Abstract
Purpose
Joint line (JL) position change in total knee arthroplasty (TKA) may alter knee biomechanics and impact function. The purpose of this study was to compare the change in JL position between robotic-assisted TKA (RA-TKA) and conventional TKA (C-TKA).
Methods
A retrospective, radiographic analysis was conducted of patients who underwent RA-TKA and C-TKA to compare JL position change. JL position was measured in consecutive RA-TKAs and C-TKAs performed by four fellowship-trained arthroplasty surgeons. Statistical analysis was done utilizing t-tests and Mann Whitney U tests, with statistical significance being defined as a p value < 0.05.
Results
Six hundred total RA-TKAs and 400 total C-TKAs were included in the analysis. There were no significant differences in patient baseline characteristics such as body mass index, range of motion, and tibiofemoral coronal alignment. RA-TKAs were associated with an average of 0.04 (2.2) mm JL position change, and C-TKAs were associated with an average 0.5 (3.2) mm JL position change (p = 0.030). There were inter-surgeon differences when comparing the change in JL position for RA-TKAs and C-TKAs between the four participating surgeons.
Conclusion
RA-TKA leads to better preservation of the JL position than C-TKA, and this seems to be dependent on the arthroplasty surgeon’s preferences and techniques during TKA. Whether this statistically significant difference is clinically relevant needs to be further investigated.
Similar content being viewed by others
Introduction
Total knee arthroplasty (TKA) is growing in use worldwide, and strategies to optimize functional outcomes and durability are important. Joint line (JL) position preservation is a goal for TKA because of how it can affect knee kinematics and outcomes [1, 2]. Elevation of the JL impacts knee biomechanics by increasing mid-flexion laxity and altering posterior cruciate ligament (PCL) tension [3, 4], which may negatively impact post-operative outcomes [5]. Several studies have shown significantly lower patient-reported outcome measures with excessive JL elevation [3,4,5,6,7]. Inferior outcomes have also been found to be associated with JL distalization in revision TKA, although its impact in primary TKA is less clear [8].
Positioning of the JL in TKA varies based on implant selection, surgical technique, PCL handling, inherent errors with conventional instruments, and severity of pre-operative flexion contractures [9, 10]. Receiving computerized information on distal femoral resection thickness, limb alignment, joint gaps, and implant orientation may facilitate re-establishment of the JL height while achieving a balanced knee in robotic-assisted TKA (RA-TKA) [11,12,13,14]. While component alignment has been studied extensively in RA-TKA, there have been fewer studies investigating JL restoration [11,12,13,14,15,16]. Several studies have found that RA-TKA leads to better preservation of the JL than conventional TKA (C-TKA), but these were performed on small sample sizes with less stringent inclusion criteria, such as the inclusion of patients with flexion contractures and tibiofemoral alignment deformities up to as much as 30°, which can impact JL position necessary to balance the extension and flexion gaps [12, 13, 15, 17].
The primary purpose of this study is to conduct a radiographic analysis of pre-operative and post-operative JL position in RA-TKA and C-TKA. Additionally, the secondary purpose is to compare whether there are differences in JL position change between the participating surgeons when using robotic-assisted or conventional techniques. We hypothesize that utilizing robotic-assistance will better maintain the pre-operative JL position and that there will be no significant differences in JL position changes between the surgeons.
Materials and methods
Study design
This study was a single-institution, retrospective radiographic analysis of patients who underwent primary TKA for osteoarthritis via a conventional or robot-assisted technique. TKAs were performed by four fellowship-trained, high volume arthroplasty surgeons. The C-TKAs were performed using a mechanical alignment technique, and the RA-TKAs were performed utilizing a restricted kinematic alignment algorithm. Despite different techniques, the goal was to achieve a balanced knee, by clinical assessment for C-TKAs, and by clinical assessment and intraoperative quantitative laxity evaluation for RA-TKAs. The JL position change from pre-operative to post-operative for 600 consecutive RA-TKAs and 400 consecutive C-TKAs were compared. This study was approved by an institutional review board (IRB).
