Abstract
There have been relatively few studies on mechanical properties of nanomaterials under high strain rates, mainly due to the lack of capable nanomechanical testing devices. Here we present a new on-chip microelectromechanical system (MEMS) for high strain-rate nanomechanical testing. The MEMS device consists of an electrostatic comb drive actuator, two capacitive displacement sensors and a load cell. The dynamic responses of the device in air and in vacuum are systematically modeled under both alternating and ramp forces. Two methods, capacitive readout and high-speed imaging, are used to measure the dynamic displacements, which agree well with the modeling results. While we demonstrate the maximum constant strain rate over 200 s−1 under ramp force, it is interesting to find that the capacitive readout used in this work can only measure strain rate up to 22 s−1 due to its limit in bandwidth. To demonstrate the utility of this new device, gold nanowires are tested at strain rates of 10−5 and 10 s−1 inside a scanning electron microscope. Increasing strain rate is found to yield higher yield strength and larger ductility.
Similar content being viewed by others
References
Yao S, Zhu Y (2015) Nanomaterial-enabled stretchable conductors: strategies, materials and devices. Adv Mater 27(9):1480–1511
Yao S, Swetha P, Zhu Y (2018) Nanomaterial-enabled wearable sensors for healthcare. Adv Healthcare Mater 7(1):1700889
Gong S, Cheng W (2017) One-dimensional nanomaterials for soft electronics. Adv Electron Mater 3(3):1600314
Zhu Y, Ke C, Espinosa H (2007) Experimental techniques for the mechanical characterization of one-dimensional nanostructures. Exp Mech 47(1):7
Park HS, Cai W, Espinosa HD, Huang H (2009) Mechanics of crystalline nanowires. MRS Bull 34(3):178–183
Zhu Y (2017) Mechanics of crystalline nanowires: an experimental perspective. Appl Mech Rev 69(1):010802
Zhu T, Li J, Ogata S, Yip S (2009) Mechanics of ultra-strength materials. MRS Bull 34(3):167–172
Cheng G, Chang T-H, Qin Q, Huang H, Zhu Y (2014) Mechanical properties of silicon carbide nanowires: effect of size-dependent defect density. Nano Lett 14(2):754–758
Richter G, Hillerich K, Gianola DS, Monig R, Kraft O, Volkert CA (2009) Ultrahigh strength single crystalline nanowhiskers grown by physical vapor deposition. Nano Lett 9(8):3048–3052
Xu F, Qin Q, Mishra A, Gu Y, Zhu Y (2010) Mechanical properties of ZnO nanowires under different loading modes. Nano Res 3(4):271–280
Agrawal R, Peng B, Gdoutos EE, Espinosa HD (2008) Elasticity size effects in ZnO nanowires− a combined experimental-computational approach. Nano Lett 8(11):3668–3674
Zhu Y, Qin Q, Xu F, Fan F, Ding Y, Zhang T, Wiley BJ, Wang ZL (2012) Size effects on elasticity, yielding, and fracture of silver nanowires: in situ experiments. Phys Rev B 85(4):045443
Zhu T, Li J, Samanta A, Leach A, Gall K (2008) Temperature and strain-rate dependence of surface dislocation nucleation. Phys Rev Lett 100:025502
Chen LY, He M-r, Shin J, Richter G, Gianola DS (2015) Measuring surface dislocation nucleation in defect-scarce nanostructures. Nat Mater 14(7):707–713
Qin Q, Yin S, Cheng G, Li X, Chang T-H, Richter G, Zhu Y, Gao H (2015) Recoverable plasticity in penta-twinned metallic nanowires governed by dislocation nucleation and retraction. Nat Commun 6:5983
Zhong L, Sansoz F, He Y, Wang C, Zhang Z, Mao SX (2017) Slip-activated surface creep with room-temperature super-elongation in metallic nanocrystals. Nat Mater 16(4):439
Bernal RA, Aghaei A, Lee S, Ryu S, Sohn K, Huang J, Cai W, Espinosa H (2014) Intrinsic Bauschinger effect and recoverable plasticity in pentatwinned silver nanowires tested in tension. Nano Lett 15(1):139–146
Chang T-H, Zhu Y (2013) A microelectromechanical system for thermomechanical testing of nanostructures. Appl Phys Lett 103(26):263114
