Introduction

The discovery of radioactivity in 1896 has led to applications in different sectors, including medicine, nuclear power, and the arms industry. Among radionuclides, Sr-90 is one of the most dangerous fission products with respect to radiotoxicity due to its long physical and biological half-life (T1/2 = 28.50 y) (Wilken and Joshi 1991) and chemical similarity to calcium (Vaida and Kim 2010). The similar metabolic behaviour to that of calcium leads to radionuclide accumulation in bone tissues (Iammarino et al. 2018).

The radioactive decay chain of Sr-90 is as follows:

$$\mathrm{Sr}-90\to \mathrm{Y}-90\to \mathrm{Zr}-90$$

where Sr-90 decays in yttrium-90 (Y-90) by emitting a pure beta particle with Eβ,max of 546 keV, while Y-90 decays in stable daughter zirconium-90 (Zr-90), emitting a beta particle with higher energy (Eβ,max = 2280 keV). Though Y-90 has a shorter half-life (T1/2 = 64.10 h) than that of Sr-90, it is particularly dangerous because it damages the bones and bone marrow (Cabrera et al. 1999).

After approximately 2 weeks, Y-90 is in secular equilibrium with its parent nuclide; this property is exploited for the indirect determination of Sr-90 activity (dell’Oro et al. 2013). Therefore, the determination of Sr-90 can be performed in two ways: direct or indirect. The first approach involves the direct separation of Sr-90 from the other components and subsequent instrumental measurement. The second approach involves the indirect determination of Sr-90 by the determination of Y-90.

The accidental or deliberate release of radioactive material into the atmosphere leads to the fall-out of radionuclides on water and soil. In Europe, the Chernobyl disaster of 1986 resulted in substantial environmental contamination with Sr-90, involving 200,000 km2 surrounding the accident. Air contamination and rainfall intensity determined the extent of radionuclide deposition on the soil (World Health Organization 2005). The most recent nuclear accident occurred in March 2011 in Fukushima (Japan), where the cooling system failed after a tsunami flooded the power station, causing core reactor meltdown, and resulting in the release of radioactive materials and enormous food safety issues worldwide (Maxwell et al. 2013). Radionuclides may be introduced into the human body by inhalation, which is the most dangerous route of exposure immediately after a nuclear accident, or by ingestion, considered the main route of introduction via the food chain (soil, water vegetation food human). This necessitates food safety controls, especially immediately after a nuclear accident, requiring reliable, robust, and reproducible procedures. A number of methods have been developed for the determination of Sr-90 in the environment and foodstuffs and for calculating transfer factors, following the migration and uptake processes, and/or estimating the dose responses (Grahek et al. 2018).

These methods are based on different techniques, e.g., beta counting (Vaida and Kim 2010; Maxwell et al. 2015), solid or liquid scintillators (Kong et al. 2018), Cerenkov counting (Jiang et al. 2017), and mass spectrometry (Russell et al. 2017). The extraction/separation of Sr-90 from environmental and/or food matrices can be performed in different ways, including the fuming nitric acid method (Livens 1990), liquid–liquid extraction (Happel et al. 2003), ion-exchange (Landstetter and Wallner 2006), and extraction chromatography (Jeter and Grob 1994). The majority of these methods are not applicable to food, while the few procedures available, applied in food analysis, have a number of limitations, including the time-consuming processes, use of hazardous reagents, low repeatability and sensitivity, and presence of interfering radionuclides. Moreover, validation procedures are lacking, and this is a strong limitation, since the laboratories in charge of official food controls need fully validated radiochemical procedures in order to fulfil the ISO 17025 requirements.

The aim of this work was to establish and validate a rapid and sensitive method for the determination of Sr-90 activity in food and feed matrices by using specific resin (Sr-Resin) for Sr-90 separation (Surman et al. 2014). The measurement of Sr-90 was obtained by ultra-low-level liquid scintillation counting (LSC). This method was fully validated and then tested on reference materials, various food matrices (meat, fish, vegetables, and cheese), and animal feeds, and submitted to proficiency test round.

