Measurements of diaphragm function by ultrasonography are affected by body position, but reference values in the seated position have not been established for an Asian population. This study aimed to determine reference values for diaphragm thickness, thickening fraction, and dome excursion by ultrasonography and to investigate the effects of sex, height, and body mass index.
Methods
Diaphragm ultrasonography was performed on 109 seated Japanese volunteers with normal respiratory function who were enrolled between March 2022 and January 2023. Thickness, thickening fraction, and excursion were measured. Reference values and the measurement success rate were calculated. Multivariate analysis adjusted for sex, height, and body mass index was performed.
Results
The measurement success rate was better for thickness than for excursion. The mean (lower limit of normal) values on the right/left sides were as follows. During quiet breathing, thickness at end expiration(mm) was 1.7 (0.9)/1.6 (0.80), thickening fraction(%) was 50 (0.0)/52 (0.0), and excursion(cm) was 1.7 (0.5)/1.9 (0.5). During deep breathing, the thickening fraction was 111 (24)/107 (22), and the excursion was 4.4 (1.7)/4.1 (2.0). In multivariate analysis, body mass index was positively associated with thickness but not with the thickening fraction.
Conclusion
The reference values in this study were smaller than those in previous reports from Europe. Considering that thickness is influenced by body mass index, using Western reference values in Asia, where the average body mass index is lower, might not be appropriate. The thickening fraction in deep breathing is unaffected by other items and can be used more universally.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Bilateral diaphragmatic dysfunction leads to symptoms such as dyspnea, chest pain, and sleep disturbance [1, 2]. Symptoms are influenced by body position, worsening in a supine position and improving when seated [1,2,3]. Diaphragmatic dysfunction may be caused by ventilator-induced injury, surgery, and respiratory, neurological, musculoskeletal, and neuromuscular junction diseases [2,3,4,5]. Ultrasonography has a sensitivity and specificity of more than 90% for diagnosis of diaphragmatic dysfunction, [2, 6,7,8] but is not a widely accepted assessment method. One potential cause for underutilization is that testing methods and reference values are not well established [2, 6, 8].
Two ultrasonographic methods can be used to measure diaphragmatic function: (1) measurement of diaphragmatic excursion during breathing and (2) measurement of the thickness of the diaphragm (Tdi) and its rate of change (i.e., the thickening fraction [TF]) during breathing. Both excursion and TF are influenced by body position. Excursion is greater in the supine position than in the seated position, whereas TF is greater in the seated position [8,9,10,11]. Therefore, there is a need to establish reference values according to the body position in which measurements are obtained. Most published reports refer to measurements obtained in the supine or semi-recumbent position, and studies of seated reference Tdi, TF, and excursion values are scarce [1, 8, 12]. In the outpatient setting, it is easier to obtain measurements in the seated position, which is preferred by patients with diaphragmatic dysfunction because of orthopnea. Establishing reference values in the seated position is important in promoting the use of diaphragm ultrasonography. Furthermore, most reports on reference values are based on Western populations, which may not be appropriate for Asians, who generally have a different physical build. The only relevant reports involving Asians are based on healthy populations with sample sizes of fewer than 30 individuals or patients with a history of stroke or spinal cord injury [13,14,15,16]. There has not been a large-scale report on reference values for healthy Asians.
This study aimed to determine reference values for Tdi, TF, and excursion of the diaphragm dome obtained by ultrasonography in healthy seated Japanese volunteers and to investigate the effects of sex, height, and body mass index (BMI) on the measurements obtained.
Methods
Study Population and Setting
Healthy adult volunteers were recruited from Tokyo Medical and Dental University Hospital (Tokyo), Iizuka Hospital (Fukuoka), and Shinai Clinic (Kanagawa), between March 2022 and January 2023 in Japan. General information such as age, sex, height, weight, and medical history was collected on the day of the examination. Respiratory function tests were performed using a spirometer (AutoSpiro507; Minato Medical Science Co., Ltd, Osaka, Japan). Volunteers with no symptoms, a percent vital capacity ≥ 80, and a forced expiratory volume in one second (FEV1)/forced vital capacity (FVC) of ≥ 70% were included. All measurements were obtained by ultrasonography technicians or physicians who were certified as instructors in point-of-care ultrasound simulation courses [17].
