Abstract
A low-cost disposable was developed for rapid detection of the protein biomarker myoglobin (Myo) as a model analyte. A screen printed electrode was modified with a molecularly imprinted material grafted on a graphite support and incorporated in a matrix composed of poly(vinyl chloride) and the plasticizer o-nitrophenyloctyl ether. The protein-imprinted material (PIM) was produced by growing a reticulated polymer around a protein template. This is followed by radical polymerization of 4-styrenesulfonic acid, 2-aminoethyl methacrylate hydrochloride, and ethylene glycol dimethacrylate. The polymeric layer was then covalently bound to the graphitic support, and Myo was added during the imprinting stage to act as a template. Non-imprinted control materials (CM) were also prepared by omitting the Myo template. Morphological and structural analysis of PIM and CM by FTIR, Raman, and SEM/EDC microscopies confirmed the modification of the graphite support. The analytical performance of the SPE was assessed by square wave voltammetry. The average limit of detection is 0.79 μg of Myo per mL, and the slope is −0.193 ± 0.006 μA per decade. The SPE-CM cannot detect such low levels of Myo but gives a linear response at above 7.2 μg · mL−1, with a slope of −0.719 ± 0.02 μA per decade. Interference studies with hemoglobin, bovine serum albumin, creatinine, and sodium chloride demonstrated good selectivity for Myo. The method was successfully applied to the determination of Myo urine and is conceived to be a promising tool for screening Myo in point-of-care patients with ischemia.
Similar content being viewed by others
References
Braunwald E, Antman EM, Beasley JW, Califf RM, Cheitlin MD, Hochman JS et al (2000) ACC/AHA guidelines for the management of patients with unstable angina and non-ST segment elevation myocardial infarction: executive summary and recommendations. Catheter Cardiovasc Interv 51:505–521
Braunwald E, Antman EM, Beasley JW, Califf RM, Cheitlin MD, Hochman JS et al (2000) ACC/AHA guidelines for the management of patients with unstable angina and non-ST-segment elevation myocardial infarction. J Am Coll Cardiol 36:970–1056
Hing E, Hall MJ, Ashman JJ, Xu J (2010) National hospital ambulatory medical care survey: 2007 outpatient department summary. Nat Health Stat Rep 28:1–32
Butler KH, Swencki SA (2006) Chest pain: a clinical assessment. Radiol Clin N Am 44:165–179
Balk EM, Ioannidis JPA, Salem D, Chew PW, Lau J (2001) Accuracy of biomarkers to diagnose acute cardiac ischemia in the emergency department: a meta-analysis. Ann Emerg Med 37:478–494
Nursalim A, Suryaatmadja M, Panggabean M (2013) Potential clinical application of novel cardiac biomarkers for acute myocardial infarction. Acta Med Indones-Indones J Intern Med 45:240–250
Matveeva EG, Gryczynski Z, Lakowicz JR (2005) Myoglobin immunoassay based on metal particle-enhanced fluorescence. J Immunol Methods 302:26–35
Castaldo AM, Ercolini P, Forino F, Basevi A, Vrenna L, Castaldo P et al (1994) Plasma myoglobin in the early diagnosis of acute myocardial-infarction. Eur J Clin Chem Clin Biochem 32:349–353
Kent L, Ahchean C, Januzzi J (2002) Cardiac markers for myocardial infarction. Am J Clin Pathol 118:S93–S99
Beuerle JR, Azzazy HME, Apple FS et al (2000) Performance characteristics of a new myoglobin microparticle enzyme immunoassay: a multicenter evaluation. Clin Biochem 33:595–598
Graff LG, Dallara J, Ross MA, Joseph AJ, Itzcovitz J, Andelman RP et al (1997) Impact on the care of the emergency department chest pain patient from the chest pain evaluation registry (CHEPER) study. Am J Cardiol 80:563–568
Gosling JP (1990) A decade of development in immunoassay methodology. Clin Chem 36:1408–1427
Jameson DM, Croney JC (2003) Fluorescence polarization: past, present and future. Comb Chem High Throughput Screen 6:167–176
Luppa PB, Sokoll LJ, Chan DW (2001) Immunosensors - principles and applications to clinical chemistry. Clin Chim Acta 314:1–26
