Background

Metacarpal fractures account for almost 18% of all fractures distal to the elbow, with non-thumb fractures accounting for 88% of this group [1]. Metacarpal shaft fractures can be treated nonoperatively, but shortening, angulation, and rotational deformity can diminish clinical function [2]. As such, surgical stabilization is commonly employed to restore anatomic alignment.

Current methods for metacarpal shaft fixation include Kirschner wires (k-wires), plating, and screws, depending on fracture orientation and physicians preference. Plating represents a strong construct for a wide variety of fracture types with the drawbacks of more extensive soft tissue dissection, higher infection risk, risk of extensor adhesions, or hardware irritation [3, 4]. Screws represent another rigid form of fixation and are an option with long oblique or spiral fractures. Although less rigid, percutaneous K-wires are appealing as they decrease soft tissue dissection. A disadvantage of k-wires is that they protrude through the skin risking infection and making early range of motion difficult [5].

With the growing active population, alternate forms of fixation to decrease tissue dissection, lower infection risk, lower risk of extensor tendon adhesion, lower hardware irritation, and allow early active motion and rapid return to function deserve consideration. Headless compression screws in hand fractures have emerged in a few clinical studies showing good results for function and healing with the advantages listed above [6,7,8].

The purpose of the study was to determine the stability of fracture fixation with intramedullary headless compression screws in two types of metacarpal shaft fractures and compare them to other common forms of rigid fixation: dorsal plating and lag screw fixation. It was hypothesized that headless compression screws would demonstrate higher load to failure than plates in transverse fractures and higher load to failure than plates and lag screws in long oblique fractures.

Methods

Specimen Preparation & Fixation

Five matched paired hands, mean age 60.9 ± 4.6 years, were obtained and underwent dual energy x-ray absorptiometry (DEXA) (Lunar DPI XQ dexascan Madison, Wis). A 1cm2 region within the metacarpal neck was measured. Metacarpals were dissected free, excluding the thumb metacarpals, yielding 40 metacarpals. Two metacarpals were disqualified from use due to previous deformity. Diaphyseal fractures were created with an oscillating saw, one millimeter thick. In fifteen, the osteotomies were transverse located 25 mm from the distal articular surface, and in twenty-three were long oblique with the osteotomy beginning dorsally 20 mm from the distal articular surface and exiting 35 mm palmarly. In the first group of matched paired hands of transverse fractures, metacarpals were fixated with dorsal non locking plating (n = 8) versus metacarpals with headless compression screws (n = 7). The second group of match paired hands with long oblique fractures fixated with dorsal non locking plating (n = 8) versus headless compressions screws (n = 8) versus two lag screws (n = 7).

In the dorsal plating groups, low profile 2.4 mm plates (Arthrex, Naples, Florida) were placed on the dorsal surface. The transverse fractures utilized a six-hole plate, and for the long oblique fractures seven-hole plates. In each, two 2.4 mm non-locking screws were inserted bicortically on each side (Fig. 1). There has not been sufficient evidence for the need of three screws on each fracture side [9].

Fig. 1
figure 1

Plate and screw construct, gross specimen (a) in transverse osteotomy and lateral fluoroscopic image in oblique osteotomy (b)

In the lag screw fixation, two 2.4 mm non-locking cortical screws (Arthrex, Naples, Florida) were inserted perpendicular to the fracture line. Each near cortex was drilled with a 2.4 mm drill bit, and the far cortex with a 1.7 mm drill bit, according to standard lag screw technique (Fig. 2).

Fig. 2
figure 2

Lateral fluoroscopic image of lag screw construct in oblique osteotomy

In the headless compression screw group, a 1.1 mm guidewire was inserted into dorsal third articular surface of the metacarpal head and subsequently into a central position of the metacarpal shaft on anteroposterior and lateral fluoroscopic imaging. The distal metacarpal articular surface was then opened with a 2.4 mm cannulated drill bit if using a 3.5 mm compression screw, and a 3.2 mm cannulated drill bit if using a 4.0 mm cannulated screw. The cannulated headless compression screw (Arthrex, Naples, Florida) was inserted over the guidewire to a depth of approximately 2 mm below the closest (dorsal) articular surface. Small and ring metacarpals were fixated with the 3.5 mm compression screw and long and index with 4.0 mm compression screws based on the diameter of the medullary canal (Fig. 3).

