Skip to main content

Advertisement

Log in

Position paper: recommendations for a digital mammography quality assurance program V4.0

  • Position Paper
  • Published:
Australasian Physical & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

In 2001 the ACPSEM published a position paper on quality assurance in screen film mammography which was subsequently adopted as a basis for the quality assurance programs of both the Royal Australian and New Zealand College of Radiologists (RANZCR) and of BreastScreen Australia. Since then the clinical implementation of digital mammography has been realised and it has become evident that existing screen-film protocols were not appropriate to assure the required image quality needed for reliable diagnosis or to address the new dose implications resulting from digital technology. In addition, the advantages and responsibilities inherent in teleradiology are most critical in mammography and also need to be addressed. The current document is the result of a review of current overseas practice and local experience in these areas. At this time the technology of digital imaging is undergoing significant development and there is still a lack of full international consensus about some of the detailed quality control (QC) tests that should be included in quality assurance (QA) programs. This document describes the current status in digital mammography QA and recommends test procedures that may be suitable in the Australasian environment. For completeness, this document also includes a review of the QA programs required for the various types of digital biopsy units used in mammography. In the future, international harmonisation of digital quality assurance in mammography and changes in the technology may require a review of this document. Version 2.0 represented the first of these updates and key changes related to image quality evaluation, ghost image evaluation and interpretation of signal to noise ratio measurements. In Version 3.0 some significant changes, made in light of further experience gained in testing digital mammography equipment were introduced. In Version 4.0, further changes have been made, most notably digital breast tomosynthesis (DBT) testing and QC have been addressed. Some additional testing for conventional projection imaging has been added in order that sites may have the capability to undertake dose surveys to confirm compliance with diagnostic reference levels (DRLs) that may be established at the National or State level. A key recommendation is that dosimetry calculations are now to be undertaken using the methodology of Dance et al. Some minor changes to existing facility QC tests have been made to ensure the suggested procedures align with those most recently adopted by the Royal Australian and New Zealand College of Radiologists and BreastScreen Australia. Future updates of this document may be provided as deemed necessary in electronic format on the ACPSEM’s website (https://www.acpsem.org.au/whatacpsemdoes/standards-position-papers and see also http://www.ranzcr.edu.au/quality-a-safety/radiology/practice-quality-activities/mqap).

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

Notes

  1. SDNR was previously referred to as the contrast to noise ratio (CNR). SDNR terminology is now preferred [6, 57].

  2. Unfortunately, with the Philips/Sectra L30/L50 systems the placement of ROIs with a processed image is not possible so this procedure must be undertaken with a raw image.

  3. Some manufacturers add a constant number to the value of the signal assigned to each pixel. This is referred to as the pixel offset value.

  4. With the Al test object of 10 mm × 10 mm the ROIs used in the analysis should be ~0.25 cm2.

  5. In principle, this recommendation does not contradict the UK, European and IAEA protocols which state that the incident air kerma should be measured at the upper surface of the PMMA phantom. It is worth noting that specific correction factors have been implicitly applied in the above protocols to relate the incident air kerma measured with PMMA phantoms to that of the equivalent breasts. In the present recommendation, however, no such corrections have been made.

  6. If the exposure time is to be measured directly then it should be done using a manual exposure that matches the mAs needed for the AEC controlled exposure. This avoids the inclusion of the trial exposure which if included will give the impression of an erroneously long exposure time.

  7. For some image receptor systems, that do not allow positioning of ROIs on the image, a quantitative measure of ghosting cannot be undertaken.

  8. This test can be performed on all units where the mean pixel value for part or all of image can be extracted. However, the detector used to monitor the air kerma may influence the measurement so it may be necessary to employ mAs as a surrogate for air kerma.

References

  1. Craig AR, Heggie J.C.P., mclean D, Coakley KS, Nicoll JJ (2001) Recommendations for a mammography quality assurance program. Australas Phys Eng Sci Med 24:1–27

    Article  Google Scholar 

  2. RANZCR (2002) Mammographic quality control manual Royal Australian and New Zealand College of Radiologists

  3. Australian Government, Department of Health, breastscreen Australia, National Accreditation Standards (2015)

  4. Digital Mammography Quality Control Manual American College of Radiology (2016)

  5. Bloomquist AK, Yaffe MJ, Pisano ET, Hendrick RE, Mawdsley GE, Bright S, Shen SZ, Mahadevappa M et al (2006) Quality control for digital mammography in the ACRIN DMIST trial: part I. Med Phys 33, 719–736

    Article  PubMed  Google Scholar 

  6. Pisano ET, Gatsonis C, Hendrick RE, Yaffe MJ, Baum JK, Acharyya S, Conant EF, Fajardo LL et al (2005) Diagnostic Performance of Digital versus Film Mammography for Breast-Cancer Screening. N Engl J Med 353:1773–1783

  7. Yaffe MJ, Bloomquist AK, Mawdsley GE, Pisano ET, Hendrick RE, Fajardo LL, Boone JM, Kanal K et al (2006) Quality control for digital mammography: part II recommendations from the ACRIN DMIST trial. Med Phys 33:737–752

    Article  PubMed  Google Scholar 

  8. EUREF (2003) Addendum on Digital Mammography: European Guidelines for Quality Assurance in Mammography Screening. European Reference Organisation for Quality Report No

  9. EUREF (2013) European Protocol for the Quality Control of the Physical and Technical Aspects of Mammography Screening as published in European guidelines for quality assurance in breast cancer screening and diagnosis. Fourth edition, Supplements

  10. EUREF (2016) Protocol for the quality control of the physical and technical aspects of digital breast tomosynthesis systems. Version 1.01

  11. IAEA (2011) Quality assurance programme for digital mammography, IAEA Human Health Series No 17, Vienna

  12. RANZCR (2012) Mammographic quality assurance program, guidelines for quality control testing for digital (CR & DR) mammography, version 3.0. Royal Australian and New Zealand College of Radiologists

  13. Pisano ET, Yaffe MJ (2005) Digital mammography. Radiology 234:353–362

    Article  PubMed  Google Scholar 

  14. American Association of Physicists in Medicine (1998) Acceptance testing and quality control of photostimulable storage phosphor imaging systems AAPM, Task Group 10 Report No. TG10

  15. Monnin P, Gutierrez D, Bulling S, Guntem D, Verdun FR (2007) A comparison of the performance of digital mammography systems. Med Phys 34:906–914

    Article  CAS  PubMed  Google Scholar 

  16. NHSBSP (2008) Technical evaluation of the Konica Minolta Regius 190 CR mammography system and three types of image plate, NHSBSP Equipment Report 0806

  17. NHSBSP (2009) Technical evaluation of agfa DX-M mammography CR reader with HM5.0 needle image plate, NHSBSP Equipment Report 0905

  18. Yaffe MJ, Boone JM, Packard N, Alonzo-Proulx O, Huang SY et al (2009) The myth of the 50–50 breast. Med Phys 36:5437–5443

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Aslund M, Cederstrom B, Lundqvist M, Danielsson M, Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors. Med Phys 34:1918–1925, 2007

    Article  PubMed  Google Scholar 

  20. Ciatto S, Houssami N, Bernardi D, Caumo F, Pellegrini M, Et (2013) al. Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14:583–589

    Article  PubMed  Google Scholar 

  21. Skanne P, Bandos AI, Gullien R, Eben EB, Ekseth U, Haakenaasen U (2013) Et al. Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56

    Article  Google Scholar 

  22. Friedewald SM, Rafferty EA, Rose SL, Durand MA, Plecha DM, Greenberg JS et al (2014) Breast cancer screening using tomosynthesis in combination with digital mammography. JAMA 311:2499–2507

    Article  CAS  PubMed  Google Scholar 

  23. Mccarthy AM, Kontos D, Synnestvedt M, Tan KS, Heitjan DF, Schnall M et al (2014) Screening outcomes following implementation of digital breast tomosynthesis in a general-population screening program. J Natl Cancer Inst 106:dju316

    Article  PubMed  PubMed Central  Google Scholar 

  24. Rafferty EA, Durand M, Conant EF, Somers Copit D, Friedewald SM, Plecha DM, Miller DP (2016) Breast cancer screening using tomosynthesis and digital mammography in dense and nondense breasts. JAMA 315:1784–1786

    Article  PubMed  Google Scholar 

  25. Mertelmeier T, Speitel J, Frumento C (2012) 3D breast tomosynthesis—intelligent technology for clear clinical benefits. White Paper, Siemens Medical Solutions

  26. Smith A, Design considerations in optimising a breast tomosynthesis system. Hologic White Paper WP-00046

  27. Reiser I, Sechopoulos I (2014) Review of digital breast tomosynthesis. Med Phys Int J 2:57–66

    Google Scholar 

  28. Skaane P, Bandos AI, Eben EB, Jebsen IN, Krager M et al (2014) Two View Digital Breast Tomosynthesis screening with synthetically reconstructed projection images: comparison with digital breast tomosynthesis with full-field digital mammographic images. Radiology 271:655–663

    Article  PubMed  Google Scholar 

  29. Zuley ML, Guo B, Catullo VJ, Chough DM, Kelly AE et al (2014) Comparison of two-dimensional synthesized mammograms versus original digital mammograms alone and in combination with tomosynthesis images. Radiology 271:664–671

    Article  PubMed  PubMed Central  Google Scholar 

  30. AAPM (2005) Assessment of display performance for medical imaging systems American Association of Physicists in Medicine, Task Group 18 Report No. TG18

