Abstract
In 2001 the ACPSEM published a position paper on quality assurance in screen film mammography which was subsequently adopted as a basis for the quality assurance programs of both the Royal Australian and New Zealand College of Radiologists (RANZCR) and of BreastScreen Australia. Since then the clinical implementation of digital mammography has been realised and it has become evident that existing screen-film protocols were not appropriate to assure the required image quality needed for reliable diagnosis or to address the new dose implications resulting from digital technology. In addition, the advantages and responsibilities inherent in teleradiology are most critical in mammography and also need to be addressed. The current document is the result of a review of current overseas practice and local experience in these areas. At this time the technology of digital imaging is undergoing significant development and there is still a lack of full international consensus about some of the detailed quality control (QC) tests that should be included in quality assurance (QA) programs. This document describes the current status in digital mammography QA and recommends test procedures that may be suitable in the Australasian environment. For completeness, this document also includes a review of the QA programs required for the various types of digital biopsy units used in mammography. In the future, international harmonisation of digital quality assurance in mammography and changes in the technology may require a review of this document. Version 2.0 represented the first of these updates and key changes related to image quality evaluation, ghost image evaluation and interpretation of signal to noise ratio measurements. In Version 3.0 some significant changes, made in light of further experience gained in testing digital mammography equipment were introduced. In Version 4.0, further changes have been made, most notably digital breast tomosynthesis (DBT) testing and QC have been addressed. Some additional testing for conventional projection imaging has been added in order that sites may have the capability to undertake dose surveys to confirm compliance with diagnostic reference levels (DRLs) that may be established at the National or State level. A key recommendation is that dosimetry calculations are now to be undertaken using the methodology of Dance et al. Some minor changes to existing facility QC tests have been made to ensure the suggested procedures align with those most recently adopted by the Royal Australian and New Zealand College of Radiologists and BreastScreen Australia. Future updates of this document may be provided as deemed necessary in electronic format on the ACPSEM’s website (https://www.acpsem.org.au/whatacpsemdoes/standards-position-papers and see also http://www.ranzcr.edu.au/quality-a-safety/radiology/practice-quality-activities/mqap).
Similar content being viewed by others
Notes
Unfortunately, with the Philips/Sectra L30/L50 systems the placement of ROIs with a processed image is not possible so this procedure must be undertaken with a raw image.
Some manufacturers add a constant number to the value of the signal assigned to each pixel. This is referred to as the pixel offset value.
With the Al test object of 10 mm × 10 mm the ROIs used in the analysis should be ~0.25 cm2.
In principle, this recommendation does not contradict the UK, European and IAEA protocols which state that the incident air kerma should be measured at the upper surface of the PMMA phantom. It is worth noting that specific correction factors have been implicitly applied in the above protocols to relate the incident air kerma measured with PMMA phantoms to that of the equivalent breasts. In the present recommendation, however, no such corrections have been made.
If the exposure time is to be measured directly then it should be done using a manual exposure that matches the mAs needed for the AEC controlled exposure. This avoids the inclusion of the trial exposure which if included will give the impression of an erroneously long exposure time.
For some image receptor systems, that do not allow positioning of ROIs on the image, a quantitative measure of ghosting cannot be undertaken.
This test can be performed on all units where the mean pixel value for part or all of image can be extracted. However, the detector used to monitor the air kerma may influence the measurement so it may be necessary to employ mAs as a surrogate for air kerma.
References
Craig AR, Heggie J.C.P., mclean D, Coakley KS, Nicoll JJ (2001) Recommendations for a mammography quality assurance program. Australas Phys Eng Sci Med 24:1–27
RANZCR (2002) Mammographic quality control manual Royal Australian and New Zealand College of Radiologists
Australian Government, Department of Health, breastscreen Australia, National Accreditation Standards (2015)
Digital Mammography Quality Control Manual American College of Radiology (2016)
Bloomquist AK, Yaffe MJ, Pisano ET, Hendrick RE, Mawdsley GE, Bright S, Shen SZ, Mahadevappa M et al (2006) Quality control for digital mammography in the ACRIN DMIST trial: part I. Med Phys 33, 719–736
Pisano ET, Gatsonis C, Hendrick RE, Yaffe MJ, Baum JK, Acharyya S, Conant EF, Fajardo LL et al (2005) Diagnostic Performance of Digital versus Film Mammography for Breast-Cancer Screening. N Engl J Med 353:1773–1783
Yaffe MJ, Bloomquist AK, Mawdsley GE, Pisano ET, Hendrick RE, Fajardo LL, Boone JM, Kanal K et al (2006) Quality control for digital mammography: part II recommendations from the ACRIN DMIST trial. Med Phys 33:737–752
EUREF (2003) Addendum on Digital Mammography: European Guidelines for Quality Assurance in Mammography Screening. European Reference Organisation for Quality Report No
EUREF (2013) European Protocol for the Quality Control of the Physical and Technical Aspects of Mammography Screening as published in European guidelines for quality assurance in breast cancer screening and diagnosis. Fourth edition, Supplements
EUREF (2016) Protocol for the quality control of the physical and technical aspects of digital breast tomosynthesis systems. Version 1.01
IAEA (2011) Quality assurance programme for digital mammography, IAEA Human Health Series No 17, Vienna
RANZCR (2012) Mammographic quality assurance program, guidelines for quality control testing for digital (CR & DR) mammography, version 3.0. Royal Australian and New Zealand College of Radiologists
Pisano ET, Yaffe MJ (2005) Digital mammography. Radiology 234:353–362
American Association of Physicists in Medicine (1998) Acceptance testing and quality control of photostimulable storage phosphor imaging systems AAPM, Task Group 10 Report No. TG10
Monnin P, Gutierrez D, Bulling S, Guntem D, Verdun FR (2007) A comparison of the performance of digital mammography systems. Med Phys 34:906–914
NHSBSP (2008) Technical evaluation of the Konica Minolta Regius 190 CR mammography system and three types of image plate, NHSBSP Equipment Report 0806
NHSBSP (2009) Technical evaluation of agfa DX-M mammography CR reader with HM5.0 needle image plate, NHSBSP Equipment Report 0905
Yaffe MJ, Boone JM, Packard N, Alonzo-Proulx O, Huang SY et al (2009) The myth of the 50–50 breast. Med Phys 36:5437–5443
Aslund M, Cederstrom B, Lundqvist M, Danielsson M, Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors. Med Phys 34:1918–1925, 2007
Ciatto S, Houssami N, Bernardi D, Caumo F, Pellegrini M, Et (2013) al. Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14:583–589
Skanne P, Bandos AI, Gullien R, Eben EB, Ekseth U, Haakenaasen U (2013) Et al. Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56
Friedewald SM, Rafferty EA, Rose SL, Durand MA, Plecha DM, Greenberg JS et al (2014) Breast cancer screening using tomosynthesis in combination with digital mammography. JAMA 311:2499–2507
Mccarthy AM, Kontos D, Synnestvedt M, Tan KS, Heitjan DF, Schnall M et al (2014) Screening outcomes following implementation of digital breast tomosynthesis in a general-population screening program. J Natl Cancer Inst 106:dju316
Rafferty EA, Durand M, Conant EF, Somers Copit D, Friedewald SM, Plecha DM, Miller DP (2016) Breast cancer screening using tomosynthesis and digital mammography in dense and nondense breasts. JAMA 315:1784–1786
Mertelmeier T, Speitel J, Frumento C (2012) 3D breast tomosynthesis—intelligent technology for clear clinical benefits. White Paper, Siemens Medical Solutions
Smith A, Design considerations in optimising a breast tomosynthesis system. Hologic White Paper WP-00046
Reiser I, Sechopoulos I (2014) Review of digital breast tomosynthesis. Med Phys Int J 2:57–66
Skaane P, Bandos AI, Eben EB, Jebsen IN, Krager M et al (2014) Two View Digital Breast Tomosynthesis screening with synthetically reconstructed projection images: comparison with digital breast tomosynthesis with full-field digital mammographic images. Radiology 271:655–663
Zuley ML, Guo B, Catullo VJ, Chough DM, Kelly AE et al (2014) Comparison of two-dimensional synthesized mammograms versus original digital mammograms alone and in combination with tomosynthesis images. Radiology 271:664–671
AAPM (2005) Assessment of display performance for medical imaging systems American Association of Physicists in Medicine, Task Group 18 Report No. TG18