Fig. 3
figure 3

Headless compression screw fixation, gross specimen (a), anteroposterior in transverse osteotomy (b), and lateral in oblique osteotomy fluoroscopic imaging (c)

Testing protocol

Each metacarpal was subjected to a three point bending, apex dorsal force, on a Materials Testing System (MTS) servohydraulic test frame (MTS Systems Corp, Eden Prairie, MN), similar to previous studies [5, 10, 11]. Each sample was placed with the dorsal surface downward toward the base plate and the loading pin centered over the fracture site (Fig. 4). The loading pin displaced toward the MTS actuator at a rate of 100 mm/min until failure. Sampling of the load displacement was recorded at a rate of 100 Hz. Failure was defined by sudden change in the load-displacement curve.

Fig. 4
figure 4

Testing setup with loading pin centered over osteotomy producing an apex dorsal force on the MTS servohydraulic frame

Data analysis

Load displacement curves were generated for each specimen. The curve was utilized to calculate the peak load to failure (N) and stiffness (N/mm). The peak load to failure was the maximum force at which an acute change in the load-displacement curve was observed. Stiffness of the construct corresponded to the slope of the load-displacement curve before the peak load was reached.

Statistics

Given the variation of testing specimens in previous metacarpal biomechanical studies [1, 5, 10, 12], an a priori power analysis was difficult. We assumed the smallest difference would be between the headless compression screws and plate fixation groups. A difference of 75 N between these groups was selected as the threshold for a relevant difference in load to failure. A standard deviation of ±50 N was assumed across groups. Assuming these parameters, a sample size of 7 per group would provide 80% power to detect a 75 N difference in load to failure at an alpha level of 0.05.

Descriptive statistics to categorize the groups were calculated using mean and standard deviation. Differences between the groups in load to failure and slope were tested with one way ANOVA with a post hoc Bonferroni correction where appropriate for longitudinal fractures and an independent t test for transverse fractures. A p value less than 0.05 was considered statistically significant.

Results

Transverse shaft fracture

For plate and screw constructs, the mean peak load to failure was 303.9 ± 94.47 N and mean stiffness was 129.02 ± 66.18 N/mm. All specimens failed by plate bending (Fig. 5a).

Fig. 5
figure 5

Failure patterns, a plate bending in a transverse osteotomy, b lag screw disengagement in oblique osteotomy, and c headless compression screw loosening in metacarpal head and displacement at osteotomy site of oblique pattern

Headless compression screws fixation showed a mean peak load to failure of 829.8 ± 116.4 N and mean stiffness was 316.9 ± 101.6 N/mm. Four specimens failed by the screw bending and three specimens failed by loosening in the metacarpal head region with resultant displacement at the fracture site.

A significant difference was observed in both peak load to failure (p < 0.001) and stiffness (p < 0.001) in favor of the headless compression screw group. There was no significant difference in bone mineral density between headless compression screw and plate specimens (p = 0.626).

Oblique shaft fractures

In plate and screw constructs, the mean peak load to failure was 518.5 ± 212.45 N and mean stiffness was 258.7 ± 99.68 N/mm. All specimens failed by plate bending.

In two lag screw constructs, the mean peak load to failure was 234.11 ± 178.06 N and mean stiffness was 172.18 ± 122.28 N/mm. All specimens failed by screw heads pulling through the near (dorsal) cortex (Fig. 5b).

For headless compression screw constructs, the mean peak load to failure was 758.4 ± 54.9 N and mean stiffness was 275.1 ± 62.6 N/mm. All specimens failed by loosening of the implant in the metacarpal head region and displacement at the fracture site (Fig. 5c).

Comparison of the groups showed a significant difference in peak load to failure of headless compression screws versus plate and screws (p = 0.023), headless compression screw versus lag screws (p < 0.001), and plate and screws versus lag screws (p = 0.009). A significant difference in stiffness was not observed between groups (p = 0.12). There were no significant differences in bone mineral density between specimens of headless compression screw and plate (p = 0.42), plate and lag screws (p = 0.46), or headless compression screw and lag screws (p = 0.94).