  31. Samei E, Badano A, Chakraborty D, Cornelius C, Flynn MJ, Hemminger B, Hangiandreou N, Johnson J et al (2005) Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report. Med Phys 32:1205–1225

    Article  PubMed  Google Scholar 

  32. Royal Australian and New Zealand College of Radiologists, Standards of Practice for Diagnostic and Interventional Radiology. Version 10.2 (2017)

  33. Norweck JT, Seibert JA, Andriole KP, Clunie DA, Curran BH et al (2013) ACR–AAPM–SIIM technical standard for electronic practice of medical imaging. J Digital Imaging 26:38–52. A more recent and slightly amended version of this document is available from https://www.acr.org/News-Publications/Publications

  34. Aslund M, Cederstrom B, Lundqvist M, Danielsson M (2005) Optimized AEC for scanning digital mammography based on local variation of scan velocity. Proc SPIE 5745:468–477

    Article  Google Scholar 

  35. Elbakri IA, Lakshminarayanan AV, Tesic MM (2005) Automatic exposure control for a slot scanning full field digital mammographic system. Med Phys 32:2763–2770

    Article  PubMed  Google Scholar 

  36. Gunn C (2002) Radiographic imaging. A practical approach, 3rd edn. Churchill Livingstone, Edinburgh

  37. International Electrotechnical Commission (2000) Evaluation and routine testing in medical imaging departments—Part 3–4: Acceptance tests - Imaging performance of dental X-ray equipment International Electrotechnical Commission Report No. IEC 61223-3-4

  38. International Electrotechnical Commission (2005) X-ray tube assemblies for medical diagnosis—characteristics of focal spots. International Electrotechnical Commission Report No. 60336, 4th edn

  39. Standards Australia & Standards New Zealand (2015) Approval and test specification—medical electrical equipment—general requirements for safety—collateral standard: requirements for radiation protection in diagnostic X-ray equipment. AS/NZS IEC Report No. 60601.1.3

  40. Strauss KJ, Rossi RP (1994) Specification, acceptance testing, and quality control of mammography imaging equipment. In: Haus AG, Yaffe MJ (eds) A categorical course in physics—technical aspects of breast imaging. RSNA, Oak Brook, pp 219–246

    Google Scholar 

  41. Standards Australia & Standards New Zealand (2015) Approval and test specification—Medical electrical equipment—General requirements for safety—Collateral Standard: Requirements for radiation protection in diagnostic X-ray equipment. AS/NZS IEC Report No. 60601.2.45

  42. Rossmann K (1969) Point spread-function, line spread-function and modulation transfer function. Radiology 93:257–272,

  43. Coltman JW (1954) The specification of imaging properties by response to a sine wave input. J Opt Soc Am 44:468–471

    Article  Google Scholar 

  44. Droege RT, Morin RL (1982) A practical method to measure the MTF of CT scanners. Med Phys 9:758–761

    Article  CAS  PubMed  Google Scholar 

  45. International Electrotechnical Commission (2007) Evaluation and routine testing in medical imaging departments—Part 3–2: Acceptance testing—imaging performance of mammographic X-ray equipment International Electrotechnical Commission Report No. 61223-3-2 Ed. 2.0

  46. Maidment A, Albert M (2003) Conditioning data for calculation of the modulation transfer function. Med Phys 30:248–252

    Article  PubMed  Google Scholar 

  47. Carton A-K, Vandenbroucke D, Struye L, Maidment A, Kao Y-H, Albert M, Bosmans H, Marchal G (2005) Validation of MTF measurement for digital mammography quality control. Med Phys 32:1684–1695

    Article  Google Scholar 

  48. International Electrotechnical Commission (2007) Medical electrical characteristics of digital X-ray imaging devices—part 1–2: determination of the quantum detection efficiency—detectors used in mammography, International Electrotechnical Commission Report No. 62220-1-2

  49. Hoeschen C, Tischenko O, Buhr E, Illers H (2005) Comparison of technical and anatomical noise in digital thorax X-ray images. Radiat Prot Dosim 114:75–80

    Article  Google Scholar 

  50. Tischenko O, Hoeschen C, Dance DR, Hunt RA, Maidment A, Bakic PR (2005) Evaluation of a novel method of noise reduction using computer-simulated mammographs. Radiat Prot Dosim 114:81–84

    Article  Google Scholar 

  51. Fletcher-Heath L, Van Metter R (2006) Quantifying the performance of human and software CDMAM phantom image observers for the qualification of digital mammography systems

  52. Van Metter R, Fletcher-Heath L (2006) Applying the “European protocol for the quality of the physical and technical aspects of mammography screening” threshold contrast visibility assessment to digital systems

  53. http://www.euref.org/downloads/software-physico-technical-protocol/cdmam-readout#CDMAM_Analyser

  54. Paquelet JR, Hendrick RE (2010) Lesion size inaccuracies in digital mammography. Am J Roentgenol 194:W115–W118

  55. NHSBSP (2009) Commissioning and routine testing of full field digital mammography systems NHSBSP Equipment report 0604 (Version 3)

  56. NHSBSP (2009) Commissioning and routine testing of full field digital mammography systems. NHSBSP Equipment report 0604 (Version 3). Appendix 4 includes appropriate spreadsheets for testing monitor and printer response

  57. dicomworks. http://dicom.online.fr/

  58. See NHSBSP technical reports available at: https://www.gov.uk/government/collections/breast-screening-professional-guidance

  59. imagej. http://rsb.info.nih.gov/ij/

  60. Sante Dicom Viewer Free http://www.santesoft.com/dicom_viewer_free.html

  61. Monnin P, Marshall NW, Bosmans H, Bochud FO, Verdun FR (2011) Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis. Phys Med Biol 56:4221–4238

    Article  CAS  PubMed  Google Scholar 

  62. International Commission on Radiation Units and Measurements (2006) Patient dosimetry for X-ray used in medical imaging ICRU Report No. Report 74

  63. Eklund S, Thilander A, Leitz W (1993) The impact of anatomic variations on absorbed radiation doses in mammography. Radiat Prot Dosim 49:167–170

    Article  Google Scholar 

  64. Cross P (1994) Doses in mammography: from the phantom to the patient. Australas Radiol 38:20–23

    Article  CAS  PubMed  Google Scholar 

  65. Geise RJ, Palchevsky A (1996) Composition of mammographic phantom materials. Radiology 198:347–350

    Article  CAS  PubMed  Google Scholar 

  66. Heggie J.C.P. (1996) Survey of doses in screening mammography. Australas Phys Eng Sci Med 19:207–216

    CAS  PubMed  Google Scholar 

  67. Klein R, Aichinger H, Dieker J, Jansen J.T.M., Joite-Barfu S, Sabel M, Schulz-Wendland R, Zoetelief J (1997) Determination of average glandular dose with modern mammography units for two large groups of patients. Phys Med Biol 42:651–671

    Article  CAS  PubMed  Google Scholar 

  68. Young KC, Ramsdale ML, Bignall F (1998) Review of dosimetric methods for mammography in the UK breast screening programme. Radiat Prot Dosim 80:183–186

    Article  Google Scholar 

  69. Beckett J, Kotre CJ (2000) Dosimetric implications of age related glandular changes in screening mammography. Phys Med Biol 45:801–813

    Article  CAS  PubMed  Google Scholar 

  70. Dance DR, Skinner CL, Young KC, Beckett J, Kotre CJ (2000) Additional factors for the estimation of mean glandular breast dose using the UK mammographic dosimetry protocol. Phys Med Biol 45:3225–3240

    Article  CAS  PubMed  Google Scholar 

  71. Vedantham S, Shi L, Karellas A, O’Connell AM (2012) Dedicated breast CT: fibroglandular volume measurements in a diagnostic population. Med Phys 39:7317–7328

    Article  PubMed  PubMed Central  Google Scholar 

  72. Dance DR (1990) Monte Carlo calculations of conversion factors for the estimation of mean glandular breast dose. Phys Med Biol 35:1211–1219

    Article  CAS  PubMed  Google Scholar 

  73. Dance DR, Young KC, van Engen RE (2011) Estimation of mean glandular dose for breast tomosynthesis: factors for use with the UK, European and IAEA breast dosimetry protocols. Phys Med Biol 56:453–471

    Article  CAS  PubMed  Google Scholar 

  74. Dance DR, Young KC, van Engen RE (2009) Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocols. Phys Med Biol 54:4361–4372

    Article  CAS  PubMed  Google Scholar 

  75. See https://medphys.royalsurrey.nhs.uk/nccpm/?s=mgd which provides access to the relevant Dance dosimetry papers and relevant tables

  76. IAEA (2007) Dosimetry in diagnostic radiology: an international code of practice, TRS 457, Vienna. Document available from http://www-pub.iaea.org/MTCD/publications/PDF/TRS457_web.pdf

  77. Wu X, Barnes GT, Tucker DM (1991) Spectral dependance of glandular tissue dose in screen-film mammography. Radiology 179:143–148

    Article  CAS  PubMed  Google Scholar 

  78. Wu X, Gingold EL, Barnes GT, Tucker DM (1994) Normalised average glandular tissue dose in molybdenum target-rhodium filter and rhodiuim-target rhodium filter mammography. Radiology 193:83–89

