In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth

  • Original Scientific Article
  • Published:
European Archives of Paediatric Dentistry Aims and scope Submit manuscript

Abstract

Aim

To compare the accuracy of three electronic apex locators in the measurement of the root length of primary teeth with the radiographic and direct visual techniques.

Methods

The sample of the in vitro study was 102 root canals of primary molars. For the direct visual technique, the root canal file was introduced into the canal until its tip passed the apical foramen, then it was moved back to the limit of the foramen or root resorption and the measurement was made with a millimetre ruler. For the radiographic measurement, a calibrated and blinded researcher (κ > 0.87) measured the root canals using the same rule. Measurements with apex locators were performed by inserting the file into the canal until the word “Apex” or audible warning to indicate the foramen or resorption. For data analysis, one-way ANOVA for repeated measurements and the Sidak post hoc tests were performed.

Results

The direct measurement (µ = 8.57) did not present statistical difference when compared to the Root ZX II measurement (µ = 8.45), as opposed to the radiographic measurements (µ = 8.12), the Endus (µ = 8.24) and iPex II (µ = 8.29).

Conclusion

The use of apex locators was superior to the radiographic method and the Root ZX II was the most effective.

This is a preview of subscription content, log in via an institution to check access.

Access this article

We’re sorry, something doesn't seem to be working properly.

Please try refreshing the page. If that doesn't work, please contact support so we can address the problem.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  • Abdullah A, Singh N, Rathore MS, Tandon S, Rajkumar B. Comparative evaluation of electronic apex locators and radiovisiography for working length determination in primary teeth in vivo. Int J Clin Pediatr Dent. 2016;2(9):118–23.

    Google Scholar 

  • Ahmad IA, Pani SC. Accuracy of electronic apex locators in primary teeth: a meta-analysis. Int Endod J. 2015;48(3):298–307.

    Article  PubMed  Google Scholar 

  • Ahmed HM. Anatomical challenges, electronic working length determination and current developments in root canal preparation of primary molar teeth. Int Endod J. 2013;46(11):1011–22.

    PubMed  Google Scholar 

  • Angwaravong O, Panitvisai P. Accuracy of an electronic apex locator in primary teeth with root resorption. Int Endod J. 2009;42(2):115–21.

    Article  PubMed  Google Scholar 

  • Beltrame AP, Triches TC, Sartori N, Bolan M. Electronic determination of root canal working length in primary molar teeth: an in vivo and ex vivo study. Int Endod J. 2011;44(5):402–6.

    Article  PubMed  Google Scholar 

  • Bhat VK, Shetty P, Anandakrishna L. A Comparative evaluation of accuracy of new-generation electronic apex locator with conventional radiography to determine working length in primary teeth: an in vivo study. Int J Clin Pediatr Dent. 2017;10(1):34–6.

    Article  PubMed  PubMed Central  Google Scholar 

  • Borges MMB, Guimarães BM, Alves JD, et al. Evaluation of the accuracy of two apex locators in apical limit: an in vitro study. Rev Odontol Bras Central. 2016;25(74):126–9.

    Google Scholar 

  • Carvalho ANL, Moura-Netto C, Moura AAM, Marques MM, Davidowicz H. Accuracy of three electronic apex locators in the presence of different irrigating solutions. Braz Oral Res. 2010;24(4):394–8.

    Article  PubMed  Google Scholar 

  • Chen X, Liu X, Zhong J. Clinical and radiographic evaluation of pulpectomy in primary teeth: a 18-months clinical randomized controlled trial. Head Face Med. 2017;13(1):12.

    Article  PubMed  PubMed Central  Google Scholar 

  • Dandempally A, Muppa R, Duddu MK, Bhupatiraju P, Nallanchakrava S. Formulating a regression equation for determination of working length in primary molars using apex locators: a clinical study. Eur Arch Paediatr Dent. 2013;14(6):369–74.

    Article  PubMed  Google Scholar 

  • Govindaraju L, Jeevanandan G, Subramanian EMG. Comparison of quality of obturation and instrumentation time using hand files and two rotary file systems in primary molars: a single-blinded randomized controlled trial. Eur J Dent. 2017;11(3):376–9.

