FormalPara Key Summary Points

The potential refractive prediction error based on alternative A-scan ultrasound and fellow-eye biometry for phacovitrectomy was analysed in macula-off RRD eyes without the preoperative IOLMaster-measured axial length (AL).

The alternative preoperative AL measured with A-scan ultrasound would be significantly underestimated in this special situation, while the AL of the fellow eye is more similar to the AL in eyes filled with silicone oil tamponade after RRD repair.

The calculated IOL power based on A-scan ultrasound was significantly greater than that based on data measured in silicone oil-filled eyes.

A real-world series of 9 eyes that faced the same preoperative situation but underwent phacovitrectomy based on fellow-eye biometry showed a postoperative myopic shift of 0.64 ± 0.78 D.

Introduction

Rhegmatogenous retinal detachment (RRD) is a common retinal disease with an incidence of one in 10,000 people per year that often causes visual field defects and moderate to severe visual impairment [1]. Surgery is the only therapeutic approach, and surgical techniques and instrumentation developed in recent decades have led to a high primary reattachment rate [2]. Indeed, pars plana vitrectomy (PPV) is one of the most effective procedures for the treatment of RRD, with a mean postoperative anatomical success rate of over 90% [3]. However, PPV has a higher incidence of cataract progression than scleral buckling. A variety of mechanisms might contribute to cataract progression: surgery can induce an inflammatory response, as demonstrated by an increase in aqueous flare values; or, in the absence of vitreous gel, molecular oxygen from the retinal vasculature can reach the lens and promote oxidative damage to the lens nucleus, an increase in light scattering, and nuclear sclerotic cataracts [4]. Due to postoperative complications involving the crystalline lens, more than half of phakic patients who underwent vitrectomy for RRD repair subsequently needed cataract surgery within 1 year [5]. Thus, combined phacovitrectomy was encouraged for patients aged 50 or older due to the lower cost of surgery, reduced complications and better visual outcome [6,7,8,9]. In 10–17% of cases of RRD with macular involvement, it was impossible to measure the axial length (AL) using optic biometry (IOLMaster) because of the limitations of the machine, such as dense media opacity, poor fixation by the patients or the lack of an available machine [10,11,12]. Rahman et al. suggested that AL measured by acoustic biometry (A-scan ultrasound, US) could provide an alternative intraocular lens (IOL) power calculation [12], and El-Khayat et al. reported that fellow-eye biometry is also recommended [13], although, to our knowledge, there has been no published report on outcomes. On the other hand, studies (including our previous work) have shown that both underestimation of the AL and anterior displacement of the IOL potentially contribute to a postoperative myopic shift after phacovitrectomy [14, 15]. Most studies that have focused on myopic shifts after phacovitrectomy have excluded cases in which the preoperative IOLMaster measurement was absent. Therefore, we conducted the present study to focus on this commonly ignored but not rare situation.

Here, we assessed the agreement in AL and predicted IOL power of macula-off RRD eyes among preoperative US, postoperative IOLMaster and fellow eyes in order to evaluate the potential refractive errors after phacovitrectomy based on alternative US or contralateral measurements when preoperative IOLMaster data were absent.

Methods

Design and Participants

This retrospective, self-controlled study was conducted in Shanghai Jing'an District Shibei Hospital between September 2018 and December 2020. All study procedures were performed in accordance with the tenets of the Declaration of Helsinki and its amendments. The ethics committee of Shanghai Jing’an District Shibei Hospital approved this study. Written informed consent was obtained from all the participants for publication of their clinical data.

Phakic macula-off RRD eyes in which ocular biometric measurements could not be obtained using IOLMaster were included; the eyes underwent 23-gauge PPV with silicone oil (SO) tamponade but not lens extraction. Eyes with a history of anisometropia and eyes with scleral buckling surgery, recurrent retinal detachment or other ocular problems that may have affected biometric measurements, including corneal scarring and lens dislocation, were excluded [16].

Other patients who faced the same preoperative situation but underwent phacovitrectomy based on fellow-eye biometry for RRD repair were recruited as controls at least 3 months after SO removal. During follow-up, after excluding those with recurrent RRD or other postoperative complications, comprehensive refractive measurements were performed to obtain the spherical dioptre. The mean spherical error was defined as the difference in spherical dioptre between the postoperative and expected refractive outcomes.