    Article  CAS  PubMed  Google Scholar 

  79. Sobol WT, Wu X (1997) Parametrization of mammography normalized average glandular dose tables. Med Phys 24:547

    Article  CAS  PubMed  Google Scholar 

  80. Liu B, Goodsitt M, Chan H-P (1995) Normalised average glandular dose in magnification mammography. Radiology 197:27–32

    Article  CAS  PubMed  Google Scholar 

  81. Robson KJ (2001) A parametric method for determining mammographic X-ray tube output and half value layer. Br J Radiol 74:335–340

    Article  CAS  PubMed  Google Scholar 

  82. ACR (1999) Stereotactic breast biopsy quality control manual. American College of Radiology, Reston

    Google Scholar 

  83. Sechopoulos I, Suryanarayanan S, Vedantham S, D’Orsi C, Karellas A (2007) Computation of the glandular radiation dose in digital tomosynthesis of the breast. Med Phys 34:221–232

    Article  PubMed  PubMed Central  Google Scholar 

  84. Sechopoulos I, D’Orsi CJ (2008) Glandular radiation dose in tomosynthesis of the breast using tungsten targets. J Appl Clin Med Phys 9:161–171

    Article  Google Scholar 

  85. Sechopoulos I, Sabol JM, Berglund J, Bolch WE, Brateman L et al (2014) Radiation dosimetry in digital breast tomosynthesis: Report of AAPM Tomosynthesis Subcommittee Task Group 223. Med Phys 41:091501

    Article  PubMed  Google Scholar 

  86. Feng SS, Sechopoulos I (2012) Clinical digital breast tomosynthesis system: dosimetric characterization. Radiology 263:35–42

    Article  PubMed  PubMed Central  Google Scholar 

  87. AAPM (1998) Acceptance testing and quality control of photostimulable storage phosphor imaging systems American Association of Physicists in Medicine, Task Group 10 Report No. TG10

Download references

Acknowledgements

We acknowledge those who assisted in the writing of this latest version of the paper through discussions and review: Alex Merchant, Stewart Midgley, Tim Ireland and Mike Irvine.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to J. C. P. Heggie.

Ethics declarations

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

Appendices

Appendix 1: Summary of recommendations for facility QC procedures for DR units in 2D mode

Procedure

Recommended control-limits/requirements

Minimum frequency

Key procedure elements

Recommendations for record keepinga

Viewing conditions

Appropriate viewing conditions

All viewbox lamps must be operational and appropriate masking available

Daily

Visual inspection of ambient lighting conditions to ensure conformance with acceptable viewing condition configuration (see text for detail)

Visual inspection of viewboxes for uniformity of brightness

Confirmation of presence and operation of masking for viewboxes

Checklist/logbook entry showing

 Date performed

 Person performing task

Full field artefact evaluation

mAs = baseline ± 10%

Mean pixel value in image = baseline ± 10%

There must be no evidence of

 Structures that are more conspicuous than the objects in the phantom used for weekly testing

 Blotches or regions of altered noise appearance

 Observable grid lines or breast support structures

 Bright or dark pixels

 Dust artefacts mimicking calcifications

 Significant stitching or registration artefacts

Daily

 Expose a uniform thickness of PMMA using clinically relevant technique factors

 Image must be acquired in “processed” or “for presentation” form

 Measure mean pixel value in 4 cm2 ROI positioned centrally along long axis of image and 6 cm from chest wall

 View image on acquisition monitor using zoom and roam to check for possible detector faults

 Print image if interpretation performed using hard copy

Records showing

 Date test was performed

 Person performing test

 Test results

 kVp, target/filter and mAs

Monitor QC (Monitors used for interpretation and attached to the acquisition workstation)

Borders must be visible, lines must be straight, squares must appear square, the ramp bars should appear continuous without any contour lines, there must be no smearing or bleeding at black-white transitions, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns and two low contrast patterns must be visible in all four corners and in the centre, the 5% and 95% pixel value squares must be clearly visible, pattern must be centred in the active area and no disturbing artefacts must be visible on the displayed TG18-QC test pattern

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11

Weekly

Display TG18-QC test pattern

Ensure viewing conditions are acceptable

Use window-width set to maximum and window-level set to half of maximum

Records showing

 Date test was performed

 Person performing test

 Monitor identification

 Monitor settings

 Test results

Monitor cleaning

Monitor screens must be free of dust, fingerprints and other marks that might interfere with image interpretation

Weekly

Clean all monitor screens gently with lint-free cloth as per manufacturer’s instructions

Checklist/logbook entry showing

 Date performed

 Person performing task

Printer area cleanliness (if applicable)

Clean and dust free environment

Weekly

Wet cleaning of printer area floor and open shelves. Inspect and clean air intake filters on the film printer

Checklist/logbook entry showing

 Date performed

 Person performing task

Image quality evaluation

mAs = baseline ± 10%

For hard copy reporting optical density = baseline ±20% and must be in the range of 1.60–2.0

The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or

Alternatively: the ability to clearly visualise 4 fibres, 3 speck groups and 4 masses in an image of the new ACR DM phantom

Weekly

Obtaining the phantom image

 Use of ACR Accreditation phantom or new ACR DM phantom

 Use of a consistent AEC detector position where this is manually selected

 Light contact between the compression paddle and the phantom surface

 Consistent positioning of the phantom

 Consistent selection of clinically relevant kVp and target/filter combinations

 Selection of the density setting in current clinical use (if applicable)

Evaluating the phantom image (preferably on reading workstation or on printed copy if hardcopy reporting used)

 Use “for presentation” image with zoom and modest adjustment of window/level functions to score fibres and specks

 Use consistent (baseline) viewing conditions that reflect those used to read actual mammograms

 Image quality scoring by the same person, if possible

 Measure optical density in reproducible part of phantom image if hardcopy reporting

 Use of a control chart to record results

Record numerical mAs values and image quality scores

Control chart showing

 Plots of mAs, image quality score/s, and OD if applicable

≥25 results

 Clearly marked control limits

 Baseline values

 Radiographic settings (kVp, target/filter combination, density setting and SID)

 Remarks e.g. corrective action

Phantom images identifying

 Date

 The X-ray system

 The technique factors

Detector calibration—flat field test

Pass or fail

Weekly or as per manufacturer’s requirements

Follow manufacturer’s specific procedure

Checklist/logbook entry showing

 Date performed

 Person performing task

Signal difference to noise ratio (SDNR)

SDNR = baseline ± 20%

Weekly

Preferably follow manufacturer procedure. Alternatively

 Use the “for presentation” image obtained with either the ACR Accreditation or ACR DM phantom for image quality purposes but with PMMA disc on paddle (if using the ACR DM phantom there is a negative contrast disc in the phantom)

 Measure the mean pixel value (MPV1) and SD in a small ROI next to PMMA disc (or negative contrast disc in the ACR DM phantom)

 Measure mean pixel value (MPV2) in ROI centred in disc

 Calculate SDNR = (MPV1 − MPV2)/SD

Records showing

 Date test was performed

 Person performing test

 X-ray system identification

 kVp, target/filter, AEC mode and mAs

 Test results

Printer QC (if applicable)

Borders must be visible, lines must be straight, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns must be visible in all four corners and the centre, the 5% and 95% pixel value squares must be clearly visible, and no disturbing artefacts must be visible on the printed TG18-QC test pattern

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11.The mid density (MD) and density difference (DD) = baseline ± 0.15

Base + fog (B+F) = baseline ± 0.03 and ≤0.25

Dmax = baseline ± 0.10 and ≥3.4

Monthly for dry lasers and daily or as used for wet lasers

Print the TG18-QC test pattern

Check visibility and distortion of several items used for evaluating the quality of the image

Check for disturbing artefacts

Measure MD, DD, B+F and Dmax

Control charts and records showing

 Date test was performed

 Person performing test

 Printer identification

 Test results

Mechanical inspection

Indicated breast thickness accurate to ±5 mm

No hazardous, inoperative, out of alignment or improperly operating items on the system

All items listed on the visual check list have received a pass

Monthly

Confirm accuracy of thickness indication

Visual inspection of the system to ensure safe and optimum operation

Checklist/logbook entry showing

 Date inspection performed

 Inspection results

 Person performing test

Repeat analysis

Repeat rate <3% (<2% preferred) [3]

Quarterly

Inclusion of images from at least 250 consecutive client examinations

The ability to determine repeat rates attributable to a range of equipment faults and positioning errors

Worksheet/logbook entries showing all results/calculations

Image receptor homogeneity

Maximum deviation in mean pixel value in ROI <±10% of mean pixel value in central ROI

Maximum variation of the mean pixel value in central ROI between successive quarterly images <±10%

Quarterly or more frequently if recommended by the manufacturer

Use manufacturer’s protocol and test block if available; otherwise

Image a standard test block at clinical settings

On the “for processing” image, draw 100 mm2 square or circular ROIs in the centre and four corners

If the mean pixel value of a ROI deviates by more than 15% from the mean pixel value in the central ROI, the detector gain map may require re-calibration

If required, to exclude failure due to non-uniformities in the standard test block, rotate latter by 180° and repeat measurement

Records showing

 Date test was performed

 Person performing test

 X-ray system identification

 kVp, target/filter, density setting and mAs

 Test results

AEC calibration test

Mean pixel value for each of 2, 4 and 6 cm PMMA within 10% of baseline values

Quarterly

Assess for both contact and magnification modes

Use PMMA thickness between 2 and 6 cm covering complete image receptor

Use clinical AEC settings (kVp, target/filter and mode)

