Abstract
Comprehensive two-dimensional gas chromatography with time-of-flight mass spectrometry (GC × GC-TOFMS) is one the most powerful analytical platforms for chemical investigations of complex biological samples. It produces large datasets that are rich in information, but highly complex, and its consistency may be affected by random systemic fluctuations and/or changes in the experimental parameters. This study details the optimization of a data processing strategy that compensates for severe 2D pattern misalignments and detector response fluctuations for saliva samples analyzed across 2 years. The strategy was trained on two batches: one with samples from healthy subjects who had undergone dietary intervention with high/low-Maillard reaction products (dataset A), and the second from healthy/unhealthy obese individuals (dataset B). The combined untargeted and targeted pattern recognition algorithm (i.e., UT fingerprinting) was tuned for key process parameters, the signal-to-noise ratio (S/N), and MS spectrum similarity thresholds, and then tested for the best transform function (global or local, affine or low-degree polynomial) for pattern realignment in the temporal domain. Reliable peak detection achieved its best performance, computed as % of false negative/positive matches, with a S/N threshold of 50 and spectral similarity direct match factor (DMF) of 700. Cross-alignment of bi-dimensional (2D) peaks in the temporal domain was fully effective with a supervised operation including multiple centroids (reference peaks) and a match-and-transform strategy using affine functions. Regarding the performance-derived response fluctuations, the most promising strategy for cross-comparative analysis and data fusion included the mass spectral total useful signal (MSTUS) approach followed by Z-score normalization on the resulting matrix.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The use of saliva in clinical diagnostics is historically under-explored compared to other bio-fluids such as urine or blood. However, emerging biotechnologies and salivary diagnostics have extended the range of saliva-based diagnostics from the oral cavity, such as for periodontal diseases and caries risk, to the physiological status, due to its biological equilibrium with plasma [1]. It has been shown that saliva metabolites closely resemble metabolic changes that take place in blood and may therefore reflect a variety of pathophysiological and nutritional changes, as well as exposure to medications and environmental factors [2,3,4,5].
The -omic analytical strategies aim to capture the variety of encrypted information of complex samples to enable to the higher level information related to the phenomena under study. Well-established methodologies in metabolomics, i.e., profiling and fingerprinting, have been developed as separate analytical techniques/approaches capable of informing about compositional differences between samples [6, 7]. If analytes of interest are identified upfront and tracked across all samples, profiling can be done specifically. However, the procedure can be extended toward a thorough evaluation of all detected components and referred to as untargeted profiling [8]. Fingerprinting, on the other hand, is a high-throughput approach that may identify compositional changes between samples [9,10,11]; however, it may not always produce precise quantitative data or provide analyte identification for all of the constituents. The goal of fingerprinting is to extract the non-obvious chemical information present in the entire instrumental signal. Chemometrics and multivariate analysis allow then the access to higher level information and understanding. In the case of comprehensive two-dimensional chromatography, fingerprinting acquires a new meaning: it is referred to as chromatographic fingerprinting and corresponds to a pattern recognition procedure where detected features are annotated and tracked across many samples based on their relative retention and spectral similarity [6].
Comprehensive two-dimensional gas chromatography coupled with time-of-flight mass spectrometry (GC × GC-TOFMS) is one of the most informative analytical platforms for complex sample analysis. It allows detailed profiling and high-resolution chromatographic fingerprinting of biofluids thanks to the high separation power, chromatographic resolution, and sensitivity. The benefits are achieved by combining two separation dimensions interfaced by a thermal modulator, which allows for an efficient compression in space of the injection band before the 2D separation [12,13,14]. However, to implement the fingerprinting approach in long-term studies, pattern realignment strategies and response normalization are needed; otherwise, data acquired with different analytical setups and/or with variable detection performances cannot be cross-compared [6, 15, 16].
This is crucial because biobanks—a term we use broadly to cover a range of structured collections, including biorepositories and databases—have become central engines of biomedical research and there are a growing number of biobanks built by directly collecting data and samples from a population or subpopulation to assemble a large-scale research resource. The re-evaluation and reanalysis of archived samples have become inevitable, as technology, bioinformatics pipelines, and medical knowledge evolve and expand over time. Both the issues, normalization of samples from different subpopulations or collection kinetics and the need for sample reanalysis, are pressing concerns, which deserve better elucidation.
In the challenging context of 2D pattern realignment between datasets, the smart template concept [13, 17] based on pattern recognition algorithms was demonstrated to be highly reliable and effective. The smart template method was developed to align 2D peaks and peak regions across many samples using a reference pattern (i.e., the template). It employs rule-based constraints (e.g., retention times windows and multispectral matching) to increase matching accuracy. Each reference peak in a smart template has constraint rules, including the spectral similarity value (NIST algorithm [18]) vs. the reference peak spectral signature, as well as arithmetic and logical operators applied to the MS spectrum. Constraints could be related to the relative intensity of a single or multiple m/z fragments, and to the identification of the base peak or molecular peak [19]. If constraint rules are verified, peaks and peak regions from the reference pattern (template) are realigned to putative candidates in an analyzed pattern. Correspondences are established, thanks to the pattern transform, even in the presence of retention times shifts or temporal inconsistencies [6, 16, 20].
In particular, if misalignments are induced by changes in oven temperature programming, they can be successfully compensated by an affine transformation, an algorithm for template matching suitable for intra-batch cross-comparative analysis [17]. On the other hand, larger misalignments such as those generating non-linear changes in retention times across the analyses (i.e., variations in the actual pressure drop [16, 21, 22], adoption of different modulation principles [23, 24], changes in the modulation period PM [16, 25]) can be successfully treated by global, low-degree polynomial transforms [17, 22, 26].
However, in a scenario of random pattern misalignments, generated by the combination of multiple concurrent variables (i.e., pressure drop, column set-up, PM), supervision and manual operations may be necessary to guide the algorithm-based realignment. The primary goal of this study was to define and validate a simple, yet effective, strategy to guide template matching algorithms for a combined untargeted and targeted (UT) fingerprinting [8, 27,28,29] exploration on datasets with severe misalignments. In addition, as a secondary objective, different response normalization methods were tested to allow for effective dataset comparative analysis. As a challenging bench test, the saliva metabolome was selected. The two salivary datasets considered were acquired in a 2-year time frame with different analytical setups. Dataset “A,” acquired in February 2020, relates to a proof-of-concept diet intervention study with food rich in Maillard reaction products (MRPs) [30], while dataset “B,” acquired in November 2017, relates to a study on two different populations of subjects with severe obesity with normal or altered metabolic parameters [14].
Materials and methods
Reference compounds and solvents
A pure standard solution of n‐alkanes (from n‐C7 to n‐C30) for linear retention indices (IT) calibration and system quality control was from Merck (Milan, Italy) and prepared in toluene at the concentration of 100 mg L−1.
The internal standard (IS) 1,4-dibromobenzene (from Merck, Milan, Italy) solution was prepared in toluene at a concentration of 10 g L−1.
