Bioabsorbable sutures versus screw fixation of displaced tibial eminence fractures: a biomechanical study

  • Original Article
  • Published:
European Journal of Orthopaedic Surgery & Traumatology Aims and scope Submit manuscript

Abstract

Background

In 1995, we developed a simple and safe arthroscopic technique of anterograde stable fixation of the displaced tibial eminence fractures using a cannulated screw and washer, allowing immediate mobilisation and weight bearing. Some authors described similar end results by arthroscopic fixation of this abruption with trans-osseous sutures. We proved with the biomechanical study that the fixation with strong trans-osseous sutures provides a comparable stable fixation the same as the cannulated screw and washer. We developed a test machine for cyclic loading and a machine for measuring of the pull-out strength for this study.

Methods

A standard osteotomy of the tibial spine was performed during the cadaveric biomechanical study. We performed an arthroscopic fixation with the cannulated screw and washer in ten knees, and in another group of ten knees, arthroscopic fixation was made using two trans-osseous Orthocord™ sutures. The operating time for each procedure was measured, and any possible technical complications were recorded. One thousand cycles of flexion from 0° to 90° were performed on a special loading device on all knees. The macroscopic dislocation of the fragment was measured. After the cyclic loading, all soft tissue was resected except the anterior cruciate ligament with the fixed fragment. The pull-out strength defined as translation of the tibia in anterior direction by breakage was measured with a custom-made measuring device.

Results

The average time for the screw and washer fixation was 20 min, and average time for the sutures fixation was 48 min. After cyclic loading, we did not see any dislocation of the fragment. Average pull-out strength at which fixation with a cannulated screw and washer fell was 253.42 N and mean pull-out strength at which fixation with trans-osseous sutures fell was 330.32 N. Fixation of the fragment with trans-osseous sutures was statistically significantly stronger—Wilcoxon–Mann–Whitney’s test, p = <0.00.

Conclusion

Fixation with strong trans-osseous sutures is stronger than fixation with a cannulated screw and washer on cadaveric knees. It appears that fixation with strong trans-osseous sutures allows immediate mobilisation and weight bearing like fixation with a cannulated screw and washer. Meanwhile, the time for fixation with trans-osseous sutures is significantly longer.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Meyers MH, McKeever FM (1959) Fractures of the intercondylar eminence of the tibia. J Bone Joint Surg Am 41:209–222

    PubMed  Google Scholar 

  2. Meyers MH, McKeever FM (1970) Fractures of the intercondylar eminence of the tibia. J Bone Joint Surg Am 52:1677–1683

    CAS  PubMed  Google Scholar 

  3. Zariczny B (1977) Avulsion fractures of the tibial eminence treated by open reduction and pinning. J Bone Joint Surg Am 59:1111–1114

    Google Scholar 

  4. McLennan JG (1982) The role of arthroscopic surgery in the treatment of fractures of the intercondylar eminence of the tibia. J Bone Joint Surg Br 64:477–480

    CAS  PubMed  Google Scholar 

  5. Lysholm J, Gillquist J (1982) Evaluation of knee ligament surgery results with special emphasis on use of a scoring scale. Am J Sports Med 10:150–154

    Article  CAS  PubMed  Google Scholar 

  6. Highgenboten CL, Jackson A, Meske NBG (1989) Genucom, KT-1000, and Stryker knee laxity measuring device comparisons. Device reproducibility and interdevice comparison in asymptomatic subjects. Am J Sports Med 17:743–746

    Article  CAS  PubMed  Google Scholar 

  7. Wroble RR, Van Ginkel LA, Grood ES, Noyes FR, Shaffer BL (1990) Repeatability of the KT-1000 arthrometer in a normal population. Am J Sports Med 18:6–9

    Google Scholar 

  8. Van Loon T, Mart RKA (1991) Fracture of the intercondylar eminence of the tibia treated by arthroscopic fixation. Arthroscopy 7:385–388

    Article  PubMed  Google Scholar 

  9. Berg EE (1993) Comminuted tibial eminence anterior cruciate ligament avulsion fractures: failure of arthroscopic treatment. Arthroscopy 9:446–450

    Article  CAS  PubMed  Google Scholar 

  10. Mathews DE, Geissler WB (1994) Arthroscopic suture fixation of displaced tibial eminence fractures. Arthroscopy 10:418–423

    Article  Google Scholar 

  11. Carro LP, Suarez GG, Cimiano FG (1994) The arthroscopic knot technique for fracture of the tibia in children. Arthroscopy 10:698–699

    Article  Google Scholar 

  12. Veselko M, Senekovič V, Tonin M (1996) Simple and safe arthroscopic placement and removal of cannulated screw and washer for fixation of tibial avulsion fracture of the anterior cruciate ligament. Arthroscopy 12:259–262