Samei E, Badano A, Chakraborty D, Cornelius C, Flynn MJ, Hemminger B, Hangiandreou N, Johnson J et al (2005) Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report. Med Phys 32:1205–1225
Royal Australian and New Zealand College of Radiologists, Standards of Practice for Diagnostic and Interventional Radiology. Version 10.2 (2017)
Norweck JT, Seibert JA, Andriole KP, Clunie DA, Curran BH et al (2013) ACR–AAPM–SIIM technical standard for electronic practice of medical imaging. J Digital Imaging 26:38–52. A more recent and slightly amended version of this document is available from https://www.acr.org/News-Publications/Publications
Aslund M, Cederstrom B, Lundqvist M, Danielsson M (2005) Optimized AEC for scanning digital mammography based on local variation of scan velocity. Proc SPIE 5745:468–477
Elbakri IA, Lakshminarayanan AV, Tesic MM (2005) Automatic exposure control for a slot scanning full field digital mammographic system. Med Phys 32:2763–2770
Gunn C (2002) Radiographic imaging. A practical approach, 3rd edn. Churchill Livingstone, Edinburgh
International Electrotechnical Commission (2000) Evaluation and routine testing in medical imaging departments—Part 3–4: Acceptance tests - Imaging performance of dental X-ray equipment International Electrotechnical Commission Report No. IEC 61223-3-4
International Electrotechnical Commission (2005) X-ray tube assemblies for medical diagnosis—characteristics of focal spots. International Electrotechnical Commission Report No. 60336, 4th edn
Standards Australia & Standards New Zealand (2015) Approval and test specification—medical electrical equipment—general requirements for safety—collateral standard: requirements for radiation protection in diagnostic X-ray equipment. AS/NZS IEC Report No. 60601.1.3
Strauss KJ, Rossi RP (1994) Specification, acceptance testing, and quality control of mammography imaging equipment. In: Haus AG, Yaffe MJ (eds) A categorical course in physics—technical aspects of breast imaging. RSNA, Oak Brook, pp 219–246
Standards Australia & Standards New Zealand (2015) Approval and test specification—Medical electrical equipment—General requirements for safety—Collateral Standard: Requirements for radiation protection in diagnostic X-ray equipment. AS/NZS IEC Report No. 60601.2.45
Rossmann K (1969) Point spread-function, line spread-function and modulation transfer function. Radiology 93:257–272,
Coltman JW (1954) The specification of imaging properties by response to a sine wave input. J Opt Soc Am 44:468–471
Droege RT, Morin RL (1982) A practical method to measure the MTF of CT scanners. Med Phys 9:758–761
International Electrotechnical Commission (2007) Evaluation and routine testing in medical imaging departments—Part 3–2: Acceptance testing—imaging performance of mammographic X-ray equipment International Electrotechnical Commission Report No. 61223-3-2 Ed. 2.0
Maidment A, Albert M (2003) Conditioning data for calculation of the modulation transfer function. Med Phys 30:248–252
Carton A-K, Vandenbroucke D, Struye L, Maidment A, Kao Y-H, Albert M, Bosmans H, Marchal G (2005) Validation of MTF measurement for digital mammography quality control. Med Phys 32:1684–1695
International Electrotechnical Commission (2007) Medical electrical characteristics of digital X-ray imaging devices—part 1–2: determination of the quantum detection efficiency—detectors used in mammography, International Electrotechnical Commission Report No. 62220-1-2
Hoeschen C, Tischenko O, Buhr E, Illers H (2005) Comparison of technical and anatomical noise in digital thorax X-ray images. Radiat Prot Dosim 114:75–80
Tischenko O, Hoeschen C, Dance DR, Hunt RA, Maidment A, Bakic PR (2005) Evaluation of a novel method of noise reduction using computer-simulated mammographs. Radiat Prot Dosim 114:81–84
Fletcher-Heath L, Van Metter R (2006) Quantifying the performance of human and software CDMAM phantom image observers for the qualification of digital mammography systems
Van Metter R, Fletcher-Heath L (2006) Applying the “European protocol for the quality of the physical and technical aspects of mammography screening” threshold contrast visibility assessment to digital systems
http://www.euref.org/downloads/software-physico-technical-protocol/cdmam-readout#CDMAM_Analyser
Paquelet JR, Hendrick RE (2010) Lesion size inaccuracies in digital mammography. Am J Roentgenol 194:W115–W118
NHSBSP (2009) Commissioning and routine testing of full field digital mammography systems NHSBSP Equipment report 0604 (Version 3)
NHSBSP (2009) Commissioning and routine testing of full field digital mammography systems. NHSBSP Equipment report 0604 (Version 3). Appendix 4 includes appropriate spreadsheets for testing monitor and printer response
dicomworks. http://dicom.online.fr/
See NHSBSP technical reports available at: https://www.gov.uk/government/collections/breast-screening-professional-guidance
imagej. http://rsb.info.nih.gov/ij/
Sante Dicom Viewer Free http://www.santesoft.com/dicom_viewer_free.html
Monnin P, Marshall NW, Bosmans H, Bochud FO, Verdun FR (2011) Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis. Phys Med Biol 56:4221–4238
International Commission on Radiation Units and Measurements (2006) Patient dosimetry for X-ray used in medical imaging ICRU Report No. Report 74
Eklund S, Thilander A, Leitz W (1993) The impact of anatomic variations on absorbed radiation doses in mammography. Radiat Prot Dosim 49:167–170
Cross P (1994) Doses in mammography: from the phantom to the patient. Australas Radiol 38:20–23
Geise RJ, Palchevsky A (1996) Composition of mammographic phantom materials. Radiology 198:347–350
Heggie J.C.P. (1996) Survey of doses in screening mammography. Australas Phys Eng Sci Med 19:207–216
Klein R, Aichinger H, Dieker J, Jansen J.T.M., Joite-Barfu S, Sabel M, Schulz-Wendland R, Zoetelief J (1997) Determination of average glandular dose with modern mammography units for two large groups of patients. Phys Med Biol 42:651–671
Young KC, Ramsdale ML, Bignall F (1998) Review of dosimetric methods for mammography in the UK breast screening programme. Radiat Prot Dosim 80:183–186
Beckett J, Kotre CJ (2000) Dosimetric implications of age related glandular changes in screening mammography. Phys Med Biol 45:801–813
Dance DR, Skinner CL, Young KC, Beckett J, Kotre CJ (2000) Additional factors for the estimation of mean glandular breast dose using the UK mammographic dosimetry protocol. Phys Med Biol 45:3225–3240
Vedantham S, Shi L, Karellas A, O’Connell AM (2012) Dedicated breast CT: fibroglandular volume measurements in a diagnostic population. Med Phys 39:7317–7328
Dance DR (1990) Monte Carlo calculations of conversion factors for the estimation of mean glandular breast dose. Phys Med Biol 35:1211–1219
Dance DR, Young KC, van Engen RE (2011) Estimation of mean glandular dose for breast tomosynthesis: factors for use with the UK, European and IAEA breast dosimetry protocols. Phys Med Biol 56:453–471
Dance DR, Young KC, van Engen RE (2009) Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocols. Phys Med Biol 54:4361–4372
See https://medphys.royalsurrey.nhs.uk/nccpm/?s=mgd which provides access to the relevant Dance dosimetry papers and relevant tables
IAEA (2007) Dosimetry in diagnostic radiology: an international code of practice, TRS 457, Vienna. Document available from http://www-pub.iaea.org/MTCD/publications/PDF/TRS457_web.pdf
Wu X, Barnes GT, Tucker DM (1991) Spectral dependance of glandular tissue dose in screen-film mammography. Radiology 179:143–148
Wu X, Gingold EL, Barnes GT, Tucker DM (1994) Normalised average glandular tissue dose in molybdenum target-rhodium filter and rhodiuim-target rhodium filter mammography. Radiology 193:83–89
Sobol WT, Wu X (1997) Parametrization of mammography normalized average glandular dose tables. Med Phys 24:547
Liu B, Goodsitt M, Chan H-P (1995) Normalised average glandular dose in magnification mammography. Radiology 197:27–32
Robson KJ (2001) A parametric method for determining mammographic X-ray tube output and half value layer. Br J Radiol 74:335–340
ACR (1999) Stereotactic breast biopsy quality control manual. American College of Radiology, Reston
Sechopoulos I, Suryanarayanan S, Vedantham S, D’Orsi C, Karellas A (2007) Computation of the glandular radiation dose in digital tomosynthesis of the breast. Med Phys 34:221–232
Sechopoulos I, D’Orsi CJ (2008) Glandular radiation dose in tomosynthesis of the breast using tungsten targets. J Appl Clin Med Phys 9:161–171
Sechopoulos I, Sabol JM, Berglund J, Bolch WE, Brateman L et al (2014) Radiation dosimetry in digital breast tomosynthesis: Report of AAPM Tomosynthesis Subcommittee Task Group 223. Med Phys 41:091501
Feng SS, Sechopoulos I (2012) Clinical digital breast tomosynthesis system: dosimetric characterization. Radiology 263:35–42
AAPM (1998) Acceptance testing and quality control of photostimulable storage phosphor imaging systems American Association of Physicists in Medicine, Task Group 10 Report No. TG10
Acknowledgements
We acknowledge those who assisted in the writing of this latest version of the paper through discussions and review: Alex Merchant, Stewart Midgley, Tim Ireland and Mike Irvine.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Ethical approval
This article does not contain any studies with human participants or animals performed by any of the authors.
