Summary
In the steady statediffusion error andO 2 consumption of the Pt electrode influence assystematic errors the measurement of the oxygen partial pressure. They depend on the geometric properties of the electrode, on the diffusion properties of the membrane, and on the properties of diffusion and convection of the measuring medium. The diffusion field in front of the Pt surface and the thus determined stationary signal of the measurement are calculated for gaseous and non-gaseous media with and without convection. These calculations give quantitative information about systematic errors. With special regard to measurements in perfused tissues (for instance in the brain cortex or in the myocardium) the influence of the “O2 consumption” of Pt electrodes on the intracapillary decrease of the oxygen pressure in the blood flow direction and on the intercapillary pressure field at the measuring point of the electrode is found. These calculations were done by means of a digital model.
Zusammenfassung
Im steady state beeinflussen Diffusionsfehler und Eigenverbrauch der Pt-Elektrode als systematische Fehler O2-Partialdruckmessungen. Sie sind abhängig von den geometrischen Eigenschaften der Elektrode, den Diffusionseigenschaften der Membran sowie den Diffusions- und Konvektionseigenschaften des Meßmediums. Das Diffusionsfeld vor der Pt-Oberfläche und das dadurch bestimmte stationäre Meßsignal werden für gasförmige und nicht gasförmige Medien mit und ohne Konvektion berechnet. Daraus resultieren quantitative Aussagen über die systematischen Fehler. Speziell für Messungen in durchbluteten Geweben (z. B. Hirnrinde und Myokard) wird der Einfluß des Eigenverbrauchs von Pt-Elektroden auf den intracapillärenpO2-Abfall in Durchblutungsrichtung und das intercapillärepO2-Feld am Meßort der Elektrode ermittelt. Diese Berechnungen erfolgten mit Hilfe eines Digitalmodells.
Similar content being viewed by others
Abbreviations
- A :
-
O2-Verbrauch des Gewebes
- α:
-
Bunsenscher Löslichkeitskoeffizient des Mediums
- α m :
-
Bunsenscher Löslichkeitskoeffizient der Membran
- C 1,C 1′,C 2,C 2′,C 3 :
-
Konstanten
- D :
-
Diffusionskoeffizient des Mediums
- DF, DF′ :
-
Diffusionsfehler bei einfacher und doppelter Membran
- DGl :
-
Differentialgleichung
- d :
-
Capillarabstand
- d h :
-
Dicke der hydrodynamischen Grenzschicht
- d m ,d m ′ :
-
Dicke der Membranen
- σ, η, δ′, η′, ε, σ:
-
dimensionslose Parameter
- exp:
-
Exponentialfunktion
- F:
-
Faradaykonstante
- grad:
-
Gradient
- I o :
-
stationäres Meßsignal in Medien ohne Konvektion
- I :
-
stationäres Meßsignal in Gasen
- I o′:
-
stationäres Meßsignal in Flüssigkeiten mit Konvektion
- J o :
-
nullte Bessel-Funktion
- K :
-
Diffusionsleitfähigkeit des Mediums
- KE, KE′ :
-
Konvektionseffekt bei einfacher und doppelter Membran
- K m ,K m ′ :
-
Diffusionsleitfähigkeit der Membranen
- l :
-
Capillarlänge
- Δl :
-
Capillarabschnitt
- μ:
-
Viscosität des Mediums
- p, pO2,p(r), p(r,z) :
-
O2-Partialdruck
- p :
-
mittlerer Partialdruck
- P a :
-
O2-Partialdruck am arteriellen Capillarende
- p c :
-
konstanter Partialdruck
- P/r o +d m :
-
O2-Partialdruck an der Grenze Membran/Medium
- P v :
-
O2-Partialdruck am venösen Capillarende
- ΔP g :
-
relativer O2-Partialdruckabfall im Gewebe
- ΔP v :
-
relativer O2-Partialdruckabfall am venösen Capillarende
- R :
-
Radius der ebenen kreisförmigen Elektrode
- RB :
-
Randbedingung
- RDF, RDF′ :
-
restlicher Diffusionsfehler einfacher und doppelter Membranen
- r o :
-
Radius der Elektrode mit halbkugelförmiger Pt-Oberfläche
- r, z :
-
Zylinderkoordinaten
- r K :
-
Capillarradius
- ΔS :
-
Sättigungsabfall im Capillarblut ohne Elektrode
- ΔS′ :
-
Sättigungsabfall im Capillarblut mit Elektrode
- u :
-
O2-Konzentration
- V :
-
Diffusionsgesamtfluß
- ΔV K :
-
Diffusionsfluß aus einem Capillarabschnitt
- v r ,v z :
-
Komponenten des Stromdichtevektors\(\overrightarrow \upsilon \) inr- bzw.z-Richtung (Zylinderkoordinaten)
- \(\bar \upsilon \) :
-
mittlere Stromdichte
- \(\vec \upsilon \) :
-
Stromdichtevektor des Flusses der O2-Moleküle
- v c :
-
konstante Geschwindigkeit des bewegten Mediums
- x, y, z :
-
Kartesische Koordinaten
- λ:
-
Integrationsvariable
- ∇2 :
-
Laplace-Operator
- \(\frac{\partial }{{\partial t}}\) :
-
partielle Ableitung nach der Zeit
Literatur
Allen, J. G.: Extracorporeal circulation, pp. 82–84. Springfield, Ill.: Ch. C. Thomas 1958.