Thickness, Thickening Fraction, and Excursion
In the seated position, Tdi was measured using a linear transducer (7.0 MHz) placed vertically around the eighth to ninth intercostal space, aligned perpendicular to the chest, avoiding ribs. The height was set to a position where a portion of the lung could be observed at the edge of the screen during inspiration. Tdi was measured at end expiration and end-inspiration during quiet breathing (QB) and deep breathing (DB). The method is depicted in Fig. 1. TF was calculated using the following formula:
Measurement of diaphragm thickness. The diaphragm was observed as a three-layer structure, consisting of a hypoechoic muscular layer sandwiched between two bright white lines (the pleural and peritoneal membranes), with a white line running in the center (considered to be the fibrous layer in the center of the diaphragm). *Measured in B-mode by placing a measurement marker from the center of the thickness of the pleural membrane to the center of the thickness of the peritoneal membrane
end-inspiration Tdi (total lung capacity [TLC]) – end-expiration Tdi (functional residual capacity [FRC])/end-expiration Tdi (FRC) × 100.
Excursion was measured on both sides during QB and DB. The phased-array transducer (2.5 MHz) was placed vertically around the eighth to ninth intercostal space, aligned perpendicular to the chest, avoiding ribs. The M-mode line was adjusted to be as perpendicular to the dome as possible, and its excursion during inspiration was measured. (Fig. 2).
One examiner measured all sites for each subject. Tdi measurement succeeded if both TLC and FRC values were obtained in one breath. Dome excursion was successful if diaphragm movement was shown in one respiratory cycle.
Statistical Analysis
Subject characteristics are shown as mean and standard deviation for continuous data, and frequency and percentage for categories. The lower limit of normal (LLN) and upper limit of normal (ULN) were calculated as mean ± 1.95 standard deviations. Diaphragm measurements’ left-right and sex differences were compared using paired and two-sample t-tests, respectively. Multiple regressions assessed associations with sex, height, and BMI. Analyses were done using STATA 17.0 (StataCorp LLC, College Station, TX, USA). A p-value < 0.05 was considered statistically significant.
Results
Study Participants
A total of 111 Japanese volunteers were recruited during the 10-month study period. Two subjects in whom respiratory function tests revealed a percent vital capacity < 80% were excluded. The remaining 109 subjects were included in the analysis. In terms of history of respiratory or neuromuscular disease, there were two cases of bronchial asthma, but neither showed any symptoms of asthma on the day of examination; both had normal FEV1/FVC and percent vital capacity values. The subject characteristics are shown in Table 1.
Measurement Success rate
Tdi was measurable in all cases with a 100% success rate. The success rate of dome excursion was generally lower on the left, especially during DB, mainly because the diaphragm was obscured by the lung. Impaired observation owing to fat attenuation was not a significant issue. Table 2 shows the success rates at each measurement site and the reasons for measurement failure.
Thickness, Thickening Fraction, and Excursion
Mean Tdi at functional residual capacity (FRC) was significantly thicker on the right side than on the left side. Furthermore, both sides were significantly thicker in men than in women. There was no significant difference in TF between the left and right sides, nor any significant sex-related differences. There was a statistically significant difference in diaphragmatic excursion between the left and right sides during QB but not during DB. Movement of the diaphragm was greater in men during DB. The mean, LLN, and ULN values for each parameter are presented in Tables 3 and 4.
Multiple Linear Regression Analyses
Multiple linear regression analyses were performed with adjustment for sex, height, and BMI. Significant associations of BMI with left and right Tdi, TF of the left diaphragm during QB, and excursion of the right diaphragm during DB were observed. In DB, a correlation was observed between the right excursion and both gender and BMI, but no correlation was found for TF with gender, height, or BMI. All results of analysis are shown in Table 5.
