Background

Limited lumbar microsurgical discectomy (LMD) to remove the free disc fragment is the gold standard treatment for patients with refractory disc herniation related radicular pain [22]. However, the failure rate of discectomy is high due to recurrent disc herniation, progression of degenerative disc disease (DDD) and continuing chronic low back pain [16]. The rate of recurrent herniation is 3%-18% in limited discectomy and correlates among other predictive factors with the annular defect area and percentage of removed disc material [24]. A defect size of > 6 mm wide has a reherniation rate of 27% [6]. Aggressive discectomy on the other hand, may lead to accelerated DDD, instability and disc height collapse with chronic low back pain [24]. Symptomatic recurrent herniation management is either conservative or with repeat surgery including discectomy or fusion [18]. The outcome after repeat surgery is often inferior to primary surgery [9]. An annular closure device (ACD) (Barricaid®, Intrinsic Therapeutics, Inc., Woburn, MA) was recently introduced and could potentially prevent recurrent herniation by blocking larger annular defects after LMD, and preserve the nucleus pulposus within the disc space [19]. The results of a European multicenter randomized controlled trial (RCT) with 550 patients have recently been published. It was shown that the frequency of symptomatic reherniation and reoperation was lower in the ACD compared to the control group—with a similar outcome over a 3-year period [12, 23]. Endplate changes (EPC), however, were more prevalent in the ACD group (84% vs. 30%) [23]. Further post hoc analysis investigating the occurrence of EPC identified mechanical stress from the ACD on the endplates as cause of the significant increase of EPC [5]. There was no correlation with outcome (low back pain or ODI) [5, 14]. The aim of the present study is to share our institution's experience with EPC following LMD with additional implantation of an ACD.

Methods

Study design

This analysis includes data from (a) European multicenter RCT study-arm of patients undergoing LMD with implantation of the ACD Barricaid® and (b) additional patients undergoing implantation of commercially available ACD at our institution. Lumbar disc herniation patients scheduled for LMD with a posterior disc height of at least 5 mm were offered ACD implantation. Further information on the RCT study protocol including inclusion and exclusion criteria can be found in the main study publication [23]. Informed consent was obtained for all RCT patients. Prior ethics committee approval was obtained for the multicenter randomized study and for the retrospective data analysis (EC numbers 2012/036; 2016/01740). This study is in accordance with the STROBE guidelines. The results from our study have been previously presented at the AANS conference 2019 [11].

Outcome measures

X-ray, computed tomography (CT) and magnetic resonance imaging (MRI) was performed prior and at the one year follow up. Clinical outcome assessment included pre- and postoperative score assessment of Visual Analogue Scale (VAS) [10], and Oswestry Disability Index (ODI) [7].

Volumetry of osteolytic endplate changes

Endplate changes (EPC) were classified as any new erosion of subchondral bone, vertebral or cartilaginous endplate. Location of EPC (lower or upper endplate) and the relation of the titanium device anchor (superior or inferior endplate) to the vertebra were determined (Fig. 1). Measurements of EPC volume (cm3) in the lumbar spine CT were performed using a commercially available software (Elements®, Brainlab, Munich, Germany) (Fig. 2). To increase reliability, all measurements were performed by two neurosurgeons. The interobserver agreement results were accepted for final analysis.

Fig. 1
figure 1

Demonstration of the nomenclature of upper and lower endplate changes in relation to the vertebra. a ACD anchor implanted in the inferior endplate of L4 with EPC in the upper endplate of L5. b ACD anchor in the superior endplate of L5 with EPC in the lower EP of L4. EPC Endplate changes, EP Endplate

Fig. 2
figure 2

Multiple examples of patients with EPC and the volume measurement technique using Elements software (Brainlab®, Munich, Germany)

Endplate and disc degeneration

All EPC were additionally classified according to Modic criteria [17] and DDD was assessed by applying Pfirrmann classification [20]. An independent neuroradiologist conducted all measurements. Disc height was assessed in the pre- and postoperative MRI within 5 mm of the posterior border. In order to rule out measurement errors, 50 randomly selected cases were reevaluated by the first author.

EPC grading system

We aimed to establish a descriptive grading classification of EPC ranging from 1 to 4 to better assess the extent of EPC and simplify radiological description. The EPC classification includes 2 criteria: localization of device anchor in the vertebra (superior or inferior) and EPC volume (< 1 cm3 or > 1 cm3).