Materials and Methods

Chemicals and Working Standard Solutions

Nitric acid (≥ 65% w/v), oxalic acid (99%), and aluminium nitrate nonahydrate (≥ 98.5%) were supplied by Sigma-Aldrich (Steinheim, Germany). Each sample was treated by using Sr-Resin columns (10 mL, particle size: 100–150 µm) supplied by Triskem International (Bruz, France). The Ultima Gold AB scintillation cocktail was purchased from PerkinElmer (Waltham, MA, USA). Strontium certified standard solution (10,000 mg L−1) in 4% nitric acid (HNO3) was purchased from CPI International (Santa Rosa, CA, USA). Sr-90 standardised solution in 0.1 N HCl (with 10 µg mL−1 of SrCl2 carrier) at a concentration of 3.95 ± 0.07 kBq g−1 (date: January 1, 2018) was obtained from Physikalisch-Technische Bundesanstalt (Braunschweig, Germany). All solutions were prepared using ultrapure water with a specific resistance of 18.2 MΩ-cm, supplied by a Milli-Q RG unit (Millipore, Bedford, MA, USA).

Analytical Equipment

Sr-90 activity was detected by using a liquid scintillation counter (Quantulus Wallac 1200; PerkinElmer, Turku, Finland). WinQ Wallac and EASY View Spectrum Analysis Software (PerkinElmer) were used for instrumental control, data acquisition, and elaboration. All acquisition parameters are specified in Table 1.

Table 1 Liquid scintillation counter and inductively coupled mass spectrometer settings

The Sr “chemical yield” was verified by ICP-MS (ELAN DRC II ICP-Mass Spectrometer (PerkinElmer, Shelton, USA). A stable Sr calibration curve was obtained by analysing a Sr-88 standard solution in 2% HNO3 at concentrations of 0.0 (blank), 0.5, 2.5, 10.0, and 50.0 µg L−1 (internal standard: Rhodium solution 0.2 mg L−1). All acquisition parameters used for determination are specified in Table 1. The strontium yield (YSr) was calculated by applying the following Eq. (1):

$${\mathrm{Y}}_{\mathrm{Sr}}=\frac{{C}_{Sr}* {V}_{1}}{{C}_{Srcarrier}* {V}_{2}}* 100$$
(1)

where CSr is the strontium concentration detected by inductively coupled plasma mass spectrometry (ICP/MS); V1 is the volume collected from the column (150 mL); CSr carrier is the concentration of Sr carrier added (1 mL of 10,000 mg L−1 standard solution in 50 mL, corresponding to 200 mg L−1); and V2 is the initial sample volume (50 mL).

Sample Preparation

As reported by several authors, the radiochemical determination of Sr-90 is a difficult task owing to its chemical affinity with other alkaline earth elements, such as Ca and Ba. This affinity makes radiochemical separation a long and laborious task. Moreover, the isotopes Pb-210 and Bi-210 may interfere radiometrically with Sr-90 (United States Environmental Protection Agency 2020) making analytical results unreliable. During the last few years, specific resins (extractants) have been developed for the radiochemical separation of several radionuclides from interfering elements present in complex matrices. A resin based on (4,4′(5′)-di-t-butylcyclohexano-18-crown-6) (Fig. 1) may be used for the successful extraction of Sr-90 from liquid (i.e., water) and solid matrices (previously mineralised) (Triskem International 2016). However, the application to food samples is limited (Lopes et al. 2010). This can be attributed to the composition of foodstuffs, rich in minerals that may interfere with separation, making Sr-90 extraction difficult (e.g., in cheese, high calcium content may saturate the active sites of the column). In this study, sample preparation was optimised for the use of specific Sr-Resin columns for food samples analysis. An effective temperature ramp was used to obtain the complete ashing of the sample. Then, different concentrations of HNO3 (3 M, 5 M, and 8 M) and volumes (25, 50, 75, and 100 mL) were tested for ash solubilisation, accounting for the compatibility with the resin. Based on the literature, different volumes of solutions composed of oxalic acid, nitric acid, and aluminium nitrate were tested for column activation (25, 50, and 100 mL of 8 M HNO3), washing solutions (25, 50, and 100 mL), and final elution (100, 150, and 200 mL of 0.05 M HNO3).