Inclusion and exclusion criteria
Patients were included if they underwent primary TKA via conventional methods or robotic-assistance and had pre-operative and post-operative weight-bearing anteroposterior (AP) radiographs. Patients were included only if they had less than a 10° flexion contracture and less than 10° of tibiofemoral coronal varus alignment pre-operatively. Patients undergoing revision or conversion arthroplasty, and those with isolated lateral compartment arthritis or recurvatum were excluded.
Data collection
Patient characteristic information were obtained from our prospectively maintained institutional database. Pre-operative electronic medical records and standing pre-operative AP radiographs were manually reviewed to determine range of motion (ROM) and coronal tibiofemoral alignment. Implant type and design was obtained from retrospective review of operative records. The identical implant system was utilized for all TKAs. Picture Archiving and Communication Software (PACS) was utilized to access the radiographs.
The adductor tubercle (AT) method [18] for measuring JL position was utilized to analyze both pre-operative to post-operative AP radiographs. These measurements were compared to determine JL position changes after RA-TKA and C-TKA. The measurements were done in duplicate, with an interval of six weeks, by two independent reviewers (N.C. and C.D.) who were blinded to the surgical technique when measuring. Any measurement discrepancies were evaluated by a third reviewer (T.D.). The differences from pre-operative to post-operative JL position were determined. Positive values indicated JL distalization and negative values indicated elevation.
Patient characteristics
There were no significant differences in sex, body mass index (BMI), or ROM between the RA-TKA and C-TKA groups (Table 1). Patients in the RA-TKA group were older than the C-TKA group (66.3 vs. 64.7; p = 0.006). The distribution of polyethylene inserts, femoral styles, and laterality are noted in Table 2.
Statistical analysis
A priori power analysis was performed, revealing a need for a total of 58 patients, with at least 29 patients in each group [17]. Descriptive analyses were performed for the full cohort. A multivariate analysis was performed to account for demographic variables and differences in implant types (cruciate-retaining (CR), ultra-congruent (UC), medial-congruent (MC), posterior-stabilized (PS), constrained-posterior-stabilized (CPS)). The PCL was resected in all cases in which a UC, PS, and CPS insert were used. With the MC insert, the PCL can be saved or resected depending on surgeon preference and intra-operative needs for soft tissue balancing. In this study, the PCL was retained in 54.2% and resected in 45.8% of cases using an MC insert. The PCL was saved in all cases in which a CR insert was used. Continuous data was compared using either t-test or Mann–Whitney U tests for the RA-TKA and C-TKA groups and ANOVA or Kruskal–Wallis when comparison were made between three or more groups for the sub-analyses. p Values less than 0.05 were deemed statistically significant. All statistical analyses were performed using R Studio (Version 3.6.3, Vienna, Austria).
Results
Radiographic measurements
There were no significant differences in pre-operative tibiofemoral coronal alignment between groups (Table 3). The RA-TKA group had an average change of − 0.04 (3.3) mm in JL position, whereas the C-TKA group had an average change of − 0.5 (3.2) mm in JL position (p = 0.030). In total, 56.2% of patients experienced JL elevation, 7.2% experienced no change, and 36.6% experienced JL distalization, with no significant differences in the directionality between groups. In the RA-TKA group, less than 15% experienced more than 4 mm of JL elevation compared to over 20% in the C-TKA group (p = 0.788); almost 26% and 30% of RA-TKAs and C-TKAs, respectively, had more than 4 mm of JL distalization (p = 0.417). Among the knees in which the JL was elevated, the RA-TKA group had an mean elevation of 2.2 (2.0) mm, which was significantly less than 2.7 (1.9) mm of for the C-TKA group (p < 0.001). Among the knees in which the JL was distalized, there was no significant difference in the amount of distalization between groups (3.2 mm vs 2.9 mm) (Table 3).