Ramachandramoorthy R, Gao W, Bernal R, Espinosa H (2015) High strain rate tensile testing of silver nanowires: rate-dependent brittle-to-ductile transition. Nano Lett 16(1):255–263
Cheng G, Miao C, Qin Q, Li J, Xu F, Haftbaradaran H, Dickey EC, Gao H, Zhu Y (2015) Large anelasticity and associated energy dissipation in single-crystalline nanowires. Nat Nanotechnol 10(8):687–691
Peng C, Zhong Y, Lu Y, Narayanan S, Zhu T, Lou J (2013) Strain rate dependent mechanical properties in single crystal nickel nanowires. Appl Phys Lett 102(8):083102
Zhu Y, Chang T-H (2015) A review of microelectromechanical systems for nanoscale mechanical characterization. J Micromech Microeng 25(9):093001
Haque M, Saif M (2002) In-situ tensile testing of nano-scale specimens in SEM and TEM. Exp Mech 42(1):123–128
Zhu Y, Espinosa HD (2005) An electromechanical material testing system for in situ electron microscopy and applications. Proc Natl Acad Sci 102(41):14503–14508
Kang W, Saif MTA (2011) A novel SiC MEMS apparatus for in situ uniaxial testing of micro/nanomaterials at high temperature. J Micromech Microeng 21(10):105017
Naraghi M, Kolluru PV, Chasiotis I (2014) Time and strain rate dependent mechanical behavior of individual polymeric nanofibers. J Mech Phys Solids 62:257–275
Naraghi M, Chasiotis I, Kahn H, Wen Y, Dzenis Y (2007) Novel method for mechanical characterization of polymeric nanofibers. Rev Sci Instrum 78(8):085108
Tsuchiya T, Ura Y, Sugano K, Tabata O (2012) Electrostatic tensile testing device with nanonewton and nanometer resolution and its application to C60 nanowire testing. J Microelectromech Syst 21(3):523–529
Zhu Y, Corigliano A, Espinosa HD (2006) A thermal actuator for nanoscale in situ microscopy testing: design and characterization. J Micromech Microeng 16(2):242
Guan C, Zhu Y (2010) An electrothermal microactuator with Z-shaped beams. J Micromech Microeng 20(8):085014
Naraghi M, Chasiotis I (2009) Optimization of comb-driven devices for mechanical testing of polymeric nanofibers subjected to large deformations. J Microelectromech Syst 18(5):1032–1046
Zhu Y, Moldovan N, Espinosa HD (2005) A microelectromechanical load sensor for in situ electron and x-ray microscopy tensile testing of nanostructures. Appl Phys Lett 86(1):013506
Pantano MF, Bernal RA, Pagnotta L, Espinosa HD (2015) Multiphysics design and implementation of a microsystem for displacement-controlled tensile testing of nanomaterials. Meccanica 50(2):549–560
Erismis MA (2013) Design and modeling of a new robust multi-mass coupled-resonator family with dynamic motion amplification. Microsyst Technol 19(8):1105–1110
Yamahata C, Sarajlic E, Krijnen GJ, Gijs MA (2010) Subnanometer translation of microelectromechanical systems measured by discrete Fourier analysis of CCD images. J Microelectromech Syst 19(5):1273–1275
King H, Warnat S, Hubbard T (2015) Effect of image degradation on nm-scale MEMS FFT optical displacement measurements. In: Electrical and Computer Engineering (CCECE), 2015 IEEE 28th Canadian Conference on. IEEE, pp 1387–1392
Ellerington N, Bschaden B, Hubbard T, Kujath M (2012) Fourier analysis of blurred images for the measurement of the in-plane dynamics of MEMS. J Micromech Microeng 22(3):035019
Boyce BL, Crenshaw TB (2005) Servohydraulic methods for mechanical testing in the sub-Hopkinson rate regime up to strain rates of 500 1/s. Sandia Natl Lab Rep:1–16
Zhu D, Rajan S, Mobasher B, Peled A, Mignolet M (2011) Modal analysis of a servo-hydraulic high speed machine and its application to dynamic tensile testing at an intermediate strain rate. Exp Mech 51(8):1347–1363
de Andrade SF, Zhu D, Mobasher B, Soranakom C, Toledo Filho RD (2010) High speed tensile behavior of sisal fiber cement composites. Mater Sci Eng A 527(3):544–552