Appendices
Appendix 1: Summary of recommendations for facility QC procedures for DR units in 2D mode
Procedure | Recommended control-limits/requirements | Minimum frequency | Key procedure elements | Recommendations for record keepinga |
---|---|---|---|---|
Viewing conditions | Appropriate viewing conditions All viewbox lamps must be operational and appropriate masking available | Daily | Visual inspection of ambient lighting conditions to ensure conformance with acceptable viewing condition configuration (see text for detail) Visual inspection of viewboxes for uniformity of brightness Confirmation of presence and operation of masking for viewboxes | Checklist/logbook entry showing Date performed Person performing task |
Full field artefact evaluation | mAs = baseline ± 10% Mean pixel value in image = baseline ± 10% There must be no evidence of Structures that are more conspicuous than the objects in the phantom used for weekly testing Blotches or regions of altered noise appearance Observable grid lines or breast support structures Bright or dark pixels Dust artefacts mimicking calcifications Significant stitching or registration artefacts | Daily | Expose a uniform thickness of PMMA using clinically relevant technique factors Image must be acquired in “processed” or “for presentation” form Measure mean pixel value in 4 cm2 ROI positioned centrally along long axis of image and 6 cm from chest wall View image on acquisition monitor using zoom and roam to check for possible detector faults Print image if interpretation performed using hard copy | Records showing Date test was performed Person performing test Test results kVp, target/filter and mAs |
Monitor QC (Monitors used for interpretation and attached to the acquisition workstation) | Borders must be visible, lines must be straight, squares must appear square, the ramp bars should appear continuous without any contour lines, there must be no smearing or bleeding at black-white transitions, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns and two low contrast patterns must be visible in all four corners and in the centre, the 5% and 95% pixel value squares must be clearly visible, pattern must be centred in the active area and no disturbing artefacts must be visible on the displayed TG18-QC test pattern The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11 | Weekly | Display TG18-QC test pattern Ensure viewing conditions are acceptable Use window-width set to maximum and window-level set to half of maximum | Records showing Date test was performed Person performing test Monitor identification Monitor settings Test results |
Monitor cleaning | Monitor screens must be free of dust, fingerprints and other marks that might interfere with image interpretation | Weekly | Clean all monitor screens gently with lint-free cloth as per manufacturer’s instructions | Checklist/logbook entry showing Date performed Person performing task |
Printer area cleanliness (if applicable) | Clean and dust free environment | Weekly | Wet cleaning of printer area floor and open shelves. Inspect and clean air intake filters on the film printer | Checklist/logbook entry showing Date performed Person performing task |
Image quality evaluation | mAs = baseline ± 10% For hard copy reporting optical density = baseline ±20% and must be in the range of 1.60–2.0 The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or Alternatively: the ability to clearly visualise 4 fibres, 3 speck groups and 4 masses in an image of the new ACR DM phantom | Weekly | Obtaining the phantom image Use of ACR Accreditation phantom or new ACR DM phantom Use of a consistent AEC detector position where this is manually selected Light contact between the compression paddle and the phantom surface Consistent positioning of the phantom Consistent selection of clinically relevant kVp and target/filter combinations Selection of the density setting in current clinical use (if applicable) Evaluating the phantom image (preferably on reading workstation or on printed copy if hardcopy reporting used) Use “for presentation” image with zoom and modest adjustment of window/level functions to score fibres and specks Use consistent (baseline) viewing conditions that reflect those used to read actual mammograms Image quality scoring by the same person, if possible Measure optical density in reproducible part of phantom image if hardcopy reporting Use of a control chart to record results | Record numerical mAs values and image quality scores Control chart showing Plots of mAs, image quality score/s, and OD if applicable ≥25 results Clearly marked control limits Baseline values Radiographic settings (kVp, target/filter combination, density setting and SID) Remarks e.g. corrective action Phantom images identifying Date The X-ray system The technique factors |
Detector calibration—flat field test | Pass or fail | Weekly or as per manufacturer’s requirements | Follow manufacturer’s specific procedure | Checklist/logbook entry showing Date performed Person performing task |
Signal difference to noise ratio (SDNR) | SDNR = baseline ± 20% | Weekly | Preferably follow manufacturer procedure. Alternatively Use the “for presentation” image obtained with either the ACR Accreditation or ACR DM phantom for image quality purposes but with PMMA disc on paddle (if using the ACR DM phantom there is a negative contrast disc in the phantom) Measure the mean pixel value (MPV1) and SD in a small ROI next to PMMA disc (or negative contrast disc in the ACR DM phantom) Measure mean pixel value (MPV2) in ROI centred in disc Calculate SDNR = (MPV1 − MPV2)/SD | Records showing Date test was performed Person performing test X-ray system identification kVp, target/filter, AEC mode and mAs Test results |
Printer QC (if applicable) | Borders must be visible, lines must be straight, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns must be visible in all four corners and the centre, the 5% and 95% pixel value squares must be clearly visible, and no disturbing artefacts must be visible on the printed TG18-QC test pattern The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11.The mid density (MD) and density difference (DD) = baseline ± 0.15 Base + fog (B+F) = baseline ± 0.03 and ≤0.25 Dmax = baseline ± 0.10 and ≥3.4 | Monthly for dry lasers and daily or as used for wet lasers | Print the TG18-QC test pattern Check visibility and distortion of several items used for evaluating the quality of the image Check for disturbing artefacts Measure MD, DD, B+F and Dmax | Control charts and records showing Date test was performed Person performing test Printer identification Test results |
Mechanical inspection | Indicated breast thickness accurate to ±5 mm No hazardous, inoperative, out of alignment or improperly operating items on the system All items listed on the visual check list have received a pass | Monthly | Confirm accuracy of thickness indication Visual inspection of the system to ensure safe and optimum operation | Checklist/logbook entry showing Date inspection performed Inspection results Person performing test |
Repeat analysis | Repeat rate <3% (<2% preferred) [3] | Quarterly | Inclusion of images from at least 250 consecutive client examinations The ability to determine repeat rates attributable to a range of equipment faults and positioning errors | Worksheet/logbook entries showing all results/calculations |
Image receptor homogeneity | Maximum deviation in mean pixel value in ROI <±10% of mean pixel value in central ROI Maximum variation of the mean pixel value in central ROI between successive quarterly images <±10% | Quarterly or more frequently if recommended by the manufacturer | Use manufacturer’s protocol and test block if available; otherwise Image a standard test block at clinical settings On the “for processing” image, draw 100 mm2 square or circular ROIs in the centre and four corners If the mean pixel value of a ROI deviates by more than 15% from the mean pixel value in the central ROI, the detector gain map may require re-calibration If required, to exclude failure due to non-uniformities in the standard test block, rotate latter by 180° and repeat measurement | Records showing Date test was performed Person performing test X-ray system identification kVp, target/filter, density setting and mAs Test results |
AEC calibration test | Mean pixel value for each of 2, 4 and 6 cm PMMA within 10% of baseline values | Quarterly | Assess for both contact and magnification modes Use PMMA thickness between 2 and 6 cm covering complete image receptor Use clinical AEC settings (kVp, target/filter and mode) Measure mean pixel value in 4 cm2 ROI positioned centrally along axis and 6 cm from chest wall Examine image for clinically significant artefacts | Records showing Date test was performed Person performing test X-ray system identification kVp, target/filter, AEC mode and mAs Test results |
Compression | Maximum motorised compression force in range 150–200 N | 6 monthly | Confirm machine indicated compression force meets requirements | Checklist/logbook entry showing Date test performed Test results Person performing test |
Test equipment quality control Densitometer calibration check | Optical density measurement accurate to within ±0.03 (0–3.0 OD) ±3% (3.0–4.0 OD) | 6 monthly | Verification of accuracy using an optical density calibration strip traceable to an accepted standard | Checklist/logbook entry showing Date test performed Test results Person performing test |
