Introduction

Over the years, intraoperative anesthetic monitoring techniques have been improved similarly to the improvement of surgical instruments. Blood pressure, heart rate, electrocardiography, and blood oxygen saturation level are mainly monitored during intraoperative anesthesia. In the present article, we review the oxygen reserve index (ORi™), which is a measurement that enables the assessment of oxygen reserve capacity, and discuss its various applications and future outlooks.

What is ORi

During anesthetic management, predicting the partial pressure of arterial oxygen (PaO2) is difficult when percutaneous oxygen saturation (SpO2) is 100%. In usual cases, in which arterial lines of the patients are secured, arterial blood gas analysis for the appropriate management of PaO2 is possible. However, as this is an invasive method, it cannot be performed on all patients. In most patients undergoing general anesthesia, tracheal intubation and artificial ventilation are carried out to manage positive-pressure ventilation. However, during surgery, atelectasis and pulmonary injury may occur owing to multiple factors [1,2,3], which may cause sudden hypoxemia during or after surgery, making it difficult to manage the patient [1, 4, 5]. Therefore, PaO2 during surgery is a useful indicator for anesthesiologists, particularly in patients with low respiratory function (e.g., obstructive/restrictive ventilatory disorder) and low oxygen reserve (e.g., pregnant women and obese patients) [6, 7]. In addition, controlling oxygen concentrations during surgery in patients with diseases in which hyperoxia should be avoided, such as interstitial pneumonia and retinopathy of prematurity [8], is a difficult problem. Although hypoxia can lead to adverse effects, hyperoxia during surgery can also induce pulmonary injury and absorption atelectasis, resulting in postsurgical respiratory complications [1,2,3]. Furthermore, hyperoxia can also increase the risk of other adverse events [9] (Table 1). Unfavorable clinical outcomes owing to a high fraction of inspired oxygen (FiO2) have been recently reported in critically ill adults, including patients with chronic obstructive pulmonary disease, myocardial infarction, cardiac arrest, stroke, and traumatic brain injury [10,11,12,13,14,15].

Table 1 Adverse events caused by exposure to high oxygen concentrations

Until recently, the pulse oximeter was the only noninvasive instrument to measure arterial blood oxygen saturation (SaO2), which is an index of blood oxygenation. During the 1970s, the world’s first pulse oximeter that was able to measure SaO2 from the fingertip was commercialized. Since then, this device has been used in a wide range of situations for the noninvasive monitoring of blood oxygenation [16]. However, this device has its flaws. Although SpO2 and PaO2 are correlated at most points, when SpO2 reaches 100%, estimating PaO2 becomes difficult [17] (Fig. 1). Thus, other methods, such as arterial blood gas analysis, are often performed to measure PaO2 for the evaluation of oxygenation.

Fig. 1
figure 1

Correlation between PaO2 and SaO2. a The oxygen dissociation curve for hemoglobin is shown. A correlation is observed up until the PaO2 is 100 mmHg, but when SpO2 reaches 100%, it becomes impossible to estimate PaO2. PaO2 partial pressure of arterial oxygen, SaO2 arterial oxygen saturation

