Abstract
Background
We evaluated the validity of calculations for refractive outcome in cataract surgery in silicone oil-filled eyes. The retrosilicone space (RSS) was included in these calculations.
Methods
In a prospective study the axial length (AL) of silicone oil-filled eyes was measured. with standardized A-scan echography (SAE) and partial coherence interferometry (PCI). Meldrum’s formula was used to transform the velocity of ultrasound within the vitreous cavity. To investigate whether refractive outcome can be calculated accurately, we assessed the difference between precalculated and final refractive outcome. Furthermore, we determined the advantages and disadvantages of SAE and PCI. A minor aim was to assess whether theAL of the two eyes differed significantly.
Results
In 85% of 117 eyes the difference between precalculated and postsurgical refraction was smaller than 1 diopter spherical and statistically not significant (p>0.2). The mean AL was 24.1 mm (range 20.0–31.4 mm). The difference in outcome between the two methods was without statistical significance: the AL difference was 0.4 (±2.6) mm on measurement with SAE and 0.04 (±0.46) mm with PCI. PCI has the advantage that it can be performed more easily, without contact, while echography is advantageous in the presence of advanced cataracts. In supine position an oil-free fluid space behind the silicone oil was detected with echography. The mean dimension of this space was 1.9 (±0.67) mm and it was taken into consideration for IOL calculation. The mean AL difference between the two eyes was 0.4 mm, but the difference was greater than 1 mm in 26% of the patients.
Conclusion
The AL of eyes filled with silicone oil can be measured reliably with SAE and PCI. In supine position the RSS has to be considered to obtain more accurate IOL calculations.
Similar content being viewed by others
References
Byrne SF, Green RL (1992) Axial eye length measurements. In: Ultrasound of the eye and orbit. Mosby, St Louis, pp 214–242
Dick B, Pavlovic S, Jacobi FK, Eisenmann D, Jacobi KW (1996) Objektive Bestimmung der Refraktionsänderungen bei silikongefüllten Augen. Einfluss der Kopfposition auf die Refraktion. Klien Monatsbl Augenheilkd 208:459–466
Drexler W, Findl O, Menapace R (1998) PCI: a novel approach to biometry in cataract surgery. Am J Ophthalmol 126:524–534
Effert R, Lommatzsch A, Wessing A (1996) Klinische Erfahrungen nach Implantation verschiedener Linsentypen bei Silikonöltamponade. Klin Monatsbl Augenheilkd 208:467–471
Ghoraba HH, El-Dorghamy AA, Atia AF, Ismail-Yassin AA (2002) The problems of biometry in combined silicone oil removal and cataract extraction. Clinical trial. Retina 22:589–596
Hauff W (1983) Die Biometrie—eine exakte Methode zur Messung der Achsenlänge des Auges. Wien Klin Wochenschr 95:271–274
Hitzenberger CK, Drexler W, Baumgartner A, Lexer F, Sattmann H, Esslinger M, Kulhavy M, Fercher AF (1997) Optical measurements of intraocular distances: a comparison of methods. Lasers Light Ophthalmol 8:85–95
Lai PC, Savage HI, Payman, AS (2003) Refractive surprise after contact ultrasonography. ARVO Annual Meeting Abstract Search and Program Planner, Abstract 212
Lege BA, Haigs W (2004) Laser interference biometry versus ultrasound biometry in certain clinical conditions. Graefes Arch Clin Exp Ophthalmol 242(1):8–12
Lucke K, Laqua H (1990) Anatomic and functional success. In: Silicone oil in the treatment of complicated retinal detachments: techniques, results and complications. Springer, Berlin Heidelberg New York, pp 45–48
Meldrum ML, Aaberg TM, Patel A, Davis J (1996) Cataract extraction after silicone oil repair of retinal detachments due to necrotizing retinitis. Arch Ophthalmol 114:885–892
Moisseiev J, Bartov E, Cahane M, Blumenthal M, Treister G (1992) Cataract extraction in eyes filled with silicone oil. Arch Ophthalmol 110:1649–1651
Murray DC, Durrani OM, Good P, Benson MT, Kirby GR (2002) Biometry of the silicone oil-filled eye. II. Eye 16 (6):727–730
Ossoinig K-C (1979) Standardized echography: basic principles, clinical applications and results. Int Ophthalmol Clin 19:127–210
Ossoinig KC (1984) How to obtain maximum measuring accuracies with standardized A-scan. In: Hillman JS, Le May MM (eds) Ophthalmic ultrasonography. Documenta ophthalmologica. Proceedings series, vol 38. Junk, The Hague, pp 197–216
Smith RC, Smith GT, Wong D (1990) Refractive changes in silicone oil filled eyes. Eye 4:230–234
Stefansson E, Tiedemann J (1988) Optics of the eye with air or silicone oil. Retina 8:10–19
Takei K, Sekine Y, Okamoto F, Hommura S (2002) Measurement of axial length of eyes with incomplete filling of silicone oil in the vitreous cavity using x ray computed tomography. Br J Ophthalmol 86:47–50
Tehrani M et al (2003) Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg 29(4):741–746
Till P (1976) Solid tissue model for the standardization of the echo ophthalmograph 7200 MA (Kretztechnik) Doc Ophthalmol 41:205–240
Till P (1980) Testung von Schallköpfen auf ihre Eignung zur Gewebsdifferenzierung mit Hilfe des Festkörper-Gewebsphantoms. Klin Monatsbl Augenheilkd 176:337–340
Verbeek AM, Bayer AL, Thijssen JM (1981). Echographic diagnosis after intraocular silicone oil injection. Doc Ophthalmol Proc Ser 29:59–66
Acknowledgement
Translations support was done carefully by Mrs. MT Untersmayr.
Author information
Authors and Affiliations
Corresponding author
Additional information
There was no private or public financial support for this study.
The authors have no proprietary or financial interest in any product mentioned in this article.
Rights and permissions
About this article
Cite this article
Nepp, J., Krepler, K., Jandrasits, K. et al. Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry. Graefe's Arch Clin Exp Ophthalmo 243, 967–972 (2005). https://doi.org/10.1007/s00417-004-1117-2
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00417-004-1117-2