TKAs performed with a PS, MC, or CPS inserts were more likely to be elevated than TKAs performed with CR and UC inserts (Table 4). There were no significant differences in JL change or the breakdown of the directionality of change when grouping insert types depending on whether the PCL was resected or retained (Table 5). The four participating arthroplasty surgeons demonstrated differences in JL position change for both RA-TKAs and C-TKAs, when controlled for implant type between RA-TKAs and C-TKAs (Table 6).
Discussion
This study found that on average RA-TKA better preserves the JL than C-TKA, although it is unclear that these differences are clinically meaningful. We found no significant differences between RA-TKAs and C-TKAs in the overall percentages of TKAs that had elevation, no change, or distalization of the JL. While there were a greater percentage of far-outliers in which the JL position was elevated more than 4 mm when using conventional methods, the differences did not reach statistical significance. Further, when looking at the changes in JL position, there were differences in JL position change depending on the tibial insert used and between the four surgeons for each technique.
To our knowledge, the current study is the largest to assess the JL change differences between primary RA-TKAs and C-TKAs. Additionally, the more restrictive inclusion criteria used eliminated some of the confounding variables that may have impacted results in prior studies. Particularly, other studies have included patients with flexion contractures as high as 30° [12, 13, 15, 17], whereas our study included patients with flexion contractures of no more than 10°. JL elevation is commonplace in knees with higher flexion contractures due to the need to proximalize the femoral component to enlarge the extension gap. Furthermore, while other studies included patients with as much as 20° of tibiofemoral coronal alignment [12, 13, 15, 17], ours included patients only if tibiofemoral varus was less than 10°. These pre-operative measurements have an influence on the results of JL position because of how they can affect operative decisions regarding orientation and depth of femoral bone resections.
The findings of the present study support the existing literature that robotic-assistance leads to better preservation of the JL than conventionally performed TKA [12, 13, 17, 19, 20]. We found that on average RA-TKA resulted in 0.04 (3.3) mm JL elevation, whereas C-TKA resulted in a mean JL elevation of 0.50 (3.2) mm. In our series, the JL was not elevated as much as other series in either C-TKA or RA-TKA, even when the PCL was resected, likely because of a concerted effort to control the JL position with both techniques. Also, with the particular implant used, the increased flexion gap resulting from PCL resection could be offset by a slightly larger femoral component, which would neutralize the flexion gap 2 mm. Therefore, offseting the increased gap observed with PCL resection and thus obviating the typical need to increase the extension gap to compensate. Vaidya et al. found an average of 0.9 (0.9) mm of elevation in RA-TKA and a 3.5 (1.6) mm of elevation for C-TKA when using PS TKAs (p < 0.001). Further, unlike our study which found 36.6% of total cases distalized, their’s found no cases of distalization [12]. Thiengwittayapron et al. reported mean JL elevations of 3.6 (3.3) mm for RA-TKA and 5.5 (0.4) mm for C-TKA using a single PS knee design (p = 0.004) [15]. When looking at the differences in JL position for the total sample of our study, 56.2% were JL elevations 2.4 (2.0) mm, 7.2% had no change, and 36.6% were JL distalizations 3.1 (2.1) mm. Similarly, Goh et al. found that after TKA there was radiographic evidence of JL elevation in 75% of cases, no change in 7%, and JL distalization in 18% [19].
When reviewing publications on JL changes, understanding how calculations are being performed and whether the differences include a factor of directionality or absolute values are being reported is useful because the direction of JL change may impact outcomes differently. Liow et al. reported a 1.9 (1.1) mm change in JL position for RA-TKA and a 3.5 (2.8) mm for C-TKA (p = 0.01) using a PS design, but these values are absolute values without specifiying directionality [17]. Likewise, Agrawal et al. found absolute JL changes of 0.334 (0.115) mm for RA-TKA and 2.304 (0.308) mm for C-TKA (p < 0.001) [13]. In that study, the authors used a combination of CR and PS knees, but did not sub-analyze their results based on implant type. Like our study, Popat et al. reported the mean changes in JL position of less than 1.0 mm for both RA-TKA and C-TKA (0.38 mm of distalization vs. 0.91 mm of elevation) [20]. It is unclear how the PCL was handled in that series. The larger variations in JL positional changes between studies likely relates to how pre-operative flexion contractures influence the JL positions after TKA, surgeon preferences on resection depths, with or without PCL resections, and the fact that some studies only reported JL elevations or average absolute differences without a function of directionality.
