Introduction

While the overall percentage of patients requiring revision TKA is low, more than 700,000 primary TKAs are performed annually in the United States alone [7], resulting in a large number of revision procedures; this carries considerable healthcare and financial implications. Schroer et al. [28] reported on the reasons for revision of more than 800 TKAs during 2010 and 2011 at six arthroplasty centers in the United States; approximately half were related to malalignment of the TKA components and wear of the polyethylene tibial insert. Malalignment before and after primary TKA have been linked to decreased implant survival [27] and inferior patient-reported outcomes [17]. Conversely, proper component alignment is strongly associated with greater stability, a lower likelihood of loosening, and less polyethylene wear [5, 13, 26], although not all studies have concurred on this finding [6, 19, 23].

Implant malalignment in primary TKA has been reported to increase the stresses placed on the bearing surfaces and the fixation interfaces between the implant components and the surrounding cement and bone [9, 10]. This has been supported by implant retrieval analysis studies of cruciate retaining (CR) [30, 32] and posterior-stabilized (PS) implants in TKAs [22], with both types of implants showing increased polyethylene damage with implant malalignment. However, major limitations of these studies include the reliance on prerevision rather than postoperative radiographs to measure alignment, the absence of preoperative radiographs before the index arthroplasty to see whether the initial deformity has a propensity to recur, the focus on a single CR or PS design, and the inclusion of results from primary and revision TKAs. A large, longitudinally maintained registry coupled with an implant retrieval program could be used to consider the effect of implant alignment on polyethylene damage while controlling for patient variables (such as age and BMI) and implant characteristics (like the design and level of constraint), a marked improvement over previous studies [22, 30, 32]. Having serial radiographs from which to measure preoperative, postoperative, and prerevision alignment would provide additional valuable information to assess how surgical alignment affects TKA outcome.

We therefore combined our institutional registry of TKA radiographs and our implant-retrieval program to ask three research questions: (1) What is the relative polyethylene damage on medial and lateral compartments of the tibial plateaus from revised TKAs? (2) Does coronal TKA alignment affect implant damage, such that TKAs aligned in varus are predisposed to experience increased polyethylene damage? (3) Does TKA alignment differ between postoperative and prerevision radiographs, and if so what does this difference suggest about the mechanical contact load on a knee with a TKA?

Patients and Methods

From May 2007 to December 2012, 18,065 primary TKAs performed in 16,083 patients at a single institution were recorded in an institutional review board-approved longitudinally maintained total joint arthroplasty registry. Patient demographics, preoperative health status, medical comorbidities, and clinical and patient-reported outcome measures were collected. If patients later underwent revision surgery at our institution, the date and reason for revision surgery were recorded as a part of the registry.

By March 2016, 417 (2%) of the primary TKAs in the registry had been revised (Fig. 1). All patients had their index and revision operations performed at our institution. We did not account for patients who had revisions performed at outside institutions. For our study, 239 TKAs that underwent revision less than 2 years after surgery were excluded because a minimum of 2 years followup typically is required for publishing clinical outcomes after joint arthroplasty, and because these patients may not have fully returned to their normal activities of daily living, and therefore the damage patterns on their retrieved polyethylene tibial inserts may not reflect implant alignment. Of the remaining 178 primary TKAs, another 18 were excluded because the tibial insert was not explanted during revision surgery, and four more were excluded because the inserts were lost or returned to the patients before the study was started. Therefore, we report on 156 retrieved polyethylene tibial inserts (in 153 patients) revised at greater than 2 years after the primary TKA.

Fig. 1
figure 1

A flow diagram shows the method for selecting retrieved TKA implants in this study.

The patients (46 males, 107 females) had a mean age of 61 years (range, 28–86 years) and a mean BMI of 31 kg/m2 (range, 18–51 kg/m2). Reasons for revision surgery included aseptic loosening in 51 TKAs (33%), instability in 46 (29%), infection in 25 (16%), stiffness in 12 (8%), periprosthetic fracture in seven (4%), and other reasons, such as persistent pain, in 15 (10%). Sixty-three (41%) of the implants were manufactured by Exactech (Gainesville, FL, USA), 50 (32%) by Smith & Nephew (Memphis, TN, USA), 30 (19%) by Zimmer (Warsaw, IN, USA), and 13 (8%) by other manufacturers. Most of the implants were PS designs (122 or 78%), while the remaining 34 (or 22%) were constrained condylar knees (CCK).