Pure reference standards used for identity confirmation of pyruvic acid, lactic acid, malonic acid, succinic acid, malic acid, 2-ketoglutaric acid, 3-hydroxybutyric acid, fumaric acid, 2-keto-3-metilvaleric acid, aspartic acid, hippuric acid, citric acid, uric acid, l-alanine, l-valine, l-leucine, l-proline, glycine, l-threonine, l-tyrosine, l-phenylalanine, l-isoleucine, l-methionine, l-cysteine, l-ornithine, l-tryptophan, xylitol, ribitol, fructose, galactose, glucose, mannitol, myo-inositol, glycerol, palmitic acid, stearic acid, and creatinine were from Merck (Milan, Italy).
Derivatizing agents O-methyl hydroxylamine hydrochloride (MOX), N,O-bis(trimethylsilyl)trifluoroacetamide (BSTFA), and LC-grade pyridine, n-hexane, dichloromethane, and toluene used as solvents were all from Merck (Milan, Italy).
Saliva samples
Subjects for dataset A were metabolically healthy German volunteers (3 females and 2 males) aged between 20 and 30, who ate unheated food virtually free of Maillard reaction products—MRPs (e.g., mainly vegetables, fruits, oils, and unroasted nuts) for 4 days. Sampling was performed from day 1 until day 5 in the morning and at 2 pm as well as 9 pm. The same group of subjects collected samples for another 5 days while eating their habitual diets and additionally including MRP-rich food. Fasting saliva was collected before breakfast and after brushing the teeth without using toothpaste and rinsing the mouth with water; quality control (QC) samples were collected from healthy Italian volunteers aged between 20 and 30 without dietary indications. Saliva was collected using Salivettes™ (Sarstedt, Germany), but the protocol was adapted to avoid stimulation: the device was placed at the center of the tongue and the collection time was 3 min; samples were stored at − 18 °C until the time of testing. The study was approved by the Ethics Committee of Technische Universität Dresden, Germany (reference: AZ 439112017). Details on the study are described in the reference paper by Manig et al. [31].
Patients from the Istituto Auxologico Italiano in Verbania, Italy, were recruited for dataset B [32, 33]. As part of the routine controls, individuals’ height, weight, and waist circumference were evaluated. For the current study, participants with a body mass index (BMI) of at least 40 kg m−2 were enrolled. Thirty-four obese men (BMI 40 kg m−2) were classified as either metabolically healthy (MHO, n = 10) or metabolically unhealthy (MUO, n = 24), depending on whether certain metabolic parameters, such as high fasting triglycerides (1.7 mmol L−1 or higher, 150 mg dL−1), decreased HDL cholesterol (1.03 mmol L−1, 40 mg dL−1), whether they were taking antihypertensive medication, had high blood pressure (130 mmHg systolic or 85 mmHg diastolic), or had fasting glucose levels above 5.6 mmol/L (100 mg dL−1). QC samples were collected from healthy Italian volunteers aged between 20 and 30 without dietary indications. The experimental procedure was approved by the ad hoc Ethical Research Committee of the Istituto Auxologico Italiano (Verbania, Italy). Written informed consent was obtained from the patients. The study protocol conformed to the guidelines of the European Convention on Human Rights and Biomedicine concerning biomedical research. Details on the study are described in the reference paper by Cialiè Rosso et al. [14].
Sample preparation
A standard derivatization protocol [34] was adjusted to comply with method sensitivity and metabolite coverage. In particular, 100 μL of saliva was gently dried with nitrogen before being mixed with 25 μL of MOX (20 g L−1 in pyridine) and allowed to react for 2 h at 60 °C preventing the formation of multiple derivatives with enols during further silylation steps. Seventy-five microliters of BSTFA was added and the solution was left at 60 °C for 1 h under a nitrogen stream. Twenty microliters of 1,4-dibromobenzene in dichloromethane 1 g L−1 was added as IS, and diluted in 80 μL of toluene to a final volume of 200 μL. Samples were immediately stored in − 18 °C and analyzed within 24 h after derivatization.
GC × GC-TOFMS instrument setup and conditions
GC × GC analyses were performed on an Agilent 7890 GC chromatograph (Agilent Technologies, Wilmington DE, USA) coupled to a Markes BenchTOF Select™ mass spectrometer featuring tandem ionization (Markes International, Llantrisant, UK). The system was equipped with a two-stage KT 2004 loop-type thermal modulator (Zoex Corporation, Houston, TX) cooled with liquid nitrogen controlled by Optimode v2.0 (SRA Intruments, Cernusco sul Naviglio, Milan, Italy).
Column settings and operative conditions for dataset A were as follows: 1D DB5 (95% polydimethylsiloxane, 5% phenyl; 30 m, 0.25 mm dc, 0.25 μm df), 2D OV1701 (86% polydimethylsiloxane, 7% phenyl, 7% cyanopropyl; 1.3 m × 0.1 mm dc, 0.10 μm df) from J&W (Agilent, Little Falls, DE, USA). The first 0.80 m of the 2D column, connected in series to the 1D column by a silTite μ-union (Trajan Scientific and Medical, Ringwood, Victoria, Australia), was wrapped in the modulator slit and used as loop-capillary for cryogenic modulation. The carrier gas was helium at 1.3 mL min−1 in constant flow mode. PM was 3.0 s operating in multi-step mode: 0–15 min. The hot jet pulse time duration was 250 ms during the first 15 min, and 350 ms during the period 15–63 min. The cold jet flow was programmed for a linear decrease from 35% of the mass flow controller (MFC) maximum flow (i.e., 40 L min−1) to 5% at the end of the run. The injector temperature was kept at 280 °C operating in split mode with a split ratio: 1:20. The oven temperature ramp was 60 °C (2′) to 120 °C at 10 °C min−1, then to 300 °C (10′) at 4 °C min−1; the injection volume was 2 μL.
The TOFMS acquisition parameters were as follows: tandem ionization™ acquisition at 70 and 12 eV with an acquisition rate of 50 Hz per channel within the mass range 35–750 m/z; the filament voltage was set at 1.7 V. The ion source and transfer line were set at 280 °C and 290 °C, respectively.
Column settings and operative conditions for dataset B were as follows: 1D DB5 (95% polydimethylsiloxane, 5% phenyl; 30 m, 0.25 mm dc, 0.25 μm df), 2D OV1701 (86% polydimethylsiloxane, 7% phenyl, 7% cyanopropyl; 2 m × 0.1 mm dc, 0.10 μm df) from J&W (Agilent, Little Falls, DE, USA). The first 0.80 m of the 2D column, connected in series to the 1D column by a silTite μ-union (Trajan Scientific and Medical, Ringwood, Victoria, Australia), was wrapped in the modulator slit and used as loop-capillary for cryogenic modulation. The carrier gas was helium at 1.6 mL/min—constant flow. PM was 5.0 s. The hot jet pulse time was 350 ms; the cold jet flow was progressively reduced with a linear function from 30% of MFC at initial conditions to 8% at the end of the run. The injector temperature was kept at 300 °C operating in split mode with a split ratio: 1:20. The oven temperature ramp was as follows: 70 °C (2′) to 120 °C at 10 °C min−1, then to 320 °C (1′) at 4 °C min−1; the injection volume was 2 μL.
TOFMS acquisition parameters were as follows: tandem ionization™ at 70 and 12 eV with an acquisition rate of 50 Hz per channel and a mass range 45–1000 m/z; the filament voltage was set at 1.6 V. The ion source and transfer line were both set at 290 °C respectively.