Fig. 1
figure 1

Optimised sequence for Sr-90 chromatographic extraction by the Sr-Resin column

Method Optimisation

Since the pulse shape depends on the vial type and scintillation cocktail/sample characteristics, the discrimination conditions should refer to the specific characteristics of the analysed sample. In this work, a 20-mL polytetrafluoroethylene (PTFE) vial was used. The vial was filled with 8 mL of 0.1 M HCl and 12 mL of Ultima Gold AB scintillation cocktail, and the optimal value of the pulse shape analysis (PSA) parameter was evaluated by marking two distinct blank samples with 1.5 Bq of Am-241 (E = 5.49 MeV) and 1.5 Bq of Sr-90/Y-90 solution at secular equilibrium. Subsequently, the percentage of alpha events on the total recorded count in the beta count window and the relative uncertainty (and vice versa) was calculated for each emitter using the following equations:

$${\tau }_{\alpha }=\frac{\beta }{\alpha +\beta } {u}_{{\tau }_{\alpha }}=\frac{\sqrt{{\alpha }^{2}{u}_{\beta }^{2}+{\beta }^{2}{u}_{\alpha }^{2}}}{{(\alpha +\beta )}^{2}}$$
(2)
$${\tau }_{\beta }=\frac{\alpha }{\alpha +\beta } {u}_{{\tau }_{\beta }}=\frac{\sqrt{{\alpha }^{2}{u}_{\beta }^{2}+{\beta }^{2}{u}_{\alpha }^{2}}}{{(\alpha +\beta )}^{2}}$$
(3)

where α is the number of events measured in the alpha window, β is the number of events measured in the beta window, and uα and uβ are the uncertainties in α and β, respectively, associated with Poisson statistics for radioactive decay.

Finally, the interference parameters τα and τβ (Eqs. (2) and (3), respectively) were plotted as a function of the PSA value. The optimal PSA in the present study was 80, corresponding to alpha interference equal to 0.84% and beta interference of approximately 0.28%.

Validation Procedure

Specific guidelines for validating radiochemical methods are not available. As demonstrated previously (dell’Oro et al. 2014; Iammarino et al. 2015), method validation can be carried out following an in-house validation model, developed according to the most representative European regulations and guidelines related to chemical methods (European Commission 2002; Thompson et al. 2002; European Commission 2017; Iammarino 2019). This approach was integrated, where necessary, with specific ISO reference documents, to complete the validation relating to other radiochemical parameters (WHO and FAO 2011).

The Sr-90 counting efficiency (ε), which is the ratio of the counts per minute (CPM) to disintegrations per minute (DPM), expressed as a percentage (CPM/DPM %), was calculated by treating the Sr-90 source at four different activity concentrations, 0.1, 0.5, 1.0, and 5.0 Bq kg−1 (six repetitions each), by the whole sample preparation, as described below (see the “Optimisation of Sample Preparation” section).

Method selectivity is guaranteed by the use of the specific Sr-Resin column, able to separate the radionuclide of interest from other elements and radionuclides that may interfere with the analysis (especially K-40 and Pb-214). Selectivity was also evaluated by treating the Sr-90 source in equilibrium with Y-90, collecting the final eluate containing only Sr-90, and verifying the growth of Y-90 by repeated measures.

The limit of quantification (LOQ) and limit of detection (LOD) were assessed using the Currie approach (IAEA 1989; De Felice et al. 2017). As indicated by ISO 11929:2019 (International Organization for Standardization 2019), Part I, §5.6, the decision limit (in Bq kg−1) of a measurand from which a background noise component (or sample blank) is subtracted must be defined by studying the uncertainty associated with the latter. The decision threshold should be chosen to assure that the measurand (with its uncertainty) is significantly higher than the background noise with its uncertainty. The problem is solved, in accordance with Berger (1985), by applying an appropriate decision theory. The decision threshold is defined as:

$$P\left(y>{y}^{*}|\tilde{y }=0\right)=\alpha$$
(4)

That is, the decision threshold is defined by the probability (α) of obtaining a measure (y) greater than the decision threshold (y*) (where the true value is equal to 0 or not detectable). Using a quadratic loss function, the decision threshold is given by the relationship:

$${y}^{*}={k}_{1-\alpha }*\tilde{u }(0)$$
(5)

where \({k}_{1-\alpha }\) represents the (1-α) quantile of the normal distribution and \(\tilde{u }\) represents the standard uncertainty of the measurand.