Statistical methods

A multivariate regression model was developed in which a linear combination of EPC was regressed on all potential predictive factors to define their significance. Univariate analysis was performed to examine the effect of defined risk factors on the location of EPC (upper vs. lower). The clinical influence of EPC on pain (VAS) and disability scores (ODI) were investigated with a univariate regression analysis of EPC and each outcome parameter. A paired t-test was performed to compare normally distributed continuous data and Wilcoxon rank sum test for non-parametric data.

Results

Patient characteristics

A total of 72 patients (37 men) underwent LMD and ACD implantation between January 2013 and October 2016 at a single institution. Twenty-nine patients had surgery and ACD implantation within the randomized controlled trial and 43 were “commercial cases”. The median age at time of surgery was 47 ± 11.63 years. Overall median follow-up was 14.67 ± 4.77 months. Forty-six patients (64%) had smoked at some point in their lives; 33 (46%) of them were still active smokers and 14 (19%) had quit the habit. Forty-seven (67%) patients had a job requiring heavy lifting. The remaining 25 (33%) had an occupation with lighter physical demands and 5 (20%) were already retired.

Intraoperative Parameters initial surgery

There were no surgical complications. The mean annular defect size was 7.77 ± 1.27 mm in width and 1.35 ± 0.77 cc of disc material was removed on average. The difficulty of surgical procedure with respect to ACD implantation was generally rated as easy or acceptable.

Endplate changes

Endplate changes were found in 71 (99%) patients included in this study. These EPC volumes were significantly larger in the lower endplate (EP) of the vertebra, with a median of 230 mm3, compared with those in the upper EP, with a median of 105 mm3 (p = 0.006) (Fig. 3). The largest EPC was 4.2 cm3. A multivariate regression analysis of risk factors for development of EPC at the upper or lower endplate only produced one significant predictor: the ACD anchor localization (p = 0.038) (Table 1). Subsequent univariate regression analysis showed that multivariate contribution of anchor localization is based on a significant effect of lower EPC after anchor implantation in the superior EP (p = 0.025) (Table 2). In the current study group, 46 (64%) ACD titanium anchors were implanted in the superior EP with the polymer mesh linked to the lower EP of the adjacent vertebra. We analyzed median upper and lower EPC volumes in association to anchor localization (Fig. 4). A significant difference between the inferior and superior anchor localization was only present at the lower EP (p = 0.022).

Fig. 3
figure 3

Median EPC volume with interquartile range (25%—75% quartiles) of all patients at the lower and upper endplate, log(10) scale. p = 0.006

Table 1 Results of the multivariate regression analysis of risk factors for EPC of the lower and upper endplate revealed only one significant parameter: anchor localization in superior endplate
Table 2 Univariate regression analysis and significance level of each predictor on lower EPC
Fig. 4
figure 4

Median lower and upper EPC volumes with interquartile range (25%—75% quartiles), split out for the two anchor locations superior and inferior, log(10) scale. p = 0.002

Descriptive score

This grading system was based on the fact that patients with ACD in inferior endplate had, on average, smaller EPC compared to superior EP fixation. Table 3 presents the descriptive criteria for the ACD—EPC scoring system. The scoring system was validated in our patient cohort and Table 4 presents the distribution in our study cohort.

Table 3 Introduction of a new classification of ACD associated EPC. The EPC score ranges from 1 to 4 and assesses anchor localization (superior/inferior) and size of EPC (< 1cm3 / > 1cm3) as criteria. Scores were defined as follows: grade 1 includes patients with ACD implanted on inferior EP and EPCs of < 1 cm3, grade 2 have an ACD in the superior EP and EPCs < 1 cm3, grade 3 received ACD in inferior EP and EPCs > 1 cm3, and finally maximum grade 4 if ACD is in the superior EP and the EPCs are > 1 cm3. ACD = Annular closure device
Table 4 Validation of the EPC classification in our study cohort. A maximum achievable score of 4 points corresponds to severe EPC and the minimum of one point indicates mild EPC. Scores of 2 to 3 points are considered intermediate

Case report

Figure 5 illustrates a case of a 52 year-old women who underwent LMD and ACD implantation at L4/5 level due to disc herniation and persistent L5 radicular pain. Preoperative CT confirmed absence of any EPC. LMD and ACD implantation in the superior EP of L5 was successful, no complications occurred intraoperative and radicular pain was resolved postoperatively. At one-year follow-up, EPC (size = 0.57 cm3) had appeared around the mesh at the lower EP of L4. Only minimal changes were noticeable below the titanium anchor at upper EP of L5. After 2 years, EPC (size = 1.15 cm3) progressed in the lower EP of L4 and at the 3 years follow-up, EPC (2.46 cm3) increased to a grade 3, with polymer mesh subsidence into the lower EP of L4. Instability caused increasing facet joint-associated low back pain and the patient underwent subsequent fusion and removal of the implant.