Measure mean pixel value in 4 cm2 ROI positioned centrally along axis and 6 cm from chest wall

Examine image for clinically significant artefacts

Records showing

 Date test was performed

 Person performing test

 X-ray system identification

 kVp, target/filter, AEC mode and mAs

 Test results

Compression

Maximum motorised compression force in range 150–200 N

6 monthly

Confirm machine indicated compression force meets requirements

Checklist/logbook entry showing

 Date test performed

 Test results

 Person performing test

Test equipment quality control

 Densitometer calibration check

Optical density measurement accurate to within

 ±0.03 (0–3.0 OD)

 ±3% (3.0–4.0 OD)

6 monthly

Verification of accuracy using an optical density calibration strip traceable to an accepted standard

Checklist/logbook entry showing

 Date test performed

 Test results

 Person performing test

Maintenance and fault logging

Separate logbooks for each imaging system including diagnostic monitors and film printer if relevant

As required

Dated entries describing fault encountered and/or maintenance performed

Logbooks with dated and initialled entries

Infection control of breast imaging equipment

Clean equipment

Before each examination

Cleaning using alcohol swipes, or as per manufacturer’s recommendations and/or suitable infection control advice

Nil

  1. aAll written/electronic QC records should be retained for a minimum of 1 year unless otherwise indicated by local Regulatory requirements. Images used to assess image quality with the ACR Accreditation or ACR DM phantom should be retained for a minimum of 1 month

Appendix 2: Summary of recommendations for facility qc procedures for CR units

Procedure

Recommended control-limits/requirements

Minimum frequency

Key procedure elements

Recommendations for record keepinga

Viewing conditions

Appropriate viewing conditions

All viewbox lamps must be operational and appropriate masking available

Daily

Visual inspection of ambient lighting conditions to ensure conformance with acceptable viewing condition configuration (see text for detail)

Visual inspection of viewboxes for uniformity of brightness

Confirmation of presence and operation of masking for viewboxes

Checklist/logbook entry showing

 Date performed

 Person performing task

Image plate erasure

Erasure of energy absorbed from scattered radiation or naturally occurring radiation by CR image plates before they are used

Daily/weekly

On a daily basis or if left unused for more than 8 hours, all CR image plates must be subjected to an erasure (following manufacturer’s instructions)

On a weekly basis all Fuji CR image plates must be subjected to a primary erasure

Logbooks with dated and initialled entries

Monitor/viewboxes cleaning

Monitor screens and viewboxes must be free of dust, fingerprints and other marks that might interfere with image interpretation

Weekly

Clean all monitor screens and viewboxes gently with lint-free cloth as per manufacturer’s instructions

Checklist/logbook entry showing

 Date performed

 Person performing task

Monitor QC (monitors used for interpretation and attached to the acquisition workstation)

Borders must be visible, lines must be straight, squares must appear square, the ramp bars should appear continuous without any contour lines, there should be no smearing or bleeding at black-white transitions, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns and two low contrast patterns must be visible in all four corners and the centre, the 5 and 95% pixel value squares must be clearly visible, pattern must be centred in the active area and no disturbing artefacts must be visible on the displayed TG18-QC test pattern

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11

Weekly

Display TG18-QC test pattern

Ensure viewing conditions are acceptable

Use window-width set to maximum and window-level set to half of maximum

Records showing

 Date test was performed

 Person performing test

 Monitor identification

 Monitor settings

 Test results

Monitor cleaning

Monitor screens must be free of dust, fingerprints and other marks that might interfere with image interpretation

Weekly

Clean all monitor screens gently with lint-free cloth as per manufacturer’s instructions

Checklist/logbook entry showing

 Date performed

 Person performing task

Printer area cleanliness (if applicable)

Clean and dust free environment

Weekly

Wet cleaning of printer area floor and open shelves. Inspect and clean air intake filters on the film printer

Checklist/logbook entry showing

 Date performed

 Person performing task

Image quality evaluation

mAs = baseline ± 10%

Dose to plate = baseline ±10%

Exposure indicator (see Appendix 6 for manufacturer dependent tolerances)

For hard copy reporting optical density = baseline ±20% and must be in the range of 1.60–2.0

The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or

The ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of an ACR DM phantom

Weekly

Obtaining the phantom image

 Use an ACR Accreditation phantom or the new ACR DM phantom

 Use of a designated test cassette and imaging plate that is in routine clinical use

 Use of a consistent AEC detector position where this is manually selected

 Light contact between the compression paddle and the phantom surface

 Consistent positioning of the phantom

 Consistent selection of clinically relevant kVp and target/filter combinations

 Selection of the density setting in current clinical use

 Consistent time delay between plate irradiation and readout

Evaluating the phantom image (preferably on reading workstation or on printed copy if hardcopy reporting used)

 Use “for presentation” image with zoom and modest adjustment of window/level functions to score fibres and specks

 Use of consistent viewing conditions that reflect those used to read actual mammograms. This applies to both soft and hard copy

 Image quality scoring by the same person, if possible

 Measure optical density in reproducible part of phantom image if hardcopy reporting

 Use of a control chart to record results

Record numerical mAs values and image quality scores

Control chart showing

 Plots of mAs, exposure indicator, image quality score/s and OD if applicable

≥25 results

 Clearly marked control limits

 Baseline values

 Radiographic settings (kVp, target/filter combination, density setting and SID)

 Remarks e.g. corrective action

Phantom images identifying

 Date

 The X-ray system

 The technique factors

Printer QC

Borders must be visible, lines must be straight, all corner patches must be visible, squares of different shades from black to white must be distinct, all high contrast resolution patterns must be visible in all four corners and the centre, the 5% and 95% pixel value squares must be clearly visible, and no disturbing artefacts must be visible on the printed TG18-QC test pattern

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11

The mid density (MD) and density difference (DD) = baseline ± 0.15

Base + fog (B+F) = baseline ± 0.03 and ≤0.25

Dmax = baseline ± 0.10 and ≥3.4

Monthly for dry lasers and daily or as used for wet lasers

Print the TG18-QC test pattern

Check visibility and distortion of several items used for evaluating the quality of the image

Check for disturbing artefacts

Measure MD, DD, B+F and Dmax

Control charts and records showing

 Date test was performed

 Person performing test

 Printer identification

 Test results

Mechanical inspection

Indicated breast thickness accurate to ±5 mm

No hazardous, inoperative, out of alignment or improperly operating items on the system

All items listed on the visual check list have received a pass

Monthly

Confirm accuracy of thickness indication

Visual inspection of the system to ensure safe and optimum operation

Checklist/logbook entry showing

 Date performed

 Inspection results

 Person performing task

Repeat analysis

Repeat rate <3% (<2% preferred) [3]

Quarterly

Inclusion of images from at least 250 consecutive client examinations

The ability to determine repeat rates attributable to a range of equipment faults and positioning errors

Worksheet/logbook entries showing all results/calculations

Image receptor homogeneity

Maximum difference in mean pixel value between any two ROIs <±10%

Maximum variation of the mean pixel value in central ROI between successive QC images <±10%

Quarterly or more frequently if recommended by the manufacturer)

Use manufacturer’s protocol and test block if available; otherwise

Image a standard test block at clinical settings

Use ‘test’ cassette

Perform measurements on the “for processing” (unprocessed) image, if possible, using a 100 mm2 square or circular ROI. Three ROIs are placed at the left, right and centre on a line 20 mm back from chest wall

If the mean pixel value of any two ROIs differ by more than 10% from each other, the CR unit’s shading correction may require re-calibration or imaging plate(s) may require replacement

If ROI analysis is not possible, do a visual inspection at narrow window width

If required, to exclude failure due to non-uniformities in the standard test block, rotate by 180° and repeat measurement

Records showing

 Date test was performed

 Person performing test

 X-ray system identification

 kVp, target/filter, density setting and mAs

 Test results

AEC calibration test

Dose to plate for each of 2, 4 and 6 cm PMMA = baseline ±10%

See Appendix 6 for equivalent manufacturer specific exposure index requirements

Quarterly

Assess for both contact and magnification modes

Use PMMA thickness between 2 and 6 cm covering complete cassette

Use clinical AEC settings (kVp, target/filter and mode including density setting)

Use a designated test cassette and imaging plate that is in routine clinical use

Use a consistent AEC detector position where this is manually selected

Consistent positioning of the PMMA

Consistent time delay between plate irradiation and readout

Records showing

 Date test was performed

 Person performing test

 X-ray system identification

 kVp, target/filter, AEC mode and mAs

 Test results

Compression

Maximum motorised compression force in range 150–200 N

6 monthly

Confirm machine indicated compression force meets requirements

Checklist/logbook entry showing

 Date test performed

 Test results

 Person performing test

Test equipment quality control

 Densitometer calibration check

Optical density measurement accurate to within

±0.03 (0–3.0 OD)

±3% (3.0–4.0 OD)

6 monthly

Verification of accuracy using an optical density calibration strip traceable to an accepted standard

Checklist/logbook entry showing

 Date test performed

 Test results

 Person performing test

Cassette/image plate condition and inter plate sensitivity variation

Clean and dust free cassettes and image plates

No major inhomogeneities on the images

See Appendix 6 for manufacturer specific tolerances on interplate variations

6 monthly

Cassette/image plate cleaning as per manufacturer’s recommendations

Image a standard test block at clinical settings

Pre-processing should be turned off as much as possible and no post processing must be applied

Evaluate for artefact on both film (if applicable) and monitor

Records showing

 Date test was performed

 Person performing test

 kVp, target/filter, AEC mode

 Exposure indicator and mAs for each plate

Maintenance and fault logging

Separate logbooks for each imaging system, including diagnostic monitors, and film printer if relevant.

As required

Dated entries describing fault encountered and/or maintenance performed

Logbooks with dated and initialled entries

Infection control of breast imaging equipment

Clean equipment

Before each examination

Cleaning using alcohol wipes, or as per manufacturer’s recommendations and/or suitable infection control advice

Nil

  1. aAll written/electronic QC records should be retained for a minimum of 1 year unless otherwise indicated by local Regulatory requirements. Images used to assess image quality with the ACR Accreditation or ACR DM phantom should be retained for a minimum of 1 month

Appendix 3: Summary of recommendations for medical physics testing only at acceptance or equipment upgrade in 2D mode

Procedure

Performance requirements/guidelines

Routine testing guidelines

Key procedure elements

Focal spot

≥11 lp/mm for line-pair bars perpendicular to anode–cathode axis and

≥13 lp/mm for line-pair bars parallel to anode–cathode axis

OR complies with IEC 60336 [38] for 0.3 and 0.1 mm focal spot sizes

Not required unless tube has been changed

As per “Focal spot size” section or IEC 60336 [38]

Leakage radiation

≤1 mGy/h at 1m from focus and ≤0.01 mGy/100 mAs @ 30 kVp and 30 cm from focus

Not required unless tube has been changed or system relocated

As per AS/NZS IEC 60601.1.30.1.3 [39]

Transmission through breast support

≤0.001 mGy @ max kVp and mAs

Not required unless change made to image receptor system

As per AS/NZS IEC 60601.1.3 [39]

Missed tissue @ chest wall

Width of missed tissue at chest wall ≤5 mm in contact mode and ≤7 mm in magnification mode

Not required unless tube has been changed or change made to image receptor system or system relocated

 

Plate fogging (CR only)

Image of coin must not be visible

Not required unless changes in storage of cassettes have occurred

Monitor during acceptance testing

MTF

Bench mark testing, compare to manufacturer’s specification

Not required unless tube has been changed or change made to image receptor system

As per IEC 62220-1-2 [48]

Threshold contrast visibility

 

Not required unless tube has been changed or change made to image receptor system

Use CDMAM phantom

Spatial linearity and geometric distortion

 

Not required unless change made to image receptor system

Use wire mesh tool

Distance calliper accuracy

Measured dimensions of ruler in image must be within 2% of true dimensions in plane specified by manufacturer

Check both contact and magnification modes

Not required unless change made to image receptor system

Use steel rulers. Determine dimensions in image, ideally using reporting workstation

  1. This test allows digital systems to be benchmarked against European standards [9]

Appendix 4: Summary of recommendations for medical physics annual testing of DR units in 2D mode

Procedure

Performance requirements/guidelines

Routine testing guidelines

Acceptance and additional tests

Mammography unit assembly evaluation

Correct and safe function of system components. Thickness display accuracy within ±5 mm, note: Flexi paddles will not comply (manufacturer recommendation varies ~11–12 mm for flexi paddles). Reproducible to 2 mm. Verify DICOM image header for correct display of parameters

Confirm function of all motorised components, warning lights, displays etc. Evaluate system for any miscellaneous safety risks etc. DICOM verification required after software upgrades

As per routine tests

Collimation and alignment assessment

 X-ray field/Image receptor alignment

The X-ray field shall irradiate the image receptor fully but not extend beyond the breast support on the chest wall edge of the image receptor by more 2 mm

Assess alignment for each target/geometry combination

As per routine tests

 Paddle/Image alignment

The chest wall edge of the compression paddle shall be aligned just beyond the chest wall edge of the image receptor such that it does not appear in the image. In addition, the compression paddle shall not extend beyond the chest wall edge of the image by more than 1% of the SID

Assess alignment for all clinically relevant Bucky/paddle/target/geometry combinations

As per routine tests

System resolution/MTF

Compare to baseline values, variation less than 10%

Measure MTF using system software if possible. Otherwise measure limiting resolution

 Use a 4cm PMMA block or equivalent

 Place resolution pattern on PMMA

 Measure both parallel and perpendicular to chest wall

 Repeat for Magnification mode if applicable

Establish base line values

AEC system performance assessment

 Reproducibility

Coefficient of variation (COV) for both absorbed dose and mAs for at least three phototimed exposures of a test object shall be better than or equal to 0.05

Use a 4cm PMMA block or equivalent

Assess COV for each AEC detector at a typical clinical kVp

As per routine tests

 Compensation and SDNR system performance assessment

Compare SDNR values to baseline and to the minimum acceptable values for 4 cm PMMA (SDNRaccept)

 SDNR2cm > 1.1 × SDNRaccept

 SDNR4cm > SDNRaccept

 SDNR6cm > 0.9 × SDNRaccept

Assess the most commonly used AEC modes for contact and magnification geometry

Use 0.2 mm Al foil as contrast test tool and measure SDNR for 2, 4 and 6 cm PMMA (also see section on glandular dose). Note: measurements are to be undertaken on “for processing” (unprocessed) images

Establish baseline values. Assess all available AEC modes for contact and magnification geometries

 Density control (if applicable)

The density control must be capable of changing the mAs from the value used normally by −25 to +50%

Assess change in mAs for at least two density settings either side of the usual clinical setting using 4 cm of PMMA

Assess change in mAs across full range of density settings

 Back-up timer/security cut-out

Security cut-out mechanisms shall be present and terminate the exposure within 50 ms or within 5 mAs, otherwise the back-up timer should terminate the exposure at ≤500 mAs and must terminate the exposure at ≤800 mAs

Use lead sheet or other heavily attenuating material to intercept beam and confirm that the back-up timer security cut-out functions within specified limits

Confirm that the back-up timer/security cut-out functions within specified limits

Image uniformity and artefact

Max. deviation of mean pixel value <±15% of mean pixel value for central ROI

Max. deviation in SNR as a function of time is ±10%

There must be no evidence of blotches or regions of altered noise appearance, observable grid lines or breast support structures, bright or dark pixels

Assess for 40 mm PMMA covering complete detector

Use five ROIs (one central, with the other 4 approximately 20 mm from any edge) each of 100 mm2

Measurements performed on unprocessed image

Exclude phantom non uniformity by rotating block 180° and repeating

Repeat in magnification mode if applicable

Assess also at 20 and 60 mm

Detector element failure

Limits currently not established. Must monitor independent of manufacturer

Inspect bad pixel map

A mammographic screen-film mesh can be used to determine if correction for bad columns successful

Bad pixel map must be available at any time, independent of manufacturer

Image quality evaluation

The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or

the ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of the ACR DM phantom

Additionally, with the ACR DM phantom the SDNR with contrast object ≥2.0

Use typical clinical settings

Measure MPVs and SDs in relevant ROIs of ACR DM phantom so that SDNR may be calculated.

As per routine testing

Ghost image evaluation

“Ghost image” factor <2.0

Assess using 40 mm PMMA (see “Ghost image evaluation” section for testing guidelines)

As per routine testing

System linearity and noise analysis

Linearity plot versus ESAK: R2 > 0.99

SD2 plot versus MPV: R2 > 0.99

Noise parameters: compare to baseline results

Use standard test block (e.g. 4 cm PMMA) at typical clinical beam settings

Measure ESAK at 6 cm from chest wall

Measure mean pixel value and SD in ROI placed 6 cm from chest wall

Plot mean pixel value as a function of ESAK

Plot SD2 as a function of MPV corrected for any pixel offset

Baseline measurements at clinical kVp, also at max and min clinical kVps for all target filter combinations

Generator performance

 kVp, reproducibility

COV ≤ 0.02 for a minimum of three exposures

Assess kVp reproducibility at a typical clinical kVp value

As per routine testing

 kVp accuracy

Measured kVp shall be within ±5% of the specified value over the clinically relevant range

Assess kVp accuracy over the clinically relevant range in, at most, 2 kVp increments Note: The kVp need only be verified for one target filter combination per kVp, however the kVp meter must be calibrated for that particular target/filter combination

Assess kVp accuracy over clinically relevant range in 1 kVp increments

Beam quality

\(\left( {{\text{kVp}}/100} \right)+0.03 \leq {\text{HVL}}<\left( {{\text{kVp}}/100} \right)+{\text{C}}\) \(\begin{aligned} {\text{where C }}&=&0.{\text{12 mm Al for Mo}}/{\text{Mo}} \hfill \\&=&0.{\text{19 mm Al for Mo}}/{\text{Rh}} \hfill \\&=&0.{\text{22 mm Al for Rh}}/{\text{Rh}} \hfill \\&=&0.{\text{23 mm Al for Rh}}/{\text{Ag}} \hfill \\&=&0.{\text{3}}0{\text{ mm Al for W}}/{\text{Rh}} \hfill \\&=&0.{\text{32 mm Al for W}}/{\text{Ag}} \hfill \\&=&0.{\text{25 mm Al for W}}/{\text{Al}}. \hfill \\ \end{aligned}\)

Measure the HVL required for mean glandular dose calculations and for establishing compliance with DRLs (see “Mean glandular dose” section for details)

As per routine tests plus measure HVL at 28 kVp for all target/filter combinations, with the compression paddle removed if unit used for biopsy purposes with open paddle

