulted in obvious distortion of the filling area. The distortion and streak artifacts tended to worsen at higher CM concentrations. Use of a minimum CM concentration based on the purpose of the examination and constant filling at the target region are effective for avoiding these artifacts.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11

Similar content being viewed by others

References

  1. Siewerdsen JH. Cone-beam CT with a flat-panel detector: from image science to image-guided surgery. Nucl Instrum Methods Phys Res A. 2011;648:S241–50.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  2. Li J, Wan F, Chen G, Ma L, Zhang G, Xu G, Gong J. The utility of angiographic CT in the diagnosis and treatment of neurovascular pathologies in the vicinity of cranial base. Neurosurg Rev. 2010;34:243–8.

    Article  PubMed  Google Scholar 

  3. Patel NV, Gounis MJ, Wakhloo AK, Noordhoek N, Blijd J, Babic D, Takhtani D, Lee SK, Norbash A. Contrast-enhanced angiographic cone-beam CT of cerebrovascular stents: experimental optimization and clinical application. AJNR Am J Neuroradiol. 2011;32:137–44.

    CAS  PubMed  Google Scholar 

  4. Miyayama S, Yamashiro M, Okuda M, Yoshie Y, Nakashima Y, Ikeno H, Orito N, Matsui O. Detection of corona enhancement of hypervascular hepatocellular carcinoma by C-arm dual-phase cone-beam CT during hepatic arteriography. Cardiovasc Intervent Radiol. 2011;34:81–6.

    Article  PubMed  Google Scholar 

  5. Koelblinger C, Schima W, Berger-Kulemann V, Wolf F, Plank C, Weber M, Lammer J. C-arm CT during hepatic arteriography tumour-to-liver contrast: intraindividual comparison of three different contrast media application protocols. Eur Radiol. 2013;23:938–42.

    Article  PubMed  Google Scholar 

  6. Kalender WA, Kyriakou Y. Flat-detector computed tomography (FD-CT). Eur Radiol. 2007;17:2767–79.

    Article  PubMed  Google Scholar 

  7. Marchant TE, Price GJ, Matuszewski BJ, Moore CJ. Reduction of motion artefacts in on-board cone beam CT by warping of projection images. Br J Radiol. 2011;84:251–64.

    Article  CAS  PubMed  Google Scholar 

  8. Bash S, Villablanca JP, Jahan R, Duckwiler G, Tillis M, Kidwell C, Saver J, Sayre J. Intracranial vascular stenosis and occlusive disease: evaluation with CT angiography, MR angiography, and digital subtraction angiography. AJNR Am J Neuroradiol. 2005;26:1012–21.

    PubMed  Google Scholar 

  9. Hosokawa S, Kawai N, Sato M, Minamiguchi H, Nakai M, Nishioku T, Sonomura T, Matsumoto H, Masuo O, Itakura T. Optimal contrast material concentration for distinguishing among carotid artery lumen, carotid stent, and neck in cone-beam computed tomography during carotid angiography: basic and clinical studies. Jpn J Radiol. 2012;30:358–64.

    Article  PubMed  Google Scholar 

  10. Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am. 1984;A1:612–9.

    Article  Google Scholar 

  11. Zellerhoff M, Scholz B, Ruehrnschopf EP, Brunner B. Low contrast 3D reconstruction from C-arm data. Proc SPIE. 2005;5745:646–55.

    Article  Google Scholar 

  12. Hunter AK, McDavid WD. Characterization and correction of cupping effect artefacts in cone beam CT. Dentomaxillofac Radiol. 2012;41:217–23.

    Article  CAS  PubMed  Google Scholar 

  13. Loffroy R, Lin M, Rao P, Bhagat N, Noordhoek N, Radaelli A, Blijd J, Geschwind JF. Comparing the detectability of hepatocellular carcinoma by C-arm dual-phase cone-beam computed tomography during hepatic arteriography with conventional contrast-enhanced magnetic resonance imaging. Cardiovasc Intervent Radiol. 2012;35:97–104.

    Article  PubMed  Google Scholar 

  14. Orth RC, Wallace MJ, Kuo MD. C-arm cone-beam CT: general principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol. 2008;19:814–20.

    Article  PubMed  Google Scholar 

  15. Ernemann UU, Gronewaller E, Duffner FB, Guervit O, Claassen J, Skalej MD. Influence of geometric and hemodynamic parameters on aneurysm visualization during three-dimensional rotational angiography: an in vitro study. AJNR Am J Neuroradiol. 2003;24:597–603.

    PubMed  Google Scholar 

  16. Brown JR, Robb JF, Block CA, Schoolwerth AC, Kaplan AV, O’Connor GT, Solomon RJ, Malenka DJ. Does safe dosing of iodinated contrast prevent contrast-induced acute kidney injury? Circ Cardiovasc Interv. 2010;3:346–50.

    Article  PubMed  Google Scholar 

  17. Ritter M, Weiss C, Rassweiler MC, Michel MS, Hacker A. Optimizing imaging quality in endourology with the Uro Dyna-CT: contrast agent dilution matters. World J Urol. 2012.

Download references

Conflict of interest

The authors declare that they have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Mitsuaki Terabe.

About this article

Cite this article

Terabe, M., Ichikawa, H., Kato, T. et al. Artifacts caused by insufficient contrast medium filling during C-arm cone-beam CT scans: a phantom study. Radiol Phys Technol 7, 25–34 (2014). https://doi.org/10.1007/s12194-013-0227-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12194-013-0227-0

Keywords

Navigation