Similarly, the detection limit is defined as the smallest true value for which the probability of an incorrect decision for the presence of the physical effect does not exceed the predefined decision threshold β (probability of a false negative):

$$P\left(y<{y}^{*}|\tilde{y }={y}^{\#}\right)=\beta$$
(6)

In this case, the decision limit y# is defined as the smallest of the solutions of the recursive equation:

$${y}^{\#}={y}^{*}+{k}_{1-\beta }*\tilde{u }({y}^{\#})$$
(7)

Generally, the situation in which α = β = 0.05 is considered, and \({k}_{1-\alpha }={k}_{1-\beta }=1.65\).

The linearity of the method was studied by building the calibration curve, using the data obtained from the counting efficiency tests. After obtaining the least squares regression equation, the coefficient of determination and the intercept significance of the calibration curve were evaluated.

The Sr chemical yield was determined by an ICP/MS analysis of six samples composed of ultrapure water (four samples) and animal feed (two samples) fortified with 1 mL of Sr carrier (10,000 mg L−1). These samples were treated by the full procedure and then the final Sr concentration was compared to the initial concentration to calculate the recovery (%).

Sr-90 activity (A) was evaluated by the following equation:

$$A=\frac{\left({r}_{g}-{r}_{0}\right)}{2*M*f\left({T}_{m}\right)*{\varepsilon }_{Sr+Y}*C}{e}^{{\lambda }_{Sr}T} Bq/kg$$
(8)

where M is the reference mass in kilogram; εSr+Y is the counting efficiency of the reference source of Sr-90 at the secular equilibrium with Y-90; λSr is the decay constant of Sr-90 in s−1; T is the time elapsed between the sampling date and the separation date; rg is the sample count rate (s−1); r0 is the blank count rate (s−1); and C is the correction factor for carrier recovery.

The previous relation is derived by integrating the Bateman equations.

Moreover:

$$f\left({T}_{m}\right)=1-\frac{{k}_{\varepsilon }}{1+{k}_{\varepsilon }}*{e}^{-{\lambda }_{Y}{T}_{m}}*\frac{1-{e}^{-{\lambda }_{Y}{T}_{g}}}{{\lambda }_{Y}{T}_{g}}$$
(9)

The equation is a function that measures the growth of Y-90 during the whole measurement period, in which:

  • Tm—is the time interval between chromatographic separation and the start of counting

  • λY—is the decay constant of Y-90 in s−1

  • Tg—is the counting time (s)

  • kε—is the ratio of the counting efficiencies of a pure Y-90 emitter and a pure Sr-90 emitter, considering that the two sources have the same geometry

The counting efficiency is a measure of the fraction of events producing a measurable pulse. Therefore, it depends not only on the scintillation efficiency (i.e., the average number of photons produced per unit of energy), but also on the photomultiplier characteristics and on the properties of the electronic measuring instrument. The total counting efficiency was evaluated by measuring a source of Sr-90/Y-90 at an activity concentration of 0.1, 0.5, 1.0, and 5.0 Bq kg−1. This efficiency is routinely used in the evaluation of the Sr-90 activity concentration. It differs significantly from the sum of the separate efficiencies for the single radionuclide Sr-90 and Y-90.

The k parameter contributes to the evaluation of the Y-90 growth factor both during the time between elution and instrumental reading and during the reading itself. The factor was evaluated by reading separately the two sources of Sr-90 and Y-90. The first was prepared by spiking ultrapure water with a certified solution of Sr-90/Y-90 followed by the specified technical procedure, and the second was prepared using the washing solution in which Y-90 is present. Ten values were obtained for each source, correcting for Y-90 ingrowth/decay.

The method accuracy was verified by estimating precision (as CV%) and trueness (as recovery %). A sample of animal feeds was prepared ad hoc for this validation step. The choice of an appropriate and representative matrix for evaluating method accuracy was based on two main requirements: (i) matrix composition; (ii) matrix contamination by Sr-90. More in depth, in order to simulate the most complex analytical conditions, a matrix characterised by both high levels of possible interfering metals and detectable amounts of Sr-90 was chosen. Since the composition and possible Sr-90 contamination is highly variable among different types of feed (Iammarino et al. 2015), a “mix” of several kinds of feed was prepared. Portions (175 g each) of forage, corn, bran, and complementary feed were ashed and carefully homogenised. This “pool” of animal feed ashes was subdivided in portions corresponding to 50.0 g of fresh weight. Two samples were analysed to evaluate the activity concentration of Sr-90 (if present). The remaining 12 portions were fortified at two levels (0.5 Bq kg−1 and 1.0 Bq kg−1, six repetitions each) and analysed.