Surgical Procedure

A standard 23-gauge PPV using the CONSTELLATION® Vision 106 System (Alcon Laboratories, Inc.) was performed in each case. During the PPV, the RESIGHT™ Fundus Viewing System (Carl Zeiss Meditec Inc.) was used. Core vitrectomy, midperipheral vitrectomy and vitreous base shaving under scleral depression were performed to remove the vitreous. Perfluorocarbon liquid (Perfluoron; Alcon Laboratories, Inc.) was used in some cases, depending on the extent of retinal detachment. Endolaser photocoagulation was performed around areas with retinal breaks, and fluid–air exchange was performed before silicone oil injection (Oxane 5700 centistokes; Bausch & Lomb Inc., Waterford, Ireland).

Measurement Techniques

Preoperative AL values of RRD eyes were measured using A-scan ultrasound biometry (UD-6000 Ultrasonic A/B scanner biometer; Tomey Corporation, Nagoya, Japan), as previously described [17], instead of using IOLMaster. Before SO removal, an optical coherence tomography scan was performed to ensure that the macula was attached in both eyes, and then the AL values of postoperative RRD eyes and fellow eyes were obtained with IOLMaster (Model 700, Carl Zeiss Meditec Ltd, Jena, Germany) using phakic and SO-filled phakic eye programs, respectively. All the AL measurements performed in this study were repeated three times by experienced examiners. The IOL power was calculated based on the constant A of the AcrySof® IQ ReSTOR® SN6AD1 IOL (Alcon Laboratories, Inc., Fort Worth, USA) with the SRK/T formula.

Statistical Analysis

Continuous variables are expressed as the mean ± standard deviation. Statistical analyses were performed using the software package SPSS Statistics 26.0 for Windows (SPSS Inc., Chicago, IL, USA). Bland–Altman plots were drawn using MedCalc 15.2.2 (MedCalc Software bvba, Ostend, Belgium). Statistical significance was set at P < 0.05.

Results

Data from 22 eyes of 22 patients (12 males and 10 females) with an age range from 36 to 65 years (mean: 51.6 ± 8.3 years) were enrolled. The mean duration of SO tamponade was 4.3 ± 0.8 months. The mean AL in eyes with RRD measured by A-scan ultrasound (AL-US) was 25.39 ± 2.14 mm (22.59–31.53 mm), the mean AL of fellow eyes (AL-FE) was 25.85 ± 2.16 mm (22.47–31.56 mm), and the mean AL in eyes with SO tamponade after RRD repair (AL-SO) was 26.08 ± 2.53 mm (22.75–32.81 mm). Independent-samples t tests revealed that there was no significant difference in AL-US, AL-FE or AL-SO between the sexes (all P > 0.05).

The mean IOL power calculated using AL-US (Power-US) was 16.81 ± 7.19 dioptres (D), the mean IOL power calculated using AL-FE (Power-FE) was 14.74 ± 6.95 D, and the mean IOL power calculated using AL-SO (Power-SO) was 13.54 ± 8.32 D. Bland–Altman plots demonstrated that the agreement between AL-US and AL-SO was good since 95.5% (21/22) of samples were included in the 95% limits of agreement (LoA), and the agreement between AL-FE and AL-SO (95.5%, 21/22 in LoA) was similar, as shown in Fig. 1A and B.

Fig. 1
figure 1

Bland–Altman plots for AL-US and AL-SO (A) and for AL–FE and AL-SO (B). The bold solid line indicates the mean difference; the dotted lines represent the 95% limits of agreement (LoA), and the thin solid lines represent the 95% CI of the upper/lower limits of the LoA. Differences in axial length (C) and IOL power estimation (D) among RRD eyes, SO-filled eyes and fellow eyes (repeated-measures ANOVA, *P < 0.05). AL-US axial length of RRD eye measured by A-scan, AL-SO axial length of eye with silicone oil tamponade for RRD repair, AL–FE axial length of fellow eye, Power-US estimated IOL power of RRD eye measured by A-scan, Power-SO estimated IOL power of eye with silicone oil tamponade for RRD repair, Power-FE estimated IOL power of fellow eye

However, repeated-measures ANOVA showed that the difference between AL-US and AL-SO was statistically significant (P < 0.05), while that between AL-FE and AL-SO was not (P > 0.05, Fig. 1C). Correspondingly, the difference between Power-US and Power-SO was statistically significant (P < 0.05), and the mean difference (Power-US − Power-SO) was 3.27 ± 3.79 D, while that between Power-FE and Power-SO was not significant (P > 0.05, Fig. 1D).