Discussion
This study determined seated reference values for Tdi, TF, and excursion in 109 healthy Japanese individuals. This is the largest-ever study to include simultaneous measurement of Tdi, TF, and excursion in seated Asian subjects and provides a benchmark for normal seated values in the Asian population.
The success rate for measurement of Tdi and TF was 100% regardless of side or breathing pattern. The main problem in measuring excursion was the overlapping of the lungs during inspiration, but fat attenuation was not a significant problem. Although diaphragm ultrasonography is more helpful in obtaining measurements during DB than during QB, [8] the success rate for excursion in DB was lower than that for Tdi and TF measurements. Therefore, measurement of TF, which has a higher success rate during DB and is less influenced by sex, height, and BMI, might be more useful than measurement of excursion.
Previous reports have indicated that it is easier to measure excursion than TF [6]. However, our study yielded different results. This inconsistency might be explained by differences in BMI between studies. The mean BMI in our study was 22.1, and the percentage of subjects with BMI ≥ 30 was 1%. Given that subcutaneous fat can hinder visibility with the linear transducer used for measuring TF, our high success rate when measuring Tdi may reflect a low obesity rate. The average BMI varies markedly from country to country. For example, the average adult BMI in Japan is 23.6 for men and 21.8 for women, 28.8 and 28.9 in the US, and 27.3 and 27.0 in the UK [18]. In Japan, only 5.4% of men and 3.6% of women have a BMI ≥ 30; these rates are considerably lower than those in Western countries [18, 19]. In populations where obesity is not a significant issue, as in our study, it might be advisable to first consider measuring the TF.
Past reports of Tdi at FRC range from 1.6 to 3.8 mm in men and 1.4 to 2.7 mm in women [1, 8, 13, 14, 20,21,22,23,24]. The results of the present study are consistent with previous reports. Our results for the LLN in men are close to those of Cardenas et al. and Boussuges et al.; however, the mean Tdi in women was more than 40% less than in the earlier studies [1, 20]. One of the criteria for a diagnosis of diaphragmatic weakness is a Tdi of < 2 mm at end expiration [2, 8, 25]; however, considering past reports and our present findings, many healthy individuals fall below this standard, suggesting that this criterion may be inappropriate for diagnosing diaphragmatic paralysis.
Our finding that the diaphragm was significantly thicker in men than in women is consistent with past reports [1, 8, 24, 26, 27]. However, in multiple linear regression analysis adjusted for sex, height, and BMI, there was no association with sex (right, p = 0.1; left, p = 0.92), but there was a statistically significant relationship with BMI (right, p < 0.01; left, p < 0.01). Other studies have also reported Tdi at FRC showed positive correlations with BMI [1, 26, 28]. Our results support these findings regarding the correlation with BMI but not with sex. It is unlikely that our failure to find a statistically significant correlation with sex was attributable to an insufficient sample size, given that the size of our study population was adequate when compared with those in the previous studies. While it appears that the diaphragm is generally thicker in men than in women, Tdi may be influenced more by BMI than by sex. As previously mentioned, the average BMI varies widely by country [18]. Considering that Tdi is influenced by BMI, standard values developed in Western countries, where the average BMI is significantly higher than that in Asian countries, may be inappropriate for use in Japan and other Asian populations.
During QB, even healthy individuals may have a change in Tdi of < 10% or no change at all [1, 27]. In our study, the TF was < 20% during QB in two cases (2%) on the right and in six cases (6%) on the left, and there was one case (1%) with a TF < 10% on the left. Therefore, TF in DB is more useful than TF in QB for evaluating diaphragmatic function and diagnosing diaphragm paralysis [1, 27].
TF is markedly affected by body position, with changes increasing in the order of supine, seated, and standing [10, 11]. In studies that examined changes in TF according to body position, the average TF values in DB were 60% in the supine position, 97% in the seated position, and 174% in the standing position [11]. Therefore, our results should be compared with previously reported measurements obtained in the seated position. However, there are limited data on normal seated TF values. In the only large-scale study of seated TF [1], which included 200 healthy individuals, the mean TF (LLN) in DB was 106% (40%)/112% (39%) for men and 116% (39%)/121% (48%) for women, with no apparent sex differences. Our mean values are close to those in the previous report. Although we calculated the LLN in the same way as the previous report [1] (i.e., the mean ± a standard deviation of 1.95), our results were about 20% lower than in their report.