Visual Assessment of Damage

Polyethylene damage modes of burnishing, pitting, scratching, delamination, surface deformation, abrasion, and third-body debris were subjectively graded on a scale of 0 to 3 to reflect the extent and severity of each damage mode [15]. Two observers (ZL, CNK) who were blinded to the reasons for revision surgery independently assessed damage on the articular surfaces in each region using a light stereomicroscope (×10 to ×32 magnification). To assess regional differences in damage to the retrieved tibial inserts, the medial and lateral compartments were divided into four regions (Fig. 2A). Because our research questions dealt with the effect of coronal alignment on polyethylene damage, only the medial and lateral surfaces (not the anterior or posterior surfaces) of the PS and CCK tibial posts were graded. The backsides of the tibial inserts also were divided into four regions (Fig. 2B). If the difference in total damage scores summed across all regions was greater than 10 points between the two observers, then a third observer (EB) scored the insert as well. The interobserver kappa statistics for the damage grading was 0.80, which is considered good agreement for interobserver variability [1]. The maximum total damage score for each of the 13 regions on the tibial insert was 21 (seven damage modes × maximum score of 3 for each mode). Therefore, the medial and lateral compartments and the backside had maximum possible scores of 84 (four regions × seven damage modes × maximum score of 3 for each mode). The posts had a maximum possible score of 42 (two regions × seven damage modes × maximum score of 3 for each mode).

Fig. 2A–B
figure 2

All TKAs in this study used posterior-stabilized implants. The (A) top surface of the polyethylene tibial insert was divided in 10 regions and the (B) backside surface was divided in four zones for damage scoring.

Radiographic Analysis

Preoperative knee alignment, postoperative TKA alignment (at approximately 6 weeks after TKA), and prerevision TKA alignment were measured on weightbearing AP-view radiographs. For our first research question, we had 151 postoperative radiographs of TKAs available (from the 156 retrieved TKA implants) to evaluate the effect of postoperative alignment on implant damage. For our second research question, radiographs were available at all three times for 140 of the 156 TKAs (Fig. 1) in our institutional picture archiving and communication system (PACS, Sectra Imtec AB, Linköping, Sweden). The radiographs from these 140 TKAs were used to evaluate alignment differences between preoperative, postoperative, and prerevision radiographs.

The overall anatomic knee alignment was defined as the angle between the femoral anatomic axis and tibial anatomic axis and was expressed as a deviation from 180° with varus being a negative and valgus being a positive value (Fig. 3) [24]. Anatomic axes of the femur and tibia were defined as a line drawn through two mid-points between the outer boundaries of the bones at distances of 100 mm and 150 mm from the joint line [29].

Fig. 3
figure 3

The same radiograph is shown twice in this figure to show the angles measured for femoral alignment (angle A), tibial alignment (angle B), and overall alignment (angle C) of a TKA. This postoperative AP radiograph is from a 52-year-old woman who had a left TKA, which was revised 3 years later for instability. The femoral component alignment (angle A) was 5° valgus, the tibial component alignment (angle B) was 90°, and overall anatomic knee alignment (angle C) was 5° valgus.

The alignments of the femoral and tibial components in the coronal plane were measured according to the protocol recommended by the Knee Society [18]. Femoral component alignment was defined as the angle between the distal surface of the femoral component and the anatomic axis of the femur and was expressed as a deviation from 90° with varus being a negative and valgus being a positive value. Tibial component alignment was defined as the angle between the inferior surface of the tibial baseplate and the anatomic axis of the tibia, with greater than 90° for a valgus component and less than 90° for a varus component. All alignment measurements were performed using the PACS system, with an accuracy of 1 mm for distance and 1° for angle.

All radiographs were reviewed by two observers (ZL, CIE), who had no preevaluation knowledge of the patients’ demographic or clinical data. The interobserver reliability for the radiographic examinations was estimated using intraclass correlation coefficients (ICCs). The interobserver kappa statistics for the radiographic alignment measurements ranged from 0.72 to 0.92, which is considered good to excellent agreement for interobserver variability [1].

