Abstract
Background
This human cadaveric study i ntroduces a laboratory model to establish and compare the fixation stability of the distal femoral locking plate (DFLP) and dynamic condylar screw (DCS) in distal femoral fracture fixation.
Materials and methods
The study was conducted on 16 fresh cadaveric femoral specimens, 8 implanted with the DCS and the other 8 with the DFLP. The construct was made unstable by removing a standard-sized medial wedge with a 1-cm base (gap osteotomy) beginning 6 cm proximal to the lateral joint line in the distal metaphyseal region with loss of the medial buttress. Each specimen underwent axial and torsional stiffness testing along with cyclic axial loading to failure. The mean DEXA value for the DFLP group was 0.82 g/cm2 and in the DCS group was 0.79 g/cm2.
Results
Axial stiffness in the DFLP group was significantly higher than in the DCS group, but no significant difference was found in torsional stiffness between the groups. A significant difference was found in the load-to-failure results between the groups. Plastic and total deformation was significantly higher in constructs in the DCS group than in those in the DFLP group. Total energy absorbed before construct failure was also significantly higher in the DFLP group than in the DCS group.
Conclusions
The DFLP construct proved stronger than the DCS in both axial stiffness and cyclic loading, but similar in torsional stiffness in biomechanical testing in a simulated A3 distal femoral fracture.
Similar content being viewed by others
References
Kolmert L, Persson BM, Romanus B. An experimental study of devices for internal fixation of the distal femoral fractures. Clin Orthop. 1982;171:290–9.
Martinet O, Cordey J, Harder Y, Maier A, Bühler M, Barraud GE. The epidemiology of fractures of the distal femur. Injury. 2000;31:C62–3.
Muller ME, Nazarian S, Koch P, Schatzker J. The comprehensive classification of fractures of long bones. Berlin: Spring; 1990. p. 116–47.
Frigg R, Appenzeller A, Christensen R, Frenk A, Gilbert S, Schavan R. The development of the distal femur less invasive stabilization system (LISS). Injury. 2001;32:SC24–31.
Krettek C, Muller M, Miclau T. Evolution of minimally invasive plate osteosynthesis (MIPO) in the femur. Injury. 2001;32:SC14–23.
Giles JB, De Lee JC, Heckman JD, Keever JE. Supracondylar-intercondylar fractures of the femur treated with a supracondylar plate and lag screw. J Bone Joint Surg Am. 1982;64:864–70.
Sanders R, Swiontkowski M, Rosen H, Helfet D. Double-plating of comminuted, unstable fractures of the distal part of the femur. J Bone Joint Surg Am. 1991;73:341–6.
Firoozbakhsh K, Behzadi K, De Coster TA, Moneim MS, Naraghi FF. Mechanics of retrograde nail versus plate fixation for supracondylar femur fractures. J Orthop Trauma. 1995;9:152–7.
Helfet DL, Lorich DG. Retrograde intramedullary nailing of supracondylar femoral fractures. Clin Orthop Relat Res. 1998;350:80–4.
Heiney JP, Barnett MD, Vrabec GA, Schoenfeld AJ, Baji A, Njus GO. Distal femoral fixation: a biomechanical comparison of trigen retrograde intramedullary (i.m.) nail, dynamic condylar screw (DCS), and locking compression plate (LCP) condylarplate. J Trauma. 2009;66:443–9.
Higgins TF, Pittman G, Hines J, Bachus KN. Biomechanical analysis of distal femur fracture fixation: fixed-angle screw-plate construct versus condylar blade plate. J Orthop Trauma. 2007;21:43–6.
Koval KJ, Hoehl JJ, Kummer FJ, Simon JA. Distal femoral fixation: a biomechanical comparison of the standard condylar buttress plate, a locked buttress plate, and the 95-degree blade plate. J Orthop Trauma. 1997;11:521–4.
Mahomed N, Harrington I, Kellam J, Maistrelli G, Hearn T, Vroemen J. Biomechanical analysis of the gamma nail and sliding hip screw. Clin Orthop Relat Res. 1994;304:280–8.
Zlowodzki M, Williamson S, Cole PA, Zardiackas LD, Kregor PJ. Biomechanical evaluation of the less invasive stabilization system, angled blade plate, and retrograde intramedullary nail for the internal fixation of distal femur fractures. J Orthop Trauma. 2004;18:494–502.
Marti A, Fankhauser C, Frenk A, Cordey J, Gasser B. Biomechanical evaluation of the less invasive stabilization system for the internal fixation of distal femur fractures. J Orthop Trauma. 2001;15:482–7.
Lucas SE, Seligson D, Henry SL. Intramedullary supracondylar nailing of femoral fractures: a preliminary report of the GSH supracondylar nail. Clin Orthop. 1993;296:200–6.
Harder Y, Martinet O, Barraud GE, Cordey J, Regazzoni P. The mechanics of internal fixation of fractures of the distal femur: a comparison of the condylar screw (DCS) with the condylar plate (CP). Injury. 1999;30:31–9.
Ito K, Grass R, Zwipp H. Internal fixation of supracondylar femoral fractures: comparative biomechanical performance of the 958 plate and two retrograde nails. J Orthop Trauma. 1998;4:259–66.
Kubiak EN, Fulkerson E, Strauss E, Egol KA. The evolution of lockedplates. J Bone Joint Surg Am. 2006;88:189–200.
Bong MR, Egol KA, Koval KJ, Kummer FJ, Su ET, Iesaka K. Comparison of the LISS and a retrograde- inserted supracondylar intramedullary nail for fixation of a periprosthetic distal femur fracture proximal to a total knee arthroplasty. J Arthroplasty. 2002;17:876–81.
Stoffel K, Dieter U, Stachowiak G, Gachter A, Kuster MS. Biomechanical testing of the LCP—how can stability in locked internal fixators be controlled? Injury. 2003;34(Suppl 2):B11–9.
Stoffel K, Lorenz KU, Kuster MS. Biomechanical considerations in plate osteosynthesis: the effect of plate-to-bone compression with and without angular screw stability. J Orthop Trauma. 2007;21:362–8.
Kolmert L, Wulff K. Epidemiology and treatment of distal femoral fractures in adults. Acta Orthop Scand. 1982;53:957–62.
Jaakkola JI, Lundy DW, Moore T, Jones B, Ganey TM, Hutton WC. Supracondylar femur fracture fixation: mechanical comparison of the 95 degrees condylar side plate and screw versus 95 degrees angled blade plate. Acta Orthop Scand. 2002;73:72–6.
Conflict of interest
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Narsaria, N., Singh, A.K., Rastogi, A. et al. Biomechanical analysis of distal femoral fracture fixation: dynamic condylar screw versus locked compression plate. J Orthop Sci 19, 770–775 (2014). https://doi.org/10.1007/s00776-014-0583-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00776-014-0583-6