Do Retrieval Analysis and Blood Metal Measurements Contribute to Our Understanding of Adverse Local Tissue Reactions?

  • Symposium: ABJS Carl T. Brighton Workshop on Implant Wear and Tribocorrosion of Total Joint Replacements
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Metal-on-metal (MoM) total hip arthroplasties (THAs) and the head-neck and neck-body junctions in modular THA are associated with a variety of local and systemic reactions to their related wear and corrosion products. Although laboratory testing is available, the relationship between laboratory values—including serum metal ion levels—and adverse local tissue reactions (ALTRs) remains controversial and incompletely characterized.

Questions/purposes

(1) What is the range of serum metal levels associated with ALTR in patients who have MoM THAs or corrosion at the head-neck and neck-body junctions in metal-on-polyethylene (or ceramic-on-polyethylene) THAs? (2) How much wear occurs in patients with MoM total hips? (3) Is there evidence of a dose-response relationship between wear and ALTR?

Methods

PubMed and Embase databases were reviewed for English-language studies assessing serum metal levels in the presence of ALTR and papers describing the results of wear measurements from revised MoM implants and ALTR histopathology were systematically reviewed. Reported linear wear data were separated into groups with ALTR and without ALTR as listed in individual papers and graphed to determine whether a dose-response relationship was present between wear and ALTR. Overall, 15 studies including 338 hips with ALTR with corresponding serum metal levels were identified and analyzed. Twelve studies reported the wear depth or volume of MoM components from patients with a variety of local reactions. Two studies investigated corrosion at the head-neck and neck-body junctions in metal-on-polyethylene THA. There was a high level of variability and study heterogeneity, and so data pooling (meta-analysis) could not be performed.

Results

Average reported metal concentrations were elevated above established normal values in patients with ALTR (cobalt concentrations ranged from 5 to 40 ppb, and chromium levels ranged from 5 to 54 ppb). Whereas several studies demonstrated that patients with ALTR had higher average linear wear of the bearing surfaces, this finding was not made in all studies that we identified in this systematic review. Because of this high degree of variability, no clear dose-response relationship between wear and ALTR could be established.

Conclusions

Serum metal level analysis and implant retrieval analysis both contribute to the understanding of ALTR. Serum metal levels generally are elevated in the presence of ALTR but should not be used in isolation for clinical decision-making. Many but not all patients with ALTR, including those with pseudotumors, demonstrate high wear, but more data and more systematic descriptions of the histopathology are needed to define the amount of wear that induces adverse reactions.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. AAOS. Current Concerns with Meta-on-metal Hip Arthroplasty. Information Statement. Available at: www.aaos.org/about/papers/advistmt/1035.asp. Accessed March 18, 2014.

  2. Almousa SA, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. The natural history of inflammatory pseudotumors in asymptomatic patients after metal-on-metal hip arthroplasty. Clin Orthop Relat Res. 2013;471:3814–3821.

    Article  PubMed  Google Scholar 

  3. Bishop N, Witt F, Pourzal R, Fischer A, Rutschi M, Michel M, Morlock M. Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings. J Orthop Res. 2013;31:1116–1122.

    Article  PubMed  Google Scholar 

  4. Bisschop R, Boomsma MF, Van Raay JJ, Tiebosch AT, Maas M, Gerritsma CL. High prevalence of pseudotumors in patients with a Birmingham Hip Resurfacing prosthesis: a prospective cohort study of one hundred and twenty-nine patients. J Bone Joint Surg Am. 2013;95:1554–1560.

    Article  CAS  PubMed  Google Scholar 

  5. Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93:608–615.

    Article  CAS  PubMed  Google Scholar 

  6. Campbell P, Ebramzadeh E, Nelson S, Takamura K, De Smet K, Amstutz HC. Histological features of pseudotumor-like tissues from metal-on-metal hips. Clin Orthop Relat Res. 2010;468:2321–2327.

    Article  PubMed Central  PubMed  Google Scholar 

  7. Chang EY, McAnally JL, Van Horne JR, Van Horne JG, Wolfson T, Gamst A, Chung CB. Relationship of plasma metal ions and clinical and imaging findings in patients with ASR XL metal-on-metal total hip replacements. J Bone Joint Surg Am. 2013;95:2015–2020.

    Article  PubMed  Google Scholar 

  8. Cooper HJ, Della Valle CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am. 2012;94:1655–1661.

    PubMed  Google Scholar 

  9. Cooper HJ, Urban RM, Wixson RL, Meneghini RM, Jacobs JJ. Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Joint Surg Am. 2013;95:865–872.

    Article  PubMed Central  PubMed  Google Scholar 

  10. De Haan R, Campbell PA, Su EP, De Smet KA. Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg Br. 2008;90:1158–1163.

    Article  PubMed  Google Scholar 

  11. De Smet K, De Haan R, Calistri A, Campbell PA, Ebramzadeh E, Pattyn C, Gill HS. Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Joint Surg Am. 2008;90(Suppl 4):202–208.