The oxygen reserve index (ORi) is a noninvasive and continuous parameter that can be measured by connecting a multi-wavelength sensor to Root® with Radical-7®, Pulse CO-Oximeter® (Masimo Corp., Irvine, CA, USA). ORi is an index measured using a non-unit scale between 0.00 and 1.00, which reflects the range of moderate hyperoxia (PaO2 between 100 and 200 mmHg), and is defined as oxygen reserve reflecting the level of venous blood oxygen saturation (SvO2). ORi remains at 0.00 in a room air environment. With supplemental oxygen, after SpO2 reaches 100%, SvO2 also increases as PaO2 does, and usually reaches a plateau at about a PaO2 of 200 mmHg [18, 19] (Fig. 2). In general, a pulse oximeter displays the SpO2 value measured from light absorption of the pulsatile blood at the site where the sensor is placed. Pulsatile changes are observed in the arteries, capillaries, and venules. Masimo developed rainbow SET® which can analyze changes in the absorption of incident light of both arterial and venous blood using a multi-wavelength sensor, and display them as ORi, which is a unitless scale between 0.00 and 1.00. Thus, ORi begins to increase from 0.00 at about a PaO2 of 100 mmHg, and reaches a plateau of 1.00 at approximately 200 mmHg [18,19,20,21]. However, there are reports stating that in cases where FiO2 is 1.0, even when oxygenation is complete, the peak value is different for each individual and reaches a plateau at about 0.60 [22]. Applegate et al. analyzed the association between ORi and PaO2 during surgery, and found that if PaO2 is below 240 mmHg, ORi and PaO2 show a positive correlation (r2 = 0.536) [23] (Fig. 3). However, if PaO2 is higher than 240 mmHg, there was no correlation (r2 = 0.0016). Furthermore, when ORi was more than 0.24, PaO2 was more than 100 mmHg. As the report showed that 96.6% of the samples had a PaO2 of ≥ 150 mmHg when ORi was above 0.55, the authors concluded that ORi can be used as an indicator of PaO2. Vos et al. investigated the correlation between PaO2 and ORi during the administration of oxygen in 20 healthy volunteers, and they also reported that the two values have a strong positive correlation [24]. In addition, regarding the correlation between PaO2 and ORi during anesthesia under oxygenation, when PaO2 was less than 240 mmHg (n = 69), linear regression analysis showed a relatively strong positive correlation (r2 = 0.706) [25].

Fig. 2
figure 2

Mechanism of ORi. a There is a correlation between SpO2 and PaO2 up to a PaO2 of 100 mmHg, and by measuring the SpO2, it is possible to estimate PaO2. For PaO2 values of more than 100 mmHg, SpO2 will have reached a maximum and will not change any further; however, SvO2 will continue to increase. A change in SvO2 causes a change in the absorption of the incident light in parallel with changes in PaO2, resulting in changes in the measurement signal. Using the Rainbow SET technology established by Masimo Japan, it is possible to extract these measurement signals, enabling the detection of PaO2 changes through changes in SvO2. The SvO2 value reaches a plateau when PaO2 exceeds more than 200 mmHg. Thus, ORi reflects changes in PaO2 in the range of 100 to 200 mmHg (orange area). (Modified and reproduced with permission) (Springer Nature; J Clin Monit Comput) [19] http://creativecommons.org/licenses/by/4.0/. b Schematic representation of arterial (red line) and venous (blue line) oxyhemoglobin dissociation curves. PaO2 partial pressure of arterial oxygen, SaO2 arterial oxygen saturation, SvO2 venous oxygen saturation, ORi oxygen reserve index

Fig. 3
figure 3

Correlation between PaO2 and ORi. a A positive correlation is observed between ORi and PaO2 below 240 mmHg (r2 = 0.536). The dotted lines indicate the 95% confidence interval of the regression line. (Modified and reproduced with permission) (Wolters Kluwer Health; Anesth Analg) [23]. PaO2 partial pressure of arterial oxygen, ORi oxygen reserve index

Suitable cases for the use of ORi with regard to anesthetic management

The following is a list of cases for which the use of ORi is suitable.

  • Rapid sequence induction

Normally in a room air environment, ORi is 0.00. After the administration of oxygen, even after SpO2 reaches 100%, ORi will continue to increase. Rapid sequence induction (RSI) is appropriate for patients in whom the aspiration of gastric contents may occur, including those with a full stomach, ileus, and expectant mothers [26,27,28]. In RSI, tracheal intubation is performed without ventilation after the occurrence of apnea. In such cases, ORi may be an indicator of the time it will take for SpO2 to decrease during tracheal intubation. Yoshida et al. investigated the usefulness of ORi during RSI [29]. They reported that in 10 out of 17 patients (77%) in whom preoxygenation was performed, ORi started to decrease 32.5 s (IQR: 18.1–51.3) before a decrease in SpO2 post-apnea. By monitoring the ORi, it is possible to predict a decrease in oxygenation about 30 s before SpO2 starts to decrease. From this finding, Yoshida concluded that it may be possible to lower the incidence rate of hypoxemia during RSI. Thus, ORi is expected to be useful for RSI.