Nine eyes of 9 patients (4 males and 5 females) with a mean age of 58.1 ± 6.5 years were recruited as controls. They were also phakic macula-off RRD eyes without preoperative IOLMaster measurements. Standard 23-gauge phacovitrectomy with SO tamponade was performed for each of them, and the implanted IOLs were all AcrySof® IQ ReSTOR® SN6AD1 IOLs, for which the power was calculated with the SRK/T formula based on fellow-eye biometry. The mean follow-up period after SO removal was 8.9 ± 3.1 months. The mean spherical error of those 9 eyes showed a postoperative myopic shift of 0.64 ± 0.78 D.

Discussion

The AL of an eye, which is measured as the distance from the cornea to the inner limiting membrane using A-scan US or the distance from the cornea to retinal pigment epithelium using IOLMaster [18], is the most crucial parameter in the IOL power calculation. A 1-mm change in the AL corresponds to an approximately 2.7-D refractive error in the IOL power [19]. Nepp et al. reported that the mean difference between the US-measured AL and the IOLMaster-measured AL in SO-filled eyes was 0.4 mm, and only 46% of cases had a value of less than 0.3 mm, while the value was greater than 1 mm in 26% [20]. Thus, we measured the AL values of these SO-filled eyes and normal fellow eyes in the present study by means of IOLMaster to achieve more accuracy and less deviation [16]. Our results showed that the AL values of SO-filled eyes after RRD repair were comparable with those of fellow eyes, so it was not surprising that the IOL power calculated using AL-SO was similarly comparable to that calculated using AL-FE. However, there were significant differences in both AL and IOL power between US-measured preoperative data for macula-off RRD eyes and IOLMaster-measured postoperative data for SO-filled eyes. The mean AL-US was approximately 0.68 mm shorter than the mean AL-SO, and the mean Power-US was accordingly 3.27 D greater than the mean Power-SO. Thus, there would be a clinically significant myopic shift after SO removal if we performed phacovitrectomy for RRD repair based on the alternative A-scan-based IOL power calculation.

Phacovitrectomy is a safe and effective procedure to treat RRD. Kang et al. and Sakamoto et al. confirmed that there is a tolerable biometric error after the treatment of RRD with phacovitrectomy in macula-sparing cases, with comparable results obtained with acoustic and optic biometry [21, 22]. Although the anatomical and functional results were comparable with those obtained with single PPV and delayed cataract surgery, the refractive outcomes were less favourable and shifted towards myopia, especially in macula-off cases [23]. When the macula was involved, there was a mean prediction error of − 1.22 ± 2.32 D for RRD eyes and a mean prediction error of − 0.01 ± 1.09 D for fellow eyes [13]. Pongsachareonnont et al. attributed this difference to an underestimation of 0.59 ± 0.90 mm in AL measurements performed by IOLMaster [24]. Kim et al. found that the underestimation was associated with the macular retinal detachment height [11]. Unlike the above studies, the present study focused on a special situation. Rahman et al. found that IOLMaster could provide AL measurements in less than one-quarter (13/54, 24.1%) of the cases [10]. Patients with worse vision, greater central macular thickness and shallow anterior chambers required more caution since their eyes were prone to inaccurate preoperative biometry [25]. Since there were no preoperative IOLMaster data to use as a reference, we could not calculate a user-adjusted AL by combining acoustic and optic biometry as previously described [12]. Therefore, delayed cataract surgery or phacovitrectomy using contralateral AL might be recommended. In the present study, we also retrospectively reviewed 9 patients who faced the same preoperative situation but underwent phacovitrectomy based on fellow-eye biometry in a real-world setting, and the mean spherical error between their postoperative and expected refractive outcomes was − 0.64 ± 0.78 D. This postoperative myopic shift was expected and comparable with eyes reported in previous studies that underwent phacovitrectomy for RRD repair based on preoperative IOLMaster data.

The results of this study should be interpreted with its limitations in mind. All the participants we recruited underwent SO tamponade, but interference with biometry (such as underfilling/overfilling or emulsification of the SO) was not entirely quantified. Also, some characteristics of retinal detachment that might be associated with the accuracy of AL measurement, such as macular detachment height, were not controlled.

Conclusion

Based on the present study, if IOLMaster cannot be performed in eyes with macula-off RRD, the use of ultrasound might lead to a refractive prediction error in IOL power. Phacovitrectomy using fellow-eye biometry as well as secondary IOL implantation could be the better choice in this situation.