Statistical Analysis

Damage scores were summarized using mean values with SDs. Damage scores were compared among regions using t tests. Overall knee alignment, femoral component alignment, and tibial component alignment were analyzed as continuous and categorical variables. We used a previously established classification system [5, 14, 28] to categorize the ranges for varus, neutral, and valgus for overall knee alignment (varus < 3°, neutral 3° to 7°, valgus > 7°) and individual alignment of the tibial (varus 80° to 86°, neutral 87° to 93°, valgus 94° to 95°) and femoral components (varus −3° to 3°, neutral 4° to 7°, valgus 8° to 18°). When analyzing alignment as a continuous variable, the relationship between alignment and implant damage was assessed using Spearman’s rank correlation coefficient. When analyzing alignment as a categorical variable, differences in implant damage among knees aligned in varus, neutral, and valgus were determined using one-way ANOVA. Subsequently, multiple linear regression analyses were performed to examine the effect of alignment on implant damage, controlling for patient baseline demographics (age, BMI, and sex) and implant characteristics (length of implantation, reason for revision, PS versus CCK, and implant manufacturer). Differences in implant damage based on reason for revision also were examined using ANOVA. Multiple logistic regression analyses were performed to examine the effect of implant damage as affected by the binary outcome variable of reason for revision surgery (aseptic loosening and instability versus all other reasons for revision), while controlling for patient baseline demographics and implant characteristics.

Results

Medial Versus Lateral Polyethylene Damage Scores

All 156 retrieved tibial inserts showed damage in the form of pitting (mean score, 20 ± 9), scratching (mean score, 16 ± 9), and burnishing (mean score, 13 ± 7). The mean total damage score for all seven damage modes for the top surface of the inserts (zones 0 through 9; Fig. 2A) was 44 ± 17 and for the backside (zones 10 through 13; Fig. 2B) was 11 ± 6. Deformation (94% of the inserts), abrasion (46%), embedded debris (19%), and delamination (8%) also were present on some of the inserts, but the scores for these damage modes were quite low (mean scores of 4 ± 3, 1 ± 1, 0.4 ± 1, 0.4 ± 1, respectively).

No difference was found in the damage scores between the medial (20 ± 8) and lateral (19 ± 7) compartments of the inserts (p = 0.06). Similarly, no differences were found between the medial and lateral regions of the backside of the tibial inserts (zones 10 and 12 versus 11 and 13; mean score for both regions, 5 ± 3; p = 0.48). However, more damage was found on the medial region of the tibial post (zone 8; mean damage score, 2.8 ± 2.3; 95% CI, 2.44–3.18) compared with the lateral region of the tibial post (zone 9; mean damage score, 2.6 ± 2.0; 95% CI, 2.26–2.88), with a mean difference of 0.24 (95% CI, 0.06–0.42; p = 0.008).

Postoperative TKA Alignment and Implant Damage

Knees with more overall varus alignment after TKA had increased total damage on the retrieved tibial inserts (Spearman’s rank correlation coefficients of −0.3 [95% CI, −0.4 to −0.1; p = 0.001]), as did those with varus femoral component alignment (Spearman’s rank correlation coefficients of −0.2 [95% CI, −0.4 to −0.06; p = 0.009]) and varus tibial component alignment (Spearman’s rank correlation coefficients of −0.2 [95% CI, −0.06 to −0.4; p = 0.017]). When defining overall knee alignment as a categorical variable (varus, neutral, or valgus), we found an increase in implant damage in all regions of the insert when a TKA had overall varus alignment (Fig. 4). Interestingly, the least damage on the polyethylene inserts occurred in TKAs with overall valgus alignment (mean total damage score 63 ± 23 for varus versus 45 ± 15 for valgus; p < 0.01) (Table 1). We evaluated damage scores for knees revised for reasons other than loosening, knees with preoperative varus alignment, or knees with preoperative valgus alignment to consider the impact of preoperative alignment or risk of loosening on damage (Table 1). Significant differences were only found when postoperative knee alignment for all knees, including those revised for loosening, were evaluated for damage.

Fig. 4
figure 4

The bar chart shows the damage scores on different regions of the tibial insert. The TKAs were divided into those that were placed with varus (red), neutral (blue), or valgus (green) postoperative alignment. Implants placed with postoperative varus alignment had higher mean damage scores on the medial compartment (21.9 for varus, 19.3 for neutral, and 16.5 for valgus; p = 0.01), lateral compartment (21.2 for varus, 18.9 for neutral, and 15.4 for valgus; p < 0.01), post (6.9 for varus, 4.5 for neutral, and 4.9 for valgus; p < 0.01), and backside surfaces (12.7 for varus, 10.0 for neutral, and 8.4 for valgus; p = 0.01) of the polyethylene. Maximum damage score is 84 for the medial compartment, lateral compartment, and backside, while the maximum score is 42 for the post.