In addition, owing to the recent outbreak of coronavirus disease 2019 (COVID-19), RSI has been recommended to prevent the infection of anesthesiologists during intubation [30, 31]. Furthermore, the use of aerosol boxes for the prevention of COVID-19 infection makes intubation difficult [32,33,34]. Thus, aerosol boxes may increase intubation times and therefore increase the risk of hypoxia in patients [32]. Using ORi may hence reduce the risk of hypoxemia and enable safe tracheal intubation.

  • Pediatric applications

In pediatric patients, oxygen consumption is higher, and the closing volume is also relatively larger than the functional residual capacity compared with adult patients. Owing to this fact, the development of hypoxemia in pediatric patients is more rapid than in adult patients during apnea. Previous studies investigating ORi in pediatric patients during anesthetic induction showed that ORi monitoring enabled detection of a reduction in oxygenation approximately 30 s prior to a rapid change in SpO2 [21]. These studies suggest that prediction of the onset of a rapid decrease in oxygen saturation enables the prompt implementation of appropriate treatments, which is an advantage. In fact, the use of ORi for tracheoplasty in children was found to enable the earlier prediction of hypoxemia [35]. In addition, as the trachea is shorter in children than in adults, when performing tracheal intubation and head bending during surgery, there is a higher possibility of unilateral intubation in children [36, 37]. Therefore, in the former case, comparison of ORi before and after intubation may be useful, and in the latter case, a sudden drop in ORi during the operation may be a useful indicator of subsequent hypoxemia. Therefore, the use of ORi may enable the earlier implementation of appropriate treatments. All of the above demonstrate that ORi is a useful indicator in pediatric anesthesia during the perioperative period.

  • Other suitable cases

ORi is also useful for patients in whom respiration management is required during one-lung ventilation (OLV). Hypoxemia may still occur in about 10% of such patients [38]. When switching from two-lung to OLV, the shunt fraction increases, oxygenation is impaired, and hypoxemia may occur [39]. Studies monitoring ORi during the anesthetic management of OLV showed that decreases in PaO2 could be detected much earlier than when only SpO2 was monitored. These studies suggest that ORi can reduce the risk of complications during OLV [40]. Another study reported that the monitoring of ORi during OLV could predict hypoxemia [41]. These findings may be helpful to adjust FiO2 individually in patients undergoing OLV, and to avoid unnecessarily high concentrations of oxygen.

In addition, the use of ORi is suitable for surgeries that have a high probability of leading to atelectasis, such as surgeries on obese patients [42, 43], surgeries in the Trendelenburg position, and laparoscopic surgeries [43,44,45]. Other studies also suggested that ORi is useful in situations in which awake intubation is performed [22]. ORi monitoring may be able to avoid hypoxemia in cases in which bronchial stenting using a rigid endoscope is performed while the patient continues to breathe spontaneously [46]. For surgeries in which patients’ respiratory statuses can vary, ORi monitoring can be an indicator to decide whether or not early intervention is required to avoid hypoxemia [47]. ORi monitoring during the induction of anesthesia can be an indicator of whether preoxygenation is being carried out appropriately.