A value of TF should be considered as indicating dysfunction if it is greater than the criterion for complete paralysis (TF < 20%) and less than the LLN [2]. However, a consensus has not yet been reached on the cut-off value for dysfunction [6]. Considering that the LLN was in the 20% range in our study, the original criterion for diaphragmatic paralysis (TF < 20%) was calculated from data obtained in the standing position, [25] and the fact that the seated TF is smaller than the standing TF [11] suggests there may be room for reconsideration of application of a TF < 20% as a criterion for diagnosis of diaphragmatic paralysis when measurements are obtained in the seated position.
Our multiple linear regression analysis adjusted for sex, height, and BMI found no association of the TF in DB with any of these three parameters. Tdi is affected by BMI, so it is difficult to refer to Western values in Asians, whose physique differs from that of Westerners. TF in DB is less affected by BMI and other factors, it may be used as a more generalized value.
Diaphragmatic dome excursion is also affected by body position, but, unlike TF, becomes larger in the supine position than in the seated position [8, 9]. Since abdominal compliance is less when seated, diaphragm dome excursion may be less for a given degree of diaphragm activation or tidal volume. Therefore, our results should be compared with those previously obtained in the seated position. However, there are few relevant reports. In the only large-scale study, Boussuges et al. measured excursion in the seated position and reported that the mean (LLN) values in QB were 1.9 cm (0.9 cm) on the right and 2.0 cm (0.9 cm) on the left in men and 1.7 cm (0.9 cm) and 1.7 cm (0.9 cm), respectively, on the right in women. In DB, the mean (LLN) values were 6.6 cm (4.1 cm) on the right and 6.7 cm (4.2 cm) on the left in men and 5.4 cm (3.3 cm) and 5.4 cm (3.2 cm), respectively, in women [12]. Our findings were equivalent to those of Boussuges et al. in QB, but were 1.5–2.0 cm smaller in DB in both sexes. Moreover, the LLN was 40–60% smaller in both QB and DB for both sexes. Multiple linear regression analysis identified a statistically significant association of excursion with female sex (p < 0.01) and BMI (p = 0.03) on the right side in DB but no association on the left side. The lack of a significant association on the left may reflect the inadequate sample size caused by the low rate of successful depiction. One reason why our excursion values in DB were smaller than those in the study by Boussuges et al. could be that the mean BMI was lower in our study subjects than in those in their study (22.5 vs. 25.0 for men and 21.7 vs. 25.0 for women). [12] It has been proposed that the criterion for diaphragm dysfunction with excursion in QB should be < 2 cm [6]. Considering past findings and our present results and that seated excursion is smaller than supine excursion, [8, 9] another criterion should be considered for the seated position.
This study has several limitations. The success rate for measurement of Tdi was higher than that for excursion, which might reflect the lower obesity rate in our study population. Therefore, it is unclear whether our results are applicable to populations with higher obesity rates. Further study is needed on a wider range of age groups, body size, and races.
In summary, we have determined reference values for use when performing diaphragm echography in a seated position. Considering that Tdi and excursion are influenced by BMI, it may not be possible to directly apply the reference values developed for the West in Asian countries, which have different obesity rates. The TF in DB does not appear to be affected by sex, height, or BMI and could be utilized more widely than other criteria. The current standards for diagnosis of diaphragmatic paralysis may need to be reconsidered for use in a seated position, given that many healthy individuals meet the criteria when measured in this manner.