Lott CD, McLain TW, Harb JN, Howell LL (2002) Modeling the thermal behavior of a surface-micromachined linear-displacement thermomechanical microactuator. Sensors Actuators A Phys 101(1–2):239–250
Shroff SS, Boer MP (2015) Constant velocity high force microactuator for stick-slip testing of micromachined interfaces. J Microelectromech Syst 24(6):1868–1877
Zhang X, Tang WC (1994) Viscous air damping in laterally driven microresonators. In: Micro Electro Mechanical Systems, 1994, MEMS'94, Proceedings, IEEE Workshop on. IEEE, pp 199–204
Bao M (2005) Analysis and design principles of MEMS devices. Elsevier, Amsterdam
Bao M, Yang H (2007) Squeeze film air damping in MEMS. Sensors Actuators A Phys 136(1):3–27
Veijola T, Kuisma H, Lahdenperä J, Ryhänen T (1995) Equivalent-circuit model of the squeezed gas film in a silicon accelerometer. Sensors Actuators A Phys 48(3):239–248
Sedlmayr A, Bitzek E, Gianola DS, Richter G, Mönig R, Kraft O (2012) Existence of two twinning-mediated plastic deformation modes in au nanowhiskers. Acta Mater 60(9):3985–3993
Park HS, Zimmerman JA (2005) Modeling inelasticity and failure in gold nanowires. Phys Rev B 72(5):054106
Cheng G, Yin S, Chang T, Richter G, Gao H, Zhu Y (2017) Anomalous tensile detwinning in twinned nanowires. Phys Rev Lett 119(25):256101
Tao W, Cao P, Park HS (2018) Atomistic simulation of the rate-dependent ductile-to-brittle failure transition in bicrystalline metal nanowires. Nano Lett 18(2):1296–1304
Acknowledgements
The authors gratefully acknowledge financial support from the National Science Foundation (NSF) through Award Nos. DMR-1410475 and CMMI-1929646. The authors would like to thank Dr. T. Fang for providing access to his high-speed camera and Dr. G. Richter for providing the Au nanowire samples.
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Appendix
Appendix
Dynamic Response of a Single-Mass-Spring System Under Ramp Force
For a single-mass-spring system as schematically shown in Fig. 14, if the input force is a ramp force F(t) = pt, the equation of motion can be written as
Solution to this equation is \( x(t)=\frac{pt}{k}-\frac{p}{k}\sqrt{\frac{m}{k}}\mathit{\sin}\left(\sqrt{\frac{k}{m}t}\right) \). This solution contains two terms; the first is a ramp term \( \frac{pt}{k}=\frac{F(t)}{k} \), and the second is a sinusoidal term \( -\frac{p}{k}\sqrt{\frac{m}{k}}\mathit{\sin}\left(\sqrt{\frac{k}{m}t}\right) \). The sinusoidal term has a frequency \( \sqrt{\frac{k}{m}} \), which is the natural frequency of the system. Its amplitude is proportional to loading rate p and can be calculated if system parameters m and k are known.
Dynamic Response of a Single-Mass-Spring-Damper System Under Ramp Force
For a single-mass-spring-damper system as schematically shown in Fig. 15, if the input force is a ramp force F(t) = pt, the equation of motion can be written as
If the system is an underdamped system, solution to this equation is \( x(t)=\frac{pt}{k}-\frac{cp}{k^2}-\frac{2 pm}{k\sqrt{4 mk-{c}^2}}\sin \left(\frac{\sqrt{4 mk-{c}^2}}{2m}t+\theta \right) \), where \( \theta =\arctan \left(\frac{c\sqrt{4 mk-{c}^2}}{c^2-2 km}\right) \). The solution contains three terms; the first is a ramp term \( \frac{pt}{k}=\frac{F(t)}{k} \), the second is a constant \( -\frac{cp}{k^2} \), and the third is a sinusoidal term \( -\frac{2 pm}{k\sqrt{4 mk-{c}^2}}\sin \left(\frac{\sqrt{4 mk-{c}^2}}{2m}t+\theta \right) \). The sinusoidal term has a frequency \( \frac{\sqrt{4 mk-{c}^2}}{2m} \), which is the damped natural frequency of the system.
Rights and permissions
About this article
Cite this article
Li, C., Zhang, D., Cheng, G. et al. Microelectromechanical Systems for Nanomechanical Testing: Electrostatic Actuation and Capacitive Sensing for High-Strain-Rate Testing. Exp Mech 60, 329–343 (2020). https://doi.org/10.1007/s11340-019-00565-5
Received:
Accepted:
Published:
Issue Date: