Skip to main content
Log in

The effect of megavoltage field size on intrafraction cone-beam CT image quality

  • Scientific Note
  • Published:
Physical and Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

To investigate the effects of scatter from a megavoltage treatment beam on intrafraction cone beam CT (CBCT) image quality. The effects of treatment beam field size and phantom geometry were investigated as well as the clinical success of IFI. Intrafraction imaging (IFI) was performed on four phantoms with four different MV field sizes using a 6 MV FFF source. The image quality of the intrafraction CBCT images was compared to that of a baseline CBCT (i.e. with no treatment beam on) and quantified using noise and low contrast visibility. Increasing the kV tube current was explored as a possible method to reduce noise induced by the MV photon scatter in the intrafraction-CBCTs. The clinical success of all IFI patients over a 2 month period was reviewed. Intrafraction-CBCT image quality and low-contrast visibility deteriorated as MV field size increased. The extent of image degradation was found to depend on the mass of the phantom resulting in a more pronounced effect for a pelvic phantom than a thoracic phantom. While increasing the tube current could reduce the noise in the intrafraction-CBCT images, increasing the current by a factor of 4 failed to reach baseline image quality. Anatomy was found to be the primary indication of clinical IFI failure with all observed failures occurring during abdominal treatments. Image quality was found to decrease with increasing MV field size and decrease with increasing treatment anatomy mass. When considering intrafraction imaging clinically, the primary indicator of IFI failure is treatment anatomy. IFI can be used during chest treatments with high success rates but care must be taken for abdominal treatments and failures should be expected.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Data availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

References

  1. Mirimanoff R-O (2015) Stereotactic ablative body radiotherapy (SABR): an alternative to surgery in stage I-II non-small-cell cancer of the lung? Chin Clin Oncol 4(4):42

    PubMed  Google Scholar 

  2. Purdie TG et al (2007) Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. Int J Radiat Oncol Biol Phys 68(1):243–252

    PubMed  Google Scholar 

  3. Bissonnette JP et al (2009) Quantifying interfraction and intrafraction tumor motion in lung stereotactic body radiotherapy using respiration-correlated cone beam computed tomography. Int J Radiat Oncol Biol Phys 75(3):688–695

    PubMed  Google Scholar 

  4. Case RB et al (2009) Inter- and intrafraction variability in liver position in non-breath-hold stereotactic body radiotherapy. Int J Radiat Oncol Biol Phys 75(1):302–308

    PubMed  Google Scholar 

  5. Pham D et al (2014) A review of kidney motion under free, deep and forced-shallow breathing conditions: implications for stereotactic ablative body radiotherapy treatment. Technol Cancer Res Treat 13(4):315–323

    CAS  PubMed  Google Scholar 

  6. Yoganathan S et al (2017) Magnitude, impact, and management of respiration-induced target motion in radiotherapy treatment: a comprehensive review. J Med Phys 42(3):101–115

    CAS  PubMed  PubMed Central  Google Scholar 

  7. Shimohigashi Y et al (2017) Tumor motion changes in stereotactic body radiotherapy for liver tumors: an evaluation based on four-dimensional cone-beam computed tomography and fiducial markers. Radiat Oncol 12(1):61

    PubMed  PubMed Central  Google Scholar 

  8. Guckenberger M et al (2008) Image-guided radiotherapy for liver cancer using respiratory-correlated computed tomography and cone-beam computed tomography. Int J Radiat Oncol Biol Phys 71(1):297–304

    PubMed  Google Scholar 

  9. Adamson J, Wu Q (2010) Prostate intrafraction motion assessed by simultaneous kilovoltage fluoroscopy at megavoltage delivery I: clinical observations and pattern analysis. Int J Radiat Oncol Biol Phys 78(5):1563–1570

    PubMed  PubMed Central  Google Scholar 

  10. Nakahara S, Tachibana M, Watanabe Y (2016) One-year analysis of Elekta CBCT image quality using NPS and MTF. J Appl Clin Med Phys 17(3):211–222