    Article  PubMed  Google Scholar 

  12. Ebramzadeh E, Campbell PA, Takamura KM, Lu Z, Sangiorgio SN, Kalma JJ, De Smet KA, Amstutz HC. Failure modes of 433 metal-on-metal hip implants: how, why, and wear. Orthop Clin North Am. 2011;42:241–250, ix.

  13. Ebreo D, Khan A, El-Meligy M, Armstrong C, Peter V. Metal ion levels decrease after revision for metallosis arising from large-diameter metal-on-metal hip arthroplasty. Acta Orthop Belg. 2011;77:777–781.

    PubMed  Google Scholar 

  14. Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94:895–900.

    Article  CAS  PubMed  Google Scholar 

  15. Glyn-Jones S, Roques A, Taylor A, Kwon YM, McLardy-Smith P, Gill HS, Walter W, Tuke M, Murray D. The in vivo linear and volumetric wear of hip resurfacing implants revised for pseudotumor. J Bone Joint Surg Am. 2011;93:2180–2188.

    Article  PubMed  Google Scholar 

  16. Grammatopoulos G, Pandit H, Kamali A, Maggiani F, Glyn-Jones S, Gill HS, Murray DW, Athanasou N. The correlation of wear with histological features after failed hip resurfacing arthroplasty. J Bone Joint Surg Am. 2013;95:e81.

    Article  PubMed  Google Scholar 

  17. Hallab NJ, Anderson S, Caicedo M, Skipor A, Campbell P, Jacobs JJ. Immune responses correlate with serum-metal in metal-on-metal hip arthroplasty. J Arthroplasty. 2004;19:88–93.

    Article  PubMed  Google Scholar 

  18. Hart AJ, Muirhead-Allwood S, Porter M, Matthies A, Ilo K, Maggiore P, Underwood R, Cann P, Cobb J, Skinner JA. Which factors determine the wear rate of large-diameter metal-on-metal hip replacements? Multivariate analysis of two hundred and seventy-six components. J Bone Joint Surg Am. 2013;95:678–685.

    Article  CAS  PubMed  Google Scholar 

  19. Hart AJ, Sabah SA, Bandi AS, Maggiore P, Tarassoli P, Sampson B, Skinner JA. Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Joint Surg Br. 2011;93:1308–1313.

    Article  CAS  PubMed  Google Scholar 

  20. Hartmann A, Hannemann F, Lutzner J, Seidler A, Drexler H, Gunther KP, Schmitt J. Metal ion concentrations in body fluids after implantation of hip replacements with metal-on-metal bearing—systematic review of clinical and epidemiological studies. PLoS One. 2013;8:e70359.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  21. Jacobs JJ, Gilbert JL, Urban RM. Corrosion of metal orthopaedic implants. J Bone Joint Surg Am. 1998;80:268–282.

    CAS  PubMed  Google Scholar 

  22. Jacobs JJ, Skipor AK, Patterson LM, Hallab NJ, Paprosky WG, Black J, Galante JO. Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study. J Bone Joint Surg Am. 1998;80:1447–1458.

    CAS  PubMed  Google Scholar 

  23. Kop AM, Keogh C, Swarts E. Proximal component modularity in THA—at what cost? An implant retrieval study. Clin Orthop Relat Res. 2012;470:1885–1894.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  24. Kwon YM, Glyn-Jones S, Simpson DJ, Kamali A, McLardy-Smith P, Gill HS, Murray DW. Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumours. J Bone Joint Surg Br. 2010;92:356–361.

    Article  PubMed  Google Scholar 

  25. Kwon YM, Ostlere SJ, McLardy-Smith P, Athanasou NA, Gill HS, Murray DW. ‘Asymptomatic’ pseudotumors after metal-on-metal hip resurfacing arthroplasty: prevalence and metal ion study. J Arthroplasty. 2011;26:511–518.

    Article  PubMed  Google Scholar 

  26. Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.

    Article  CAS  PubMed  Google Scholar 

  27. Levine BR, Hsu AR, Skipor AK, Hallab NJ, Paprosky WG, Galante JO, Jacobs JJ. Ten-year outcome of serum metal ion levels after primary total hip arthroplasty: a concise follow-up of a previous report*. J Bone Joint Surg Am. 2013;95:512–518.

    Article  PubMed Central  PubMed  Google Scholar 

  28. Lohmann CH, Meyer H, Nuechtern JV, Singh G, Junk-Jantsch S, Schmotzer H, Morlock MM, Pfluger G. Periprosthetic tissue metal content but not serum metal content predicts the type of tissue response in failed small-diameter metal-on-metal total hip arthroplasties. J Bone Joint Surg Am. 2013;95:1561–1568.

    Article  CAS  PubMed  Google Scholar 

  29. Matthies AK, Racasan R, Bills P, Blunt L, Cro S, Panagiotidou A, Blunn G, Skinner J, Hart AJ. Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements. J Orthop Res. 2013;31:1677–1685.

    PubMed  Google Scholar 

  30. Matthies AK, Skinner JA, Osmani H, Henckel J, Hart AJ. Pseudotumors are common in well-positioned low-wearing metal-on-metal hips. Clin Orthop Relat Res. 2012;470:1895–1906.