Ruggedness under major changes condition was evaluated using different matrix types (two repetitions for each type). The recovery % and CV% were compared with those obtained during the validation using animal feeds. The following samples were utilised (50 g each): cheese pool, including hard cheese, spreadable fresh cheese, and slices of sweet cheese; meat pool, including hamburger, fresh sausage, and mortadella; fish pool, including sea bream, sea bass, mussels, clams, prawns, and cuttlefish; and vegetable pool (frozen vegetable soup). Each sample was analysed before fortification in duplicate to ensure the absence of Sr-90 (< LOQ).

The measurement uncertainty was calculated according to UNI CEI ENV 13,005:2000 (Ente Italiano di Normazione 2000) and ISO 21748:2010 (International Organization for Standardization 2010). The following sources of uncertainty were evaluated and included in the final calculation: net counting, sample weight, Sr chemical yield, counting efficiency, Y-90 decay time, and repeatability.

Proficiency Test Round

In May 2021, this radioanalytical method was tested within a Proficiency Test organised by ERA (Waters™, Golden, CO, USA, an accredited ISO/IEC 17,043:2010 agency), involving 32 laboratories from different countries. A sample composed of a mix of different vegetables was analysed for the quantification of Sr-90. The measured value was 36.7 ± 4.8 Bq kg−1, against a true value of 43.5 Bq kg−1 (satisfactory range: 31.2–55.9 Bq kg−1), resulting in a Z-score equal to − 1.11 (Table 3).

Results and Discussion

Optimisation of Sample Preparation

The proposed method includes three steps: sample ashing, Sr-90 separation by a Sr-Resin column, and determination of Sr-90 activity by LSC. The first two steps were optimised with respect to ashing times and temperatures (to minimise the times and obtain clear ashes without residues) and Sr-Resin column usage was optimised. The optimised sample preparation procedure was as follows: 50 g of homogenised sample was ashed using the following temperature ramp, 250 °C (5 h), 550 °C (14 h), 600 °C (5 h). This procedure is faster than the conventional approach (Iammarino et al. 2017), which requires almost 3 days for completion. This improvement was achieved by using containers composed of more resistant material (zirconium) in place of porcelain. After cooling, the ashes were dissolved in 50 mL of 65% (w/v) HNO3, directly in the container, transferred in a 500-mL glass flask, and 1 mL of 10,000 mg L−1 strontium standard solution (Sr carrier) was added. Then, a first aliquot of 50 µL of the solution was withdrawn, transferred to a 50-mL volumetric flask, and diluted to 50 mL by the addition of 0.1 M HNO3. This step is used for determining the Sr chemical yield by inductively coupled plasma mass spectrometry (ICP/MS). The traditional procedure used for chemical yield verification, performed by weighing the final counting form of the element, as a precipitate recovered on filter papers or on counting discs composed of inert metals, is based on a series of analytical steps that may compromise method reproducibility. Inversely, ICP/MS guarantees high sensitivity, selectivity, and accuracy (Iammarino et al. 2016a, 2015). If necessary, the mixture was treated by leaching at 320 °C on a hot plate for a few minutes to completely solubilise the ashes. The mixture was then filtered under a vacuum (~ 0.3 bar) using Purabind 0.45 filters (0.45 µm, 47 mm; Whatman, Dassel, Germany) and 25 mL of 2 M aluminium nitrate was added. A Sr-Resin column was placed on a vacuum chamber (a maximum vacuum of 0.3 bar may be applied, if necessary) and Sr-90 extraction was performed following the optimised scheme presented in Fig. 1. Indeed, the complete elution of stable Sr was achieved using 150 mL of 0.05 M HNO3. This result was confirmed by the experimental construction of the elution curve using stable Sr. A second aliquot of 50 µL of the final solution eluted from the column was withdrawn, transferred into a 50-mL volumetric flask, and diluted to 50 mL by the addition of 0.1 M HNO3 for subsequent Sr-90 chemical yield determination by ICP/MS, comparing the mass concentrations of the two withdrawn aliquots. The solution was then evaporated to dryness on a hot plate; the residue was dissolved in 8 mL of 0.05 M HNO3 (to avoid any loss of counting efficiency due to the high quenching effect of high molar HNO3 solutions) and transferred in a scintillation vial, together with 12 mL of scintillation cocktail. This mixture was used for the determination of Sr-90 by ultra-low-level LSC. Counting was started 2 h after the addition of the scintillation cocktail, since this time gap ensures the resetting of both chemiluminescence and photoluminescence.