UT fingerprinting workflow principles and parameters
A template is a pattern of 2D peaks and/or graphic objects that is built over a reference image(s) (single or cumulative image [35]) and then used to recognize similar patterns of 2D peaks in an analyzed image(s) [36]. Dedicated matching functions or transforms guide this process while effectively compensating for inconsistent retention times and variable peak detection [29]. Each template object (2D-peak/blob or peak region/area graphic) can carry various metadata including compound chemical name, retention times, mass spectra, informative ions (qualifier and quantifier ions) and their relative ratios, constraint functions to limit peak correspondences above certain thresholds, and qualifier functions.
The challenging task of multi-chromatogram fingerprinting in the presence of temporal inconsistencies and detector fluctuations was addressed successfully by introducing the concept of peak region features [8]. Peak regions attempt to define one chromatographic region around each individual peak thereby achieving the one-feature-to-one-analyte selectivity that is the goal of peak features approaches [8, 13], but with an implicit matching of regional features and greater robustness than can be achieved with peak detection [29]. 2D peaks and peak-regions are features adopted in the UT fingerprinting strategy [27, 35, 37, 38].
UT fingerprinting establishes a set of reliable peaks, positively matched across all or most of a set of chromatograms [39], and then uses them to align chromatograms [22] for their combination into a single, composite chromatogram. The algorithm for the determination of reliable peaks is the Consistent Cliques Method (CCM) [39] algorithm: it picks peaks that are consistently matched across several chromatograms, but possibly not all of them. The operator can define the most appropriate minimum threshold for the selection of reliable peaks: either peaks that are pairwise matched across all but one (most constrained) chromatogram or peaks pairwise matched in 50% + 1 (most relaxed) of the processed chromatograms. The first option will result in a fewer number but more reliable peaks, while the latter will result in a larger number but less reliable peaks. The choice of the minimum threshold is driven by the goal of the study: if realignment is the aim, the issue of a small number of reliable peaks might not be a concern as long as enough reliable peaks are found; but if the comparative analysis is the goal, certain important peaks might be lost together with the information they carry. The most constrained option does not work well for large sample sets, since even for very consistent features, the risk of a failed match increases exponentially as the number of chromatograms increases.
Once the composite chromatogram is built (i.e., the combination of the realigned responses in the 2D retention-times plane), 2D peaks are detected and their outlines are recorded to define peak-region objects. The set of reliable 2D peaks and peak-regions objects are then collected in a so-called feature template, or consensus template, covering the whole sample-set compositional diversity and capable of cross-aligning chemical feature patterns among samples [6].
Within all detected analytes (reliably matched or not over the chromatogram set), the sub-group of targeted compounds can be highlighted by completing their metadata fields (compound name, ion ratios, ITS) and computed together with untargeted features during the data processing.
Untargeted features comprehensively map the sample’s chemical dimensionality [40] and are automatically generated through a dedicated workflow. On the other hand, targeted features require supervised processing for the reliable identification of analytes and the definition of specific ions for response isolation and/or informative ion ratios to add constraints to the template matching process.
Identification of targeted features was carried out by matching candidate EI-MS fragmentation patterns at 70 eV (NIST MS Search algorithm, version 2.0) with those collected in commercial and in-house databases (subject to a DMF threshold of 900 and RMF threshold of 950). In addition, linear retention indices (IT) were adopted as a further constraint; experimental values were compared with NIST reference indices using a tolerance of ± 10 units.
Data acquisition and processing
Data were acquired by TOF-DS software (Markes International, Llantrisant, UK) and processed using GC Image GC × GC Software ver. 2021r2 (GC Image, LLC, Lincoln, NE, USA). Data mining was performed using Matlab R2021a (The MathWorks, Inc., Natick, MA, USA) with the following packages: PCA toolbox (v1.5) [41] and Classification toolbox (v6.0) [42], and XLSTAT 2014 by Addinsoft (New York, USA).
Results and discussion
This section illustrates the step-by-step procedure (i.e., workflow) adopted to evaluate the impact of key-processing parameters on template matching accuracy and template transformation in a context of severe misalignment. At first, template construction parameters are tested for their impact on matching accuracy (type-I and type-II errors; “Template construction parameters” section). Subsequently, the combined untargeted targeted (UT) template is applied to dataset A exploring saliva metabolite patterns correlated to a diet rich in MRPs (“Dataset A—Saliva samples from diet intervention with Maillard reaction products–rich food” section). As the third step, the UT template built on dataset A is transformed to match on dataset B retention pattern logic (“Template transformation for reliable template matching between severely misaligned patterns” section). Realigned datasets are then fused after suitable response normalization and further mined to explore the possibility of reliable re-investigation of data after their fusion (“Response normalization for consistent cross-comparison between datasets” section).
Template construction parameters
The best combination of processing parameters to build a reliable template with consistent reference metadata (i.e., template object metadata) was first evaluated in light of its adoption as a key-tool for pattern realignment, even in a context of dramatic chromatographic shifts.
Signal-to-noise ratio (S/N) thresholds, which is the ratio of the total intensity count (TIC) of the apex spectrum to the standard deviation of background noise TIC [43], and MS similarity threshold, expressed as NIST similarity match factor (direct match factor, DMF) [18], were first considered since both contribute to the matching specificity. The S/N threshold allows the exclusion of 2D peaks with spectral information deteriorated by the noise contribution, while the MS similarity threshold applies additional constraints to retention times alignment and improves matching specificity.
To achieve the highest consistency in cross-alignment, a sub-group of analytes within those detected in the saliva samples was selected. For low-intensity and trace peaks, in particular, spectral quality fluctuations were hypothesized based on the variances in the absolute response [44]. Saliva 2D contour plots were carefully inspected and candidate 2D peaks with suitable characteristics were selected and grouped into three classes as a function of their normalized responses (i.e., TIC current/detector response normalized to that of the IS). Analytes used for performance evaluation are highlighted in an exemplary 2D contour plot illustrated in Fig. 1. The class “high” includes 2D peaks with a normalized response above 0.3, for “medium” it was between 0.15 and 0.3, and for “low” it was below 0.15.
Inconsistent MS spectral fingerprints may result in false negative matches (i.e., type-II errors when a peak that is present is not matched) for 2D peaks having S/N values lower than 50, as shown by Stilo et al. [16] since when it comes to these peaks, neither the reference nor the peak spectra are consistent enough to convey reliable information for identity confirmation. Regarding the spectral similarity threshold, a total of 20 combinations of processing parameters were tested and the number of false positive (i.e., type-I error when a peak that is not present is matched) and false negative (type-II error) matches were computed. The range of variation for S/N threshold was between 0, thus without any filter on the peak spectral signature, and 200, which was the maximum value still enabling the detection of low-intensity analytes. S/N was arbitrarily step-wise varied by 50 counts within the range. Spectral similarity, by DMF, was step-wise varied by 100 counts between the lowest value of 600 and the highest 900. The range was chosen based on previous studies [9, 16] which indicated that template matching accuracy for targeted peaks achieves its maximum with DMF around 750 if accompanied by S/N threshold values for reference template peaks above 100. In the current dataset, due to the different MS analyzers used (i.e., a TOFMS vs. qMS adopted in the Kiefl et al. [9]), the range 600–900 has been set to check the actual accuracy achievable by TOFMS detection.