Fig. 5
figure 5

Case report of a patient that underwent LMD and ACD implantation at L4/5 level due to disc herniation. a Preoperative baseline CT: No EPC visible. b One-year follow-up: EPC (0.57 cm3) appear at the lower EP of L4. c 2-year follow-up: EPC (1.15 cm3) progressed at lower EP of L4. d 3-year follow-up: EPC (2.46 cm3) accelerated to a grade 3, with polymer mesh subsidence into lower EP of L4

Reoperation due to reherniation or device failure

At the time of study analysis in 2018, 65 patients (90%) were still living with the implant and in 7 (10%) patients the whole implant (n = 4), or only the polymer mesh (n = 3) had been removed. A total of 17 (24%) patients revealed reherniation in the postoperative follow-up MRI, 10 patients (13.8%) were symptomatic and 7 (10%) asymptomatic. In addition to conservative treatments (n = 4), reoperation was necessary in 6 (8.3%) patients in the symptomatic cohort. The procedures included recurrent LMD only (n = 3), LMD with ACD removal (n = 1) and LMD with mesh removal (n = 2). A device failure occurred in 19 (26.4%) patients. "Failure" was defined as: dislocation of the whole device > 2 mm (n = 5, 6.9%), device anchor-head breakage (n = 1, 1.3%), or posterior dislocation of the mesh into spinal canal (n = 13, 18%). Mesh subsidence into EPC was documented in 15 (20.8%) patients (Table 5 and Fig. 6). Overall, 7 patients (9.7%) underwent reoperation due to device failure within the follow-up period. In terms of reoperation techniques, the following surgical approaches were used in a total of 13 reoperations (18.1%): Re-LMD only (n = 3, 4%), removal of polymer mesh and LMD (n = 2, 3%), LMD with mesh and anchor-head removal (n = 1, 1%), explant of the whole ACD and LMD (n = 1, 1%), transpedicular fusion with ACD in place (n = 3, 4%) and fusion with ACD removal (n = 3, 4%) (Table 6). All reoperations occurred on average 18 ± 11 months after LMD and ACD implantation. According to the surgeons, reoperation and ACD removal was more difficult than re-LMD only. An unintended durotomy during reoperation occurred in 31% of cases. During all revision surgeries, portions of the mesh, bone and disc material were sent for bacteriological growth testing, and any ongoing infection was excluded.

Table 5 Types of surgical and device failure
Fig. 6
figure 6

Several types of device failure are shown. a Dislocation of the whole device posterior into spinal canal. Patient underwent rediscectomy and ACD removal. b Device material failure: fracture of the titanium anchor-head and dislocation toward spinal canal. Rediscectomy and anchor head and mesh removal was performed. c Posterior dislocation of polymer mesh into spinal canal. A LMD and mesh removal was performed during revision surgery. d Intraoperative picture of removed ACD device. e Radiological appearance of mesh subsidence into EPC and vertebra osteolysis. Clinically asymptomatic finding, no reoperation necessary to-date. f ACD failure: symptomatic recurrent disc herniation at the index level leading to a reoperation with LMD only

Table 6 The patient subgroup that underwent reoperation due to recurrent disc herniation or device failure. Details of reoperation, implant approach and any complications are presented

Clinical outcome

In summary, the surgery achieved significant improvements with good clinical outcome. These findings are most likely independent from the ACD implantation and a normal consequence of LMD surgery. Mean preoperative ODI of 57.27 decreased significantly postoperative to a mean of 17.58 (95% CI [35.71, 46.99] p < 0.001). With regard to low back pain VAS, a preoperative mean of 63.53 showed significant postoperative reduction to 19.80 (95% CI [35.32, 57.26] p < 0.001). Left leg pain VAS, decreased significant from a mean of 45.66 preoperative to 11.46 postoperative (95% CI [23.72, 49.26] p < 0.001). Similarly, right leg pain VAS reduced from a preoperative mean of 42.97 to postoperative 10.79 (95% CI [21.45, 49.72] p < 0.001) (Fig. 7). We also investigated if the magnitude of EPC had any influence on postoperative outcome. In particular, lower EPC showed significant correlation with postoperative ODI at the last follow-up only (p = 0.01) and a trend was present with lower EPC and last postoperative low back pain VAS (p = 0.6). In fact, EPC do not seem to affect postoperative clinical outcome in a significant manner.

