Does transcranial stimulation for motor evoked potentials (TcMEP) worsen seizures in epileptic patients following spinal deformity surgery?

  • Original Article
  • Published:
European Spine Journal Aims and scope Submit manuscript

Abstract

Purpose

To investigate the effect of Transcranial Motor Evoked Potentials (TcMEP) in increasing the severity or frequency of post-operative seizures in patients undergoing deformity corrective spine surgery with a known history of seizures pre-operatively.

Methods

The information on all patients with history of epilepsy/seizures who underwent spinal TcMEP cord monitoring for deformity correction surgery was retrospectively collected through a review of the hospital notes. The benefits of TcMEP in the early detection of potential cord ischemia were deemed by the operating surgeon to outweigh the increased risks of seizures, tongue biting, etc. Data on age, gender, pre-operative diagnosis, curve type, intra-operative monitoring alerts, duration of hospital stay, and post-operative in-hospital seizures were collected. Additionally, the patients were contacted following discharge and data on any change in the frequency of the seizures or an alteration in seizure-related medication post-operatively was also collected.

Results

The records of 449 consecutively monitored patients were reviewed and 12 (2.7 %) patients with a history of seizures pre-operatively were identified. The mean age was 23 (9–59) years, 7 females, 11 scoliosis corrections (4 neuromuscular, 1 degenerative, 6 idiopathic adolescent), and one sagittal balance correction surgery. Intra-operatively, all patients had TcMEP monitoring, were catheterised, and had no neuromonitoring alerts or record of tongue biting or laceration. Post-operatively, the mean hospital stay was 12 (4–25) days with no recorded seizures. At a mean of 23 (12–49) months post-discharge, none of the patients reported a worsening of seizures (pattern or frequency) or required an alteration in the seizure-related medications.

Conclusion

TcMEP does not appear to trigger intra-operative or post-operative seizures and is not associated with deterioration in the seizure control of patients suffering seizures pre-operatively.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Burke D, Hicks RG (1998) Surgical monitoring of motor pathways. J Clin Neurophysiol Off Publ Am Electroencephalogr Soc 15:194–205

    CAS  Google Scholar 

  2. Schwartz DM, Auerbach JD, Dormans JP, Flynn J, Drummond DS, Bowe JA, Laufer S, Shah SA, Bowen JR, Pizzutillo PD, Jones KJ, Drummond DS (2007) Neurophysiological detection of impending spinal cord injury during scoliosis surgery. J Bone Joint Surg Am 89:2440–2449. doi:10.2106/JBJS.F.01476

    Article  PubMed  Google Scholar 

  3. Pajewski TN, Arlet V, Phillips LH (2007) Current approach on spinal cord monitoring: the point of view of the neurologist, the anesthesiologist and the spine surgeon. Eur Spine J Off Publ Eur Spine Soc Eur Spinal Deform Soc Eur Sect Cerv Spine Res Soc 16(Suppl 2):S115–S129. doi:10.1007/s00586-007-0419-6

    Article  Google Scholar 

  4. Kelm GH, LuÈders HO, Jasper HH, Elger C (1999) The ten-twenty electrode system of the International Federation. The International Federation of Clinical Neurophysiology. Electroencephalogr Clin Neurophysiol Suppl 52:3–6

    Google Scholar 

  5. Lenke LG, Betz RR, Harms J, Bridwell KH, Clements DH, Lowe TG, Blanke K (2001) Adolescent idiopathic scoliosis: a new classification to determine extent of spinal arthrodesis. J Bone Joint Surg Am 83-a:1169–1181

    CAS  PubMed  Google Scholar 

  6. Vauzelle C, Stagnara P, Jouvinroux P (1973) Functional monitoring of spinal cord activity during spinal surgery. Clin Orthop Relat Res 93:173–178

    Article  PubMed  Google Scholar 

  7. Padberg AM, Raynor BL, Thuet ED, Bolon SM (2012) Spinal cord and nerve root monitoring. In: Bridwell KH, Dewald RL (eds) The textbook of spinal surgery. Lippincott Williams & Wilkins, Philadelphia, pp 7–40