    Article  PubMed Central  PubMed  Google Scholar 

  31. Mayo Clinic MCL. Metal-on-metal Wear of Orthopedic Implants: Role of Serum Cobalt and Chromium Testing. October 2011. Available at: www.mayomedicallaboratories.com/articles/features/metal-ortho/index.html. Accessed July 8, 2014.

  32. MHRA. Medical Device Alert. Available at: www.mhra.gov.uk/home/groups/dts-bs/documents/medicaldevicealert/con155767.pdf. Accessed March 18, 2014.

  33. Milosev I, Trebse R, Kovac S, Cor A, Pisot V. Survivorship and retrieval analysis of Sikomet metal-on-metal total hip replacements at a mean of seven years. J Bone Joint Surg Am. 2006;88:1173–1182.

    Article  PubMed  Google Scholar 

  34. Mokka J, Junnila M, Seppanen M, Virolainen P, Polonen T, Vahlberg T, Mattila K, Tuominen EK, Rantakokko J, Aarimaa V, Kukkonen J, Makela KT. Adverse reaction to metal debris after ReCap-M2A-Magnum large-diameter-head metal-on-metal total hip arthroplasty. Acta Orthop. 2013;84:549–554.

    Article  PubMed Central  PubMed  Google Scholar 

  35. Munro JT, Masri BA, Duncan CP, Garbuz DS. High complication rate after revision of large-head metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2014;472:523–528.

    Article  PubMed  Google Scholar 

  36. Nawabi DH, Nassif NA, Do HT, Stoner K, Elpers M, Su EP, Wright T, Potter HG, Padgett DE. What causes unexplained pain in patients with metal-on metal hip devices? A retrieval, histologic, and imaging analysis. Clin Orthop Relat Res. 2014;472:543–554.

    Article  PubMed  Google Scholar 

  37. Nich C, Hamadouche M. Cup loosening after cemented Metasul® total hip replacement: a retrieval analysis. Int Orthop. 2011;35:965–970.

    Article  PubMed Central  PubMed  Google Scholar 

  38. Pelt CE, Erickson J, Clarke I, Donaldson T, Layfield L, Peters CL. Histologic, serologic, and tribologic findings in failed metal-on-metal total hip arthroplasty: AAOS exhibit selection. J Bone Joint Surg Am. 2013;95:e163.

    Article  PubMed  Google Scholar 

  39. Takamura KM, Amstutz HC, Lu Z, Campbell PA, Ebramzadeh E. Wear analysis of 39 conserve plus metal-on-metal hip resurfacing retrievals. J Arthroplasty. 2014;29:410–415.

    Article  PubMed  Google Scholar 

  40. Van Der Straeten C, Grammatopoulos G, Gill HS, Calistri A, Campbell P, De Smet KA. The 2012 Otto Aufranc Award: The interpretation of metal ion levels in unilateral and bilateral hip resurfacing. Clin Orthop Relat Res. 2013;471:377–385.

    Article  Google Scholar 

  41. Wiley KF, Ding K, Stoner JA, Teague DC, Yousuf KM. Incidence of pseudotumor and acute lymphocytic vasculitis associated lesion (ALVAL) reactions in metal-on-metal hip articulations: a meta-analysis. J Arthroplasty. 2013;28:1238–1245.

    Article  PubMed  Google Scholar 

  42. Willert HG, Buchhorn GH, Gobel D, Koster G, Schaffner S, Schenk R, Semlitsch M. Wear behavior and histopathology of classic cemented metal on metal hip endoprostheses. Clin Orthop Relat Res. 1996;329(Suppl):S160–186.

    Article  PubMed  Google Scholar 

  43. Witzleb WC, Hanisch U, Ziegler J, Guenther KP. In vivo wear rate of the Birmingham Hip Resurfacing arthroplasty. A review of 10 retrieved components. J Arthroplasty. 2009;24:951–956.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

We thank Dr Edward Ebramzadeh for his statistical advice related to the analysis of these papers.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Patricia A. Campbell PhD.

Additional information

One of the authors (JJJ) certifies that he has or may receive payments or benefits, during the study period, an amount of USD 10,000 to USD 100,000 from Medtronics Sofamor Danek (Memphis, TN, USA), an amount of USD 10,000 to USD 100,000 from Nuvasive (San Diego, CA, USA), and an amount of USD 10,000 to USD 100,000 from Zimmer (Warsaw, IN, USA). One of the authors (PAC) certifies that she has or may receive payments or benefits, during the study period, an amount of USD 10,000 to USD 100,000 from DePuy (Warsaw, IN, USA). The institution of one of the authors (PAC) has received, during the study period, funding from DePuy.

All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research ® editors and board members are on file with the publication and can be viewed on request.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Campbell, P.A., Kung, M.S., Hsu, A.R. et al. Do Retrieval Analysis and Blood Metal Measurements Contribute to Our Understanding of Adverse Local Tissue Reactions?. Clin Orthop Relat Res 472, 3718–3727 (2014). https://doi.org/10.1007/s11999-014-3893-2

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11999-014-3893-2

Keywords

Navigation