Abbreviations
- Tdi:
-
thickness of the diaphragm
- BMI:
-
body mass index
- DB:
-
deep breathing
- FEV1 :
-
forced expiratory volume in one second
- FRC:
-
functional residual capacity
- FVC:
-
forced vital capacity
- LLN:
-
lower limit of normal
- QB:
-
quiet breathing
- TF:
-
thickening fraction
- TLC:
-
total lung capacity
- ULN:
-
upper limit of normal
References
Boussuges A, Rives S, Finance J et al (2021) Ultrasound assessment of diaphragm thickness and thickening: reference values and limits of normality when in a seated position. Front Med (Lausanne) 8:742703
McCool FD, Tzelepis GE (2012) Dysfunction of the diaphragm. N Engl J Med 366(10):932–942
McCool FD, Manzoor K, Minami T (2018) Disorders of the diaphragm. Clin Chest Med 39(2):345–360
Minami T, Manzoor K, McCool FD (2018) Assessing diaphragm function in chest wall and neuromuscular Diseases. Clin Chest Med 39(2):335–344
Perrin C, Unterborn JN, Ambrosio CD, Hill NS (2004) Pulmonary Complications of chronic neuromuscular Diseases and their management. Muscle Nerve 29(10):5–27
Haaksma ME, Smit JM, Boussuges A et al (2022) EXpert consensus on Diaphragm UltraSonography in the critically ill (EXODUS): a Delphi consensus statement on the measurement of diaphragm ultrasound-derived parameters in a critical care setting. Crit Care 26(1):99
Li C, Li X, Han H, Cui H, Wang G, Wang Z (2018) Diaphragmatic ultrasonography for predicting ventilator weaning: a meta-analysis. Med (Baltim) 97(22):e10968
Boussuges A, Rives S, Finance J, Brégeon F (2020) Assessment of diaphragmatic function by ultrasonography: current approach and perspectives. World J Clin Cases 8(12):2408–2424
Houston JG, Angus RM, Cowan MD, McMillan NC, Thomson NC (1994) Ultrasound assessment of normal hemidiaphragmatic movement: relation to inspiratory volume. Thorax 49(5):500–503
Hellyer NJ, Andreas NM, Bernstetter AS et al (2017) Comparison of diaphragm thickness measurements among postures via ultrasound imaging. PM R 9(1):21–25
Brown C, Tseng SC, Mitchell K, Roddey T (2018) Body position affects ultrasonographic measurement of diaphragm contractility. Cardiopulm Phys Ther J 29(4):166–172
Boussuges A, Finance J, Chaumet G, Brégeon F (2021) Diaphragmatic motion recorded by M-mode ultrasonography: limits of normality. ERJ Open Res 7(1):00714–2020
Zhu Z, Li J, Yang D, Gao F, Du L, Yang M (2021) Ultrasonographic evaluation of diaphragm thickness and excursion in patients with cervical spinal cord injury. J Spinal Cord Med 44(5):742–747
Kim M, Lee K, Cho J, Lee W (2017) Diaphragm thickness and inspiratory muscle functions in chronic Stroke patients. Med Sci Monit 23:1247–1253
Jung KJ, Park JY, Hwang DW, Kim JH, Kim JH (2014) Ultrasonographic diaphragmatic motion analysis and its correlation with pulmonary function in hemiplegic Stroke patients. Ann Rehabil Med 38(1):29–37
Oguri M, Okanishi T, Ikeguchi T et al (2022) Influence of gender on diaphragm thickness using a method for determining intima media thickness in healthy young adults. BMC Med Imaging 22(1):26
Yamada T, Minami T, Soni NJ et al (2018) Skills acquisition for novice learners after a point-of-care ultrasound course: does clinical rank matter? BMC Med Educ 18(1):202
World Health Organization. The Global Health Observatory: Mean BMI (kg/m²) (age-standardized estimate). [cited 2023 Jul 28]. Available from: https://www.who.int/data/gho/data/indicators/indicator-details/GHO/mean-bmi-(kg-m-)-(age-standardized-estimate)
Ministry of Health, Labour and Welfare, Government of Japan. Obesity status (BMI), age group and sex. [cited 2023 Jul 19]. Available from: https://www.mhlw.go.jp/toukei/youran/indexyk_2_1.html