Results are reported in Table 1: with DMF between 800 and 900, a significantly higher number of false negative (FN) matches occurred; type-II error mainly affects medium and low-intensity analytes, as expected. All the combinations with DMF of 800 and 900 were therefore excluded. At the same time, using a DMF of 600 increased the number of false positive (FP) matches (type-I errors), thus leading to the exclusion of such combinations of parameters. By focusing on the five combinations adopting a DMF of 700, the ones with S/N thresholds of 100, 150, and 200 were excluded because of the high rate of FN matches, of 6%, 11%, and 15%, respectively. The same results were achieved with “DMF 700 & S/N 0” and “DMF 700 & S/N 50”.
Results comply with previous studies and related pieces of evidence on the crucial role of processing parameters in the construction of a reliable template [16, 22, 43]. However, in the current investigation, spanning a wider dynamic range of responses, the challenge posed by trace and sub-trace peaks (i.e., “low” group of tested analytes) was effectively handled by lowering the S/N threshold. To note, a S/N of 0 would result in the detection of too many 2D peaks in noisy regions because of false detections. Based on the outcome of this exploration, the combination of S/N threshold of 50 and DMF of 700 was chosen to further process saliva samples’ raw data and to build a reliable template for the effective realignment of datasets.
Dataset A—Saliva samples from diet intervention with Maillard reaction products–rich food
Optimized data processing parameters from the “Template construction parameters” section were used to generate a reliable template, thus created with peaks present in 50% + 1 of the chromatograms (n = 25). The list of reliable peaks (including both targeted and untargeted peak features) is reported in Table 2. The total number of reliable peaks observed was 68, 43 of which were putatively identified with the aid of IT and spectrum similarity, 7 were recognized as carbohydrates but their identity was not univocally assigned, while 18 could not be associated with confidence for any putative identity.
Unsupervised and supervised statistics were conducted on both reliable peaks and peak regions after normalization and scaling of the absolute response to find distinctive patterns of metabolites associated with diet intervention with MRPs-rich food. The PCA on reliable peaks (25 samples × 68 variables)—score plot shown in Fig. 2A—explained the differences between the QC samples and the saliva samples from MRPs-rich/low diets. Even though no natural clusterization was evident between the two conditions under study, a PLS-DA was conducted to maximize the differences for descriptive modeling as well as for discriminative variable selection—score plot shown in Fig. 2B. The latter objective was achieved by looking at the variable importance in the projection (VIP) scores, and by considering discriminating variables as those with a VIPs > 1. Glutamic acid, citric acid, aspartic acid, hydroxyglutaric acid, butanoic acid, oxoproline, stearic acid, and palmitic acid were identified as the most discriminant, together with five carbohydrates and 7 unknowns (i.e., also referred to as knowns unknowns according to Stilo et al. [45])—Table 2. The same approach was conducted with peak regions with comparable results.
Both stearic and palmitic acids were previously identified as possible markers of dietary lipid intake and correlated with obese patients affected by hepatic steatosis [46, 47] and showed a relative increase in diabetic patients and periodontal diseases [48]. However, the intake of the high-MRP diet was accompanied by a higher intake of fats. Butyric acid, a locally produced metabolite of pathogenic periodontal bacteria, has been shown to cause caspase-dependent apoptosis in gingival fibroblasts and could potentially operate as a chemotaxonomic indicator of bacterial metabolism in the oral environment [49]. Five non-identified monosaccharides were typical of high-MRP diets, as expected. The non-identified monosaccharides leave space for speculations. Beside MRPs, the high-MRP diet contained more sugar and starch as expected from the food questionnaires. Structural similar sugar metabolites such as 3-deoxyglucosone (3-DG) or other dicarbonyl compounds may be taken into account. The presence of carbonyls, in particular the reactive dicarbonyl species, e.g., methylglyoxal, reacts with amino compounds of endogenous tissues in the body, leading to that what is known as “carbonyl stress,” which causes an increase in protein and DNA alteration that contribute to cell and tissue malfunction in aging and illness [50, 51]. The role of dietary MRPs in this scenario is not yet clear.
Template transformation for reliable template matching between severely misaligned patterns
Experimental settings between dataset A and dataset B create large pattern differences with inconsistent misalignments due to the concurrent effect of several parameters: (a) PM 3 s vs. 5 s, (b) different column lengths/dimensions, (c) carrier gas nominal flow differences (1.3 mL/min vs. 1.6 mL/min) and variable average velocities (19.0 cm/s and 201.9 cm/s vs. 16.3 cm/s and 164.9 cm/s for 1D and 2D, respectively), (d) oven temperature programming, and (e) MS detector optimization. Figure 1A, B shows the contour plot of saliva metabolites from dataset A (Fig. 1A) and dataset B (Fig. 1B). Highlighted compounds are those considered for the template parameters optimization (the “Template construction parameters” section).
To guide the correct strategy for template transform, e.g., global or local, affine or low-degree polynomial function, pattern discrepancies generated by the two experimental setups were computed. To simplify the process, the reference standard mix including selected amino acids, organic acids, carbohydrates, and additional salivary metabolites, analyzed within datasets A and B, was cross-aligned in the temporal domain.
Figure 3 shows the pronounced, non-linear 1D and 2D retention time shifts generated by the experimental conditions. In particular, analytes are represented in the Cartesian space corresponding to the 1D and 2D retention, with relative position over two reference peaks. Serine 3TMS derivative, which elutes in the middle retention times region of the chromatographic plane, was assigned as a centroid, while stearic acid, the last-eluting analyte, was used to normalize the 1D relative position [38, 52]. 1D and 2D relative retention (RR) values were calculated by Eqs. 1 and 2:
where \({}^{\mathit1}t_{\mathit R\mathit i}\) corresponds to the 1D retention time expressed in minutes for target peak i, \({}^{\mathit2}{\mathit t}_{\mathit R\mathit i}\) corresponds to the 2D retention time expressed in seconds for the target peak i, and PM is the modulation period. The two patterns showed dramatic misalignments mainly due to changes (a), (b), (c), (d), which led to opposite elution order in both 1D and 2D across the first and third quadrant for compounds belonging to the same chemical class for both amino acids and organic acids respectively indicated in red and blue.
To define the best strategy to solve the misalignment issue, QC samples from the two datasets were compared. Rules for template peak thresholds and reference spectra were then applied to build a reference targeted template for known analytes. By examining patterns from dataset A, a template of 52 2D peaks was created; reference peaks inclusion was restricted to analytes with an S/N of 50 or higher; reference spectra were taken from peak spectrum; the MS constraint was set at 700 DMF and 700 reverse match factor (RMF).
Different template transformation strategies, within those available in the software, were tested to solve the significant misalignment between patterns on datasets A and B. They are (a) local match-and-transform algorithm which applies the affine transform function to locally realign template peaks to candidate peaks in the chromatogram; (b) translation-and-scale algorithm which applies the affine transform-and-rescale according to positively matched peaks; (c) global polynomial transform which applies 2nd order degree polynomial functions for template transformation; and (d) supervised multi-centroid transformation which includes the analyst supervision guiding template subdivision and local translation followed by a local affine transformation.
