Abstract
Objective
Continuous assessment of central venous oxygen saturation (ScevoxO2) with the CeVOX device (Pulsion Medical Systems, Munich, Germany) was evaluated against central venous oxygen saturation (ScvO2) determined by co-oximetry.
Methods
In 20 cardiac surgical patients, a CeVOX fiberoptic probe was introduced into a standard central venous catheter placed in the right internal jugular vein and advanced 2–3 cm beyond the catheter tip. After in vivo calibration of the probe, ScevoxO2, ScvO2, mixed venous oxygen saturation (SmvO2) haemoglobin (Hb), body temperature, heart rate, central venous and mean arterial pressure, and cardiac index were assessed simultaneously at 30 min intervals during surgery and at 60 min intervals during recovery in the intensive care unit. Agreement between ScevoxO2, and ScvO2 was determined by Bland-Altman analysis. Simple regression analysis was used to assess the correlation of ScevoxO2, and ScvO2 to Hb, body temperature and haemodynamic parameters.
Results
Values of ScevoxO2 and ScvO2 (84 data pairs during surgery and 106 in the intensive care unit) ranged between 45–89% and 43–90%, respectively. Mean bias and limits of agreement of ScevoxO2 and ScvO2 were –0.9 (−7.9/+6.1)% during surgery and −1.2 (−10.5/+8.1)% in the intensive care unit. In 37.9% of all measured data pairs, the difference between ScevoxO2 and ScvO2 was beyond clinically acceptable limits (≥1 s.d.). Mean bias was significantly influenced by cardiac index. Sensitivity and specificity of ScevoxO2 to detect substantial (≥1 s.d.) changes in ScvO2 were 89 and 82%, respectively.
Conclusions
In adult patients during and after cardiac surgery, the current version of the CeVOX device might not be the tool to replace ScvO2 determined by co-oxymetry, although sensitivity and specificity of ScevoxO2 to predict substantial changes in ScvO2 were acceptable.
Similar content being viewed by others
References
Wilkinson AR, Phibbs RH, Gregory GA. Continuous measurement of oxygen saturation in sick newborn infants. J Pediatr 1978; 93: 1016–1019
Wilson RF. Get the venous blood too. Am Surg 1978; 44: 396–400
Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M. Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med 2001; 345: 1368–1377
Rivers EP, Ander DS, Powell D. Central venous oxygen saturation monitoring in the critically ill patient. Curr Opin Crit Care 2001; 7: 204–211
Ladakis C, Myrianthefs P, Karabinis A, Karatzas G, Dosios T, Fildissis G, Gogas J, Baltopoulos G. Central venous and mixed venous oxygen saturation in critically ill patients. Respiration 2001; 68: 279–285
Rady MY, Rivers EP, Martin GB, Smithline H, Appelton T, Nowak RM. Continuous central venous oximetry and shock index in the emergency department: use in the evaluation of clinical shock. Am J Emerg Med 1992; 10: 538–541
Collaborative Study Group on Perioperative ScvO2 Monitoring. Multicentre study on peri- and postoperative central venous oxygen saturation in high-risk surgical patients. Crit Care 2006; 10: R158
Reinhart K, Kuhn HJ, Hartog C, Bredle DL. Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill. Intensive Care Med 2004; 30: 1572–1578
Huber D, Osthaus WA, Optenhofel J, Breymann T, Marx G, Piepenbrock S, Sümpelmann R. Continuous monitoring of central venous oxygen saturation in neonates and small infants: in vitro evaluation of two different oximetry catheters. Paediatr Anaesth 2006; 16: 1257–1261
Muller M, Lohr T, Scholz S, Thul J, Akinturk H, Hempelmann G. Continuous SvO2 measurement in infants undergoing congenital heart surgery—first clinical experiences with a new fiberoptic probe. Paediatr Anaesth 2007; 17: 51–55
Molnar Z, Umgelter A, Toth I, Livingstone D, Weyland A, Sakka SG, Meier-Hellmann A. Continuous monitoring of ScvO(2) by a new fibre-optic technology compared with blood gas oximetry in critically ill patients: a multicentre study. Intensive Care Med 2007; [Epub ahead of print]
Levett JM, Replogle RL. Thermodilution cardiac output: a critical analysis and review of the literature. J Surg Res 1979; 27: 392–404
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; i: 307–310
Hecker BR, Brown DL, Wilson D. A comparison of two pulmonary artery mixed venous oxygen saturation catheters during the changing conditions of cardiac surgery. J Cardiothorac Anesth 1989; 3: 269–275
Scuderi PE, MacGregor DA, Bowton DL, James RL. A laboratory comparison of three pulmonary artery oximetry catheters. Anesthesiology 1994; 81: 245–253
Burchell SA, Yu M, Takiguchi SA, Ohta RM, Myers SA. Evaluation of a continuous cardiac output and mixed venous oxygen saturation catheter in critically ill surgical patients. Crit Care Med 1997; 25: 388–391
Hofer CK, Ganter MT, Fodor P, Genoni M, Zollinger A. Validation of CeVOX for continuous central venous oxygenation measurement in patients undergoing off-pump coronary artery bypass grafting. Eur J Anaesthesiol 2006; 23(Suppl 38): 131
Parolari A, Colli S, Mussoni L, Eligini S, Naliato M, Wang X, Gandini S, Tremoli E, Biglioli P, Alamanni F. Coagulation and fibrinolytic markers in a two-month follow-up of coronary bypass surgery. J Thorac Cardiovasc Surg 2003; 125: 336–343
Spiess BD. Ischemia—a coagulation problem?. J Cardiovasc Pharmacol 1996; 27(Suppl 1): S38–S41
Baulig W, Dullenkopf A, Hasenclever P, Schmid ER, Weiss M. In vitro evaluation of the CeVOX continuous central venous oxygenation monitoring system. Anaesthesia 2008; 63: 412–417
Dueck MH, Klimek M, Appenrodt S, Weigand C, Boerner U. Trends but not individual values of central venous oxygen saturation agree with mixed venous oxygen saturation during varying hemodynamic conditions. Anesthesiology 2005; 103: 249–257
Edwards JD, Mayall RM. Importance of the sampling site for measurement of mixed venous oxygen saturation in shock. Crit Care Med 1998; 26: 1356–1360
Reinhart K, Rudolph T, Bredle DL, Hannemann L, Cain SM. Comparison of central-venous to mixed-venous oxygen saturation during changes in oxygen supply/demand. Chest 1989; 95: 1216–1221
There were no financial or non-financial competing interests in the accomplishment of this study.
Author information
Authors and Affiliations
Corresponding author
Additional information
Baulig W, Dullenkopf A, Kobler A, Baulig B, Roth HR, Schmid ER. Accuracy of Continuous Central Venous Oxygen Saturation Monitoring in Patients Undergoing Cardiac Surgery.
Rights and permissions
About this article
Cite this article
Baulig, W., Dullenkopf, A., Kobler, A. et al. Accuracy of Continuous Central Venous Oxygen Saturation Monitoring in Patients Undergoing Cardiac Surgery. J Clin Monit Comput 22, 183–188 (2008). https://doi.org/10.1007/s10877-008-9123-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10877-008-9123-2