Discussion

In the current study, biomechanical stability of various fixation methods for diaphyseal metacarpal fractures were compared. In transverse fractures, intramedullary headless compression screw was compared to dorsal plating. The former was found to be superior in both stiffness and peak load to failure. In the oblique fractures, intramedullary headless compression screw, dorsal plating, and two lag screws were compared. While the intramedullary compression screw had significantly higher peak to load failure compared to the other two methods, there was no significant difference in stiffness.

Cannulated headless compression screws have been widely used in hand surgery since Herbert and Fisher [13] introduced its utility in scaphoid fractures. However, its use as an intramedullary device for metacarpal fractures is relatively new [6, 14, 15].

The intramedullary compression screw has several advantages over conventional fixation methods. It avoids the risk of pin tract infection or need for removal associated with percutaneous K-wire fixation. Unlike dorsal plating which requires a more extensive dissection, the intramedullary screw can be inserted through a small dorsal incision. Furthermore, with the compression screw, the entire hardware is buried inside the metacarpal and thus unlikely to cause irritation of surrounding soft tissue or tendon ruptures. Finally, the compression screw allows for early active motion exercises to minimize stiffness when compared to K-wires [7, 8, 16]. These advantages make intramedullary compression screw an attractive option for fixation of metacarpal fractures. Some downsides of intramedullary screw fixation have to be mentioned on the other hand as extensor tendon disruption and difficult explantation in case of infection or refracture is possible [10, 17].

The results from the current study support that compression screw fixation provides excellent biomechanical stability that is comparable to dorsal plating or lag screws. Analogous to plate fixation, it may provide an option for rapid return to sport [18] with a lower risk of adverse effects. All construct exceeded by far the invivo forces generated by the flexor tendons of 30 N and therefore can be considered possible options [19] from a biomechanically point of view.

In comparison to existing biomechanical data by Oh et al. we could demonstrate higher peak loads for all tested devices [20]. This may be due to the effect of a more stable plate and larger diameter intramedullary compression screws with a different design. Especially the size of the intramedullary compression screw has to fit perfectly in order to generate optimize fixation strength. Oftentimes the available screw size may not be large enough to archive endosteal purchase [21].

Reported clinical results for compression screw fixation have been good. Del Pinal et al. [14] reported a series of 48 metacarpal fractures, all of which healed with total active motion (sum of distal phalangeal joint, proximal phalangeal joint, and metacarpophalangeal joint motion) averaging 249 degrees. Ruchelsman et al. [6] first reports a series of 20 metacarpal fractures in which all fractures healed with full flexion and grip strength measuring 105% (range 58 to 230%) compared to the contralateral side. In their second series of now 91 patients they could continue to show favorable results without the disadvantages of open reduction and fixation [7].

There are several limitations of the study. Failure of the constructs was tested only in one dimension and without cyclic loads demonstrating a possible loosening during the early postoperative phase. We chose to load the construct with an apex-dorsal force based on previous biomechanical studies [5, 10, 11]. Axial, rotational, or cyclical loading may reveal clinically important differences. Also, only 3-point bending instead of cantilever bending was used that may not reflect individual anatomy and physiology of the hand and metacarpal loading.

Particularly with plate fixation, other methods or construct configurations are possible and are not accounted for, such as locking plate constructs, which could demonstrate different results. Additionally, the current study was performed on cadavers and does not account for any differences in biological healing the different constructs induce. However, differences in bone mineral density between cadaver specimens was tested and found to be insignificant.

It should also be noted that the retrograde placement of intramedullary screw requires an arthrotomy of MCP joint as it is inserted through the articular surface of the metacarpal head which may put the joint at an increased risk for future arthritis [22]. Furthermore perfect anatomic reduction may be more difficult percutaneously than in open techniques, but clinical results remain good [6, 7, 14].

Conclusion

Intramedullary fixation of diaphyseal metacarpal fractures with a headless compression screw provides excellent biomechanical stability. Coupled with lower risks for adverse effects, headless compression screws may be a preferable option for those requiring rapid return to sport or work.