    Article  PubMed  PubMed Central  Google Scholar 

  • Jenkins JA, Walker WA, Schindler WG, Flores CM. An in vitro evaluation of the accuracy of the Root ZX in the presence of various irrigants. J Endod. 2001;27(3):209–11.

    Article  PubMed  Google Scholar 

  • Kielbassa AM, Muller U, Munz I, Monting JS. Clinical evaluation of the measuring accuracy of ROOT ZX in primary teeth. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;95(1):94–100.

    Article  PubMed  Google Scholar 

  • Kumar LV, Sreelakshmi N, Reddy ER, et al. Clinical evaluation of conventional radiography, radiovisiography, and an electronic apex locator in determining the working length in primary teeth. Pediatr Dent. 2016;38(1):37–41.

    PubMed  Google Scholar 

  • Leonardo MR, Silva LA, Nelson-Filho P, Silva RA, Raffaini MS. Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth. Int Endod J. 2008;41(4):317–21.

    Article  PubMed  Google Scholar 

  • Mello-Moura AC, Moura-Netto C, Araki AT, Guedes-Pinto AC, Mendes FM. Ex vivo performance of five methods for root canal length determination in primary anterior teeth. Int Endod J. 2010;43(2):142–7.

    Article  PubMed  Google Scholar 

  • Mente J, Seidel J, Buchalla W, Koch MJ. Electronic determination of root canal length in primary teeth with and without root resorption. Int Endod J. 2002;35(5):447–52.

    Article  PubMed  Google Scholar 

  • Navit S, Jaiswal N, Khan SA, et al. Antimicrobial efficacy of contemporary obturating materials used in primary teeth—an in-vitro study. journal of clinical and diagnostic research. J Clin Diagn Res. 2016;10(9):ZC09–12.

    PubMed  PubMed Central  Google Scholar 

  • Nelson-Filho P, Romualdo PC, Bonifácio KC, et al. Accuracy of the iPex multi-frequency electronic apex locator in primary molars: an ex vivo study. Int Endod J. 2011;44(4):303–6.

    Article  PubMed  Google Scholar 

  • Oznurhan F, Ünal M, Kapdan A, Ozturk C, Aksoy S. Clinical evaluation of apex locator and radiography in primary teeth. Int J Paediatr Dent. 2015;25(3):199–203.

    Article  PubMed  Google Scholar 

  • Parisay I, Ghoddusi J, Forghanib M. A review on vital pulp therapy in primary teeth. Iran Endod J. 2015;10(1):6–15.

    PubMed  Google Scholar 

  • Patiño-Marín N, Zavala-Alonso NV, Martínez-Castañón GA, et al. Clinical evaluation of the accuracy of conventional radiography and apex locators in primary teeth. Pediatr Dent. 2011;33(1):19–22.

    PubMed  Google Scholar 

  • Saritha S, Uloopi KS, Vinay C, Chandra SR, Rao VV. Clinical evaluation of Root ZX II electronic apex locator in primary teeth. Eur Arch Paediatr Dent. 2012;13(1):32–5.

    Article  PubMed  Google Scholar 

  • Tosun G, Erdemir A, Eldeniz AU, Sermet U, Sener Y. Accuracy of two electronic apex locators in primary teeth with and without apical resorption: a laboratory study. Int Endod J. 2008;41(5):436–41.

    Article  PubMed  Google Scholar 

  • Vasconcelos BC, Bueno MM, Luna-Cruz SM, Duarte MA, Fernandes CA. Accuracy of five electronic foramen locators with different operating systems: an ex vivo study. J Appl Oral Sci. 2013;21(2):132–7.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors are grateful to the Coordination for the Improvement of Higher Education Personnel (CAPES) for financial support (finance code: 001).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to J. C. Barasuol.

Ethics declarations

Conflict of interest

The authors declare that they have no conflicts of interest related to this study.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee (Process no. 2572237) and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Adriano, L.Z., Barasuol, J.C., Cardoso, M. et al. In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth. Eur Arch Paediatr Dent 20, 403–408 (2019). https://doi.org/10.1007/s40368-018-00413-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s40368-018-00413-5

Keywords

Navigation