Maintenance and fault logging | Separate logbooks for each imaging system including diagnostic monitors and film printer if relevant | As required | Dated entries describing fault encountered and/or maintenance performed | Logbooks with dated and initialled entries |
Infection control of breast imaging equipment | Clean equipment | Before each examination | Cleaning using alcohol swipes, or as per manufacturer’s recommendations and/or suitable infection control advice | Nil |
Appendix 2: Summary of recommendations for facility qc procedures for CR units
Procedure | Recommended control-limits/requirements | Minimum frequency | Key procedure elements | Recommendations for record keepinga |
---|---|---|---|---|
Viewing conditions | Appropriate viewing conditions All viewbox lamps must be operational and appropriate masking available | Daily | Visual inspection of ambient lighting conditions to ensure conformance with acceptable viewing condition configuration (see text for detail) Visual inspection of viewboxes for uniformity of brightness Confirmation of presence and operation of masking for viewboxes | Checklist/logbook entry showing Date performed Person performing task |
Image plate erasure | Erasure of energy absorbed from scattered radiation or naturally occurring radiation by CR image plates before they are used | Daily/weekly | On a daily basis or if left unused for more than 8 hours, all CR image plates must be subjected to an erasure (following manufacturer’s instructions) On a weekly basis all Fuji CR image plates must be subjected to a primary erasure | Logbooks with dated and initialled entries |
Monitor/viewboxes cleaning | Monitor screens and viewboxes must be free of dust, fingerprints and other marks that might interfere with image interpretation | Weekly | Clean all monitor screens and viewboxes gently with lint-free cloth as per manufacturer’s instructions | Checklist/logbook entry showing Date performed Person performing task |
Monitor QC (monitors used for interpretation and attached to the acquisition workstation) | Borders must be visible, lines must be straight, squares must appear square, the ramp bars should appear continuous without any contour lines, there should be no smearing or bleeding at black-white transitions, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns and two low contrast patterns must be visible in all four corners and the centre, the 5 and 95% pixel value squares must be clearly visible, pattern must be centred in the active area and no disturbing artefacts must be visible on the displayed TG18-QC test pattern The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11 | Weekly | Display TG18-QC test pattern Ensure viewing conditions are acceptable Use window-width set to maximum and window-level set to half of maximum | Records showing Date test was performed Person performing test Monitor identification Monitor settings Test results |
Monitor cleaning | Monitor screens must be free of dust, fingerprints and other marks that might interfere with image interpretation | Weekly | Clean all monitor screens gently with lint-free cloth as per manufacturer’s instructions | Checklist/logbook entry showing Date performed Person performing task |
Printer area cleanliness (if applicable) | Clean and dust free environment | Weekly | Wet cleaning of printer area floor and open shelves. Inspect and clean air intake filters on the film printer | Checklist/logbook entry showing Date performed Person performing task |
Image quality evaluation | mAs = baseline ± 10% Dose to plate = baseline ±10% Exposure indicator (see Appendix 6 for manufacturer dependent tolerances) For hard copy reporting optical density = baseline ±20% and must be in the range of 1.60–2.0 The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or The ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of an ACR DM phantom | Weekly | Obtaining the phantom image Use an ACR Accreditation phantom or the new ACR DM phantom Use of a designated test cassette and imaging plate that is in routine clinical use Use of a consistent AEC detector position where this is manually selected Light contact between the compression paddle and the phantom surface Consistent positioning of the phantom Consistent selection of clinically relevant kVp and target/filter combinations Selection of the density setting in current clinical use Consistent time delay between plate irradiation and readout Evaluating the phantom image (preferably on reading workstation or on printed copy if hardcopy reporting used) Use “for presentation” image with zoom and modest adjustment of window/level functions to score fibres and specks Use of consistent viewing conditions that reflect those used to read actual mammograms. This applies to both soft and hard copy Image quality scoring by the same person, if possible Measure optical density in reproducible part of phantom image if hardcopy reporting Use of a control chart to record results | Record numerical mAs values and image quality scores Control chart showing Plots of mAs, exposure indicator, image quality score/s and OD if applicable ≥25 results Clearly marked control limits Baseline values Radiographic settings (kVp, target/filter combination, density setting and SID) Remarks e.g. corrective action Phantom images identifying Date The X-ray system The technique factors |
Printer QC | Borders must be visible, lines must be straight, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns must be visible in all four corners and the centre, the 5% and 95% pixel value squares must be clearly visible, and no disturbing artefacts must be visible on the printed TG18-QC test pattern The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11 The mid density (MD) and density difference (DD) = baseline ± 0.15 Base + fog (B+F) = baseline ± 0.03 and ≤0.25 Dmax = baseline ± 0.10 and ≥3.4 | Monthly for dry lasers and daily or as used for wet lasers | Print the TG18-QC test pattern Check visibility and distortion of several items used for evaluating the quality of the image Check for disturbing artefacts Measure MD, DD, B+F and Dmax | Control charts and records showing Date test was performed Person performing test Printer identification Test results |
Mechanical inspection | Indicated breast thickness accurate to ±5 mm No hazardous, inoperative, out of alignment or improperly operating items on the system All items listed on the visual check list have received a pass | Monthly | Confirm accuracy of thickness indication Visual inspection of the system to ensure safe and optimum operation | Checklist/logbook entry showing Date performed Inspection results Person performing task |
Repeat analysis | Repeat rate <3% (<2% preferred) [3] | Quarterly | Inclusion of images from at least 250 consecutive client examinations The ability to determine repeat rates attributable to a range of equipment faults and positioning errors | Worksheet/logbook entries showing all results/calculations |
Image receptor homogeneity | Maximum difference in mean pixel value between any two ROIs <±10% Maximum variation of the mean pixel value in central ROI between successive QC images <±10% | Quarterly or more frequently if recommended by the manufacturer) | Use manufacturer’s protocol and test block if available; otherwise Image a standard test block at clinical settings Use ‘test’ cassette Perform measurements on the “for processing” (unprocessed) image, if possible, using a 100 mm2 square or circular ROI. Three ROIs are placed at the left, right and centre on a line 20 mm back from chest wall If the mean pixel value of any two ROIs differ by more than 10% from each other, the CR unit’s shading correction may require re-calibration or imaging plate(s) may require replacement If ROI analysis is not possible, do a visual inspection at narrow window width If required, to exclude failure due to non-uniformities in the standard test block, rotate by 180° and repeat measurement | Records showing Date test was performed Person performing test X-ray system identification kVp, target/filter, density setting and mAs Test results |
AEC calibration test | Dose to plate for each of 2, 4 and 6 cm PMMA = baseline ±10% See Appendix 6 for equivalent manufacturer specific exposure index requirements | Quarterly | Assess for both contact and magnification modes Use PMMA thickness between 2 and 6 cm covering complete cassette Use clinical AEC settings (kVp, target/filter and mode including density setting) Use a designated test cassette and imaging plate that is in routine clinical use Use a consistent AEC detector position where this is manually selected Consistent positioning of the PMMA Consistent time delay between plate irradiation and readout | Records showing Date test was performed Person performing test X-ray system identification kVp, target/filter, AEC mode and mAs Test results |
Compression | Maximum motorised compression force in range 150–200 N | 6 monthly | Confirm machine indicated compression force meets requirements | Checklist/logbook entry showing Date test performed Test results Person performing test |
Test equipment quality control Densitometer calibration check | Optical density measurement accurate to within ±0.03 (0–3.0 OD) ±3% (3.0–4.0 OD) | 6 monthly | Verification of accuracy using an optical density calibration strip traceable to an accepted standard | Checklist/logbook entry showing Date test performed Test results Person performing test |
