Abstract
AIM: To evaluate the accuracy of Root ZX II electronic apex locator (EAL) for root canal working length determination in primary teeth. METHODS: Forty primary maxillary incisor teeth were included in the study. Root ZX II EAL was used to determine the electronic working length (EL). The working length obtained with EAL was evaluated using digital radiography with the measuring file set to EL. The samples were categorised into three groups. Group 1 (acceptable): file tip 0–1 mm short of the radiographic apex; Group 2 (short): file tip ≫ 1mm short of the apex and Group 3 (long): file tip beyond the apex. Digital radiographic working length (RL) was derived by adjusting EL to the radiographic apex. The assigned calibrations were tabulated and statistically analysed. RESULTS: Group 1 included 28 out of 40 teeth indicating a clinical accuracy of 70%. Group 2 contained 10 (25%) teeth whereas group 3 had only 2 (5%) teeth. STATISTICS: Pearson correlation coefficient statistical analysis showed a high correlation (r = +0.82; p =s < 0.001) among the groups. CONCLUSION: Root ZX II EAL can be used as a reliable device for obtaining root canal length in primary maxillary incisor teeth.
Similar content being viewed by others
References
Angwaravong O, Panitvisai P. Accuracy of electronic apex locator in primary teeth with resorption. Int Endod J 2009; 42:115–121.
Athar A, Angelopoulos C, Katz JO, Williams KB, Spencer P. Radiographic endodontic working length estimation: comparison of three digital image receptors. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008; 106:604–608.
Ebrahim AK, Yoshioka T, Kobayashi C, Suda H. The effects of file size, sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals: an in vitro study. Aust Dent J 2006; 51:153–157.
Elayouti A, Dima E, Ohmer J. Consistency of Apex Locator Function: A clinical Study. J Endod 2009; 35:2.
Ghaemmaghami S, Eberle J, Duperon D. Evaluation of RootZX Apex Locator in primary teeth. Pediatr Dent 2008; 30:496–498.
Goldberg F, De Silvio AC, Manfre S, Nastri N. In vitro measurement accuracy of an electronic apex locator in teeth with simulated apical root resorption. J Endod 2002; 28:461–463.
Gordon MPJ, Chandler NP. Electronic apex locators. Int Endod J 2004; 37:408–414.
Hoer D, Attin T. The accuracy of electronic working length determination. Int Endod J 2004; 37:125–131.
Horner K, Shearer AC, Walker A, Wilson NHF. “Radiovisiography”: An initial evaluation. Br Dent J 1990; 168:244–428.
Katz A, Mass E, Kaufman AY. Electronic apex locator: a useful tool for root canal treatment in the primary dentition. ASDC J Dent Child 1996; 63:414–417.
Kaufman AY, Keila S, Yoshpe M. Accuracy of a new apex locator: an in vitro study. Int Endod J 2002; 35:186–192.
Kielbassa AM, Muller U, Munz I, Monting JS. Clinical evaluation of the measuring accuracy of Root ZX in primary teeth. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003; 95:94–100.
Leonardo MR, Silva LAB, Nelson-Filho P, Silva RAB, Raffaini MSGG. Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth. Int Endod J 2008; 41:317–321.
Leonardo MR, Da Silva LA, Nelson-Filho P, Da Silva RA, Lucisano MP. In vitro comparative study of conventional radiography and Root ZX apex locator in determining root canal working length. N Y State Dent J 2009; 754:48–51.
Moshonov, Goldberg IS. Apex Locators: Updates and Prospects for the future. Int J Computerized Dent 2004; 7:359–370.
Nelson-Filho P, Lucisano MP, Leonardo MR, da Silva RA, da Silva LA. Electronic working length determination in primary teeth by ProPex and Digital Signal Procession. Aust Endod J. 2010; 36:105–108.
Odabas ME, Bodur H, Tulunoglu O, Alacam A. Accuracy of an electronic apex locator: A clinical evaluation in primary molars with and without resorption. J Clin Pediatr Dent 2011; 35:255–258.
Ounsi HF, Naaman A. In vitro evaluation of the reliability of the Root ZX electronic apex locator. Int Endod J 1999; 32:120–123.
Pratten DH, McDonald NJ. Comparison of radiographic and electronic working lengths. J Endod 1996; 22:173–176.
Sanabe ME, Basso MD, Goncalves MA, Cordiero RDCL, Santos Pinto L. Digital versus conventional radiography for determination of primary incisor length. Braz J Oral Sci 2009; 8:101–104.
Shabahang S, Goon WW, Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod 1996; 22:616–618.
Sim S. Modern endodontic practice: instruments and techniques. Dent Clin North Am 2004; 48:1–9.
Subramaniam P, Konde S, Mandanna DK. An in vitro comparison of root canal measurement in primary teeth. J Indian Soc Pedod Prev Dent. 2005; 23:124–125.
Sunada I. New method for measuring the Length of the root canal. J. Dent. Res 1962; 41:375–387.
Versteeg KH, Sanderink GC, Van Ginkel FC, Van der Stelt PF. Estimating distances on direct digital images and conventional radiographs. J Am Dent Assoc 1997; 128:439–443.
Weiger R, John C, Geigle H, Lost C. An in vitro comparison of two modern apex locators. J Endod 1999; 25:765–768.
Wrbas KT, Ziegler AA, Altenburger MJ, Schirrmeister JF. In vivo comparison of working length determination with two electronic apex locators. Int Endod J 2007; 40:133–138.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Saritha, S., Uloopi, K.S., Vinay, C. et al. Clinical evaluation of Root ZX II electronic apex locator in primary teeth. Eur Arch Paediatr Dent 13, 32–35 (2012). https://doi.org/10.1007/BF03262838
Published:
Issue Date:
DOI: https://doi.org/10.1007/BF03262838