Mean glandular dose

≤2.0 mGy for a 4.2 cm 50% adipose, 50% glandular breast (i.e. ACR Accreditation phantom or ACR DM phantom)

<1 mGy for 2.0 cm PMMA (2.3 cm 50% adipose, 50% glandular breast)

<4.5 mGy for 6.0 cm PMMA, (6.5 cm 50% adipose, 50% glandular breast)

the displayed MGD values must agree with calculated values to ≤25%

Assess for an AEC controlled exposure using typical clinical settings using ACR Accreditation phantom (or ACR DM phantom) and also for 20 mm and 60 mm PMMA

Additional dose (and HVL) measurements may be necessary to confirm compliance with DRLs (see “Mean glandular dose”)

As per routine tests

Exposure time

For all clinically relevant SID settings the maximum exposure time when irradiating 6 cm PMMA must be less than 3.5 and 2 s for fine and broad focus, respectively

Assess for both contact and magnification modes

Use 6 cm of PMMA

Use clinically relevant technique factors for this PMMA thickness consistent with SDNR and MGD measurements

Record mAs and infer the exposure time from tube rating or measure directly using a manual exposure matched to mAs needed for AEC initiated exposure

As per routine tests

Viewbox luminance and room illuminance (hardcopy only)

Viewing area illuminance ≤50 lx

Viewbox luminance ≥3000 cd/m2

Assess viewing conditions for all viewers

As per routine tests

Monitor luminance and viewing conditions

Image interpretation must not be done on a monitor of less than 4.2 mega pixels

Luminance ratio approximately 350:1

Maximum luminance >450 cd/m2 and maximum luminance of paired monitors matched to ≤5%

Minimum luminance preferably not less than 1 cd/m2

Ambient light <20 lx

In PACS situations images must be stored with lossless compression

Measure luminance ratio under clinical lighting conditions

Confirm luminance uniformity

Confirm no cross-talk and pixel defects

As per routine testing with the additional requirement of checking GSDF

Monitor or workstation may have comprehensive QC program which needs to be validated

Monitor performance

No smearing artefact, ramps without terracing

Lines straight, boxes square, active display centred, borders complete

Squares of different shades from black to white must be distinct and small squares in corners of each clearly discernible

Free from artefact

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11

Test patterns to be displayed at full resolution

Test under clinical lighting conditions

Use TG18-QC test pattern

As per routine testing. Monitor or workstation may have comprehensive QA program

Printer (hardcopy only)

B + F = baseline ± 0.03 and ≤0.25 OD

Dmax = baseline ± 0.10 and ≥3.4 OD

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11

Print TG18-QC test pattern as per weekly printer QC test

As per routine tests

Appendix 5: Summary of recommendations for medical physics annual testing of CR units

Procedure

Performance requirements/guidelines

Routine testing guidelines

Acceptance and additional tests

Mammography unit assembly evaluation

Correct and safe function of system components. Thickness display accuracy within ±5 mm, reproducible to 2 mm. Verify DICOM image header for correct display of parameters

Confirm function of all motorised components, warning lights, displays etc. Evaluate system for any miscellaneous safety risks etc. DICOM verification required after software upgrades

As per routine tests

Collimation and alignment assessment

 X-ray field/image/breast-support alignment

The X-ray field shall irradiate the image receptor fully but not extend beyond the breast support on the chest wall edge of the image receptor by more than 2 mm

Assess alignment for largest collimator in clinical use for each Bucky/target combination. For magnification geometry only assess chest wall alignment

As per routine tests

 Paddle/image alignment

The chest wall edge of the compression paddle shall be aligned just beyond the chest wall edge of the image receptor such that the chest wall compression paddle does not appear in the image. In addition the compression paddle shall not extend beyond the chest wall edge of the image receptor by more than 1% of the SID

Assess alignment for all clinically relevant Bucky/paddle/geometry combinations

As per routine tests

System resolution/MTF

Compare to baseline values, variation less than 10%

Measure MTF using system software if possible. Otherwise measure limiting resolution

 Use a 4cm PMMA block or equivalent

 Place resolution pattern on PMMA

 Measure both parallel and perpendicular to chest wall

 Repeat for Magnification mode if applicable

Establish base line values

AEC system performance assessment

 Reproducibility

Coefficient of variation (COV) for both absorbed dose and mAs for at least three phototimed exposures of a test object shall be better than or equal to 0.05

Use a 4cm PMMA block or equivalent

Assess COV for each AEC detector at a typical clinical kVp

As per routine tests

 Compensation and SDNR system performance assessment

Compare SDNR values to baseline and to the minimum acceptable values for 4 cm PMMA (SDNRaccept)

 SDNR2cm > 1.1 × SDNRaccept

 SDNR4cm > SDNRaccept

 SDNR6cm > 0.9 × SDNRaccept

Note: for magnification mode this last requirement is relaxed to

 SDNR6 cm > 0.65 × SDNRaccept

Assess the most commonly used AEC modes for contact and magnification geometry

Use clinical AEC settings (kVp, target/filter and mode including density setting)

Use a designated test cassette and imaging plate that is in routine clinical use

Use a consistent AEC detector position where this is manually selected

Use 0.2 mm Al foil as contrast test tool and measure SDNR for 2, 4 and 6 cm PMMA (also see section on glandular dose). Note: measurements are to be undertaken on “for processing” (unprocessed) image

Consistent time delay between plate irradiation and readout

Record exposure indicator for each PMMA thickness

Measure film density for each image if applicable

Establish baseline values. Assess all available AEC modes for contact and magnification geometries. Assess both 18 × 24 cm2 and 24 × 30 cm2 Buckys

 Density control

The density control must be capable of changing the mAs from the value used normally by −25 to +50%

Assess change in mAs for at least two density settings either side of the usual clinical setting using 4 cm of PMMA

Assess change in mAs across full range of density settings

 Back-up timer/security cut-out

Security cut-out mechanisms shall be present and terminate the exposure within 50 ms or within 5 mAs, otherwise the back-up timer should terminate the exposure at ≤500 mAs and must terminate the exposure at ≤800 mAs

Use lead sheet or other heavily attenuating material to intercept beam and confirm that the back-up timer/security cut-out functions within specified limits

Confirm that the back-up timer/security cut-out functions within specified limits

Image uniformity and artefact

Max. deviation of mean pixel value <±10% of mean pixel value for central ROI

Max. deviation in SNR of central ROI as a function of time is ±10%.

No major inhomogeneities on the images

Assess for 40 mm PMMA covering complete CR plate

Use three ROIs each of ~100 mm2 placed on a line parallel to and approximately 20 mm from chest wall

Assess also at 20 and 60 mm

Uniformity of cassette/image plate response

Maximum mAs variation <±5% between all plates of one size

Maximum mAs variation <±20% between plates of different sizes

See Appendix 6 for manufacturer dependent allowed tolerances on the exposure indicator

Assess for 40 mm PMMA covering complete CR plate

As per routine testing

Image quality evaluation

The ability to clearly visualise 5 fibres, 3.5 speck groups (4 is desirable) and 4 masses in an image of an ACR Accreditation phantom or the ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of the ACR DM phantom

Additionally, with the ACR DM phantom the SDNR with contrast object ≥2.0

Use typical clinical settings

Measure MPVs and SDs in relevant ROIs of ACR DM phantom so that SDNR may be calculated

As per routine testing

Ghost image evaluation

“Ghost image” factor <2.0

Assess using 40 mm PMMA (see “Ghost image evaluation” section for testing guidelines)

As per routine testing

System linearity and noise analysis

Compare to baseline results and note requirement for linearity (see text and Appendix 6) has R2 > 0.99

Noise analysis remains optional

Use standard test block (e.g. 4 cm PMMA) at typical clinical beam settings

Use the same cassette/image plate for all exposures

Record exposure indicator

Plot exposure indicator as a function of ESAK (see Appendix 6)

Baseline measurements at clinical kVp, also at max and min clinical kVps for all target filter combinations

Generator performance

 kVp, reproducibility

COV ≤ 0.02 for a minimum of three exposures

Assess kVp reproducibility at a clinical kVp value

As per routine testing

 kVp accuracy

Measured kVp shall be within ±5% of the specified value over the clinically relevant range

Assess kVp accuracy over the clinically relevant range in, at most, 2 kVp increments

Assess kVp accuracy over clinically relevant range in 1 kVp increments

Beam quality

\(\left[ {\left( {{\text{kVp}}/100} \right)+0.03} \right] \leq {\text{HVL}}<\left[ {\left( {{\text{kVp}}/100} \right)+{\text{C}}} \right]\) \(\begin{aligned} {\text{where C }}&=&0.{\text{12 mm Al for Mo}}/{\text{Mo}}\\ &=&0.{\text{19 mm Al for Mo}}/{\text{Rh}}\\&=&0.{\text{22 mm Al for Rh}}/{\text{Rh}}\\&=&0.{\text{23 mm Al for Rh}}/{\text{Ag}}\\&=&0.{\text{3}}0{\text{ mm Al for W}}/{\text{Rh}} \\ &=&0.{\text{32 mm Al for W}}/{\text{Ag}} \\ &=&0.{\text{25 mm Al for W}}/{\text{Al}}. \\ \end{aligned}\)

Measure the HVL required for Mean Glandular Dose calculations and for establishing compliance with DRLs (see “Mean glandular dose” for details)

As per routine tests plus measure HVL at 28 kVp for all target/filter combinations, with the compression paddle removed if unit used for biopsy purposes with open paddle