Cassette/image plate condition and inter plate sensitivity variation | Clean and dust free cassettes and image plates No major inhomogeneities on the images See Appendix 6 for manufacturer specific tolerances on interplate variations | 6 monthly | Cassette/image plate cleaning as per manufacturer’s recommendations Image a standard test block at clinical settings Pre-processing should be turned off as much as possible and no post processing must be applied Evaluate for artefact on both film (if applicable) and monitor | Records showing Date test was performed Person performing test kVp, target/filter, AEC mode Exposure indicator and mAs for each plate |
Maintenance and fault logging | Separate logbooks for each imaging system, including diagnostic monitors, and film printer if relevant. | As required | Dated entries describing fault encountered and/or maintenance performed | Logbooks with dated and initialled entries |
Infection control of breast imaging equipment | Clean equipment | Before each examination | Cleaning using alcohol wipes, or as per manufacturer’s recommendations and/or suitable infection control advice | Nil |
Appendix 3: Summary of recommendations for medical physics testing only at acceptance or equipment upgrade in 2D mode
Procedure | Performance requirements/guidelines | Routine testing guidelines | Key procedure elements |
---|---|---|---|
Focal spot | ≥11 lp/mm for line-pair bars perpendicular to anode–cathode axis and ≥13 lp/mm for line-pair bars parallel to anode–cathode axis OR complies with IEC 60336 [38] for 0.3 and 0.1 mm focal spot sizes | Not required unless tube has been changed | As per “Focal spot size” section or IEC 60336 [38] |
Leakage radiation | ≤1 mGy/h at 1m from focus and ≤0.01 mGy/100 mAs @ 30 kVp and 30 cm from focus | Not required unless tube has been changed or system relocated | As per AS/NZS IEC 60601.1.30.1.3 [39] |
Transmission through breast support | ≤0.001 mGy @ max kVp and mAs | Not required unless change made to image receptor system | As per AS/NZS IEC 60601.1.3 [39] |
Missed tissue @ chest wall | Width of missed tissue at chest wall ≤5 mm in contact mode and ≤7 mm in magnification mode | Not required unless tube has been changed or change made to image receptor system or system relocated | |
Plate fogging (CR only) | Image of coin must not be visible | Not required unless changes in storage of cassettes have occurred | Monitor during acceptance testing |
MTF | Bench mark testing, compare to manufacturer’s specification | Not required unless tube has been changed or change made to image receptor system | As per IEC 62220-1-2 [48] |
Threshold contrast visibility | Not required unless tube has been changed or change made to image receptor system | Use CDMAM phantom | |
Spatial linearity and geometric distortion | Not required unless change made to image receptor system | Use wire mesh tool | |
Distance calliper accuracy | Measured dimensions of ruler in image must be within 2% of true dimensions in plane specified by manufacturer Check both contact and magnification modes | Not required unless change made to image receptor system | Use steel rulers. Determine dimensions in image, ideally using reporting workstation |
Appendix 4: Summary of recommendations for medical physics annual testing of DR units in 2D mode
Procedure | Performance requirements/guidelines | Routine testing guidelines | Acceptance and additional tests |
---|---|---|---|
Mammography unit assembly evaluation | Correct and safe function of system components. Thickness display accuracy within ±5 mm, note: Flexi paddles will not comply (manufacturer recommendation varies ~11–12 mm for flexi paddles). Reproducible to 2 mm. Verify DICOM image header for correct display of parameters | Confirm function of all motorised components, warning lights, displays etc. Evaluate system for any miscellaneous safety risks etc. DICOM verification required after software upgrades | As per routine tests |
Collimation and alignment assessment | |||
X-ray field/Image receptor alignment | The X-ray field shall irradiate the image receptor fully but not extend beyond the breast support on the chest wall edge of the image receptor by more 2 mm | Assess alignment for each target/geometry combination | As per routine tests |
Paddle/Image alignment | The chest wall edge of the compression paddle shall be aligned just beyond the chest wall edge of the image receptor such that it does not appear in the image. In addition, the compression paddle shall not extend beyond the chest wall edge of the image by more than 1% of the SID | Assess alignment for all clinically relevant Bucky/paddle/target/geometry combinations | As per routine tests |
System resolution/MTF | Compare to baseline values, variation less than 10% | Measure MTF using system software if possible. Otherwise measure limiting resolution Use a 4cm PMMA block or equivalent Place resolution pattern on PMMA Measure both parallel and perpendicular to chest wall Repeat for Magnification mode if applicable | Establish base line values |
AEC system performance assessment | |||
Reproducibility | Coefficient of variation (COV) for both absorbed dose and mAs for at least three phototimed exposures of a test object shall be better than or equal to 0.05 | Use a 4cm PMMA block or equivalent Assess COV for each AEC detector at a typical clinical kVp | As per routine tests |
Compensation and SDNR system performance assessment | Compare SDNR values to baseline and to the minimum acceptable values for 4 cm PMMA (SDNRaccept) SDNR2cm > 1.1 × SDNRaccept SDNR4cm > SDNRaccept SDNR6cm > 0.9 × SDNRaccept | Assess the most commonly used AEC modes for contact and magnification geometry Use 0.2 mm Al foil as contrast test tool and measure SDNR for 2, 4 and 6 cm PMMA (also see section on glandular dose). Note: measurements are to be undertaken on “for processing” (unprocessed) images | Establish baseline values. Assess all available AEC modes for contact and magnification geometries |
Density control (if applicable) | The density control must be capable of changing the mAs from the value used normally by −25 to +50% | Assess change in mAs for at least two density settings either side of the usual clinical setting using 4 cm of PMMA | Assess change in mAs across full range of density settings |
Back-up timer/security cut-out | Security cut-out mechanisms shall be present and terminate the exposure within 50 ms or within 5 mAs, otherwise the back-up timer should terminate the exposure at ≤500 mAs and must terminate the exposure at ≤800 mAs | Use lead sheet or other heavily attenuating material to intercept beam and confirm that the back-up timer security cut-out functions within specified limits | Confirm that the back-up timer/security cut-out functions within specified limits |
Image uniformity and artefact | Max. deviation of mean pixel value <±15% of mean pixel value for central ROI Max. deviation in SNR as a function of time is ±10% There must be no evidence of blotches or regions of altered noise appearance, observable grid lines or breast support structures, bright or dark pixels | Assess for 40 mm PMMA covering complete detector Use five ROIs (one central, with the other 4 approximately 20 mm from any edge) each of 100 mm2 Measurements performed on unprocessed image Exclude phantom non uniformity by rotating block 180° and repeating Repeat in magnification mode if applicable | Assess also at 20 and 60 mm |
Detector element failure | Limits currently not established. Must monitor independent of manufacturer Inspect bad pixel map | A mammographic screen-film mesh can be used to determine if correction for bad columns successful | Bad pixel map must be available at any time, independent of manufacturer |
Image quality evaluation | The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or the ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of the ACR DM phantom Additionally, with the ACR DM phantom the SDNR with contrast object ≥2.0 | Use typical clinical settings Measure MPVs and SDs in relevant ROIs of ACR DM phantom so that SDNR may be calculated. | As per routine testing |
Ghost image evaluation | “Ghost image” factor <2.0 | Assess using 40 mm PMMA (see “Ghost image evaluation” section for testing guidelines) | As per routine testing |
System linearity and noise analysis | Linearity plot versus ESAK: R2 > 0.99 SD2 plot versus MPV: R2 > 0.99 Noise parameters: compare to baseline results | Use standard test block (e.g. 4 cm PMMA) at typical clinical beam settings Measure ESAK at 6 cm from chest wall Measure mean pixel value and SD in ROI placed 6 cm from chest wall Plot mean pixel value as a function of ESAK Plot SD2 as a function of MPV corrected for any pixel offset | Baseline measurements at clinical kVp, also at max and min clinical kVps for all target filter combinations |
Generator performance | |||
kVp, reproducibility | COV ≤ 0.02 for a minimum of three exposures | Assess kVp reproducibility at a typical clinical kVp value | As per routine testing |
kVp accuracy | Measured kVp shall be within ±5% of the specified value over the clinically relevant range | Assess kVp accuracy over the clinically relevant range in, at most, 2 kVp increments Note: The kVp need only be verified for one target filter combination per kVp, however the kVp meter must be calibrated for that particular target/filter combination | Assess kVp accuracy over clinically relevant range in 1 kVp increments |