    Google Scholar 

  8. Nash CL, Jr., Lorig RA, Schatzinger LA, Brown RH (1977) Spinal cord monitoring during operative treatment of the spine. Clini Orthop Relat Res 126:100–105

    Google Scholar 

  9. Dinner DS, Luders H, Lesser RP, Morris HH, Barnett G, Klem G (1986) Intraoperative spinal somatosensory evoked potential monitoring. J Neurosurg 65:807–814. doi:10.3171/jns.1986.65.6.0807

    Article  CAS  PubMed  Google Scholar 

  10. Hilibrand AS, Schwartz DM, Sethuraman V, Vaccaro AR, Albert TJ (2004) Comparison of transcranial electric motor and somatosensory evoked potential monitoring during cervical spine surgery. J Bone Joint Surg Am 86-a:1248–1253

    PubMed  Google Scholar 

  11. Calancie B, Harris W, Broton JG, Alexeeva N, Green BA (1998) “Threshold-level” multipulse transcranial electrical stimulation of motor cortex for intraoperative monitoring of spinal motor tracts: description of method and comparison to somatosensory evoked potential monitoring. J Neurosurg 88:457–470. doi:10.3171/jns.1998.88.3.0457

    Article  CAS  PubMed  Google Scholar 

  12. Langeloo DD, Journee HL, Polak B, de Kleuver M (2001) A new application of TCE-MEP: spinal cord monitoring in patients with severe neuromuscular weakness undergoing corrective spine surgery. J Spinal Disord 14:445–448

    Article  CAS  PubMed  Google Scholar 

  13. Quraishi NA, Lewis SJ, Kelleher MO, Sarjeant R, Rampersaud YR, Fehlings MG (2009) Intraoperative multimodality monitoring in adult spinal deformity: analysis of a prospective series of one hundred two cases with independent evaluation. Spine 34:1504–1512. doi:10.1097/BRS.0b013e3181a87b66

    Article  PubMed  Google Scholar 

  14. Macdonald DB (2006) Intraoperative motor evoked potential monitoring: overview and update. J Clin Monit Comput 20:347–377. doi:10.1007/s10877-006-9033-0

    Article  PubMed  Google Scholar 

  15. (2009) Intraoperative neurophysiological monitoring of spinal cord function during spinal deformity surgery. In: SRS Information Statement. Scoliosis Research Society, p 6

  16. MacDonald DB (2002) Safety of intraoperative transcranial electrical stimulation motor evoked potential monitoring. J Clin Neurophysiol Off Publ Am Electroencephalogr Soc 19:416–429

    Google Scholar 

  17. Sackeim HA, Devanand DP, Prudic J (1991) Stimulus intensity, seizure threshold, and seizure duration: impact on the efficacy and safety of electroconvulsive therapy. Psychiatr Clin N Am 14:803–843

    CAS  Google Scholar 

  18. Modica PA, Tempelhoff R, White PF (1990) Pro- and anticonvulsant effects of anesthetics (Part I). Anesth Analg 70:303–315

    Article  CAS  PubMed  Google Scholar 

  19. Modica PA, Tempelhoff R, White PF (1990) Pro- and anticonvulsant effects of anesthetics (Part II). Anesth Analg 70:433–444

    Article  CAS  PubMed  Google Scholar 

  20. Legatt AD (2002) Current practice of motor evoked potential monitoring: results of a survey. J Clin Neurophysiol Off Publ Am Electroencephalogr Soc 19:454–460

    Google Scholar 

Download references

Conflict of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Khalid M. I. Salem.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Salem, K.M.I., Goodger, L., Bowyer, K. et al. Does transcranial stimulation for motor evoked potentials (TcMEP) worsen seizures in epileptic patients following spinal deformity surgery?. Eur Spine J 25, 3044–3048 (2016). https://doi.org/10.1007/s00586-015-3993-z

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00586-015-3993-z

Keywords

Navigation