Cardenas LZ, Santana PV, Caruso P, Ribeiro de Carvalho CR, Pereira de Albuquerque AL (2018) Diaphragmatic ultrasound correlates with inspiratory muscle strength and pulmonary function in healthy subjects. Ultrasound Med Biol 44(4):786–793
Boon AJ, Sekiguchi H, Harper CJ et al (2014) Sensitivity and specificity of diagnostic ultrasound in the diagnosis of phrenic neuropathy. Neurology 83(14):1264–1270
Baria MR, Shahgholi L, Sorenson EJ et al (2014) B-mode ultrasound assessment of diaphragm structure and function in patients with COPD. Chest 146(3):680–685
Scarlata S, Mancini D, Laudisio A, Raffaele AI (2019) Reproducibility of diaphragmatic thickness measured by M-mode ultrasonography in healthy volunteers. Respir Physiol Neurobiol 260:58–62
Carrillo-Esper R, Pérez-Calatayud ÁA, Arch-Tirado E et al (2016) Standardization of sonographic diaphragm thickness evaluations in healthy volunteers. Respir Care 61(7):920–924
Gottesman E, McCool FD (1997) Ultrasound evaluation of the paralyzed diaphragm. Am J Respir Crit Care Med 155(5):1570–1574
Boon AJ, Harper CJ, Ghahfarokhi LS, Strommen JA, Watson JC, Sorenson EJ (2013) Two-dimensional ultrasound imaging of the diaphragm: quantitative values in normal subjects. Muscle Nerve 47(6):884–889
Harper CJ, Shahgholi L, Cieslak K, Hellyer NJ, Strommen JA, Boon AJ (2013) Variability in diaphragm motion during normal breathing, assessed with B-mode ultrasound. J Orthop Sports Phys Ther 43(12):927–931
van Doorn JLM, Wijntjes J, Saris CGJ, Ottenheijm CAC, van Alfen N, Doorduin J (2022) Association of diaphragm thickness and echogenicity with age, sex, and body mass index in healthy subjects. Muscle Nerve 66(2):197–202
Acknowledgements
The authors are grateful to Heita Nakayama, Hiroshi Imura, Hiroyuki Ide, Megumi Honda, Namie Hattori, Riko Oda, Sachio Muramatsu, Takuya Tobari, and Yosuke Matsuzaki for their efforts in data collection and management of the research environment. They also thank Edanz (https://jp.edanz.com/ac) for editing a draft of this manuscript.
Funding
The study was supported by a JSPS KAKENHI grant (number JP20K16543).
Author information
Authors and Affiliations
Contributions
Toru Yamada, Taro Minami, Syumpei Yoshino, Ken Emoto, Suguru Mabuchi, and Masayoshi Hashimoto contributed to the design of the study. Toru Yamada, Syumpei Yoshino, Ken Emoto, and Suguru Mabuchi contributed to the data collection. The data were analyzed and interpreted by Toru Yamada, Ryoichi Hanazawa, and Akihiro Hirakawa. Toru Yamada, Taro Minami, and Masayoshi Hashimoto supervised the study. Toru Yamada wrote the first draft with input from Taro Minami. All authors contributed to the writing and review of the main manuscript, had full access to all the data in the study, and had final responsibility for the decision to submit for publication.
Corresponding author
Ethics declarations
Conflict of Interest
Taro Minami is a consultant for FUJIFILM and AA Health Dynamics in relation to a project funded by the Ministry of Economy, Trade and Industry, Japan.
Ethics Approval
The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Review Board of Tokyo Medical and Dental University (protocol code M2020-112, date of approval: December 21th, 2021).
Consent to Participate
Written informed consent was obtained from all participants involved in the study.
Consent to Publish
The authors affirm that human research participants provided informed consent for publication.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Yamada, T., Minami, T., Yoshino, S. et al. Diaphragm Ultrasonography: Reference Values and Influencing Factors for Thickness, Thickening Fraction, and Excursion in the Seated Position. Lung 202, 83–90 (2024). https://doi.org/10.1007/s00408-023-00662-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00408-023-00662-2