Beam quality | \(\left( {{\text{kVp}}/100} \right)+0.03 \leq {\text{HVL}}<\left( {{\text{kVp}}/100} \right)+{\text{C}}\) \(\begin{aligned} {\text{where C }}&=&0.{\text{12 mm Al for Mo}}/{\text{Mo}} \hfill \\&=&0.{\text{19 mm Al for Mo}}/{\text{Rh}} \hfill \\&=&0.{\text{22 mm Al for Rh}}/{\text{Rh}} \hfill \\&=&0.{\text{23 mm Al for Rh}}/{\text{Ag}} \hfill \\&=&0.{\text{3}}0{\text{ mm Al for W}}/{\text{Rh}} \hfill \\&=&0.{\text{32 mm Al for W}}/{\text{Ag}} \hfill \\&=&0.{\text{25 mm Al for W}}/{\text{Al}}. \hfill \\ \end{aligned}\) | Measure the HVL required for mean glandular dose calculations and for establishing compliance with DRLs (see “Mean glandular dose” section for details) | As per routine tests plus measure HVL at 28 kVp for all target/filter combinations, with the compression paddle removed if unit used for biopsy purposes with open paddle |
Mean glandular dose | ≤2.0 mGy for a 4.2 cm 50% adipose, 50% glandular breast (i.e. ACR Accreditation phantom or ACR DM phantom) <1 mGy for 2.0 cm PMMA (2.3 cm 50% adipose, 50% glandular breast) <4.5 mGy for 6.0 cm PMMA, (6.5 cm 50% adipose, 50% glandular breast) the displayed MGD values must agree with calculated values to ≤25% | Assess for an AEC controlled exposure using typical clinical settings using ACR Accreditation phantom (or ACR DM phantom) and also for 20 mm and 60 mm PMMA Additional dose (and HVL) measurements may be necessary to confirm compliance with DRLs (see “Mean glandular dose”) | As per routine tests |
Exposure time | For all clinically relevant SID settings the maximum exposure time when irradiating 6 cm PMMA must be less than 3.5 and 2 s for fine and broad focus, respectively | Assess for both contact and magnification modes Use 6 cm of PMMA Use clinically relevant technique factors for this PMMA thickness consistent with SDNR and MGD measurements Record mAs and infer the exposure time from tube rating or measure directly using a manual exposure matched to mAs needed for AEC initiated exposure | As per routine tests |
Viewbox luminance and room illuminance (hardcopy only) | Viewing area illuminance ≤50 lx Viewbox luminance ≥3000 cd/m2 | Assess viewing conditions for all viewers | As per routine tests |
Monitor luminance and viewing conditions | Image interpretation must not be done on a monitor of less than 4.2 mega pixels Luminance ratio approximately 350:1 Maximum luminance >450 cd/m2 and maximum luminance of paired monitors matched to ≤5% Minimum luminance preferably not less than 1 cd/m2 Ambient light <20 lx In PACS situations images must be stored with lossless compression | Measure luminance ratio under clinical lighting conditions Confirm luminance uniformity Confirm no cross-talk and pixel defects | As per routine testing with the additional requirement of checking GSDF Monitor or workstation may have comprehensive QC program which needs to be validated |
Monitor performance | No smearing artefact, ramps without terracing Lines straight, boxes square, active display centred, borders complete Squares of different shades from black to white must be distinct and small squares in corners of each clearly discernible Free from artefact The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11 | Test patterns to be displayed at full resolution Test under clinical lighting conditions Use TG18-QC test pattern | As per routine testing. Monitor or workstation may have comprehensive QA program |
Printer (hardcopy only) | B + F = baseline ± 0.03 and ≤0.25 OD Dmax = baseline ± 0.10 and ≥3.4 OD The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11 | Print TG18-QC test pattern as per weekly printer QC test | As per routine tests |
Appendix 5: Summary of recommendations for medical physics annual testing of CR units
Procedure | Performance requirements/guidelines | Routine testing guidelines | Acceptance and additional tests |
---|---|---|---|
Mammography unit assembly evaluation | Correct and safe function of system components. Thickness display accuracy within ±5 mm, reproducible to 2 mm. Verify DICOM image header for correct display of parameters | Confirm function of all motorised components, warning lights, displays etc. Evaluate system for any miscellaneous safety risks etc. DICOM verification required after software upgrades | As per routine tests |
Collimation and alignment assessment | |||
X-ray field/image/breast-support alignment | The X-ray field shall irradiate the image receptor fully but not extend beyond the breast support on the chest wall edge of the image receptor by more than 2 mm | Assess alignment for largest collimator in clinical use for each Bucky/target combination. For magnification geometry only assess chest wall alignment | As per routine tests |
Paddle/image alignment | The chest wall edge of the compression paddle shall be aligned just beyond the chest wall edge of the image receptor such that the chest wall compression paddle does not appear in the image. In addition the compression paddle shall not extend beyond the chest wall edge of the image receptor by more than 1% of the SID | Assess alignment for all clinically relevant Bucky/paddle/geometry combinations | As per routine tests |
System resolution/MTF | Compare to baseline values, variation less than 10% | Measure MTF using system software if possible. Otherwise measure limiting resolution Use a 4cm PMMA block or equivalent Place resolution pattern on PMMA Measure both parallel and perpendicular to chest wall Repeat for Magnification mode if applicable | Establish base line values |
AEC system performance assessment | |||
Reproducibility | Coefficient of variation (COV) for both absorbed dose and mAs for at least three phototimed exposures of a test object shall be better than or equal to 0.05 | Use a 4cm PMMA block or equivalent Assess COV for each AEC detector at a typical clinical kVp | As per routine tests |
Compensation and SDNR system performance assessment | Compare SDNR values to baseline and to the minimum acceptable values for 4 cm PMMA (SDNRaccept) SDNR2cm > 1.1 × SDNRaccept SDNR4cm > SDNRaccept SDNR6cm > 0.9 × SDNRaccept Note: for magnification mode this last requirement is relaxed to SDNR6 cm > 0.65 × SDNRaccept | Assess the most commonly used AEC modes for contact and magnification geometry Use clinical AEC settings (kVp, target/filter and mode including density setting) Use a designated test cassette and imaging plate that is in routine clinical use Use a consistent AEC detector position where this is manually selected Use 0.2 mm Al foil as contrast test tool and measure SDNR for 2, 4 and 6 cm PMMA (also see section on glandular dose). Note: measurements are to be undertaken on “for processing” (unprocessed) image Consistent time delay between plate irradiation and readout Record exposure indicator for each PMMA thickness Measure film density for each image if applicable | Establish baseline values. Assess all available AEC modes for contact and magnification geometries. Assess both 18 × 24 cm2 and 24 × 30 cm2 Buckys |
Density control | The density control must be capable of changing the mAs from the value used normally by −25 to +50% | Assess change in mAs for at least two density settings either side of the usual clinical setting using 4 cm of PMMA | Assess change in mAs across full range of density settings |
Back-up timer/security cut-out | Security cut-out mechanisms shall be present and terminate the exposure within 50 ms or within 5 mAs, otherwise the back-up timer should terminate the exposure at ≤500 mAs and must terminate the exposure at ≤800 mAs | Use lead sheet or other heavily attenuating material to intercept beam and confirm that the back-up timer/security cut-out functions within specified limits | Confirm that the back-up timer/security cut-out functions within specified limits |
Image uniformity and artefact | Max. deviation of mean pixel value <±10% of mean pixel value for central ROI Max. deviation in SNR of central ROI as a function of time is ±10%. No major inhomogeneities on the images | Assess for 40 mm PMMA covering complete CR plate Use three ROIs each of ~100 mm2 placed on a line parallel to and approximately 20 mm from chest wall | Assess also at 20 and 60 mm |
Uniformity of cassette/image plate response | Maximum mAs variation <±5% between all plates of one size Maximum mAs variation <±20% between plates of different sizes See Appendix 6 for manufacturer dependent allowed tolerances on the exposure indicator | Assess for 40 mm PMMA covering complete CR plate | As per routine testing |
Image quality evaluation | The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or the ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of the ACR DM phantom Additionally, with the ACR DM phantom the SDNR with contrast object ≥2.0 | Use typical clinical settings Measure MPVs and SDs in relevant ROIs of ACR DM phantom so that SDNR may be calculated | As per routine testing |
Ghost image evaluation | “Ghost image” factor <2.0 | Assess using 40 mm PMMA (see “Ghost image evaluation” section for testing guidelines) | As per routine testing |
System linearity and noise analysis | Compare to baseline results and note requirement for linearity (see text and Appendix 6) has R2 > 0.99 Noise analysis remains optional | Use standard test block (e.g. 4 cm PMMA) at typical clinical beam settings Use the same cassette/image plate for all exposures Record exposure indicator Plot exposure indicator as a function of ESAK (see Appendix 6) | Baseline measurements at clinical kVp, also at max and min clinical kVps for all target filter combinations |
Generator performance | |||
kVp, reproducibility | COV ≤ 0.02 for a minimum of three exposures | Assess kVp reproducibility at a clinical kVp value | As per routine testing |
kVp accuracy | Measured kVp shall be within ±5% of the specified value over the clinically relevant range | Assess kVp accuracy over the clinically relevant range in, at most, 2 kVp increments | Assess kVp accuracy over clinically relevant range in 1 kVp increments |
