Background

We evaluated the validity of calculations for refractive outcome in cataract surgery in silicone oil-filled eyes. The retrosilicone space (RSS) was included in these calculations.

Methods

In a prospective study the axial length (AL) of silicone oil-filled eyes was measured. with standardized A-scan echography (SAE) and partial coherence interferometry (PCI). Meldrum’s formula was used to transform the velocity of ultrasound within the vitreous cavity. To investigate whether refractive outcome can be calculated accurately, we assessed the difference between precalculated and final refractive outcome. Furthermore, we determined the advantages and disadvantages of SAE and PCI. A minor aim was to assess whether theAL of the two eyes differed significantly.

Results

In 85% of 117 eyes the difference between precalculated and postsurgical refraction was smaller than 1 diopter spherical and statistically not significant (p>0.2). The mean AL was 24.1 mm (range 20.0–31.4 mm). The difference in outcome between the two methods was without statistical significance: the AL difference was 0.4 (±2.6) mm on measurement with SAE and 0.04 (±0.46) mm with PCI. PCI has the advantage that it can be performed more easily, without contact, while echography is advantageous in the presence of advanced cataracts. In supine position an oil-free fluid space behind the silicone oil was detected with echography. The mean dimension of this space was 1.9 (±0.67) mm and it was taken into consideration for IOL calculation. The mean AL difference between the two eyes was 0.4 mm, but the difference was greater than 1 mm in 26% of the patients.

Conclusion

The AL of eyes filled with silicone oil can be measured reliably with SAE and PCI. In supine position the RSS has to be considered to obtain more accurate IOL calculations.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Byrne SF, Green RL (1992) Axial eye length measurements. In: Ultrasound of the eye and orbit. Mosby, St Louis, pp 214–242

    Google Scholar 

  2. Dick B, Pavlovic S, Jacobi FK, Eisenmann D, Jacobi KW (1996) Objektive Bestimmung der Refraktionsänderungen bei silikongefüllten Augen. Einfluss der Kopfposition auf die Refraktion. Klien Monatsbl Augenheilkd 208:459–466

    CAS  Google Scholar 

  3. Drexler W, Findl O, Menapace R (1998) PCI: a novel approach to biometry in cataract surgery. Am J Ophthalmol 126:524–534

    Article  PubMed  CAS  Google Scholar 

  4. Effert R, Lommatzsch A, Wessing A (1996) Klinische Erfahrungen nach Implantation verschiedener Linsentypen bei Silikonöltamponade. Klin Monatsbl Augenheilkd 208:467–471

    PubMed  CAS  Google Scholar 

  5. Ghoraba HH, El-Dorghamy AA, Atia AF, Ismail-Yassin AA (2002) The problems of biometry in combined silicone oil removal and cataract extraction. Clinical trial. Retina 22:589–596

    Article  PubMed  Google Scholar 

  6. Hauff W (1983) Die Biometrie—eine exakte Methode zur Messung der Achsenlänge des Auges. Wien Klin Wochenschr 95:271–274

    PubMed  CAS  Google Scholar 

  7. Hitzenberger CK, Drexler W, Baumgartner A, Lexer F, Sattmann H, Esslinger M, Kulhavy M, Fercher AF (1997) Optical measurements of intraocular distances: a comparison of methods. Lasers Light Ophthalmol 8:85–95

    Google Scholar 

  8. Lai PC, Savage HI, Payman, AS (2003) Refractive surprise after contact ultrasonography. ARVO Annual Meeting Abstract Search and Program Planner, Abstract 212

  9. Lege BA, Haigs W (2004) Laser interference biometry versus ultrasound biometry in certain clinical conditions. Graefes Arch Clin Exp Ophthalmol 242(1):8–12

    Article  PubMed  CAS  Google Scholar 

  10. Lucke K, Laqua H (1990) Anatomic and functional success. In: Silicone oil in the treatment of complicated retinal detachments: techniques, results and complications. Springer, Berlin Heidelberg New York, pp 45–48

    Google Scholar 

  11. Meldrum ML, Aaberg TM, Patel A, Davis J (1996) Cataract extraction after silicone oil repair of retinal detachments due to necrotizing retinitis. Arch Ophthalmol 114:885–892

    PubMed  CAS  Google Scholar 

  12. Moisseiev J, Bartov E, Cahane M, Blumenthal M, Treister G (1992) Cataract extraction in eyes filled with silicone oil. Arch Ophthalmol 110:1649–1651

    PubMed  CAS  Google Scholar 

  13. Murray DC, Durrani OM, Good P, Benson MT, Kirby GR (2002) Biometry of the silicone oil-filled eye. II. Eye 16 (6):727–730

    Article  PubMed  CAS  Google Scholar 

  14. Ossoinig K-C (1979) Standardized echography: basic principles, clinical applications and results. Int Ophthalmol Clin 19:127–210

    Article  PubMed  CAS  Google Scholar 

  15. Ossoinig KC (1984) How to obtain maximum measuring accuracies with standardized A-scan. In: Hillman JS, Le May MM (eds) Ophthalmic ultrasonography. Documenta ophthalmologica. Proceedings series, vol 38. Junk, The Hague, pp 197–216

    Google Scholar 

  16. Smith RC, Smith GT, Wong D (1990) Refractive changes in silicone oil filled eyes. Eye 4:230–234

    PubMed  Google Scholar 

  17. Stefansson E, Tiedemann J (1988) Optics of the eye with air or silicone oil. Retina 8:10–19

    PubMed  CAS  Google Scholar 

  18. Takei K, Sekine Y, Okamoto F, Hommura S (2002) Measurement of axial length of eyes with incomplete filling of silicone oil in the vitreous cavity using x ray computed tomography. Br J Ophthalmol 86:47–50

    Article  PubMed  Google Scholar 

  19. Tehrani M et al (2003) Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg 29(4):741–746

    Article  PubMed  Google Scholar 

  20. Till P (1976) Solid tissue model for the standardization of the echo ophthalmograph 7200 MA (Kretztechnik) Doc Ophthalmol 41:205–240

    Article  PubMed  CAS  Google Scholar 

  21. Till P (1980) Testung von Schallköpfen auf ihre Eignung zur Gewebsdifferenzierung mit Hilfe des Festkörper-Gewebsphantoms. Klin Monatsbl Augenheilkd 176:337–340

    Article  PubMed  CAS  Google Scholar 

  22. Verbeek AM, Bayer AL, Thijssen JM (1981). Echographic diagnosis after intraocular silicone oil injection. Doc Ophthalmol Proc Ser 29:59–66

    Google Scholar 

Download references

Acknowledgement

Translations support was done carefully by Mrs. MT Untersmayr.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Johannes Nepp.

Additional information

There was no private or public financial support for this study.

The authors have no proprietary or financial interest in any product mentioned in this article.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Nepp, J., Krepler, K., Jandrasits, K. et al. Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry. Graefe's Arch Clin Exp Ophthalmo 243, 967–972 (2005). https://doi.org/10.1007/s00417-004-1117-2

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-004-1117-2

Keywords

Navigation