Table 1 Damage scores for overall postoperative knee alignment groups

After controlling for differences in patient baseline demographics and implant characteristics we found that varus TKA alignment was associated with increased total implant damage (1.2 point increase in total damage for every 1° increase in varus alignment; p < 0.01), lateral damage (0.3 point increase in total damage for every 1° increase in varus alignment; p = 0.02), post damage (0.2 point increase in total damage for every 1° increase in varus alignment; p < 0.01), and backside damage (0.4 point increase in total damage for every 1° increase in varus alignment; p < 0.01) (Table 2). Total damage on the insert was greater for CCK as opposed to PS designs (15.6 point increase in total damage for CCK implants compared with PS implants; p < 0.01), and with decreased BMI (0.4 point increase in total damage for every 1 unit increase in BMI; p = 0.03). Total damage scores on the articulating surface were higher in implants removed for loosening compared with those removed for stiffness (regression coefficient −33.6 stiffness versus loosening; p< 0.01) or infection (regression coefficient −21.8 infection versus loosening; p < 0.01). Comparisons to other reasons for revision (Table 2) and the mean total damage scores for different reasons for revision surgery are shown (Fig. 5).

Table 2 Multiple linear regression analyses comparing patient and implant characteristics with implant damage
Fig. 5
figure 5

A bar chart shows the mean total damage scores on the surfaces of tibial inserts revised for different reasons.

Variability in Alignment With Time

Preoperative knee deformity tended to be corrected to neutral postoperatively and then recur before revision surgery, with the average measurements (mean ± SD) for preoperative, postoperative, and prerevision overall alignment being valgus 2.5° ± 8.4°, valgus 3.6° ± 4.0°, and valgus 1.7° ± 6.4°, respectively (p = 0.04). Similarly, differences also were found among preoperative, postoperative, and prerevision femoral alignment (valgus 8.6° ± 2.8°, valgus 5.2° ± 3.0°, and valgus 4.4° ± 4.5°, respectively; p < 0.01) and tibial alignment (valgus 3.9° ± 3.7°, valgus 1.3° ± 2.4°, and valgus 2.4° ± 3.6°, respectively; p < 0.01). As expected, postoperative TKA alignment showed less variability (range, 9° varus to 16° valgus) and more correction toward neutral compared with preoperative TKA alignment (range, 20° varus to 26° valgus). However, by the time the patients had experienced complications with their TKA and were facing revision surgery, the prerevision TKA alignment (Fig. 6) had an average of 2° more varus compared with postoperative alignment measurements (range, 22° more varus to 11° more valgus; p < 0.01). Therefore, prerevision alignment drifted back toward varus, with an average overall alignment of 1.7° valgus compared with 3.7° valgus postoperatively (Fig. 6). Of the 74 patients who had neutral postoperative overall TKA alignment (3° to 7°) and radiographs at all three times available for analysis, 38 (51%) had greater varus alignment and 29 (39%) had more valgus alignment before revision surgery (Fig. 7A). The changes in alignment were influenced by the reason for revision surgery of the TKAs. TKAs tended to drift back into varus, with 32 of 45 (71%) TKAs revised for aseptic loosening, 18 of 45 (40%) revised for instability, four of 12 (33%) revised for stiffness (33%), and three of 22 (14%) revised for infection drifting back into varus (Fig. 7B). TKAs with varus postoperative alignment tended to maintain varus alignment, with 41 of the 49 (84%) postoperative varus TKAs having varus alignment before revision surgery. The same was found with valgus TKA alignment (> 7°), with 16 of 17 postoperative valgus TKAs (94%) having valgus alignment before revision surgery.

Fig. 6A–D
figure 6

The bar charts show distribution of (A) preoperative, (B) postoperative, and (C) prerevision overall TKA alignment in patients undergoing TKA. (D) The distribution for preoperative, postoperative, and prerevision alignment changed, with a shift back toward varus alignment from postoperative to prerevision times. PreP = Preoperative; PostP = Postoperative; PreR = Prerevision.

Fig. 7A–B
figure 7

(A) A scatterplot shows the preoperative, postoperative, and prerevision overall TKA alignment for each patient. On the y-axis, each patient is represented at one mark, such that each horizontal line on the graph should have two dots showing the alignment for the patient postoperatively and before revision surgery. The data were sorted according to postoperative TKA alignment, such that the red dots on the graph represent where the TKA was originally aligned. Most prerevision alignment measurements (yellow dots) are to the left of the postoperative measurements (red dots), showing that most TKAs fall into varus. (B) The scatterplot uses the same patients as those used for Illustration A but the data were sorted according to reason for revision surgery. Most TKAs revised for loosening fall into varus, as seen by the prerevision alignment measurements (yellow dots) being to the left of the postoperative measurements (red dots).