Mishra SK, Srivastava AK, Kumar D (2014) Bio-functionalized Pt nanoparticles based electrochemical impedance immunosensor for human cardiac myoglobin. Rsc Adv 4:21267–21276
Dürkop A, Lehmann F, Wolfbeis OS (2002) Polarization immunoassays using reactive ruthenium metal-ligand complexes as luminescent labels. Anal Bioanal Chem 372:688–694
Hayes MA, Petkus MM, Garcia AA et al (2009) Demonstration of sandwich and competitive modulated supraparticle fluoroimmunoassay applied to cardiac protein biomarker myoglobin. Analyst 3:533–541
Cho J-H, Kim M-H, Mok R-S (2014) Two-dimensional paper chromatography-based fluorescent immunosensor for detecting acute myocardial infarction markers. J Chromatogr B-Anal Technol Biomed Life Sci 967:139–146
Kim D-H, Seo S-M, Cho H-M, Hong S-J, Lim D-S, Paek S-H (2014) Continuous immunosensing of myoglobin in human serum as potential companion diagnostics technique. Biosens Bioelectron 62:234–241
Hashimoto M, Tsukagoshi K, Nakajima R, Kondo K (1999) Chemiluminescence detection of heme proteins separated by capillary isoelectric focusing. J Chromatogr A 852:597–601
Liang HR, Scott MK, Murry DJ, Sowinski SM (2001) Determination of albumin and myoglobin in dialysate and ultrafiltrate samples by high-performance size-exclusion chromatography. J Chromatogr B 754:141–151
Shen L, Huang R, Hu NF (2002) Myoglobin in polyacrylamide hydrogel films: direct electrochemistry and electrochemical catalysis. Talanta 56:1131–1139
Shumiantseva VV, Bulko TV, Vagin MI, Suprun EV, Archakov AI (2010) Electrochemical immunoanalysis of cardiac myoglobin. Biomed Khim 56:758–768
Suprun EV, Shilovskaya AL, Lisitsa AV, Bulko TV, Shumyantseva VV, Archakov AI (2011) Electrochemical immunosensor based on metal nanoparticles for cardiac myoglobin detection in human blood plasma. Electroanalysis 23:1051–1057
Moreira FTC, Sharma S, Dutra RAF, Noronha JPC, Cass AEG, Sales MGF (2013) Smart plastic antibody material (SPAM) tailored on disposable screen printed electrodes for protein recognition: application to myoglobin detection. Biosens Bioelectron 45:237–244
Wulff G (2013) Fourty years of molecular imprinting in synthetic polymers: origin, features and perspectives. Microchim Acta 15–16:1359–1370
Moreira FTC, Dutra RAF, Noronha JPC, Sales MGF (2011) Myoglobin-biomimetic electroactive materials made by surface molecular imprinting on silica beads and their use as ionophores in polymeric membranes for potentiometric transduction. Biosens Bioelectron 26:4760–4766
Moreira FTC, Dutra RAF, Noronha JPC, Sales MGF (2013) Electrochemical biosensor based on biomimetic material for myoglobin detection. Electrochim Acta 107:481–487
Pijanowska DG, Torbicz W (2008) Ion selective and semi-permeable membranes for biosensors in biomedical applications. Biocybernetics Biomed Eng 28:11–19
Piletsky SA, Turner APF (2002) Electrochemical sensors based on molecularly imprinted polymers. Electroanalysis 14:317–323
Guerreiro JRL, Sales MGF, Moreira FTC, Rebelo TSR (2011) Selective recognition in potentiometric transduction of amoxicillin by molecularly imprinted materials. Eur Food Res Technol 232:39–50
Hong L, Chen NP (2000) Proton-conducting polymer membrane based on sulfonated polystyrene microspheres and an amphiphilic polymer blend. J Pol Sci Part B-Pol Phys 38:1530–1538
Kudin KN, Ozbas B, Schniepp HC, Prud’homme RK, Aksay IA, Car R (2008) Raman spectra of graphite oxide and functionalized graphene sheets. Nano Lett 8:36–41
Miles AB, Compton RG (2000) The theory of square wave voltammetry at uniformly accessible hydrodynamic electrodes. J Electroanal Chem 487:75–89
Acknowledgments
One of the authors (FTCM) gratefully acknowledges Fundação para a Ciência e Technologia for the financial support (PhD grant reference SFRH/BD/66735/2009).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Moreira, F.T.C., Sharma, S., Dutra, R.A.F. et al. Detection of cardiac biomarker proteins using a disposable based on a molecularly imprinted polymer grafted onto graphite. Microchim Acta 182, 975–983 (2015). https://doi.org/10.1007/s00604-014-1409-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00604-014-1409-0