Is it safe to inject contrast through the side arm of an introducer sheath? An in vitro study

  • Original Article
  • Published:
Emergency Radiology Aims and scope Submit manuscript

Abstract

Purpose

Occasionally, a patient requiring computed tomography (CT) with poor venous access or in an unstable condition may have a sheath as their only form of IV access. Many institutions prohibit contrast power injection through the side ports, for concern for rupture. We hypothesize that under conditions encountered in CT scanning, the side arm of the sheath introducer is safe for power injection of contrast material.

Methods

In this in vitro study, we injected contrast at different flow rates varying from 2 to 8 ml/s, through the side port of the sheath. Time-pressure graphs were obtained for each injection from the injector display. The assembly was observed for any signs of leakage or material failure.

Results

There was not a single event of leak, rupture, or displacement. Maximum pressures were within the range of conventional contrast administrations.

Conclusion

This study suggests that iodinated contrast can safely be injected through the side port of a venous sheath.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Heshmatzadeh Behzadi A, Farooq Z, Newhouse JH, Prince MR (2018) MRI and CT contrast media extravasation: a systematic review. Medicine (Baltimore) 97(9):e0055

    Article  Google Scholar 

  2. Bae KT (2010) Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology [Internet]. 1;256(1):32–61. Available from: https://doi.org/10.1148/radiol.10090908

  3. Sou V, McManus C, Mifflin N, Frost SA, Ale J, Alexandrou E (2017) A clinical pathway for the management of difficult venous access. BMC Nurs. 16:64. doi: https://doi.org/10.1186/s12912-017-0261-z. PMID: 29176933; PMCID: PMC5693534

  4. Coyle D, Bloomgarden D, Beres R, Patel S, Sane S, Hurst E (2004) Power injection of contrast media via peripherally inserted central catheters for CT. J Vasc Interv Radiol 15(8):809–814

    Article  Google Scholar 

  5. Buijs SB, Barentsz MW, Smits MLJ, Gratama JWC (2017) Spronk PE. Systematic review of the safety and efficacy of contrast injection via venous catheters for contrast-enhanced computed tomography. Eur J Radiol open [Internet]. 4:118–22. Available from: https://www.ncbi.nlm.nih.gov/pubmed/29034281

  6. Herts BR, O’Malley CM, Wirth SL, Lieber ML, Pohlman B (2001) Power Injection of contrast media using central venous catheters. Am J Roentgenol [Internet]. 1;176(2):447–53. Available from: https://doi.org/10.2214/ajr.176.2.1760447

  7. Andersen HW, Benumof JL, Trousdale FR, Ozaki GT (1982) Increasing the functional gauge on the side port of large catheter sheath introducers. Anesthesiology. 56(1):57–59

    Article  CAS  Google Scholar 

  8. Benumof JL, Trousdale FR, Alfery DD, Ozaki GT (1981 Apr) Larger catheter sheath introducers and their side port functional gauge. Anesth Analg 60(4):216–217

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jeffrey D. Robinson.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Khatri, G.D., Robinson, J.D. Is it safe to inject contrast through the side arm of an introducer sheath? An in vitro study. Emerg Radiol 27, 303–306 (2020). https://doi.org/10.1007/s10140-020-01768-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10140-020-01768-3

Keywords

Navigation