In addition, hyperoxia can be detected by monitoring ORi in patients under general anesthesia, and thus the administration of unnecessarily high concentrations of oxygen can possibly be avoided. In fact, there are several reports demonstrating that the use of ORi in managing oxygen levels during and after surgery, and for patients in the intensive care unit (ICU) can prevent exposure to excess oxygen [25, 48, 49]. In general, high oxygen concentrations are more harmful to infants, children, and older patients than to adults [50]. Oxidative stress that occurs as a consequence of treatment with high oxygen concentrations during the perinatal period notably affects the lung and brain development of premature and newborn infants. For instance, bronchopulmonary dysplasia is caused by hyperoxic lung injury [51,52,53]. In older patients, the production and accumulation of reactive oxygen species (ROS) are largely increased, and antioxidative functions are decreased, resulting in various aging-associated diseases [54, 55]. Hyperoxia promotes ROS production, suggesting that perioperative hyperoxia might enhance the adverse effects of aging-associated diseases, particularly in lung and brain diseases. Therefore, adequate oxygenation using ORi may lead to a reduction in complications.

Methods and cautions of the use of ORi

ORi is measured by placing an RD rainbow Lite SET® sensor, which is a multiwavelength sensor, on a patient’s fingertip and connecting it to Root® with Radical-7®. In addition to ORi, this sensor can measure a patient’s SpO2, pulse rate (PR), Perfusion Index (Pi), and Pleth Variability Index (PVi®) (Fig. 4). In addition, ORi can be measured by a RD rainbow SET®-2 sensor, which can measure Noninvasive and Continuous Hemoglobin (SpHb®). However, it is necessary to block any outside light when using the SpHb sensor, as this sensor is easily affected by light owing to its high sensitivity to long wavelengths (Fig. 4). Furthermore, special attention is needed to make sure that the SpO2 is greater than 98% because if this criterion is not established, ORi does not show a positive number. One caution during monitoring is that measurement is difficult when the patient’s body is moving or when the blood perfusion rate is very low. Furthermore, ORi values may fluctuate if a pigmented drug is administered intravenously. There is a report showing that ORi rapidly decreased after indigo carmine was administered intravenously [56]. Another report showed that when indocyanine green (ICG) was administered to 10 patients, ORi increased in all of the patients [57]. The authors of this study suggested that it is hence important to take into account an increase in ORi when using ICG during surgery. ORi values can also fluctuate for other unknown reasons, and hence it is also important to make a comprehensive assessment of the patient’s condition and possible effects. In addition, the Radical-7 software version for the calculation of ORi and the RD rainbow Lite SET sensor were updated in January 2018, and although the peak is different for each individual, a plateau is reached at about 1.00 in most cases, and the artifact elimination function has been improved compared with the previous version. It is hence important to note that the obtained ORi values may differ depending on the version of the device used, so care must be taken when interpreting the results of previous studies.

Fig. 4
figure 4

ORi function built into Root® with Radical-7®. a ORi, b SpO2, c blockage of light is necessary to measure ORi using a SpHb sensor. ORi oxygen reserve index, SpO2 percutaneous oxygen saturation

Outlooks of ORi

As mentioned above, in recent years, ORi has been used as an index of oxygenation during surgery. However, in the near future, ORi may also be used in situations in which changes in oxygenation occur frequently, such as in patients in the ICU and general wards. Yoshida et al. suggested the usefulness of ORi for patients in the ICU, when suctioning through an endotracheal tube is performed during ventilation [58]. Furthermore, the fact that the ORi device is noninvasive and easy-to-wear will enable its use in various clinical situations. Recently, COVID-19 has become a worldwide epidemic. ORi might be useful to prevent infections, such as COVID-19, by reducing the frequency of contact of clinical workers with patient blood when performing blood gas analysis or securing an arterial line. The use of ORi will not only enable the early prediction of hypoxemia, but may also be useful for avoiding hyperoxia. In recent years, there have been reports that the mortality rate of patients managed for hyperoxia is high [59]. Thus, in the near future, ORi monitoring may become an effective method to lower the adverse effects caused by high oxygen concentrations. ORi not only determines the optimal inhaled oxygen concentration during the perioperative period, but also acts as an index of proper oxygenation, and has the potential to help avoid complications caused by hypoxemia and hyperoxia. Studies on how the use of ORi can change patient prognoses should be performed in the near future.