Optimisation of Measurement Conditions

The optimal alpha–beta discrimination conditions can be identified by the PSA parameter value such that the PSA circuit better discriminates alpha from beta events. Two samples were measured at different PSA, one traced with a pure alpha emitter and the other traced with a pure beta emitter, to evaluate the percentage of events recorded in the wrong counting window. The optimal operating condition corresponds to the situation in which both alpha interference (% of alpha events recorded in the beta counting window) and beta interference (% of beta events recorded in the alpha counting window) are minimised. The optimal condition corresponds to the intersection point of two curves. The optimised acquisition parameters of both LSC and ICP-MS are reported in Table 1. The detector channels used for spectral acquisition were selected to exclude the counting rate due to photoluminescence and emission of H-3 (which may be present in both reagents and samples). In these conditions, the total efficiency was 0.996 ± 0.036. The efficiencies of Sr-90 and Y-90 were 0.51 and 0.61, respectively, with a calculated k factor of 1.19.

Validation Parameters

Critical validation parameters are summarised in Table 2.

Table 2 Validation parameters

Selectivity

The selectivity of a method represents the ability to determine the analyte of interest in the presence of interference. Selectivity was conferred by the specific Sr-resin chromatographic column capable of separating the analyte of interest from matrix interferences (e.g., Ca, K-40, and Pb/Bi isotopes).

A separation test of a Sr-90 source in secular equilibrium with Y-90 was carried out to evaluate the selectively of the chromatographic column. A feed sample with a high K-40 content (about 750 Bq, measured by HPGe gamma spectrometry) was fortified with this source, and the ingrowth of Y-90 in the final eluate containing only Sr-90 and the decay of Y-90 in the waste were evaluated. A series of 15 successive readings excluded the presence of potassium in the final eluate and confirmed the presence of a strong signal from K-40 in the waste. In the latter case, as expected, the final CPM values were in agreement with the blank value of 1.840 ± 0.096 CPM (Fig. 2). The selectivity of the method was confirmed following the growth of Y-90 obtained from repeated measures on the final eluate containing only Sr-90 (Fig. 3A).

Fig. 2
figure 2

Time evolution of measured Sr-90 (alone) and Y-90 (alone) sources and the theoretical trend

Fig. 3
figure 3

Spectral overlay: A pure Sr-90 source (red) and Y-90 growth over 15 days (green); B three samples, IAEA-154 (blue), spiked feed (red), and blank sample (green)

Linearity

The linearity of the method was studied by building a calibration curve using the levels adopted during the counting efficiency tests (0.1, 0.5, 1.0, and 5.0 Bq kg−1). The coefficient of determination (r2) was 0.9995, confirming linearity within the range: LOQ (0.024, see the “Characteristic Limits” section) to 5.0 Bq kg−1. Obviously, the measurement range may be expanded by opportune sample dilution.

Sr Chemical Yield

The Sr chemical yield was assessed by ICP/MS using six samples (four water and two feed) fortified with 1 mL of Sr carrier. The yield was 89.2 ± 11.8%.

Accuracy

The accuracy of the method was evaluated by analysing the validation matrix (see the “Validation Procedure” section) fortified at two levels (0.5 and 1.0 Bq kg−1, 6 repetitions each). The recovery % and CV% values are reported in Table 2. The mean recovery percentage was 100.1% and the CV% was less than 13.1%. These parameters are comparable to those of other methods (Iammarino et al. 2016b), supporting the accuracy of the method. Accuracy was also evaluated by five replicated measurements of the IAEA-154 certified matrix (whey powder), whose recommended value on August 31, 1987, was 6.90 Bq kg−1 (95% confidence interval: 6.00–8.00), and an average value of 6.14 ± 0.80 Bq kg−1 was obtained. The good agreement with the declared value for the reference material (Z-score: 0.69, En-score: 0.47) confirmed the accuracy of the method. Figure 3B shows three representative spectra for samples analysed during validation (Table 3).

