Internal carotid artery volume flow correlates to rCBF measurements

  • Published:
Acta Neurochirurgica Aims and scope Submit manuscript

Summary

Background. There is no simple, cost effective bedside method available for measuring global cerebral blood flow (CBF) rapidly and repeatedly.

Method. Based on the ultrasound technique a device was developed to measure flow volume per unit of time in the internal carotid artery. The system utilizes dual beam, angle-independent Doppler technology, and employs simultaneous sampling and full spectrum determination. The aim of this study was to evaluate this device in comparison with cerebral blood flow measurements using the well established Xenon133 clearence technique.

Findings. 10 patients suffering from SAH were included. 20 measurements were performed. CBF was measured employing the Xenon133 clearence technique. Velocity profiles across the ICA were obtained with the high-resolution FlowGuard Doppler ultrasound flow system (Cardiosonix Ltd., Israel). According to the statistical analysis the ICA flow correlated significantly to the mean rCBF15 index of the hemisphere (p<0.0001).

Conclusions. This new device seems to be promising for monitoring cerebral blood flow in critically ill neurological and neurosurgical patients.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Published online September 18, 2003

Rights and permissions

Reprints and permissions

About this article

Cite this article

Rothoerl, R., Schebesch, K., Woertgen, C. et al. Internal carotid artery volume flow correlates to rCBF measurements. Acta Neurochir 145, 943–947 (2003). https://doi.org/10.1007/s00701-003-0108-0

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00701-003-0108-0

Navigation