Discussion

Although primary TKA has shown its effectiveness by reducing knee pain and increasing knee function in activities of daily living, a general consensus regarding best alignment of the femoral and tibial components is lacking. In this study, which combined the strength of a registry and an implant-retrieval program, we sought to determine whether postoperative overall varus TKA alignment increases polyethylene damage in modern primary PS and CCK implant designs. We found TKAs aligned in postoperative overall varus are predisposed to experience increased polyethylene damage, but the damage was not higher in the medial compartment than the lateral compartment. In addition, we found that revised TKAs tended to drift back into greater varus as shown by comparing postoperative with prerevision radiographs of the same patients (Figs. 6D and 7A), despite preoperative limb alignment or surgical efforts to correct the knee to neutral mechanical alignment.

Limitations

Inherent limitations of any retrieval study are that such studies are retrospective for which certain variables such as reason for revision cannot be controlled and that the cohort includes implants that by definition have undergone revisions and thus may not reflect well-functioning devices. The patients who returned to our institution may not have been representative of the overall population who underwent TKAs done during this time, and patients who underwent revision surgery may be more or less likely to return to the original surgeon. However, we sought to address the relationship between revision and tibial insert damage, therefore focusing on implants retrieved at revision surgery was necessary. We did not take into account all the implant design differences among manufacturers that may have contributed to polyethylene damage (including post geometry, tray surface finishes, and polyethylene material properties. However, 92% of the implants (143 of 156) were manufactured by one of three companies and all were condylar TKA designs. In addition, we used implants retrieved from a large registry, therefore we had nearly complete clinical, demographic, and radiographic information for our patients. Like most studies of alignment after TKA, our study was focused on the static, coronal plane alignment and we were limited to measuring knee alignment on conventional two-dimensional radiographs, since full-length weightbearing radiographs or CT scans are not routinely ordered at our institution owing to the greater cost and increased radiation exposure. Therefore, we did not evaluate sagittal plane or rotational alignment, the combined effects of coronal, sagittal, and rotational alignment, or the dynamic effect of gait and other activities on alignment. Third, we do not know if the damage present on the surface of the implant was related to wear (removal of polyethylene from the surface in the form of debris) or how much of an increase in damage leads to an increase in revision rate. While we have associated damage with malalignment, it remains unclear whether the damage contributed to poor implant survivorship or whether the damage was a byproduct of the contact mechanics caused by malalignment. However, this is an inherent limitation to all retrieval studies of revised implants.

Influence of Alignment on Polyethylene Damage

We found that varus TKA alignment was associated with increased polyethylene damage, which is supported by prior retrieval studies [8, 22]. We theorized that this relationship may exist, as Ritter et al. [27] showed coronal TKA alignment as measured on a prerevision AP radiograph reliably predicted the risk of revision surgery, with varus overall alignment associated with collapse under the medial tibial plateau, and valgus overall alignment associated with instability. Given the increased revision rates that they reported of overall tibiofemoral alignment were outside the range of 2.5° or greater varus and 7.4° or less valgus, we anticipated that polyethylene damage would be highest with varus or valgus malalignment. Interestingly, we found varus alignment was associated with increased damage, but valgus malalignment (mean damage score, 45) showed the least damage compared with varus or neutral alignment groups (mean damage scores, 63 and 53, respectively; Table 1). Varus tibial component alignment may not be detrimental as long as it is paired with a femoral component aligned in valgus. This finding also can be explained by considering the reduction in the adduction moment during gait if the lower limb is in valgus [3]. While overall alignment was found to be associated with increased polyethylene damage, we did not find a substantial increase in polyethylene damage when the tibial component was in varus. Finding a correlation between overall TKA alignment, but not tibial alignment, and higher damage was reported in another retrieval study of 83 PS TKA inserts [22]. While a particular postoperative tibiofemoral knee alignment in the coronal plane may be found to be more advantageous in limiting tibial bearing polyethylene damage (or tibial component loosening), how such a postoperative tibiofemoral knee alignment may affect collateral soft tissue balancing, bone resections, kinematic function of the knee, patellofemoral tracking, and comfort of the patient also must be considered by the operating surgeon. Furthermore, properly balancing the soft tissues about the knee arthroplasty remains important beyond simply reconstructing the knee to a preferred tibiofemoral alignment.