Mean glandular dose

≤2.0 mGy for a 4.2 cm 50% adipose, 50% glandular breast (i.e. ACR Accreditation phantom or ACR DM phantom)

<1 mGy for 2.0 cm PMMA (2.3 cm 50% adipose, 50% glandular breast)

<4.5 mGy for 6.0 cm PMMA, (6.5 cm 50% adipose, 50% glandular breast)

Assess for an AEC controlled exposure using typical clinical settings using ACR Accreditation phantom (or ACR DM phantom) and also for 20 mm and 60 mm PMMA

Additional dose measurements may be necessary to confirm compliance with DRLs (see “Mean glandular dose”)

As per routine tests

Exposure time

For all clinically relevant SID settings the maximum exposure time when irradiating 6 cm PMMA must be less than 3.5 and 2 s for fine and broad focus, respectively

Assess for both contact and magnification modes

Use 6 cm of PMMA

Use clinically relevant technique factors for this PMMA thickness consistent with SDNR and MGD measurements

Record mAs and infer the exposure time from tube rating or measure directly using a manual exposure matched to mAs needed for AEC initiated exposure

As per routine tests

Viewbox luminance and room illuminance (hardcopy only)

Viewing area illuminance ≤50 lx

Viewbox luminance ≥3000 nit

Assess viewing conditions for all viewers

As per routine tests

Monitor luminance and viewing conditions

Image interpretation must not be done on a monitor of less than 4.2 mega pixels

Luminance ratio approximately 350:1

Maximum luminance >450 cd/m2 and maximum luminance of paired monitors matched to ≤5%

Minimum luminance preferably not less than 1 cd/m2

Ambient light <20 lx

In PACS situations images must be stored with lossless compression

Measure luminance ratio under clinical lighting conditions

Confirm luminance uniformity

Confirm no cross-talk and pixel defects

As per routine testing with the additional requirement of checking GSDF

As per routine testing. Monitor or workstation may have comprehensive QC program which needs to be validated

Monitor performance

No smearing artefact, ramps without terracing

Lines straight, boxes square, active display centred, borders complete

Squares of different shades from black to white must be distinct and small squares in corners of each clearly discernible

Free from artefact

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11

Test patterns to be displayed at full resolution

Test under clinical lighting conditions

Use TG18-QC test pattern

As per routine testing. Monitor or workstation may have comprehensive QA program

Printer (hardcopy only)

B + F = baseline ± 0.03 and ≤0.25 OD

Dmax = baseline ± 0.10 and ≥3.4 OD

The number of letters visible in the phrase “Quality Control” for the dark, mid-gray and light renditions must be ≥11

Print TG18-QC test pattern as per weekly printer QC test

As per routine tests

Appendix 6: Summary of recommendations for facility QC for biopsy units

Procedure

Minimum frequency

Fully integrated digital/biopsy unit

Digital mammography with add on image system

Stand alone biopsy system

Viewing conditions

Weekly

Previously covereda

Previously covered

See “Viewing conditions” section

Monitor QC

Weekly

Previously covered

ssab

See “Monitor QC” section

Monitor cleaning

Weekly

Previously covered

ssa

See “Monitor/viewbox cleaning” section

Image quality evaluation

Weekly

ssa

ssa

See “Image quality evaluation” section Note may use ACR ‘mini’ digital stereotactic phantom—see text

Printer QC (if applicable)

Weekly

Previously covered

ssa

See “Printer QC” section

Mechanical inspection

Monthly

ssa

ssa

See “Mechanical inspection and breast thickness indication” section. Note additionally image receptor and compression plate/ biopsy window must be free of wobble; Vernier drive and needle guide rigid and wobble free, localisation system zeros and biopsy device properly immobilised—see text

Repeat analysis

Quarterly

ssa

ssa

See “Repeat analysis” sectin

Image receptor homogeneity

Quarterlyc

Previously covered

ssa

See “Image receptor homogeneity” section. Note; procedure should be modified as seen in text

AEC calibration test

Quarterly

ssa

ssa

See “AEC calibration test” section. Procedure may vary for different types of units—see text

Compression

6 monthly

ssa

ssa

See “Compression” section

Test equipment quality control

Densitometer calibration check

6 monthly

Previously covered

Previously covered

See “Test equipment calibration” section

Maintenance and fault logging

As required

ssa

ssa

See “Maintenance and fault logging” section

Infection control of breast imaging equipment

Before each examination

Previously covered

Previously covered

See “Infection control of breast imaging equipment” section

Stereotactic Accuracy confirmation

Prior to first use on day of procedures

ssa

ssa

Localisation within ±1 mm. Procedure as per manufacturer’s recommendations; Checklist/logbook entry showing

 Date test performed

 Test results

 Person performing test

  1. aTest previously completed as part of mammography tests
  2. b Ssa see stand alone biopsy units
  3. cOr more frequently if recommended by the manufacturer

Appendix 7: Summary of recommendations for medical physics testing for biopsy units

Procedure

Frequency

Fully integrated digital/biopsy unit

Digital mammography with add on image system

Stand alone biopsy system

Focal spot

Acceptance

Previously covereda

Previously covered

See “Focal spot size” section

Leakage radiation

Acceptance

Previously covered

Previously covered

See “Leakage radiation” section

MTF

Acceptance

Previously covered

ssab

See “Modulation transfer function” section

Spatial linearity and geometric distortion

Acceptance

Previously covered

ssa

See “Spatial linearity and geometric distortion” section

Distance calliper accuracy

Acceptance

Previously covered

ssa

See “Distance calliper accuracy” section

Mammography unit assembly evaluation

Annual

ssa

ssa

See “Mammography unit assembly evaluation” section. Note additionally ensure X-ray tube angular locations positively locked; image receptor and compression plate/ biopsy window free of wobble; Vernier drive and needle guide rigid and wobble free, localisation system zeros; biopsy device properly immobilised and AEC chart displayed—see text

Collimation assessment

Annual

ssa

ssa

FOV defined by biopsy window and is aligned centrally with digital image receptor, with tolerances of ±5 mm—see text

System resolution

Annual

Previously covered

ssa

See “System resolution/MTF” section

AEC/SDNR

Annual

Previously covered

ssa

See “Automatic exposure control system performance assessment/signal difference to noise ratio” section. Note: technique charts should be consulted for correct factor settings. Minimum PMMA thickness of 2 cm used for SDNR see text

Image uniformity and artefact evaluation

Annual

Previously covered

ssa

See “Image uniformity and artefact evaluation” section .Note—ROIs to be in corners of image, 10 mm from edge

Image quality evaluation

Annual

Previously covered

ssa

See “Image quality evaluation” section .Note may use ACR ‘mini’ digital stereotactic phantom—see text for revised scoring

Ghost image evaluation

Annual

Previously covered

ssa

See “Ghost image evaluation” section

System linearity and noise analysis

Annual

Previously covered

Previously covered

See “System linearity and noise analysis” section

kVp performance

Annual

Previously covered

Previously covered

See “Generator performance” section

HVL

Annual

Previously covered

Previously covered

See “Beam quality or half value layer” section

Mean glandular dose

Annual

ssa

ssa

See “Mean glandular dose” section—note; see technique chart for factors used in dose calculations

Exposure time

Annual

Previously covered

Previously covered

See “Exposure time” section

Viewbox and room luminance

Annual

ssa

ssa

See “Viewbox luminance and room illuminance (hardcopy only)” section

Monitor performance

Annual

Previously covered

ssa

See “Monitor luminance and viewing conditions” section

Printer (hardcopy)

Annual

Previously covered

ssa

See “Printer (hardcopy)” section

Localisation accuracy test

Annual

ssa

ssa

 
  1. aTest previously completed as part of mammography tests
  2. bSee stand alone biopsy units

Appendix 8: Summary of criteria in terms of CR exposure indicators

A number of companies currently manufacture CR units for use in mammography and they have developed unique exposure indicators. Reviews of these indicators, with a comparison between the different manufacturers, have been reported in the literature [9, 87]. The table below can be used to indicate the test criteria that should be applied in terms of the current CR exposure indicators.