Diemer, K.: Über die Sauerstoffdiffusion im Gehirn. Pflügers Arch. ges. Physiol.285, 99–108 (1965).
Gleichmann, U., Lübbers, D. W.: Die Messung des Sauerstoffdruckes in Gasen und Flüssigkeiten mit der Pt-Elektrode unter besonderer Berücksichtigung der Messung im Blut. Pflügers Arch. ges. Physiol.271, 431–455 (1960).
Grote, J.: Die Sauerstoffdiffusionskonstanten im Lungengewebe und Wasser und ihre Temperaturabhängigkeit. Pflügers Arch. ges. Physiol.295, 245–254 (1967)
—, Thews, G.: Die Bedingungen für die Sauerstoffversorgung des Herzmuskelgewebes. Pflügers Arch. ges. Physiol.276, 142–165 (1962).
Grunewald, W.: Tur Theorie der Ausgleichsvorgänge an Pt-Elektroden und ihre mathematischen Grundlagen. Diss., Marburg 1966.
—: Die Beeinflussung der pO2-Verteilung im Gewebe durch diepO2-Messung mit der Pt-Elektrode. Pflügers Arch.312, R 144–145 (1969).
Grunewald, W.: Digitale Simulation eines räumlichen Diffusionsmodelles der O2-Versorgung biologischer Gewebe. Pflügers Arch.309, 266–284 (1969).
Hudson, J. A.: Measurement of oxygen tension by the oxygen cathode. Med. biol. Engng.5, 207–223 (1967).
Kolthoff, I. M., Lingane, J. J.: Polarography. New York: Interscience 1952.
Krogh, A.: The number and the distribution of capillaries in muscles with the calculation of the oxygen pressure head necessary for supplying the tissue. J. Physiol. (Lond.)52, 409–415 (1918/19).
Lübbers, D. W.: The meaning of the tissue oxygen distribution curve and its measurement by means of Pt-electrodes. In: Oxygen pressure recording in gases, fluids, and tissues, ed. F. Kreuzer. Progr. Resp. Res.3, 112–123 (1969).
—, Ingvar, D., Betz, E., Fabel, H., Kessler, M., Schmahl, F. W.: Sauerstoffverbrauch der Großhirnrinde in Schlaf- und Wachzustand beim Hund. Pflügers Arch. ges. Physiol.281, R 58 (1964).
Neitzert, W. A.: Schweißen, und Heißsiegeln von Kunststoffolien. Die Kunststoffbücherei, Bd. 7, S. 36. Speyer-Wien-Zürich: Zechner 1964.
Paulussen, F., Hübner, G., Grebe, D., Bretschneider, H. J.: Die Feinstruktur des Herzmuskels während einer Ischämie mit Senkung des Energiebedarfs durch spezielle Kardioplegie. Klin. Wschr.46, 165–171 (1968).
Saito, Y.: A theoretical study on the diffusion current at the stationary electrodes of circular and narrow band types. Rev. Polargr. (Japan)15, 177–187 (1968).
Schuler, R.: Evaluation and design of rapid polarographic in vivo oxygen catheter electrode. Diss., Nijmegen 1966.
—, Kreuzer, F.: Rapid polarographic in vivo catheter electrodes. Respir. Physiol.3, 90–110 (1967).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Grunewald, W. Diffusionsfehler und Eigenverbrauch der Pt-elektrode beipO2-Messungen im steady state. Pflugers Arch. 320, 24–44 (1970). https://doi.org/10.1007/BF00588455
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00588455