We did not find regional differences in polyethylene damage on the surfaces (for example, increased medial damage in TKAs with excessive varus alignment). This is because TKAs do not solely experience varus and valgus moments during activities of daily living [4, 16]. Another explanation could be the effect of implant design. For example, Ritter et al. [27] reported on the effect of knee alignment on implant survival by evaluating a large series of TKAs of the AGC® knee implant design (Biomet, Warsaw, IN, USA). That specific implant design has considerably flatter mediolateral geometries on the condylar surfaces of the femoral components and the articular surfaces on the plateaus of the tibial component compared with the curve-on-curve toroidal articulating surfaces used in the PS and CCK knee designs in our study. Under varus and valgus moments, the flatter design of the AGC® knee implant could result in edge loading of the femoral component on a relatively small contact area on the outer edges of either the medial or lateral tibial compartments [33]. Although long-term clinical studies have not found edge loading to be a problem with the AGC® knee implant [14, 26], the geometries of the articular surfaces may influence regionally (medial versus lateral) damage on the plateau and how well a specific implant design may perform when malaligned. A more recent study showed greater damage in the lateral compartment of varus-aligned TKAs and attributed this result to lateral condylar lift-off inducing impact and shear loading in these knees [31]. Our results do not support this finding.

Interestingly, we found increased polyethylene damage on the medial and lateral surfaces of the tibial posts in primary TKAs with CCK implants compared with primary TKAs with PS implants. That the tibial posts, regardless of design, showed damage is evidence that the post was contacting the intercondylar box of the femoral component during in vivo use. This added constraint provides another load transfer pathway in addition to the articular surfaces, and also could partially explain the lack of findings of regional (medial versus lateral) differences. This additional constraint was not present in the AGC® cruciate retaining design used by Ritter et al. [27]. The difference between the CCK and PS designs reflects differences in the amount of constraint, with the CCK providing additional constraint through a thickened tibial post with tighter spacing in the intercondylar box [11]. The higher damage on the CCK implants shows that this constraint mechanism was functional, suggesting that the use of a CCK design in some primary TKAs is reasonable [2]; however, a disadvantage of the added constraint may be increased interfacial stresses around the femoral component leading to early loosening [21].

Other factors influenced the damage observed on our retrieved tibial inserts. For example, damage scores were associated with reasons for revision, with TKAs revised for loosening or instability having tibial inserts with more damage than those revised for stiffness or infection (Fig. 5). Patients with stiffness or infection are likely to be less demanding of their implant, and the reduced mechanical burden translates to less surface damage to the polyethylene. In addition, we found increased implant damage associated with increased BMI. Increasing BMI was shown to increase the risk of TKA failure [27], suggesting that poor implant alignment combined with a high BMI represents a greater risk to implant survival than either risk factor alone.

Changes in Alignment With Time

We found a tendency for knees that underwent revision to have drifted back into varus in the period between implantation and revision. While surgeons attempt to correct TKA alignment, and in most cases are successful (Figs. 6B and 7A), patients nonetheless continue to place high demands on their implants. For example, high adduction moments occur even during normal gait as the ground reaction force passes medial to the center of the knee [16]. Previous nonsurgical and surgical interventions have targeted the reduction of the adduction moment to slow progression of osteoarthritis [12, 16, 25]. However, despite restoration of static knee alignment, the adduction moment remains high after surgery [20], just as in a healthy native human knee. While we found a predisposition toward recurrence of the preoperative varus deformity, we did not find increased medial polyethylene damage. Therefore, the biomechanical consequences of the change in alignment from postoperatively to prerevision may be better reflected by collapse of underlying cancellous bone, the weak link in the load transfer pathway from the articular surface to the tibial cortex, than by increased polyethylene damage. This can be seen in the greater tendency for a drift back to varus in TKAs that experienced loosening compared with those revised for other reasons (Fig. 7B).

Conclusions

We found that damage was higher in TKAs with postoperative varus alignment. While understanding there may be other operative considerations, such as soft tissue balancing, our findings support restoring limb alignment as a way of reducing implant polyethylene damage and the subsequent risk of osteolysis caused by biological reaction to the released particulate wear debris. Like most studies of alignment after TKA, our study was focused on the static, coronal plane alignment and did not evaluate sagittal plane or rotational alignment, the combined effects of coronal, sagittal, and rotational alignment, or the dynamic effect of gait and other activities on alignment. We have a prospective study underway to examine the relationship between three-dimensional static alignment and dynamic function in patients who have had TKA.