Prior studies comparing robotic-assistance and conventional techniques used a variety of methods of measuring JL position, relying on different anatomical landmarks when performing measurements [12, 13, 15, 17, 20]. These differences challenge the comparisons between studies and may explain variations in the absolute JL values. Multiple methods exist for quantifying the JL [20,21,22,23], but the present study utilized the AT method on both pre-operative and post-operative radiographs because of its strong reliability [18, 24, 25]. Further, the alignment and balancing techniques used for the RA-TKAs and C-TKAs must be accounted for because of their impact on JL position. Prior studies utilized various techniques between their RA-TKAs and C-TKAS such as mechanical axis alignment, measured resection technique, gap balancing technique, kinematic alignment, and functional alignment [12, 13, 15, 17, 20]. In our study, the RA-TKAs were performed with a restricted kinematic alignment algorithm in which the femoral component was positioned between 0° and 2° of varus relative to the mechanical axis, whereas the C-TKAs were performed with intra-articular guides, referencing a cut between 4° and 5° of anatomic valgus in all cases. It is unclear to what extent these differences in the orientation of the distal femoral resections had on joint line position between our two groups.
Implant design and PCL resection can influence JL position, and we analyzed whether the use of robotic assistance can mitigate their impact. Consistent with other studies [26], our findings show that on average TKAs performed with PS, CPS, and MC implants in which the PCL is resected are more likely to be elevated—in both RA-TKA and C-TKA—than those in which the PCL was preserved and CR implants utilized.
The effect of JL changes on functional and clinical outcomes remains a topic that warrants further investigation. While some studies argue that JL changes do not affect patient reported and clinical outcomes [27, 28], others have found significant relationships between JL changes and outcomes, particularly when the JL is elevated more than 4–5 mm [29]. In our study, 14.7% of RA-TKAs and 20.4% of C-TKAs had elevations more than 4 mm, and 25.8% of RA-TKAs and 29.7% of C-TKAs had distalizations more than 4 mm, so whether these outlier groups have compromised outcomes is the topic of ongoing investigation at our center.
Strengths of this study include a larger sample size and more stringent inclusion criteria, including minimizing coronal deformities and flexion contractures, all which could otherwise influence JL position. However, this was a retrospective, non-randomized study, which may be subject to inherent limitations such as variable techniques, patient selection, etc. Nonetheless, our groups were equally compared. Additionally, although it is a strength for this study that only a single robot and implant design were used, the results may not be generalizable to other robot systems or implant designs or individual surgeon techniques which may alter the outcomes of a similar analysis. Further, the subtle differences in establishing femoral component alignment and balance between the mechanical and restricted kinematic alignment techniques used in this study may have a subtle impact on JL positions. While a number of differences in JL position between groups in this study reached statistical significance, these differences were often less than 1 mm and are likely not clinically important. Future study should be performed to determine how JL position change affects functional outcomes and the need for manipulation under anaesthesia.
Conclusion
RA-TKA better preserves the native joint line position than C-TKA. These findings are influenced by multiple factors, such as surgeon-specific techniques of femoral resections irrespective of use of robotics, the handling of the PCL and choice of polyethylene insert.
Data availability
The data will not be deposited in a public repository. Access to the data is limited by the privacy and data processing teams.
Code availability
Not applicable.