To obtain a genuine representation of the real-world application, instead of evaluating the potential reliable template transformations over the standard mix of amino acids, organic acids, carbohydrates, and selected salivary metabolites, QC samples from dataset B, consisting of saliva collected from healthy Italian volunteers, were used. This allows assessment of the quality of the transformation in a chromatographic plane where the complexity of a real-world sample consists of a greater number of 2D peaks and interferents, a condition that would not be considered if a standard mixture was used.
The targeted template obtained from dataset A with the parameters evaluated in the “Template construction parameters” section was applied over four QC analyses from dataset B (i.e., REF1, REF2, REF3, and REF4). A graphical schematization of the workflow is illustrated in Figure SF1.
Strategy 1 involved the following procedure: the template was linearly transformed on the 2D to realign the IS peak (1,4-dibromobenzene). Iterating the process of matching the template and transform it based on the matched locations allows the template to be tailored to the new pattern. This procedure gradually increases the number of available alignment points, enhancing the quality of the global template transformation at every stage up to a plateau. For instance, with sample REF1 the first matching ratio resulted in 46.15% of positive matches corresponding to 24/52 matched peaks. After reiterating the transform and match for 3 additional times, the total number of matched peaks resulted 28/52 corresponding to 53.84% positive matches. Strategy 2 is similar to 1, but includes exponential scaling transformations, while strategy 3 includes polynomial ones. The latter was effective in solving complex realignment issues like with translated methods from thermal to flow modulated platforms and showed positive results when realigning chromatograms acquired with different operative pressures [16, 22].
Strategy 4 involves the following steps: (a) split the reference template into multiple subsections, (b) identify a reliable peak with high intensity for each subsection as a centroid, (c) manually realign the centroids, and (d) reiterate the match-and-transform procedure until a plateau. To realign the two datasets, 1,4-dibromobenzene, proline 2TMS, oxoproline 2TMS, and palmitic acid were chosen as local realignment points. Template matching results for the four approaches are summarized in Table 3. Results with a template consisting of only reliable peaks showed that the manual multi-centroid transformation (strategy 4) performed equally to the global polynomial transformation (strategy 3), both superior to the local match-and-transform (strategy 1) and the affine algorithms (strategy 2). The main shortcoming of reiterated global polynomial transformation is the possible deformation of the peak regions template on the chromatographic plane as a result of the overfitting, thus making it the optimal solution with templates constituted by only reliable peaks.
To cross-aligning datasets A and B, strategy 4 was successfully adopted with the following template transformation settings: 0–14.5 min was shifted of − 18 datapoints (dp) on 1D and 0 dp on 2D, 14.5–21.5 min was shifted of − 13 dp on 1D and 0 dp on 2D, 21.5–34 min was shifted of 0 dp on 1D and − 6 dp on 2D, and 34–66 min was shifted of + 11 dp on 1D and − 16 dp on 2D. Afterward, strategy 2 was adopted to finely adapt the subregions; the transformation described as strategy 4 is shown on the testing template in Fig. 4A, and the final template applied over a test sample is illustrated in Fig. 4B.
Response normalization for consistent cross-comparison between datasets
The reduction of features into a similar range, known as data normalization, is a crucial pre-processing step that prevents larger numeric feature responses from dominating smaller numeric ones. The major objective is to reduce the bias of those features in pattern classes that contribute numerically more than others, so that the variables are given equal weight when computing statistical analysis but still avoiding the implication that these features are equally relevant [53, 54].
Many strategies are available to model the distribution of the original values into a new set of values. Within these, the most common mathematical ones consist of min–max normalization, Z-score normalization, and logarithmic transformation. The min–max normalization rescales the data within a new minimum and maximum value preventing highly concentrated peaks from dominating the other ones in a dataset with a significant min–max difference [55]. With Z-score normalization, each variable’s mean is subtracted from the data values individually and followed by dividing data values by standard deviation to give the variables a variance of one [56]; compared to min–max normalization, Z-score normalization is less prone to create bias deriving from outliers. The latter, i.e., logarithmic transformation, consists of replacing each value in the dataset with its logarithm, thus allowing to make skewed distributions more symmetric [57].
Besides mathematical approaches, such as those listed above, scaling transformation strategies can be also applied to the raw response data. In the mass spectral total useful signal (MSTUS) [58], for example, each variable is normalized over a selection of ions/features (e.g., total volume of reliable peaks and peak regions features) to get a more accurate normalization by avoiding the inclusion of noise or artifact peaks [59, 60].
Moreover, before data mining, missing values can be replaced with random values of at least half of the smallest response since they are typically caused by the signal intensity being below the LOD; the replacement with random values instead of 0 or a fixed number is generally performed to avoid bias-introduction for further unsupervised and supervised statistical approaches.
Absolute response and background noise intensity are directly impacted by MS detector response oscillations caused by MS tuning, optimization, or other causes. During this study, datasets A and B were acquired in a 2-year time frame; thus, differences in terms of MS performances are expected. Output signals exhibited different absolute TIC responses (i.e., different actual sensitivity) and background noise intensities. MS performances between the two datasets were evaluated by comparing two QC analyses picked from those collected during the two analytical batches. Background noise was sampled in the middle of the chromatogram and reported an average intensity of 8484 counts (1D RSD% = 2.91%, 2D RSD% = 1.77%) for the dataset A and 24,522 counts (1D RSD% = 22.16%, 2D RSD% = 8.86%) for dataset B before any background correction was applied. After the baseline removal, the noise intensity averaged 127 and 2000 counts, respectively. Experimental results indicate that dataset B had a greater absolute noise (2.9 times) compared to dataset A while background noise correction greatly affected dataset A reducing the noise signal by 66.5 times compared to the 12.26 times of the dataset B. The number of detected peaks in dataset A with S/N > 50 was 137, compared to 498 in dataset B, indicating a higher relative sensitivity of the method in the second case.
Different combinations of data pretreatment were performed on the fused dataset (A + B), after which peaks and peak region features were mined with unsupervised statistics (PCA). Results are illustrated in Fig. 5. Figure 5A shows the combination of MSTUS normalized responses on the included 2D peaks volume with Z-score normalization on the resulting matrix. Although no differences arise between the healthy and unhealthy obese patients from dataset B and the diets high and low in MRPs foods from dataset A, the two datasets are considered comparable given the realignment of the QC samples obtained from the two datasets. On the other hand, Fig. 5B shows the issue arising by omitting the Z-score normalization after MSTUS normalization: no clear clustering is achieved due to an imbalanced min–max distribution of the data matrix. Figure 5C and D both show the necessity in metabolomics studies to apply MSTUS normalization when comparing different datasets: despite having different approaches (raw responses and logarithmically scaled responses respectively), comparison between the samples cannot be performed as the detector performance fluctuations heavily affect the quality of the data. It can be concluded that such issues/inconsistencies can be only overcome with MSTUS normalization.
The hierarchical clustering (HC) of the fused dataset obtained on MSTUS normalized responses on the included blob volume is illustrated in Fig. 6 (data available on request—see “Data availability”). Interesting biological outcomes can be retrieved from this comparison: saliva samples from QC samples were richer (higher relative distribution) in amino acids such as oxoproline, aspartic acid, and glutamic acid, which were found to be discriminant variables characteristic of the low MRP diet group, as a witness of the different eating habits of the two populations. On the other end, both metabolically healthy and unhealthy obese patients were clustered together and characterized by higher levels of butyric acid and hydroxybutyric acid, which were previously confirmed as obesity markers detected in both urine and blood [61]. Further investigations on dataset 2 were already carried out by Cialiè Rosso and coworkers [14] and will not be further discussed here.