Beam quality | \(\left[ {\left( {{\text{kVp}}/100} \right)+0.03} \right] \leq {\text{HVL}}<\left[ {\left( {{\text{kVp}}/100} \right)+{\text{C}}} \right]\) \(\begin{aligned} {\text{where C }}&=&0.{\text{12 mm Al for Mo}}/{\text{Mo}}\\ &=&0.{\text{19 mm Al for Mo}}/{\text{Rh}}\\&=&0.{\text{22 mm Al for Rh}}/{\text{Rh}}\\&=&0.{\text{23 mm Al for Rh}}/{\text{Ag}}\\&=&0.{\text{3}}0{\text{ mm Al for W}}/{\text{Rh}} \\ &=&0.{\text{32 mm Al for W}}/{\text{Ag}} \\ &=&0.{\text{25 mm Al for W}}/{\text{Al}}. \\ \end{aligned}\) | Measure the HVL required for Mean Glandular Dose calculations and for establishing compliance with DRLs (see “Mean glandular dose” for details) | As per routine tests plus measure HVL at 28 kVp for all target/filter combinations, with the compression paddle removed if unit used for biopsy purposes with open paddle |
Mean glandular dose | ≤2.0 mGy for a 4.2 cm 50% adipose, 50% glandular breast (i.e. ACR Accreditation phantom or ACR DM phantom) <1 mGy for 2.0 cm PMMA (2.3 cm 50% adipose, 50% glandular breast) <4.5 mGy for 6.0 cm PMMA, (6.5 cm 50% adipose, 50% glandular breast) | Assess for an AEC controlled exposure using typical clinical settings using ACR Accreditation phantom (or ACR DM phantom) and also for 20 mm and 60 mm PMMA Additional dose measurements may be necessary to confirm compliance with DRLs (see “Mean glandular dose”) | As per routine tests |
Exposure time | For all clinically relevant SID settings the maximum exposure time when irradiating 6 cm PMMA must be less than 3.5 and 2 s for fine and broad focus, respectively | Assess for both contact and magnification modes Use 6 cm of PMMA Use clinically relevant technique factors for this PMMA thickness consistent with SDNR and MGD measurements Record mAs and infer the exposure time from tube rating or measure directly using a manual exposure matched to mAs needed for AEC initiated exposure | As per routine tests |
Viewbox luminance and room illuminance (hardcopy only) | Viewing area illuminance ≤50 lx Viewbox luminance ≥3000 nit | Assess viewing conditions for all viewers | As per routine tests |
Monitor luminance and viewing conditions | Image interpretation must not be done on a monitor of less than 4.2 mega pixels Luminance ratio approximately 350:1 Maximum luminance >450 cd/m2 and maximum luminance of paired monitors matched to ≤5% Minimum luminance preferably not less than 1 cd/m2 Ambient light <20 lx In PACS situations images must be stored with lossless compression | Measure luminance ratio under clinical lighting conditions Confirm luminance uniformity Confirm no cross-talk and pixel defects | As per routine testing with the additional requirement of checking GSDF As per routine testing. Monitor or workstation may have comprehensive QC program which needs to be validated |
Monitor performance | No smearing artefact, ramps without terracing Lines straight, boxes square, active display centred, borders complete Squares of different shades from black to white must be distinct and small squares in corners of each clearly discernible Free from artefact The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11 | Test patterns to be displayed at full resolution Test under clinical lighting conditions Use TG18-QC test pattern | As per routine testing. Monitor or workstation may have comprehensive QA program |
Printer (hardcopy only) | B + F = baseline ± 0.03 and ≤0.25 OD Dmax = baseline ± 0.10 and ≥3.4 OD The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11 | Print TG18-QC test pattern as per weekly printer QC test | As per routine tests |
Appendix 6: Summary of recommendations for facility QC for biopsy units
Procedure | Minimum frequency | Fully integrated digital/biopsy unit | Digital mammography with add on image system | Stand alone biopsy system |
---|---|---|---|---|
Viewing conditions | Weekly | Previously covereda | Previously covered | See “Viewing conditions” section |
Monitor QC | Weekly | Previously covered | ssab | See “Monitor QC” section |
Monitor cleaning | Weekly | Previously covered | ssa | See “Monitor/viewbox cleaning” section |
Image quality evaluation | Weekly | ssa | ssa | See “Image quality evaluation” section Note may use ACR ‘mini’ digital stereotactic phantom—see text |
Printer QC (if applicable) | Weekly | Previously covered | ssa | See “Printer QC” section |
Mechanical inspection | Monthly | ssa | ssa | See “Mechanical inspection and breast thickness indication” section. Note additionally image receptor and compression plate/ biopsy window must be free of wobble; Vernier drive and needle guide rigid and wobble free, localisation system zeros and biopsy device properly immobilised—see text |
Repeat analysis | Quarterly | ssa | ssa | See “Repeat analysis” sectin |
Image receptor homogeneity | Quarterlyc | Previously covered | ssa | See “Image receptor homogeneity” section. Note; procedure should be modified as seen in text |
AEC calibration test | Quarterly | ssa | ssa | See “AEC calibration test” section. Procedure may vary for different types of units—see text |
Compression | 6 monthly | ssa | ssa | See “Compression” section |
Test equipment quality control Densitometer calibration check | 6 monthly | Previously covered | Previously covered | See “Test equipment calibration” section |
Maintenance and fault logging | As required | ssa | ssa | See “Maintenance and fault logging” section |
Infection control of breast imaging equipment | Before each examination | Previously covered | Previously covered | See “Infection control of breast imaging equipment” section |
Stereotactic Accuracy confirmation | Prior to first use on day of procedures | ssa | ssa | Localisation within ±1 mm. Procedure as per manufacturer’s recommendations; Checklist/logbook entry showing Date test performed Test results Person performing test |
Appendix 7: Summary of recommendations for medical physics testing for biopsy units
Procedure | Frequency | Fully integrated digital/biopsy unit | Digital mammography with add on image system | Stand alone biopsy system |
---|---|---|---|---|
Focal spot | Acceptance | Previously covereda | Previously covered | See “Focal spot size” section |
Leakage radiation | Acceptance | Previously covered | Previously covered | See “Leakage radiation” section |
MTF | Acceptance | Previously covered | ssab | See “Modulation transfer function” section |
Spatial linearity and geometric distortion | Acceptance | Previously covered | ssa | See “Spatial linearity and geometric distortion” section |
Distance calliper accuracy | Acceptance | Previously covered | ssa | See “Distance calliper accuracy” section |
Mammography unit assembly evaluation | Annual | ssa | ssa | See “Mammography unit assembly evaluation” section. Note additionally ensure X-ray tube angular locations positively locked; image receptor and compression plate/ biopsy window free of wobble; Vernier drive and needle guide rigid and wobble free, localisation system zeros; biopsy device properly immobilised and AEC chart displayed—see text |
Collimation assessment | Annual | ssa | ssa | FOV defined by biopsy window and is aligned centrally with digital image receptor, with tolerances of ±5 mm—see text |
System resolution | Annual | Previously covered | ssa | See “System resolution/MTF” section |
AEC/SDNR | Annual | Previously covered | ssa | See “Automatic exposure control system performance assessment/signal difference to noise ratio” section. Note: technique charts should be consulted for correct factor settings. Minimum PMMA thickness of 2 cm used for SDNR see text |
Image uniformity and artefact evaluation | Annual | Previously covered | ssa | See “Image uniformity and artefact evaluation” section .Note—ROIs to be in corners of image, 10 mm from edge |
Image quality evaluation | Annual | Previously covered | ssa | See “Image quality evaluation” section .Note may use ACR ‘mini’ digital stereotactic phantom—see text for revised scoring |
Ghost image evaluation | Annual | Previously covered | ssa | See “Ghost image evaluation” section |
System linearity and noise analysis | Annual | Previously covered | Previously covered | See “System linearity and noise analysis” section |
kVp performance | Annual | Previously covered | Previously covered | See “Generator performance” section |
HVL | Annual | Previously covered | Previously covered | See “Beam quality or half value layer” section |
Mean glandular dose | Annual | ssa | ssa | See “Mean glandular dose” section—note; see technique chart for factors used in dose calculations |
Exposure time | Annual | Previously covered | Previously covered | See “Exposure time” section |
Viewbox and room luminance | Annual | ssa | ssa | See “Viewbox luminance and room illuminance (hardcopy only)” section |
Monitor performance | Annual | Previously covered | ssa | See “Monitor luminance and viewing conditions” section |
Printer (hardcopy) | Annual | Previously covered | ssa | See “Printer (hardcopy)” section |
Localisation accuracy test | Annual | ssa | ssa |
Appendix 8: Summary of criteria in terms of CR exposure indicators
A number of companies currently manufacture CR units for use in mammography and they have developed unique exposure indicators. Reviews of these indicators, with a comparison between the different manufacturers, have been reported in the literature [9, 87]. The table below can be used to indicate the test criteria that should be applied in terms of the current CR exposure indicators.