References
Selvarajah E, Hooper G (2009) Restoration of the joint line in total knee arthroplasty. J Arthroplasty 24:1099–1102. https://doi.org/10.1016/j.arth.2008.06.030
Fornalski S, McGarry MH, Bui CNH et al (2012) Biomechanical effects of joint line elevation in total knee arthroplasty. Clin Biomech (Bristol Avon) 27:824–829. https://doi.org/10.1016/j.clinbiomech.2012.05.009
Cross MB, Nam D, Plaskos C et al (2012) Recutting the distal femur to increase maximal knee extension during TKA causes coronal plane laxity in mid-flexion. Knee 19:875–879. https://doi.org/10.1016/j.knee.2012.05.007
Luyckx T, Vandenneucker H, Ing LS et al (2018) Raising the joint line in TKA is associated with mid-flexion laxity: a study in cadaver knees. Clin Orthop Relat Res 476:601–611. https://doi.org/10.1007/s11999.0000000000000067
van Lieshout WAM, Valkering KP, Koenraadt KLM et al (2019) The negative effect of joint line elevation after total knee arthroplasty on outcome. Knee Surg Sports Traumatol Arthrosc 27:1477–1486. https://doi.org/10.1007/s00167-018-5099-8
Clavé A, Le Henaff G, Roger T et al (2016) Joint line level in revision total knee replacement: assessment and functional results with an average of seven years follow-up. Int Orthop 40:1655–1662. https://doi.org/10.1007/s00264-015-3096-9
Partington PF, Sawhney J, Rorabeck CH et al (1999) Joint line restoration after revision total knee arthroplasty. Clin Orthop Relat Res 165–171
Hofmann AA, Kurtin SM, Lyons S et al (2006) Clinical and radiographic analysis of accurate restoration of the joint line in revision total knee arthroplasty. J Arthroplasty 21:1154–1162. https://doi.org/10.1016/j.arth.2005.10.026
Figgie HE, Goldberg VM, Heiple KG et al (1986) The influence of tibial-patellofemoral location on function of the knee in patients with the posterior stabilized condylar knee prosthesis. J Bone Joint Surg Am 68:1035–1040
Kucera T, Urban K, Karpas K, Sponer P (2007) Restricted motion after total knee arthroplasty. Acta Chir Orthop Traumatol Cech 74:326–331
Zambianchi F, Bazzan G, Marcovigi A et al (2021) Joint line is restored in robotic-arm-assisted total knee arthroplasty performed with a tibia-based functional alignment. Arch Orthop Trauma Surg 141:2175–2184. https://doi.org/10.1007/s00402-021-04039-z
Vaidya NV, Deshpande AN, Panjwani T et al (2022) Robotic-assisted TKA leads to a better prosthesis alignment and a better joint line restoration as compared to conventional TKA: a prospective randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 30:621–626. https://doi.org/10.1007/s00167-020-06353-2
Agrawal VO, Gadekar AP, Vaidya N (2022) Does robotic technology successfully restore the joint line after total knee arthroplasty? A retrospective analysis. Arthroplasty 4:6. https://doi.org/10.1186/s42836-021-00103-6
Onggo JR, Onggo JD, De Steiger R, Hau R (2020) Robotic-assisted total knee arthroplasty is comparable to conventional total knee arthroplasty: a meta-analysis and systematic review. Arch Orthop Trauma Surg 140:1533–1549. https://doi.org/10.1007/s00402-020-03512-5
Thiengwittayaporn S, Uthaitas P, Senwiruch C et al (2021) Imageless robotic-assisted total knee arthroplasty accurately restores the radiological alignment with a short learning curve: a randomized controlled trial. Int Orthop 45:2851–2858. https://doi.org/10.1007/s00264-021-05179-y
Kim Y-H, Yoon S-H, Park J-W (2020) Does robotic-assisted TKA result in better outcome scores or long-term survivorship than conventional TKA? A randomized, controlled trial. Clin Orthop Relat Res 478:266–275. https://doi.org/10.1097/CORR.0000000000000916
Liow MHL, Xia Z, Wong MK et al (2014) Robot-assisted total knee arthroplasty accurately restores the joint line and mechanical axis. A prospective randomised study. J Arthroplasty 29:2373–2377. https://doi.org/10.1016/j.arth.2013.12.010
Yeh K-T, Chen I-H, Wang C-C et al (2019) The adductor tubercle can be a radiographic landmark for joint line position determination: an anatomic-radiographic correlation study. J Orthop Surg Res 14:189. https://doi.org/10.1186/s13018-019-1221-y
