Objective
To evaluate the value of Bayesian penalized likelihood (BPL) reconstruction for improving lesion conspicuity of malignant lung tumors on 18F-fluoro-2-deoxy-d-glucose (FDG) positron emission tomography computed tomography (PET/CT) as compared with the ordered subset expectation maximization (OSEM) reconstruction incorporating time-of-flight (TOF) model and point-spread-function (PSF) correction.
Methods
Twenty-nine patients with primary or metastatic lung cancers who underwent 18F-FDG PET/CT were retrospectively studied. PET images were reconstructed with OSEM + TOF, OSEM + TOF + PSF, and BPL with noise penalty strength β-value of 200, 400, 600, and 800. The signal-to-noise ratio (SNR) was determined in normal liver parenchyma. Lung lesion conspicuity was evaluated in 50 lung lesions by using a 4-point scale (0, no visible; 1, poor; 2, good; 3, excellent conspicuity). Two observers were independently asked to choose the most preferred reconstruction for detecting the lung lesions on a per-patient level. The maximum standardized uptake value (SUVmax) was measured in each of the 50 lung lesions.
Results
Liver SNR on the images reconstructed by BPL with β-value of 600 and 800 (17.8 ± 3.7 and 22.5 ± 4.6, respectively) was significantly higher than that by OSEM + TOF + PSF (15.0 ± 3.4, p < 0.0001). BPL with β-value of 600 was chosen most frequently as the preferred reconstruction algorithm for lung lesion assessment by both observers. The conspicuity score of the lung lesions < 10 mm in diameter on images reconstructed by BPL with β-value of 600 was significantly greater than that with OSEM + TOF + PSF (2.2 ± 0.8 vs 1.6 ± 0.9, p < 0.0001), while the conspicuity score of the lesions ≥ 10 mm in diameter was not significantly different between BPL with β-value of 600 and OSEM + TOF + PSF. The mean SUVmax was increased by BPL with β-value of 600 for the lung lesions with < 10 mm in diameter, compared to OSEM + TOF + PSF (3.4 ± 3.1 to 4.2 ± 3.5, p = 0.001). In contrast, BPL with β-value of 600 did not provide increased SUVmax for the lesions ≥ 10 mm in diameter.
Conclusion
BPL reconstruction significantly improves the detection of small inconspicuous malignant tumors in the lung, improving the diagnostic performance of PET/CT.
Similar content being viewed by others
References
Maffione AM, Grassetto G, Rampin L, Chondrogiannis S, Marzola MC, Ambrosini V, et al. Molecular imaging of pulmonary nodules. AJR Am J Roentgenol. 2014;202:W217–W223223.
van Tinteren H, Hoekstra OS, Smit EF, van den Bergh JH, Schreurs AJ, Stallaert RA, et al. Effectiveness of positron emission tomography in the preoperative assessment of patients with suspected non-small-cell lung cancer: the PLUS multicentre randomised trial. Lancet. 2002;359:1388–93.
Gould MK, Maclean CC, Kuschner WG, Rydzak CE, Owens DK. Accuracy of positron emission tomography for diagnosis of pulmonary nodules and mass lesions: a meta-analysis. JAMA. 2001;285:914–24.
Lois C, Jakoby BW, Long MJ, Hubner KF, Barker DW, Casey ME, et al. An assessment of the impact of incorporating time-of-flight information into clinical PET/CT imaging. J Nucl Med. 2010;51:237–45.
Boellaard R, van Lingen A, Lammertsma AA. Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. J Nucl Med. 2001;42:808–17.
Alessio AM, Stearns CW, Tong S, Ross SG, Kohlmyer S, Ganin A, et al. Application and evaluation of a measured spatially variant system model for PET image reconstruction. IEEE Trans Med Imaging. 2010;29:938–49.
Varrone A, Sjoholm N, Eriksson L, Gulyas B, Halldin C, Farde L. Advancement in PET quantification using 3D-OP-OSEM point spread function reconstruction with the HRRT. Eur J Nucl Med Mol Imaging. 2009;36:1639–50.
Tong S, Alessio AM, Kinahan PE. Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation. Phys Med Biol. 2010;55:1453–73.
Akamatsu G, Mitsumoto K, Taniguchi T, Tsutsui Y, Baba S, Sasaki M. Influences of point-spread function and time-of-flight reconstructions on standardized uptake value of lymph node metastases in FDG-PET. Eur J Radiol. 2014;83:226–30.
Prieto E, Dominguez-Prado I, Garcia-Velloso MJ, Penuelas I, Richter JA, Marti-Climent JM. Impact of time-of-flight and point-spread-function in SUV quantification for oncological PET. Clin Nucl Med. 2013;38:103–9.
Tong S, Alessio AM, Kinahan PE. Image reconstruction for PET/CT scanners: past achievements and future challenges. Imaging Med. 2010;2:529–45.
Adams MC, Turkington TG, Wilson JM, Wong TZ. A systematic review of the factors affecting accuracy of SUV measurements. AJR Am J Roentgenol. 2010;195:310–20.
Ahn S, Ross SG, Asma E, Miao J, Jin X, Cheng L, et al. Quantitative comparison of OSEM and penalized likelihood image reconstruction using relative difference penalties for clinical PET. Phys Med Biol. 2015;60:5733–51.
Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV. Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT system. J Nucl Med. 2015;56:1447–522.
Teoh EJ, McGowan DR, Bradley KM, Belcher E, Black E, Moore A, et al. 18F-FDG PET/CT assessment of histopathologically confirmed mediastinal lymph nodes in non-small cell lung cancer using a penalised likelihood reconstruction. Eur Radiol. 2016;26:4098–106.
Parvizi N, Franklin JM, McGowan DR, Teoh EJ, Bradley KM, Gleeson FV. Does a novel penalized likelihood reconstruction of 18F-FDG PET-CT improve signal-to-background in colorectal liver metastases? Eur J Radiol. 2015;84:1873–8.
Fukukita H, Suzuki K, Matsumoto K, Terauchi T, Daisaki H, Ikari Y, et al. Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 2.0. Ann Nucl Med. 2014;28:693–705.
Teoh EJ, McGowan DR, Bradley KM, Belcher E, Black E, Gleeson FV. Novel penalised likelihood reconstruction of PET in the assessment of histologically verified small pulmonary nodules. Eur Radiol. 2016;26:576–84.
Howard BA, Morgan R, Thorpe MP, Turkington TG, Oldan J, James OG, et al. Comparison of Bayesian penalized likelihood reconstruction versus OS-EM for characterization of small pulmonary nodules in oncologic PET/CT. Ann Nucl Med. 2017;31:623–8.
Bettinardi V, Presotto L, Rapisarda E, Picchio M, Gianolli L, Gilardi MC. Physical performance of the new hybrid PETCT Discovery-690. Med Phys. 2011;38:5394–411.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that there is no conflict of interest or industry financial support for this research.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Kurita, Y., Ichikawa, Y., Nakanishi, T. et al. The value of Bayesian penalized likelihood reconstruction for improving lesion conspicuity of malignant lung tumors on 18F-FDG PET/CT: comparison with ordered subset expectation maximization reconstruction incorporating time-of-flight model and point spread function correction. Ann Nucl Med 34, 272–279 (2020). https://doi.org/10.1007/s12149-020-01446-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12149-020-01446-x