    Article  CAS  PubMed  Google Scholar 

  13. Kogan MG, Marks P, Amendola A (1997) Technique for arthroscopic suture fixation of displaced tibial intercondylar eminence fractures. Arthroscopy 13:301–306

    Article  CAS  PubMed  Google Scholar 

  14. Mah JY, Adili A, Otsuka NY, Ogilvie R (1998) Follow-up study of arthroscopic reduction and fixation of type III tibial eminence fractures. J Pediatr Orthop 18:475–477

    CAS  PubMed  Google Scholar 

  15. Iborra JP, Mazeau P, Louahem D, Dimeglio A (1999) Fractures of the intercondylar eminence of the tibia in children. Apropos of 25 cases with a 1–20 year follow up. Rev Chir Orthop Reparatrice Appar Mot 85:563–573

    CAS  PubMed  Google Scholar 

  16. Osti L, Merlo F, Liu SH, Bocchi L (2000) A simple modified arthroscopic procedure for fixation of displaced tibial eminence fractures. Arthroscopy 16:379–382

    Article  CAS  PubMed  Google Scholar 

  17. Binnet MS, Gurkan I, Ylmaz C et al (2001) Arthroscopic fixation of intercondylar eminence fractures using 4-portal technique. Arthroscopy 17:450–460

    Article  CAS  PubMed  Google Scholar 

  18. Kocher MS, Foreman ES, Micheli LJ (2003) Laxity and functional outcome after arthroscopic reduction and internal fixation of displaced tibial spine fractures in children. Arthroscopy 19:1085–1090

    Article  PubMed  Google Scholar 

  19. Senekovič V, Veselko M (2003) Anterograde arthroscopic fixation of avulsion fractures of the tibial eminence with a cannulated screw: five-year results. Arthroscopy 19:54–61

    Article  PubMed  Google Scholar 

  20. Hunter RE, Willis JA (2004) Arthroscopic fixation of avulsion fractures of the tibial eminence: technique and outcome. Arthroscopy 20:113–121

    Article  PubMed  Google Scholar 

  21. Ahn JH, Yoo JC (2005) Clinical outcome of arthroscopic reduction and suture for displaced acute and chronic tibial spine fractures. Knee Surg Sports Traumatol Arthrosc 13:116–121

    Article  PubMed  Google Scholar 

  22. Bong MR, Romero A, Kubiak E, Iesaka K, Heywood CS, Kummer F, Rosen J, Jazrawi L (2005) Suture versus screw fixation of displaced tibial eminence fractures: a biomechanical comparison. Arthroscopy 21:1172–1176

    Article  PubMed  Google Scholar 

  23. Tsukada H, Ishibashi Y, Tsuda E, Hiraga Y, Toh SA (2005) Biomechanical comparison of repair techniques for anterior cruciate ligament tibial avulsion fracture under cyclic loading. Arthroscopy 10:1197–1201

    Article  Google Scholar 

  24. Zhao J, Huangfu X (2007) Arthroscopic treatment of nonunited anterior cruciate ligament tibial avulsion fracture with figure-of-8 suture fixation technique. Arthroscopy 23:405–410

    Article  PubMed  Google Scholar 

  25. Mahar AT, Duncan D, Oka R, Lowry A, Gillingham B, Chambers H (2008) Biomechanical comparison of four different fixation techniques for pediatric tibial eminence avulsion fractures. J Pediatr Orthop 28(2):159–162

    Article  PubMed  Google Scholar 

  26. Wilfinger C, Castellani C, Raith J, Pilhatsch A, Höllwarth ME, Weinberg AM (2009) Nonoperative treatment of tibial spine fractures in children-38 patients with a minimum follow-up of 1 year. J Orthop Trauma 23(7):519–524

    Article  PubMed  Google Scholar 

  27. Tudisco C, Giovarruscio R, Febo A, Savarese E, Bisicchia S (2010) Intercondylar eminence avulsion fracture in children: long-term follow-up of 14 cases at the end of skeletal growth. J Pediatr Orthop B 19(5):403–408

    Article  PubMed  Google Scholar 

  28. Su WR, Wang PH, Wang HN, Lin CJ (2011) A simple, modified arthroscopic suture fixation of avulsion fracture of the tibial intercondylar eminence in children. J Pediatr Orthop B 20(1):17–21

    Article  PubMed  Google Scholar 

Download references

Conflict of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Vladimir Senekovic.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Senekovic, V., Balazic, M. Bioabsorbable sutures versus screw fixation of displaced tibial eminence fractures: a biomechanical study. Eur J Orthop Surg Traumatol 24, 209–216 (2014). https://doi.org/10.1007/s00590-013-1176-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00590-013-1176-3

Keywords

Navigation