Goh GS-H, Bin Abd Razak HR, Tan JY-W, Yeo S-J (2017) Intraoperative measurements of joint line changes using computer navigation do not correlate with postoperative radiographic measurements in total knee arthroplasty. J Arthroplasty 32:61–65. https://doi.org/10.1016/j.arth.2016.06.018
Popat R, Albelooshi A, Mahapatra P et al (2022) Improved joint line and posterior offset restoration in primary total knee replacement using a robotic-assisted surgical technique: An international multi-centre retrospective analysis of matched cohorts. PLoS One 17:e0272722. https://doi.org/10.1371/journal.pone.0272722
Pereira GC, von Kaeppler E, Alaia MJ et al (2016) Calculating the position of the joint line of the knee using anatomical landmarks. Orthopedics 39:381–386. https://doi.org/10.3928/01477447-20160729-01
Han H, Zhang X (2020) A new method for evaluation of patellar height and the position of the joint line before and after total knee arthroplasty. BMC Musculoskelet Disord 21:768. https://doi.org/10.1186/s12891-020-03794-1
Sato T, Koga Y, Sobue T et al (2007) Quantitative 3-dimensional analysis of preoperative and postoperative joint lines in total knee arthroplasty: a new concept for evaluation of component alignment. J Arthroplasty 22:560–568. https://doi.org/10.1016/j.arth.2006.06.020
Chen I-H, Wu W-T, Wang C-C et al (2016) An unambiguous technique for locating the adductor tubercle and using it to identify the joint line. Knee 23:960–963. https://doi.org/10.1016/j.knee.2016.05.014
Iacono F, Raspugli GF, Bruni D et al (2014) The adductor tubercle as an important landmark to determine the joint line level in total knee arthroplasty: from radiographs to surgical theatre. Knee Surg Sports Traumatol Arthrosc 22:3034–3038. https://doi.org/10.1007/s00167-013-2809-0
Bin Abd Razak HR, Pang HN, Yeo SJ et al (2013) Joint line changes in cruciate-retaining versus posterior-stabilized computer-navigated total knee arthroplasty. Arch Orthop Trauma Surg 133:853–859. https://doi.org/10.1007/s00402-013-1738-1
D’Amato M, Kosse NM, Wymenga AB (2021) Restoration of pre-operative joint line orientation and alignment does not affect KSS and KOOS 1 year after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 29:3170–3177. https://doi.org/10.1007/s00167-020-06097-z
Calek A-K, Ladurner A, Jud L et al (2022) Tibial joint line orientation has no effect on joint awareness after mechanically aligned total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 30:389–396. https://doi.org/10.1007/s00167-021-06696-4
Babazadeh S, Dowsey MM, Swan JD et al (2011) Joint line position correlates with function after primary total knee replacement: a randomised controlled trial comparing conventional and computer-assisted surgery. J Bone Joint Surg Br 93:1223–1231. https://doi.org/10.1302/0301-620X.93B9.26950
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Material preparation and data collection were performed by Nicholas F. Cozzarelli, BS, Cristian A. DeSimone, BS, and Taylor D’Amore, MD. Statistical analysis was performed by Matthew B. Sherman, BS. The senior author (Jess H. Lonner, MD) provided study oversight. The first draft of the manuscript was written by Nicholas F. Cozzarelli, BS, and all authors edited the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval
Institutional review board (IRB) approval was received prior to initiating the study (iRISID2023-1789).
Informed consent
The research presents no more than minimal risk of harm to subjects and involves no procedures for which written consent is normally required outside the research context.
Consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
The senior author is a consultant for the company who developed the robot utilized for the robotic-assisted total knee arthroplasties in this study, but he has conducted this study with good ethical and clinical practice.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Cozzarelli, N.F., DeSimone, C.A., D’Amore, T. et al. Joint line position change in primary total knee arthroplasty: a radiographic analysis comparing conventional and robotic techniques. International Orthopaedics (SICOT) 48, 1023–1030 (2024). https://doi.org/10.1007/s00264-023-06031-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-023-06031-1