Table 3 Proficiency test and reference material analysis

Robustness

Briefly, robustness refers to the ability of the method to yield consistent results, irrespective of variation introduced during its execution. Robustness was evaluated under major changes condition on alternative matrices: cheese, meat, fish, and vegetables. Each alternative matrix was formulated by pooling various products (see the “Validation Procedure” section).

Each pooled sample was first analysed to verify the absence of Sr-90 (activity concentration lower than the decision threshold). Following this verification, each pooled sample was spiked with Sr-90 at an activity concentration of 0.5 Bq kg−1. The resulting samples were processed according to the method described above. CV% values for each pool, as an indicator of robustness, were lower than 10.0% (Table 2). This result supports the broad application of the method to solid food matrices (cheese, meat, seafood, and vegetables) and animal feed, since CV% calculated were lower than the estimated value obtained during validation (10.7%). Moreover, considering the ductility of the Sr-Resin, it is possible that the method can also be applied to environmental solid matrices, such as soil and solid waste, in addition to food liquid matrices (e.g., milk), after proper sample solidification before ashing.

Characteristic Limits

To evaluate the characteristic limits, a Bayesian approach was used.

By replacing the quantities \(\tilde{u }(y)\) with the uncertainties determined during the validation and solving the associated equations recursively, the following numerical results were obtained (10) and (11):

$$LOD= {y}^{*}=0.011 {Bq kg}^{-1}$$
(10)
$$LOQ= {y}^{\#}=0.024 {Bq kg}^{-1}$$
(11)

The sensitivity of this procedure was higher than those of various previously established approaches for the determination of Sr-90 in solid matrices (Heilgeist 2000; Kabai et al. 2011; Shuzhong et al. 2015; Temba et al. 2011) and comparable to several others (Gerstmann and Tschöpp 2006; Lopes et al. 2010; Suomela et al. 1993).

Assessment of Measurement Uncertainty

All sources of uncertainty with substantial contributions to overall uncertainty were evaluated, as proposed by the UNI CEI ENV 13,005:2000 standard and in accordance with the ISO 21748:2010 standard, which accounts for repeatability. The relative contributions of sources uncertainties were as follows: net counts (2.80%), sample mass (0.01%), strontium yield (11.2%), counting efficiency (3.7%), ingrowth of Y-90 (3.0%), and repeatability (3.9%). Thus, the overall measurement uncertainty was 13.1%.

Owing to the established selectivity of the method, its extension to proportional counting is straightforward, given the lower background noise (counting efficiency generally lower than that for the liquid scintillator) and suitability for routine controls. In fact, the sample in the final reduction step can be dried directly on a stainless-steel counting planchet and checked for Sr-90. In this case, the sensitivity should be slightly lower, since proportional counting is less efficient than LSC.

Proficiency Test Round

The sample of vegetables analysed within the Proficiency test gave a Sr-90 activity concentration equal to 31.1 Bq kg−1. The Z-score value obtained was satisfactory, confirming the accuracy of this radioanalytical method for the quantitative determination of Sr-90 in solid foods.

Application to Commercial Samples

During validation, several solid food samples were collected from local stores (stored in the laboratory at − 18 °C) and analysed to verify the applicability of the method to routine radio-surveillance in food safety. Pooled samples of solid foods were prepared. The Sr-90 activity concentrations were lower than the LOQ (11.0 mBq kg−1), confirming the negligible level of Sr-90 in solid foods. Several animal feed samples were also analysed during official radio-surveillance activity at Italian National Reference Center for the Detection of Radioactivity in Feed and Foodstuffs. The Sr-90 activity concentrations detected, in the range of LOQ to 4.22 Bq kg−1, are reported in Table 3. The mean activity concentrations of Sr-90 detected in animal feeds were comparable to those reported by Iammarino et al. (2015) which registered mean levels equal to 0.77, 2.07, and 2.93 Bq kg−1 in animal feed, silage, and hay, respectively. This brief monitoring confirmed that animal feeds (especially not processed feed, such as hay) might be contaminated by not negligible levels of Sr-90. Another confirmation of this important aspect of food safety and animal wellness is the low percentage of samples with not quantifiable activity concentrations of Sr-90 (only 1 sample out of 24 analysed, 0.04%). So, the application of this novel radiochemical approach for the determination of Sr-90 in solid matrices confirmed its reliability, particularly for the quantification of this radionuclide in animal feed.