Test

Test criteria

Tolerance in terms of CR exposure indicator

Fuji, Philips and Konica

Kodak (Carestream)

Agfaa

Image quality evaluation

Air kerma (dose) to the plate must not change by greater than ±10%

±10% in S# of baseline

±40 units in EI of baseline

±5% in SAL or ±430 in SAL log or ±580 in PVL log16 of baseline

AEC calibration test

Air kerma (dose) to the plate for each of the three thicknesses of PMMA be within ±10% of the baseline value for each thickness

±10% in S# of baseline for each thickness

±40 units in EI of baseline for each thickness

±5% in the SAL, or ±430 in SAL log or ±580 in PVI log16 of baseline for each thickness

Cassette image plate condition and interplate sensitivity variation (also “Image uniformity and artefact evaluation

Air kerma (dose) to individual plate must differ from mean for that size by less than ±5%

Difference in mean air kerma (dose) to plates of different sizes <20%

S# for individual plates must be within ±5% of mean for same size

S# difference for two different plate sizes <20%

EI for individual plate must be within ±20 units of mean for same size

EI difference for two different plate sizes <100 units

SAL for individual plates must be within ±2.5% or SAL log must be within ±220 or PVI log16 must be within ±290 of mean for same size

SAL difference <10% or SAL log difference <1000, or PVI log16 <1300 for two different plate sizes

System linearity and noise analysis

R2 value of appropriate plot of exposure indicator versus ESAK must be >0.99

Plot S# versus reciprocal of ESAK

Plot EI versus log (ESAK)

Plot SAL versus SQRT(ESAK) or SAL log versus log(ESAK) or PVI log 16 versus log(ESAK)

Exposure indicator calibration and image fading

Under specified conditions (see Table 4) Exposure Indicator must meet criteria outlined in columns to right

S# = 120 ± 20

EI = 2300 ± 100

SAL = 1130 ± 100

SAL log = 21,600 ± 1000

PVI log 16 = 41,100 ± 1300

  1. aAgfa have indicated that the preferred exposure indices to use with mammography plates are SAL log (sometimes called PVI log15) or PVI log16, the actual choice being dictated by software version and plate type. SAL may be used in some old software versions and LgM should not be used

Appendix 9: Summary of recommendations for facility QC for DBT units

Procedure

Recommended control-limits/requirements

Minimum frequency

Key procedure elements

Recommendations for record keepinga

Full field artefact evaluation

mAs = baseline ± 10%

There must be no evidence of

 Clinically significant structures that are more conspicuous than the objects in the phantom used for weekly testing

 Blotches or regions of altered noise appearance

 Observable grid lines or breast support structures

 Bright or dark pixels

 Dust artefacts mimicking calcifications

 Significant stitching or registration artefacts

Daily

Expose a uniform thickness of PMMA using clinically relevant technique factors under AEC

Central projection image and central reconstructed image should be inspected closely for potential artefacts

View images on acquisition monitor using zoom and roam to check for possible detector faults. The magnification should be sufficient to achieve at least 1:1 resolution

Records showing

 Date test was performed

 Person performing test

 Test results

 kVp, target/filter and mAs

Image quality evaluation

mAs = baseline ± 10%

The ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of an ACR accreditation phantom OR

The ability to clearly visualise 2 fibres, 1 speck group and 2 masses in an image of an ACR DM phantom

The position of the reconstructed slice used for scoring the phantom must not change by more than ±1 mm

Weekly

Obtaining the phantom image

 Use an ACR accreditation phantom or the new ACR DM phantom

 Light contact between the compression paddle and the phantom surface

 Consistent positioning of the phantom

 Consistent selection of clinically relevant kVp and target/filter combinations

Evaluating the phantom image

 Scroll through the reconstructed images until the slice displaying the speck details most clearly is reached. Use zoom and modest adjustment of window/level functions to score fibres and specks

 Use of consistent viewing conditions that reflect those used to read actual mammograms

 Image quality scoring by the same person, if possible

 Use of a control chart to record results

Record radiographic settings (kVp, target/filter combination, mAs values) and image quality scores, position of slice used for scoring

Control chart showing

 Plots of mAs, image quality scores, slice position

 ≥25 results

 Clearly marked control limits

 Baseline values

 Remarks e.g. corrective action

Phantom images identifying

 Date

 The X-ray system

 The technique factors

Detector calibration—flat field test

Pass or fail

Weekly or as per manufacturer’s requirements

Follow manufacturer’s specific procedure

Checklist/logbook entry showing

 Date performed

 Person performing task

AEC calibration test

mAs = baseline ± 10% for same target/filter combination for each thickness of PMMA

Quarterly

Use PMMA thicknesses of 2, 4 and 6 cm covering complete image receptor

Use clinical AEC settings (kVp, target/filter and mode)

Records showing

 Date test was performed

 Person performing test

 X-ray system identification

 kVp, target/filter, AEC mode and mAs

Compressed breast thickness

For units using special Bucky’s for DBT indicated breast thickness accurate to ±5 mm

Monthly

Confirm accuracy of thickness indication under conditions as indicated by the manufacturer

Checklist/logbook entry showing

 Date performed

 Inspection results

 Person performing task

  1. aAll written/electronic QC records should be retained for 1 year unless otherwise indicated by local Regulatory requirements. Images used to assess image quality with the ACR Accreditation or ACR DM phantom should be retained for a minimum of 1 month

Appendix 10: Summary of recommendations for medical physics testing for DBT units

Procedure

Performance requirements/guidelines

Routine testing guidelines

Acceptance and additional tests

Collimation and alignment assessment

 X-ray field/Image receptor alignment

The X-ray field shall irradiate the image receptor fully but not extend beyond breast support on the chest wall edge of the image receptor by more than 2 mm or beyond the Bucky support on the other three margins

Assess alignment for each target/geometry combination

Note: When a special Bucky is used for DBT that may also be used to acquire normal 2D projection images the requirements outlined in “Collimation and alignment assessment” section must be met

As per routine tests

 Paddle/Image alignment

The chest wall edge of the compression paddle shall be aligned just beyond the chest wall edge of the image receptor such that it does not appear in the image. In addition, the compression paddle shall not extend beyond the chest wall edge of the image by more than 1% of the SID

As per routine tests

Compressed breast thickness

For units using special Bucky’s for DBT, indicated breast thickness accurate to ±5 mm

Confirm accuracy of thickness indication under conditions as indicated by the manufacturer

As per routine tests

Missed tissue

The missing tissue must be ≤5 mm

Full thickness of breast tissue must be imaged

Confirm in DBT mode of acquisition even if previously confirmed in projection imaging mode

As per routine tests

Distance calliper accuracy

Measured dimensions of object within reconstructed image plane must be within 2% of true dimensions

Use the ACR accreditation (or ACR DM) phantom as per image quality test below

Select the reconstructed slice which best displays the speck details for image scoring and perform in plane distance measurements The outside location of the detail insert is useful for this purpose

Compare with actual dimensions

Confirm accuracy of measurements at reporting workstation if possible and also confirm accuracy in more than one slice

AEC system performance assessment

mAs = baseline ± 10% for same target/filter combination for each thickness of PMMA

Use PMMA thicknesses of 2, 4 and 6 cm covering complete image receptor

Use clinical AEC settings (kVp, target/filter and mode)

As per routine tests

Image uniformity and artefact evaluation

mAs = baseline ± 10%

There must be no evidence of

 Clinically significant structures that are more conspicuous than the objects in the phantom used for weekly testing

 Blotches or regions of altered noise appearance

 Observable grid lines or breast support structures

 Bright or dark pixels

 Dust artefacts mimicking calcifications

 Significant stitching or registration artefacts

Assess for 40 mm PMMA covering complete detector using clinically relevant technique factors under AEC

Central projection image and central reconstructed image should be inspected closely for potential artefacts

View images on acquisition monitor using zoom and roam to check for possible detector faults. The magnification should be sufficient to achieve at least 1:1 resolution

Assess all projection and reconstructed images in the acquisition for all available target/filter combinations

Image quality evaluation

mAs = baseline ± 10% for same target/filter combination

Slice used for scoring should be 37 ± 2 mm (ACR accreditation phantom) or 34 ± 2 mm (ACR DM phantom) above breast support and must not change by more than ±1 mm from previous measurement

The ability to clearly visualise 4 fibres, 3 speck groups and 3 masses in an image of an ACR accreditation phantom or the ability to clearly visualise 2 fibres, 1 speck group and 2 masses in an image of the ACR DM phantom

Use typical clinical acquisition parameters selected under AEC. Note that the acquisition may be combined with FFDM mode using “Combo mode”

Select the reconstructed slice which best displays the speck details for image scoring. This is typically 37 ± 2 mm or 34 ± 2 mm above the breast support with the ACR accreditation phantom and ACR DM phantom, respectively

As per routine testing

Beam quality

\(\left( {{\text{kVp}}/100} \right)+0.03 \leq {\text{HVL}}<\left( {{\text{kVp}}/100} \right)+{\text{C}}\) \(\begin{aligned} {\text{where C }}&=&0.{\text{12 mm Al for Mo}}/{\text{Mo}}\\ &=&0.{\text{19 mm Al for Mo}}/{\text{Rh}}\\&=&0.{\text{22 mm Al for Rh}}/{\text{Rh}}\\&=&0.{\text{23 mm Al for Rh}}/{\text{Ag}}\\&=&0.{\text{3}}0{\text{ mm Al for W}}/{\text{Rh}} \\ &=&0.{\text{32 mm Al for W}}/{\text{Ag}} \\ &=&0.{\text{31 mm Al for W}}/{\text{Al}}. \\ \end{aligned}\)

Measure the HVL required for mean glandular dose calculations

As per routine tests

Mean glandular dose

≤2.0 mGy for a 4.2 cm 50% adipose, 50% glandular breast (i.e. ACR accreditation phantom or ACR DM phantom)

<1.2 mGy for 2.0 cm PMMA (2.3 cm 50% adipose, 50% glandular breast)

<4.5 mGy for 6.0 cm PMMA, (6.5 cm 50% adipose, 50% glandular breast)

Displayed MGD values must agree with calculated values within ±25%

Assess for an AEC controlled exposure using typical clinical settings using ACR phantom (or ACR DM phantom) and also for 20 mm and 60 mm PMMA

Confirm displayed and calculated MGDs agree to ± 25%

As per routine tests

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Heggie, J.C.P., Barnes, P., Cartwright, L. et al. Position paper: recommendations for a digital mammography quality assurance program V4.0. Australas Phys Eng Sci Med 40, 491–543 (2017). https://doi.org/10.1007/s13246-017-0583-x

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13246-017-0583-x

Keywords

Navigation