Test | Test criteria | Tolerance in terms of CR exposure indicator | ||
---|---|---|---|---|
Fuji, Philips and Konica | Kodak (Carestream) | Agfaa | ||
Air kerma (dose) to the plate must not change by greater than ±10% | ±10% in S# of baseline | ±40 units in EI of baseline | ±5% in SAL or ±430 in SAL log or ±580 in PVL log16 of baseline | |
Air kerma (dose) to the plate for each of the three thicknesses of PMMA be within ±10% of the baseline value for each thickness | ±10% in S# of baseline for each thickness | ±40 units in EI of baseline for each thickness | ±5% in the SAL, or ±430 in SAL log or ±580 in PVI log16 of baseline for each thickness | |
Cassette image plate condition and interplate sensitivity variation (also “Image uniformity and artefact evaluation” | Air kerma (dose) to individual plate must differ from mean for that size by less than ±5% Difference in mean air kerma (dose) to plates of different sizes <20% | S# for individual plates must be within ±5% of mean for same size S# difference for two different plate sizes <20% | EI for individual plate must be within ±20 units of mean for same size EI difference for two different plate sizes <100 units | SAL for individual plates must be within ±2.5% or SAL log must be within ±220 or PVI log16 must be within ±290 of mean for same size SAL difference <10% or SAL log difference <1000, or PVI log16 <1300 for two different plate sizes |
R2 value of appropriate plot of exposure indicator versus ESAK must be >0.99 | Plot S# versus reciprocal of ESAK | Plot EI versus log (ESAK) | Plot SAL versus SQRT(ESAK) or SAL log versus log(ESAK) or PVI log 16 versus log(ESAK) | |
Under specified conditions (see Table 4) Exposure Indicator must meet criteria outlined in columns to right | S# = 120 ± 20 | EI = 2300 ± 100 | SAL = 1130 ± 100 SAL log = 21,600 ± 1000 PVI log 16 = 41,100 ± 1300 |
Appendix 9: Summary of recommendations for facility QC for DBT units
Procedure | Recommended control-limits/requirements | Minimum frequency | Key procedure elements | Recommendations for record keepinga |
---|---|---|---|---|
Full field artefact evaluation | mAs = baseline ± 10% There must be no evidence of Clinically significant structures that are more conspicuous than the objects in the phantom used for weekly testing Blotches or regions of altered noise appearance Observable grid lines or breast support structures Bright or dark pixels Dust artefacts mimicking calcifications Significant stitching or registration artefacts | Daily | Expose a uniform thickness of PMMA using clinically relevant technique factors under AEC Central projection image and central reconstructed image should be inspected closely for potential artefacts View images on acquisition monitor using zoom and roam to check for possible detector faults. The magnification should be sufficient to achieve at least 1:1 resolution | Records showing Date test was performed Person performing test Test results kVp, target/filter and mAs |
Image quality evaluation | mAs = baseline ± 10% The ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of an ACR accreditation phantom OR The ability to clearly visualise 2 fibres, 1 speck group and 2 masses in an image of an ACR DM phantom The position of the reconstructed slice used for scoring the phantom must not change by more than ±1 mm | Weekly | Obtaining the phantom image Use an ACR accreditation phantom or the new ACR DM phantom Light contact between the compression paddle and the phantom surface Consistent positioning of the phantom Consistent selection of clinically relevant kVp and target/filter combinations Evaluating the phantom image Scroll through the reconstructed images until the slice displaying the speck details most clearly is reached. Use zoom and modest adjustment of window/level functions to score fibres and specks Use of consistent viewing conditions that reflect those used to read actual mammograms Image quality scoring by the same person, if possible Use of a control chart to record results | Record radiographic settings (kVp, target/filter combination, mAs values) and image quality scores, position of slice used for scoring Control chart showing Plots of mAs, image quality scores, slice position ≥25 results Clearly marked control limits Baseline values Remarks e.g. corrective action Phantom images identifying Date The X-ray system The technique factors |
Detector calibration—flat field test | Pass or fail | Weekly or as per manufacturer’s requirements | Follow manufacturer’s specific procedure | Checklist/logbook entry showing Date performed Person performing task |
AEC calibration test | mAs = baseline ± 10% for same target/filter combination for each thickness of PMMA | Quarterly | Use PMMA thicknesses of 2, 4 and 6 cm covering complete image receptor Use clinical AEC settings (kVp, target/filter and mode) | Records showing Date test was performed Person performing test X-ray system identification kVp, target/filter, AEC mode and mAs |
Compressed breast thickness | For units using special Bucky’s for DBT indicated breast thickness accurate to ±5 mm | Monthly | Confirm accuracy of thickness indication under conditions as indicated by the manufacturer | Checklist/logbook entry showing Date performed Inspection results Person performing task |
Appendix 10: Summary of recommendations for medical physics testing for DBT units
Procedure | Performance requirements/guidelines | Routine testing guidelines | Acceptance and additional tests |
---|---|---|---|
Collimation and alignment assessment | |||
X-ray field/Image receptor alignment | The X-ray field shall irradiate the image receptor fully but not extend beyond breast support on the chest wall edge of the image receptor by more than 2 mm or beyond the Bucky support on the other three margins | Assess alignment for each target/geometry combination Note: When a special Bucky is used for DBT that may also be used to acquire normal 2D projection images the requirements outlined in “Collimation and alignment assessment” section must be met | As per routine tests |
Paddle/Image alignment | The chest wall edge of the compression paddle shall be aligned just beyond the chest wall edge of the image receptor such that it does not appear in the image. In addition, the compression paddle shall not extend beyond the chest wall edge of the image by more than 1% of the SID | As per routine tests | |
Compressed breast thickness | For units using special Bucky’s for DBT, indicated breast thickness accurate to ±5 mm | Confirm accuracy of thickness indication under conditions as indicated by the manufacturer | As per routine tests |
Missed tissue | The missing tissue must be ≤5 mm Full thickness of breast tissue must be imaged | Confirm in DBT mode of acquisition even if previously confirmed in projection imaging mode | As per routine tests |
Distance calliper accuracy | Measured dimensions of object within reconstructed image plane must be within 2% of true dimensions | Use the ACR accreditation (or ACR DM) phantom as per image quality test below Select the reconstructed slice which best displays the speck details for image scoring and perform in plane distance measurements The outside location of the detail insert is useful for this purpose Compare with actual dimensions | Confirm accuracy of measurements at reporting workstation if possible and also confirm accuracy in more than one slice |
AEC system performance assessment | mAs = baseline ± 10% for same target/filter combination for each thickness of PMMA | Use PMMA thicknesses of 2, 4 and 6 cm covering complete image receptor Use clinical AEC settings (kVp, target/filter and mode) | As per routine tests |
Image uniformity and artefact evaluation | mAs = baseline ± 10% There must be no evidence of Clinically significant structures that are more conspicuous than the objects in the phantom used for weekly testing Blotches or regions of altered noise appearance Observable grid lines or breast support structures Bright or dark pixels Dust artefacts mimicking calcifications Significant stitching or registration artefacts | Assess for 40 mm PMMA covering complete detector using clinically relevant technique factors under AEC Central projection image and central reconstructed image should be inspected closely for potential artefacts View images on acquisition monitor using zoom and roam to check for possible detector faults. The magnification should be sufficient to achieve at least 1:1 resolution | Assess all projection and reconstructed images in the acquisition for all available target/filter combinations |
Image quality evaluation | mAs = baseline ± 10% for same target/filter combination Slice used for scoring should be 37 ± 2 mm (ACR accreditation phantom) or 34 ± 2 mm (ACR DM phantom) above breast support and must not change by more than ±1 mm from previous measurement The ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of an ACR accreditation phantom or the ability to clearly visualise 2 fibres, 1 speck group and 2 masses in an image of the ACR DM phantom | Use typical clinical acquisition parameters selected under AEC. Note that the acquisition may be combined with FFDM mode using “Combo mode” Select the reconstructed slice which best displays the speck details for image scoring. This is typically 37 ± 2 mm or 34 ± 2 mm above the breast support with the ACR accreditation phantom and ACR DM phantom, respectively | As per routine testing |
Beam quality | \(\left( {{\text{kVp}}/100} \right)+0.03 \leq {\text{HVL}}<\left( {{\text{kVp}}/100} \right)+{\text{C}}\) \(\begin{aligned} {\text{where C }}&=&0.{\text{12 mm Al for Mo}}/{\text{Mo}}\\ &=&0.{\text{19 mm Al for Mo}}/{\text{Rh}}\\&=&0.{\text{22 mm Al for Rh}}/{\text{Rh}}\\&=&0.{\text{23 mm Al for Rh}}/{\text{Ag}}\\&=&0.{\text{3}}0{\text{ mm Al for W}}/{\text{Rh}} \\ &=&0.{\text{32 mm Al for W}}/{\text{Ag}} \\ &=&0.{\text{31 mm Al for W}}/{\text{Al}}. \\ \end{aligned}\) | Measure the HVL required for mean glandular dose calculations | As per routine tests |
Mean glandular dose | ≤2.0 mGy for a 4.2 cm 50% adipose, 50% glandular breast (i.e. ACR accreditation phantom or ACR DM phantom) <1.2 mGy for 2.0 cm PMMA (2.3 cm 50% adipose, 50% glandular breast) <4.5 mGy for 6.0 cm PMMA, (6.5 cm 50% adipose, 50% glandular breast) Displayed MGD values must agree with calculated values within ±25% | Assess for an AEC controlled exposure using typical clinical settings using ACR phantom (or ACR DM phantom) and also for 20 mm and 60 mm PMMA Confirm displayed and calculated MGDs agree to ± 25% | As per routine tests |
Rights and permissions
About this article
Cite this article
Heggie, J.C.P., Barnes, P., Cartwright, L. et al. Position paper: recommendations for a digital mammography quality assurance program V4.0. Australas Phys Eng Sci Med 40, 491–543 (2017). https://doi.org/10.1007/s13246-017-0583-x
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13246-017-0583-x