Comparison to Other Available Methods

The proposed radiochemical method presents several aspects of novelty, representing an improvement from the current state-of-the-art. Indeed, some rapid radiochemical procedures for the determination of Sr-90, which use extraction chromatographic resins, are already available in the literature (Maxwell et al. 2010; 2016; Maxwell 2018). However, these methods were applied to other types of matrix than food (i.e., limestone, marble, vegetation, etc.). In food analysis, method selectivity is one of the most significant parameters to ascertain during validation, since it is high variable among different foods, and it must be assured for avoiding “false positive” responses. The main effort of this study was the ascertainment of method applicability to the highest number of wide-consumed foods, such as meat, cheese, seafood, and vegetables, through the fundamental verification of validation parameters such as selectivity and accuracy.

The main advantage of this method over other methods used for foodstuffs analysis (Lopes et al. 2010; Iammarino et al. 2016a) is the ability to obtain fast and accurate results without waiting for Y-90 ingrowth. Thus, the method enables the final quantification of Sr-90 in 48 h (excluding the ashing step), while the Y-90 analysis requires more than 15 days to reach secular equilibrium. As a consequence, the times and cost are significantly reduced. Comprehensive validation, with analyses of all radiochemical parameters, was performed in this study. Lopes et al. (2010) reported some relevant parameters, such as the CV% and recovery percentage. However, only three repetitions were performed for each level, using very high concentration activities, not common in the routine control of food and feed. Compared to the validation parameters reported by Iammarino et al. (2016a), the CV% is similar, the mean recovery is higher, and the measurement uncertainty is slightly lower for the newly proposed method. For samples of equal weights, the LOD and LOQ values are lower. Accordingly, the present method is more sensitive due to the higher counting efficiency and chemical recovery, even compared with the method published by Pan et al. (2017). Using this method, it is important to control the precise measurement of the time between Sr-90 elution and instrumental counting. In fact, in this method, the determination of the contribution of Y-90 is crucial for the precise quantification of the Sr-90 activity concentration.

In Table 4, the comparison of main analytical performances of radiochemical methods for the determination of 90Sr in food, actually available, is reported. From this table, the main advantages of the proposed method are well appreciable. The most important is the application field, which allows applying the procedure for all most important solid matrices exploiting, as stated above, the ductility of the Sr-Resin. If compared to the only method comparable under this point of view (Iammarino et al. 2016a), there is an evident advantage in terms of time requested for a single analysis. Another substantial novelty of this study is the proposed approach for the development of a full validation procedure in radiochemical determinations. This approach allows the verification of most important analytical parameters and of measurement uncertainty that is a rarely estimated parameter within this context Table 5.

Table 4 Evaluation of Sr-90 contamination in different types of animal feed
Table 5 Comparison of main analytical performances of radiochemical methods for the determination of 90Sr in food

Conclusion

The proposed method for the determination of Sr-90 in solid foodstuffs is very simple and suitable for routine analyses of food samples. The method is based on Sr-90 extraction/separation by specific Sr-Resin and detection/quantification by ultra-low-level LSC. If compared to others available, this approach is more manageable in terms of times, costs, operator safety, reagents consumption, and waste. The method was optimised and validated following an in-house validation model according to relevant legislation. Good analytical performance was obtained in terms of specificity, linearity, sensitivity (minimal detectable activity: 11.0 mBq kg−1), mean repeatability (CV%: 10.7%), mean recovery (89.2%), and measurement uncertainty (13.1%). The procedure was also applied to reference materials and commercial samples, confirming its reliability. All these verifications assured method applicability in routinely control activity of Sr-90 levels in solid foods and animal feed. Finally, the successful application of the procedure to highly complex matrices, like feed and forages, suggests that it can be extended to environmental samples, like soil and waste, even in the case of nuclear emergencies.