Abstract
Purpose
Occasionally, a patient requiring computed tomography (CT) with poor venous access or in an unstable condition may have a sheath as their only form of IV access. Many institutions prohibit contrast power injection through the side ports, for concern for rupture. We hypothesize that under conditions encountered in CT scanning, the side arm of the sheath introducer is safe for power injection of contrast material.
Methods
In this in vitro study, we injected contrast at different flow rates varying from 2 to 8 ml/s, through the side port of the sheath. Time-pressure graphs were obtained for each injection from the injector display. The assembly was observed for any signs of leakage or material failure.
Results
There was not a single event of leak, rupture, or displacement. Maximum pressures were within the range of conventional contrast administrations.
Conclusion
This study suggests that iodinated contrast can safely be injected through the side port of a venous sheath.
Similar content being viewed by others
References
Heshmatzadeh Behzadi A, Farooq Z, Newhouse JH, Prince MR (2018) MRI and CT contrast media extravasation: a systematic review. Medicine (Baltimore) 97(9):e0055
Bae KT (2010) Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology [Internet]. 1;256(1):32–61. Available from: https://doi.org/10.1148/radiol.10090908
Sou V, McManus C, Mifflin N, Frost SA, Ale J, Alexandrou E (2017) A clinical pathway for the management of difficult venous access. BMC Nurs. 16:64. doi: https://doi.org/10.1186/s12912-017-0261-z. PMID: 29176933; PMCID: PMC5693534
Coyle D, Bloomgarden D, Beres R, Patel S, Sane S, Hurst E (2004) Power injection of contrast media via peripherally inserted central catheters for CT. J Vasc Interv Radiol 15(8):809–814
Buijs SB, Barentsz MW, Smits MLJ, Gratama JWC (2017) Spronk PE. Systematic review of the safety and efficacy of contrast injection via venous catheters for contrast-enhanced computed tomography. Eur J Radiol open [Internet]. 4:118–22. Available from: https://www.ncbi.nlm.nih.gov/pubmed/29034281
Herts BR, O’Malley CM, Wirth SL, Lieber ML, Pohlman B (2001) Power Injection of contrast media using central venous catheters. Am J Roentgenol [Internet]. 1;176(2):447–53. Available from: https://doi.org/10.2214/ajr.176.2.1760447
Andersen HW, Benumof JL, Trousdale FR, Ozaki GT (1982) Increasing the functional gauge on the side port of large catheter sheath introducers. Anesthesiology. 56(1):57–59
Benumof JL, Trousdale FR, Alfery DD, Ozaki GT (1981 Apr) Larger catheter sheath introducers and their side port functional gauge. Anesth Analg 60(4):216–217
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Khatri, G.D., Robinson, J.D. Is it safe to inject contrast through the side arm of an introducer sheath? An in vitro study. Emerg Radiol 27, 303–306 (2020). https://doi.org/10.1007/s10140-020-01768-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10140-020-01768-3