    PubMed  PubMed Central  Google Scholar 

  11. Hynds S et al (2011) Assessing the daily consistency of bladder filling using an ultrasonic Bladderscan device in men receiving radical conformal radiotherapy for prostate cancer. Br J Radiol 84(1005):813–818

    CAS  PubMed  PubMed Central  Google Scholar 

  12. Nakagawa K et al (2007) Verification of in-treatment tumor position using kilovoltage cone-beam computed tomography: a preliminary study. Int J Radiat Oncol Biol Phys 69(4):970–973

    PubMed  Google Scholar 

  13. Li R et al (2013) Clinical implementation of intrafraction cone beam computed tomography imaging during lung tumor stereotactic ablative radiation therapy. Int J Radiat Oncol Biol Phys 87(5):917–923

    PubMed  PubMed Central  Google Scholar 

  14. Keall P et al (2017) Stereotactic prostate adaptive radiotherapy utilising kilovoltage intrafraction monitoring: the TROG 15.01 SPARK trial. BMC Cancer 17(1):180

    PubMed  PubMed Central  Google Scholar 

  15. Arumugam S et al (2016) An online x-ray based position validation system for prostate hypofractionated radiotherapy. Med Phys 43(2):961–974

    PubMed  Google Scholar 

  16. van Herk M, Ploeger L, Sonke J-J (2011) A novel method for megavoltage scatter correction in cone-beam CT acquired concurrent with rotational irradiation. Radiother Oncol 100(3):365–369

    PubMed  Google Scholar 

  17. Ling C et al (2011) Acquisition of MV-scatter-free kilovoltage CBCT images during RapidArc or VMAT. Radiother Oncol 100(1):145–149

    PubMed  Google Scholar 

  18. Gomà C et al (2016) The role of a microDiamond detector in the dosimetry of proton pencil beams. Zeitschrift für Medizinische Physik 26(1):88–94

    PubMed  Google Scholar 

  19. Muralidhar K, Murthy P, Kumar R (2008) Commissioning and quality assurance of the X-ray volume Imaging system of an image-guided radiotherapy capable linear accelerator. J Med Phys 33(2):72–77

    CAS  PubMed  PubMed Central  Google Scholar 

  20. Stock M et al (2009) Image quality and stability of image-guided radiotherapy (IGRT) devices: a comparative study. Radiother Oncol 93(1):1–7

    PubMed  PubMed Central  Google Scholar 

  21. Lim SY, Hafiz MZ (2017) Quantitative image quality evaluation for kV cone-beam CT-based IGRT. J Phys Conf Ser 851(1):012029

    Google Scholar 

  22. Schindelin J et al (2012) Fiji: an open-source platform for biological-image analysis. Nat Methods 9:676

    CAS  PubMed  Google Scholar 

  23. Le NTT, Robinson J, Lewis SJ (2015) Obese patients and radiography literature: what do we know about a big issue? J Med Radiat Sci 62(2):132–141

    PubMed  PubMed Central  Google Scholar 

  24. Goyal S, Kataria T (2014) Image guidance in radiation therapy: techniques and applications. Radiol Res Pract 2014:705604

    PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The author would like to acknowledge the radiation therapists on PA4 at the Princess Alexandra Hospital who helped immensely with data collection for this project. The author would also like to acknowledge Dr Elizabeth Brown for helping with ethics approval and reviewing the manuscript from a radiation therapist perspective, and Luke K Webb who brainstormed ideas with me, talked through problems and generally provided moral support and guidance throughout this project.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Georgia Friend.

Ethics declarations

Ethical approval

Granted by Metro South HREC Number LNR/2018/QMS/47578.

Conflict of interest

The authors declare that they have no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Friend, G., O’Connor, P. & Charles, P. The effect of megavoltage field size on intrafraction cone-beam CT image quality. Phys Eng Sci Med 43, 711–717 (2020). https://doi.org/10.1007/s13246-020-00870-7

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13246-020-00870-7

Keywords

Navigation