Fig. 7
figure 7

Clinical outcome a + b Regarding right and left leg pain VAS, a significant postoperative improvement could be shown. c Mean low back pain VAS improved significantly after surgery. d Significant improvement of mean postoperative ODI

Disc degeneration

MRI findings prior and after surgery were compared with a mean postoperative follow-up of 22.7 ± 9.7 months to assess DDD. Preoperative mean disc height measured 6.2 ± 1 mm and decreased postoperatively to a mean of 5.5 mm ± 1 mm. This 0.7 ± 0.6 mm reduction in disc height was non-significant. Disc degeneration at index level was Pfirrmann grade III in 49% preoperative and 66% postoperative (Fig. 8). A third of patients (31%) had no Modic changes before surgery, and Type 1 increased from 14 to 44% (Fig. 9).

Fig. 8
figure 8

Assignment of our study cohort according to Pfirrmann classification. The majority intervertebral discs were classified as grade III before and after surgery

Fig. 9
figure 9

Assessment of endplate changes according to Modic classification in the current study patients. The amount of patients having Modic changes and specifically Type 1 Modic changes, increased after surgery. Number of Type 2 Modic changes was stable, and no Type 3 Modic changes were seen

Discussion

The findings of this study support nascent evidence that ACD implantation after LMD might accelerate EPC and lead to osteolysis of the surrounding bone. In this study, EPC was documented in almost all patients during follow-up. Nevertheless, no impact on clinical outcome was apparent.

The present study revealed a reherniation rate of 14%, which is in the 3%-18% range following LMD reported in the literature [2]. The 3 year follow-up results from the multicenter randomized controlled European trial (RCT LMD—ACD), comparing LMD and ACD versus LMD only showed a symptomatic reherniation rate of 14.8% in the ACD group, versus 29.5% in the control group [12].

Our reoperation rate of 18% during the 18-months postoperative period was above reported average between 1 and 11.5% [1, 26]. In comparison, the RCT LMD—ACD trial showed that index level reoperations were significantly less frequent with ACD (9% vs. 16% after 2 years and 11% vs. 19.3% after 3 years) [12, 23].

Other studies have also reported the appearance of EPC after LMD and ACD [14]. One case report described ACD removal and fusion 1.5 months after surgery due to resorbed bone in vicinity of the device and inflammatory changes in the adjacent tissue [13]. In the RCT LMD – ACD study, EPC were significantly more prevalent in the ACD group compared to the control group (84% vs. 30% after 2 years and 89% vs. 41% after 3 years) [12, 23]. At baseline, the number of patients showing EPC was similar in both groups [5]. The number and total EPC volume increased more significantly in the ACD group over time [5]. Barth et al. had reported earlier that ACD implantation in the superior EP causes more frequent and larger EPC in the opposite lower EP due to applied pressure from the mesh [5]. Indeed, the lower endplate seems to be more affected by DDD and ongoing tissue damage.[3].

Endplate changes themselves are a common finding after LMD only, with a reported incidence of 43%—57% [4, 25]. There might be an association in LMD between surgical technique and aggressive treatment of sensitive EP. Earlier studies have shown, that endplate changes and DDD are related [8, 21]. In the present study, no relevant increase of DDD grade was seen during follow up.

In cases with ACD, the flexible polymer mesh might cause friction, leading to EPC around the mesh. More research and improvement of implant design is needed. Although anatomy mainly determines the anchor position, in general, ACD implantation in the superior EP should be avoided.

Lange et al. recently reported on a symptomatic ACD loosening due to infection with Propionibacterium acnes, which they confirmed after revision surgery [15]. In our series, disc material cultures and device sonification ruled out any low-grade infection as a possible cause of EPC.

A descriptive classification was introduced to facilitate interprofessional communication about ACD-associated EPC. This scoring system was validated in our patient cohort but further external score validation will be necessary.

Finally, the risk of EPC must be balanced against a possible lower reherniation rate. Based on our experience, we would suggest that only patients with a large annular defect (> 6 mm) and high remaining disc volume should be considered for ACD implantation on the inferior EP. Of higher importance is an improvement of surgical techniques to achieve smaller annular incisions (< 6 mm) in order to prevent reherniation.

Study limitations

Limitations of our current study are the fact that this is a non-comparative single center study with a small patient cohort.

Conclusion

There was significant postoperative clinical improvement after limited LMD and ACD implantation. The high incidence and volume of EPC did not affect clinical outcome. The ACD might prevent disc reherniation despite implant failure rates. Long-term clinical and radiological assessments is necessary to evaluate the consequences of these findings.