Abstract
In transfusion-dependent thalassemia major, iron-induced cardiomyopathy is the predominant cause of morbidity and mortality. Assessment of myocardial iron loading using MRI gradient echo T2* measurements have been described, but has only been performed at one centre in London. We assessed the transferability of this method by comparing the results from three different MR scanners in three different countries. Ten patients with thalassemia major underwent myocardial T2* assessment using a Siemens Sonata Scanner in London. Patients were also scanned with either a similar T2* sequence on a GE Systems CVI scanner in Athens, or a GE Systems signa echospeed scanner in Cagliari. Two scans were performed at the respective site in all patients to assess interstudy reproducibility at each site. The mean difference and coefficient of variability for the heart between scanners was 0.08 ms and 9.7% between London and Athens; and 0.30 ms and 1.6% between London and Cagliari. The interstudy mean difference and coefficient of variability for the heart in Athens was 0.6 ms and 3.5%, and 0.2 ms and 2.4% in Cagliari. In conclusion, the myocardial iron estimations were consistent between the three centres with scanners of differing manufacture, suggesting that this technique may have widespread application in the assessment of patients with iron overload conditions such as thalassaemia.
Similar content being viewed by others
References
DJ. Weatherall (1996) ArticleTitleAnaemia as a World Health Problem Oxford Textbook of Medicine 22 3463–3482
DJ Weatherall (1997) ArticleTitleFortnightly review. The thalassaemias Br Med J 314 1675–1678
AS Schafer RG Cheron R Dluky B Cooper R Gleason JS Soeldner FH Bunn (1981) ArticleTitleClinical consequences of acquired transfusional iron overload in adults N Engl J Med 304 319–324 Occurrence Handle6777701
B Modell M Khan M Darlison (2000) ArticleTitleSurvival in beta thalassaemia major in the UK: data from the UK thalassaemia register Lancet 355 2051–2052 Occurrence Handle10.1016/S0140-6736(00)02357-6 Occurrence Handle10885361
LJ Anderson S Holden B Davies et al. (2001) ArticleTitleCardiovascular T2* (T2 star) magnetic resonance for the early diagnosis of myocardial iron overload Eur Heart J 22 2171–2179 Occurrence Handle10.1053/euhj.2001.2822 Occurrence Handle11913479
LJ Anderson B Wonke E Prescott S Holden JM Walker DJ Pennell (2002) ArticleTitleComparison of effects of oral deferiprone and subcutaneous desferrioxamine on myocardial iron levels and ventricular function in beta thalassemia Lancet 360 516–520 Occurrence Handle10.1016/S0140-6736(02)09740-4 Occurrence Handle12241655
MA Westwood LJ Anderson DN Firmin PD Gatehouse C Lorenz B Wonke D Pennell (2003) ArticleTitleInterscanner reproducibility of cardiovascular magnetic resonance T2* measurements of tissue iron in thalassemia J Magn Reson Imaging 18 616–620
MA Westwood LJ Anderson DN Firmin PD Gatehouse CC Charrier B Wonke DJ Pennell (2003) ArticleTitleA single breath-hold multiecho T2* magnetic resonance technique for diagnosis of myocardial iron overload J Magn Reson Imaging 18 33–39
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Westwood, M.A., Firmin, D.N., Gildo, M. et al. Intercentre Reproducibility of Magnetic Resonance T2* Measurements of Myocardial Iron in Thalassaemia. Int J Cardiovasc Imaging 21, 531–538 (2005). https://doi.org/10.1007/s10554-005-0651-2
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s10554-005-0651-2