Conclusions
This paper examined 2D data processing strategies based on pattern recognition algorithms, which are capable of overcoming 2D pattern severe misalignments and detector response fluctuations generated in analyses acquired within wide time frames. The strategy described in the current paper allows the creation of a template capable to adapt to different datasets without the necessity to reiterate the preprocessing of the additional batches and to preserve the information encrypted in the metadata (i.e., reliable peaks) to extract biological information once the data fusion is performed. Focusing on the chromatographic aspect, severe misalignments derived by the concurring effect of different analytical setups (different nominal flows and carrier velocities, variations in the actual pressure drop, column dimensions, and PM) can be solved by a relatively simple strategy consisting of a manual multi-centroid transformation. Regarding the response fluctuation caused by differences in the detector performances, the most promising strategy to cross compare different datasets was MSTUS normalization based on the included blob volume followed by a Z-score normalization on the resulting matrix.
Data availability
Data has been uploaded to the Open Science Framework (OSF) website in a dedicated repository: https://osf.io/jbdxe/?view_only=603945dcc6e547b981b3615bc1c3cf6e. The access is made available on request.
Change history
18 February 2023
A Correction to this paper has been published: https://doi.org/10.1007/s00216-023-04572-3
References
Spielmann N, Wong D. Saliva: diagnostics and therapeutic perspectives. Oral Dis. 2011;17:345–54. https://doi.org/10.1111/j.1601-0825.2010.01773.x.
de Gier J, ’t Hart B, Wilderink P, Nelemans F. Comparison of plasma and saliva levels of diazepam. Br J Clin Pharmacol. 1980;10:151–5. https://doi.org/10.1111/j.1365-2125.1980.tb01732.x.
Yoshizawa JM, Schafer CA, Schafer JJ, Farrell JJ, Paster BJ, Wong DTW. Salivary biomarkers: toward future clinical and diagnostic utilities. Clin Microbiol Rev. 2013;26:781–91. https://doi.org/10.1128/CMR.00021-13.
Lehmann-Kalata A, Miechowicz I, Korybalska K, Swora-Cwynar E, Czepulis N, Łuczak J, Orzechowska Z, Grzymisławski M, Surdacka A, Witowski J. Salivary fingerprint of simple obesity. Cytokine. 2018;110:174–80. https://doi.org/10.1016/j.cyto.2018.05.006.
Knaś M, Maciejczyk M, Sawicka K, Hady HR, Niczyporuk M, Ładny JR, Matczuk J, Waszkiel D, Żendzian-Piotrowska M, Zalewska A. Impact of morbid obesity and bariatric surgery on antioxidant/oxidant balance of the unstimulated and stimulated human saliva. J Oral Pathol Med. 2016;45:455–64. https://doi.org/10.1111/jop.12383.
Stilo F, Bicchi C, Jimenez-Carvelo AM, Cuadros-Rodriguez L, Reichenbach SE, Cordero C. Chromatographic fingerprinting by comprehensive two-dimensional chromatography: fundamentals and tools. TrAC Trends Anal Chem. 2021;134:116133. https://doi.org/10.1016/j.trac.2020.116133.
Stilo F, Cordero C, Sgorbini B, Bicchi C, Liberto E. Highly informative fingerprinting of extra-virgin olive oil volatiles: the role of high concentration-capacity sampling in combination with comprehensive two-dimensional gas chromatography. Separations. 2019;6:34. https://doi.org/10.3390/separations6030034.
Cordero C, Liberto E, Bicchi C, Rubiolo P, Reichenbach SE, Tian X, Tao Q. Targeted and non-targeted approaches for complex natural sample profiling by GC×GC-qMS. J Chromatogr Sci. 2010;48:251–61. https://doi.org/10.1093/chromsci/48.4.251.
Kiefl J, Cordero C, Nicolotti L, Schieberle P, Reichenbach SE, Bicchi C. Performance evaluation of non-targeted peak-based cross-sample analysis for comprehensive two-dimensional gas chromatography-mass spectrometry data and application to processed hazelnut profiling. J Chromatogr A. 2012;1243:81–90. https://doi.org/10.1016/j.chroma.2012.04.048.
Purcaro G, Tranchida PQ, Mondello L, Comprehensive gas chromatography methodologies for the analysis of lipids, Elsevier Inc., 2017. https://doi.org/10.1016/b978-0-12-811732-3.00011-x
Fiehn O. Combining genomics, metabolome analysis, and biochemical modelling to understand metabolic networks. Comp Funct Genomics. 2001;2:155–68. https://doi.org/10.1002/cfg.82.
Pollo BJ, Teixeira CA, Belinato JR, Furlan MF, de Matos Cunha IC, Vaz CR, Volpato GV, Augusto F. Chemometrics, comprehensive two-dimensional gas chromatography and “omics” sciences: basic tools and recent applications. TrAC Trends Anal Chem. 2021;134:116111. https://doi.org/10.1016/j.trac.2020.116111.
Reichenbach SE, Tian X, Tao Q, Ledford EB, Wu Z, Fiehn O. Informatics for cross-sample analysis with comprehensive two-dimensional gas chromatography and high-resolution mass spectrometry (GCxGC-HRMS). Talanta. 2011;83:1279–88. https://doi.org/10.1016/j.talanta.2010.09.057.
Cialiè Rosso M, Stilo F, Squara S, Liberto E, Mai S, Mele C, Marzullo P, Aimaretti G, Reichenbach SE, Collino M, Bicchi C, Cordero C. Exploring extra dimensions to capture saliva metabolite fingerprints from metabolically healthy and unhealthy obese patients by comprehensive two-dimensional gas chromatography featuring Tandem Ionization mass spectrometry. Anal Bioanal Chem. 2021;413:403–18. https://doi.org/10.1007/s00216-020-03008-6.
Stilo F, Bicchi C, Reichenbach SE, Cordero C. Comprehensive two-dimensional gas chromatography as a boosting technology in food-omic investigations. J Sep Sci. 2021;44:1592–611. https://doi.org/10.1002/jssc.202100017.
Stilo F, Liberto E, Reichenbach SE, Tao Q, Bicchi C, Cordero C. Untargeted and targeted fingerprinting of extra virgin olive oil volatiles by comprehensive two-dimensional gas chromatography with mass spectrometry: challenges in long-term studies. J Agric Food Chem. 2019;67:5289–302. https://doi.org/10.1021/acs.jafc.9b01661.
Reichenbach SE, Carr PW, Stoll DR, Tao Q. Smart templates for peak pattern matching with comprehensive two-dimensional liquid chromatography. J Chromatogr A. 2009;1216:3458–66. https://doi.org/10.1016/j.chroma.2008.09.058.
Scientific N, Databases T, NIST/EPA/NIH mass spectral library with search program: (data version: NIST 08, Software version 2.0f), 2010 (2005)
Reichenbach SE, Kottapalli V, Ni M, Visvanathan A, Computer language for identifying chemicals with comprehensive two-dimensional gas chromatography and mass spectrometry, in: J. Chromatogr. A, 2005: pp. 263–269. https://doi.org/10.1016/j.chroma.2004.08.125
Ramos L, Comprehensive analytical chemistry. Comprehensive two dimensional gas chromatography, 2009
Nicolotti L, Cordero C, Bressanello D, Cagliero C, Liberto E, Magagna F, Rubiolo P, Sgorbini B, Bicchi C. Parallel dual secondary column-dual detection: a further way of enhancing the informative potential of two-dimensional comprehensive gas chromatography. J Chromatogr A. 2014;1360:264–74. https://doi.org/10.1016/j.chroma.2014.07.081.
Rempe DW, Reichenbach SE, Tao Q, Cordero C, Rathbun WE, Zini CA. Effectiveness of global, low-degree polynomial transformations for GCxGC data alignment. Anal Chem. 2016;88:10028–35. https://doi.org/10.1021/acs.analchem.6b02254.
Cordero C, Rubiolo P, Reichenbach SE, Carretta A, Cobelli L, Giardina M, Bicchi C. Method translation and full metadata transfer from thermal to differential flow modulated comprehensive two dimensional gas chromatography: profiling of suspected fragrance allergens. J Chromatogr A. 2017;1480:70–82. https://doi.org/10.1016/j.chroma.2016.12.011.
Stilo F, Gabetti E, Bicchi C, Carretta A, Peroni D, Reichenbach SE, Cordero C, McCurry J. A step forward in the equivalence between thermal and differential-flow modulated comprehensive two-dimensional gas chromatography methods. J Chromatogr A. 2020;1627:461396. https://doi.org/10.1016/j.chroma.2020.461396.
Stilo F, Jiménez-Carvelo AM, Liberto E, Bicchi C, Reichenbach SE, Cuadros-Rodríguez L, Cordero C. Chromatographic fingerprinting enables effective discrimination and identitation of high-quality Italian extra-virgin olive oils. J Agric Food Chem. 2021;69:8874–89. https://doi.org/10.1021/acs.jafc.1c02981.
Reichenbach SE, Rempe DW, Tao Q, Bressanello D, Liberto E, Bicchi C, Balducci S, Cordero C. Alignment for comprehensive two-dimensional gas chromatography with dual secondary columns and detectors. Anal Chem. 2015;87:10056–63. https://doi.org/10.1021/acs.analchem.5b02718.
Magagna F, Valverde-Som L, Ruíz-Samblás C, Cuadros-Rodríguez L, Reichenbach SE, Bicchi C, Cordero C. Combined untargeted and targeted fingerprinting with comprehensive two-dimensional chromatography for volatiles and ripening indicators in olive oil. Anal Chim Acta. 2016;936:245–58. https://doi.org/10.1016/j.aca.2016.07.005.
Hollingsworth BV, Reichenbach SE, Tao Q, Visvanathan A. Comparative visualization for comprehensive two-dimensional gas chromatography. J Chromatogr A. 2006;1105:51–8. https://doi.org/10.1016/j.chroma.2005.11.074.
Reichenbach SE, Tian X, Cordero C, Tao Q. Features for non-targeted cross-sample analysis with comprehensive two-dimensional chromatography. J Chromatogr A. 2012;1226:140–8. https://doi.org/10.1016/j.chroma.2011.07.046.
Manig F, Hellwig M, Pietz F, Henle T. Quantitation of free glycation compounds in saliva. PLoS One. 2019;14:e0220208. https://doi.org/10.1371/journal.pone.0220208.
Manig F, Hellwig M, Pietz F, Henle T. Studies about the dietary impact on “free” glycation compounds in human saliva. Foods. 2022;11:2112. https://doi.org/10.3390/foods11142112.
Collotta D, Cordero C, Gerlach JQ, Liberto E, Chiazza F, Cialiè Rosso M, Mele C, Mai S, Reichenbach SE, Tao Q, Le Berre M, Mastrocola R, Aragno M, Joshi L, Bicchi C, Marzullo P, Aimaretti G, Collino M Pilot study on comparative profiling of biofluids (plasma, urine and saliva) from metabolically healthy and metabolically unhealthy obese subject, in: Proc. 39° Congr. Naz. Della Soc. Ital. Di Farmacol. Firenze, 20–23 Novembre, 2019, 2019
Mai S, Grugni G, Mele C, Vietti R, Vigna L, Sartorio A, Aimaretti G, Scacchi M, Marzullo P. Irisin levels in genetic and essential obesity: clues for a potential dual role. Sci Rep. 2020;10:1–9. https://doi.org/10.1038/s41598-020-57855-5.
Zhang Q, Wang G, Du Y, Zhu L, Jiye A. GC/MS analysis of the rat urine for metabonomic research. J Chromatogr B Anal Technol Biomed Life Sci. 2007;854:20–5. https://doi.org/10.1016/j.jchromb.2007.03.048.
Bressanello D, Liberto E, Collino M, Chiazza F, Mastrocola R, Reichenbach SE, Bicchi C, Cordero C. Combined untargeted and targeted fingerprinting by comprehensive two-dimensional gas chromatography: revealing fructose-induced changes in mice urinary metabolic signatures. Anal Bioanal Chem. 2018;410:2723–37. https://doi.org/10.1007/s00216-018-0950-9.
GC ImageTM, GC image GCxGC edition users’ guide, 2017
Magagna F, Guglielmetti A, Liberto E, Reichenbach SE, Allegrucci E, Gobino G, Bicchi C, Cordero C. Comprehensive chemical fingerprinting of high-quality cocoa at early stages of processing: effectiveness of combined untargeted and targeted approaches for classification and discrimination. J Agric Food Chem. 2017;65:6329–41. https://doi.org/10.1021/acs.jafc.7b02167.
Magagna F, Liberto E, Reichenbach SE, Tao Q, Carretta A, Cobelli L, Giardina M, Bicchi C, Cordero C. Advanced fingerprinting of high-quality cocoa: challenges in transferring methods from thermal to differential-flow modulated comprehensive two dimensional gas chromatography. J Chromatogr A. 2018;1536:122–36. https://doi.org/10.1016/j.chroma.2017.07.014.
Reichenbach SE, Tian X, Boateng AA, Mullen CA, Cordero C, Tao Q. Reliable peak selection for multisample analysis with comprehensive two-dimensional chromatography. Anal Chem. 2013;85:4974–81. https://doi.org/10.1021/ac303773v.
Giddings JC. Sample dimensionality: a predictor of order-disorder in component peak distribution in multidimensional separation. J Chromatogr A. 1995;703:3–15. https://doi.org/10.1016/0021-9673(95)00249-M.
Ballabio D. A MATLAB toolbox for principal component analysis and unsupervised exploration of data structure. Chemom Intell Lab Syst. 2015;149:1–9. https://doi.org/10.1016/j.chemolab.2015.10.003.
Ballabio D, Consonni V. Classification tools in chemistry. Part 1: Linear models. PLS-DA Anal Methods. 2013;5:3790–8. https://doi.org/10.1039/c3ay40582f.
Cordero C, Guglielmetti A, Bicchi C, Liberto E, Baroux L, Merle P, Tao Q, Reichenbach SE. Comprehensive two-dimensional gas chromatography coupled with time of flight mass spectrometry featuring tandem ionization: challenges and opportunities for accurate fingerprinting studies. J Chromatogr A. 2019;1597:132–41. https://doi.org/10.1016/j.chroma.2019.03.025.
Reichenbach SE, Zini CA, Nicolli KP, Welke JE, Cordero C, Tao Q. Benchmarking machine learning methods for comprehensive chemical fingerprinting and pattern recognition. J Chromatogr A. 2019;1595:158–67. https://doi.org/10.1016/j.chroma.2019.02.027.
Stilo F, Bicchi C, Robbat A, Reichenbach SE, Cordero C. Untargeted approaches in food-omics: the potential of comprehensive two-dimensional gas chromatography/mass spectrometry. TrAC Trends Anal Chem. 2021;135:116162. https://doi.org/10.1016/j.trac.2020.116162.
Actis AB, Perovic NR, Defagó D, Beccacece C, Eynard AR. Fatty acid profile of human saliva: a possible indicator of dietary fat intake. Arch Oral Biol. 2005;50:1–6. https://doi.org/10.1016/j.archoralbio.2004.08.001.
Troisi J, Belmonte F, Bisogno A, Pierri L, Colucci A, Scala G, Cavallo P, Mandato C, Di Nuzzi A, Di Michele L, Bovi APD, Nuzio SG, Vajro P, Metabolomic salivary signature of pediatric obesity related liver disease and metabolic syndrome, Nutrients. 11 (2019). https://doi.org/10.3390/nu11020274
Barnes VM, Kennedy AD, Panagakos F, Devizio W, Trivedi HM, Jönsson T, Guo L, Cervi S, Scannapieco FA. Global metabolomic analysis of human saliva and plasma from healthy and diabetic subjects, with and without periodontal disease. PLoS ONE. 2014;9:1–8. https://doi.org/10.1371/journal.pone.0105181.
Zgorzynska E, Wierzbicka-Ferszt A, Dziedzic B, Witusik-Perkowska M, Zwolinska A, Janas A, Walczewska A. Docosahexaenoic acid attenuates oxidative stress and protects human gingival fibroblasts against cytotoxicity induced by hydrogen peroxide and butyric acid. Arch Oral Biol. 2015;60:144–53. https://doi.org/10.1016/j.archoralbio.2014.09.009.
Izuhara Y, Miyata T, Ueda Y, Kurokawa K. Accumulation of carbonyls accelerates the formation of two advanced glycation endproducts: carbonyl stress in uremia. Int Congr Ser. 2002;1245:381–2. https://doi.org/10.1016/S0531-5131(02)00901-9.
Rabbani N, Thornalley PJ. Dicarbonyl stress in cell and tissue dysfunction contributing to ageing and disease. Biochem Biophys Res Commun. 2015;458:221–6. https://doi.org/10.1016/j.bbrc.2015.01.140.
Cordero C, Rubiolo P, Cobelli L, Stani G, Miliazza A, Giardina M, Firor R, Bicchi C. Potential of the reversed-inject differential flow modulator for comprehensive two-dimensional gas chromatography in the quantitative profiling and fingerprinting of essential oils of different complexity. J Chromatogr A. 2015;1417:79–95. https://doi.org/10.1016/j.chroma.2015.09.027.
García S, Luengo J, Herrera F. Data preprocessing in data mining, Springer International Publishing. Cham. 2015. https://doi.org/10.1007/978-3-319-10247-4.
Singh D, Singh B. Investigating the impact of data normalization on classification performance. Appl Soft Comput. 2020;97:105524. https://doi.org/10.1016/j.asoc.2019.105524.
Gajera V, Shubham, Gupta R, Jana PK. An effective multi-objective task scheduling algorithm using min-max normalization in cloud computing. Proc 2016 2nd Int Conf Appl Theor Comput Commun Technol ICATccT. 2017;2016:812–6. https://doi.org/10.1109/ICATCCT.2016.7912111.
Cheadle C, Vawter MP, Freed WJ, Becker KG. Analysis of microarray data using Z score transformation. J Mol Diagnostics. 2003;5:73–81. https://doi.org/10.1016/S1525-1578(10)60455-2.
van den Berg RA, Hoefsloot HCJ, Westerhuis JA, Smilde AK, van der Werf MJ. Centering, scaling, and transformations: improving the biological information content of metabolomics data. BMC Genomics. 2006;7:1–15. https://doi.org/10.1186/1471-2164-7-142.
Gagnebin Y, Tonoli D, Lescuyer P, Ponte B, de Seigneux S, Martin PY, Schappler J, Boccard J, Rudaz S. Metabolomic analysis of urine samples by UHPLC-QTOF-MS: impact of normalization strategies. Anal Chim Acta. 2017;955:27–35. https://doi.org/10.1016/j.aca.2016.12.029.
Franchina FA, Zanella D, Dubois LM, Focant JF. The role of sample preparation in multidimensional gas chromatographic separations for non-targeted analysis with the focus on recent biomedical, food, and plant applications. J Sep Sci. 2021;44:188–210. https://doi.org/10.1002/jssc.202000855.
Mack CI, Egert B, Liberto E, Weinert CH, Bub A, Hoffmann I, Bicchi C, Kulling SE, Cordero C, Robust markers of coffee consumption identified among the volatile organic compounds in human urine, Mol. Nutr. Food Res. 63 (2019). https://doi.org/10.1002/mnfr.201801060.
Kasubuchi M, Hasegawa S, Hiramatsu T, Ichimura A, Kimura I. Dietary gut microbial metabolites, short-chain fatty acids, and host metabolic regulation. Nutrients. 2015;7:2839–49. https://doi.org/10.3390/nu7042839.
Funding
Open access funding provided by Università degli Studi di Torino within the CRUI-CARE Agreement. This research was funded by “ERA-NET Biomarkers for Nutrition and Health Implementing the JPI HDHL objectives”—Project Acronym “SALIVAGES”.
The MRPs diet intervention study was funded by the German Federal Ministry for Education and Research (Bundesministerium für Bildung und Forschung, BMBF), grant number 01EA1703.
Author information
Authors and Affiliations
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
Dataset A: the study was approved by the Ethics Committee of Technische Universität Dresden, Germany (reference: AZ 439112017). Dataset B: the experimental procedure was approved by the ad hoc Ethical Research Committee of the Istituto Auxologico Italiano (Verbania, Italy). Written informed consent was obtained from the patients. The study protocol conformed to the guidelines of the European Convention on Human Rights and Biomedicine concerning biomedical research.
Conflict of interest
The authors declare the following competing financial interest(s): Stephen E. Reichenbach and Qingping Tao have a financial interest in GC Image, LLC. The other authors declare no conflict of interest.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Published in the topical collection Comprehensive 2D Chromatography with guest editors Peter Q. Tranchida and Luigi Mondello.
The original online version of this article was revised: Author Qingping Tao should be affiliated to affiliation 5 and author Massimo Collino should be affiliated to affiliation 6.
Supplementary information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Squara, S., Manig, F., Henle, T. et al. Extending the breadth of saliva metabolome fingerprinting by smart template strategies and effective pattern realignment on comprehensive two-dimensional gas chromatographic data. Anal Bioanal Chem 415, 2493–2509 (2023). https://doi.org/10.1007/s00216-023-04516-x
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-023-04516-x