Introduction

Chronic kidney disease is a major life-threatening disease worldwide [1, 2]. Glomerular filtration rate (GFR) assessment is extremely important in the management of chronic kidney disease; abnormal GFR is closely related to cardiovascular morbidity and premature mortality [3, 4]. The inulin clearance test is considered the gold standard test for GFR [1], but technical difficulties inherent in the test have resulted in researchers simplifying the original protocol of the test [5, 6] or seeking other filtration agents, whether radioactive or non-radioactive [1, 7]. Estimated GFR (eGFR) from the Modification of Diet in Renal Disease (MDRD) Study reflecting serum creatinine (sCr) level, age, gender and ethnicity seems to be the most frequently advocated substitute method for measured GFR [3, 8]. Currently, an eGFR of less than 60 mL/min/1.73 m2 of the body surface area (BSA) is considered to indicate kidney disease that requires attention from both patients and physicians to prevent renal morbidity and mortality [2, 4]. However, some calibration issues in the measurement of sCr and methodological issues in the calculation of eGFR have not yet been properly addressed [810]. Furthermore, ethnic differences can affect the eGFR, requiring the adjustment of coefficients for different ethnic populations [6, 11]. Thus, the current consensus regarding the utility of eGFR is that it plays a limited role in evaluating kidney function, in particular for patients with a GFR greater than 60 mL/min/1.73 m2 [12, 13].

Recent advancements in surgical techniques have enabled nephron-sparing surgery, even in patients with malignant renal tumours. Robot-assisted or laparoscopic partial nephrectomy has emerged as a promising surgical technique, especially in patients with small renal tumours [14, 15]. Accordingly, the need for accurate and facile evaluation of preoperative and postoperative kidney function has drastically increased [16, 17]. However, accurate evaluation of perioperative GFR changes is challenging because the expected GFR is usually above the range in which eGFR gives reliable estimates. Furthermore, direct measurement of GFR was not possible in most imaging studies using indirect GFR surrogates [1820]. To our knowledge, the GFR of remnant renal tissue in individual kidneys after partial nephrectomy has never been directly assessed by any other method.

Gates’ method involves gamma camera imaging-based measurement of GFR using Tc-99m DTPA (diethylenetriaminepentaacetic acid) [21, 22]. Tc-99m DTPA is almost completely excreted through glomerular filtration without reabsorption or secretion at tubular levels. There is a specific time period when the renal uptake of Tc-99m DTPA is closely correlated with true GFR (i.e. a 2- to 3-min period post-intravenous injection of Tc-99m DTPA) [21]. Individual and total GFRs are readily obtained using Gates’ method because the percentage renal uptake of Tc-99m DTPA, the main parameter of Gates’ method, can be easily calculated using a region of interest (ROI) analysis over simple dynamic gamma camera imaging. Since its introduction in 1982, Gates’ method has been available as commercial software built in to most current gamma cameras [22, 23]. However, GFR measurement using Gates’ method has been seriously criticised, mainly because of its unreliability and inaccuracy, especially for high GFR levels [2427]. Therefore, even when accurate and facile measurement of both individual and total GFRs had become a major interest, particularly for urologists who are performing nephron-sparing surgeries, the lack of refinement in Gates’ original method has limited its utility for urologists and nuclear medicine physicians.

Thus, in the current study, we aimed to improve Gates’ method using the Cr-51 EDTA (ethylenediaminetetraacetic acid) GFR test as a gold standard. In brief, we first investigated the reproducibility of Tc-99m DTPA GFR scintigraphy, which is conducted in our institute, to address the reliability in comparison with sCr-based eGFR. Second, we revised Gates’ original GFR equation from the %renal uptake of Tc-99m DTPA using the Cr-51 EDTA GFR test as the gold standard. Last, perioperative GFR changes after partial nephrectomy were evaluated using the revised equation in another cohort of patients with renal tumours.

Materials and methods

Patients

This study was conducted with the permission of the institutional ethics review board. All subjects provided written informed consent. Fourteen healthy volunteers (male/female = 9:5; age, 26.07 ± 6.43 years) with normal blood test and urinalysis results (Table 1), who represented the normal-to-high GFR cohort, were prospectively recruited first. Of these 14 volunteers, 12 (male/female = 7:5; age, 25.0 ± 6.65 years) underwent Tc-99m DTPA GFR scintigram twice within 7 days to evaluate the reproducibility of GFR measurements. In addition, 31 patients who had undergone unilateral nephrectomy (male/female = 24:7; age, 54.9 ± 13.0 years) for renal cell cancer (n = 25), transitional cell cancer (n = 3), kidney donation (n = 2) and renal angiomyolipoma (n = 1) were prospectively enrolled to represent a cohort with moderately impaired GFRs. Of these 31 patients, 20 underwent partial and 11 had undergone total nephrectomy 2.7 ± 5.0 months before the study. There were 16 hypertensive and 3 diabetic patients in the nephrectomy cohort. However, they had only mild-to-moderate hypertension requiring minimal medication, such as a single-agent calcium channel blocker, and none of the patients received diuretics or angiotensin-converting enzyme inhibitors that are known to affect GFR. No diabetic patients were under insulin treatment and no clinically confirmed vasculopathies, such as diabetic retinopathy or coronary artery disease, were reported in the diabetic patients. The volunteers (n = 14) and nephrectomised patients (n = 31) were not statistically different from each other in terms of gender proportion, blood biochemical markers (total protein, albumin, aspartate aminotransferase and alanine aminotransferase), and urinalysis results (albuminuria and glycosuria; P > 0.05). However, the nephrectomised patients were significantly older (P < 0.0001) and had more impaired renal function (P = 0.0062 for sCr, and P < 0.0001 for eGFR) than the healthy volunteers (Table 1). These two patient cohorts, described above, were used for the generation of the revised GFR equation.

Table 1 Characteristics of the study participants

In addition, we retrospectively recruited a different patient cohort (n = 25; male/female = 19:6; age = 53.2 ± 11.6 years) who had undergone unilateral partial nephrectomy resulting from renal cell carcinoma (n = 23), angiomyolipoma (n = 1) and a renal cyst (n = 1). Tc-99m DTPA GFR scintigrams were performed as a baseline 1 month (35.0 ± 5.2 days; range, 30–49 days) before surgery. Follow-up Tc-99m DTPA GFR scintigrams were done twice at 3 months (95.4 ± 10.6 days; range, 73–114 days) and 6 months (185.6 ± 10.2 days; range 164–207 days) after surgery. In the retrospective study, we intended to test the utility of the revised GFR equation.

Tc-99m DTPA GFR scintigraphy

Thirty minutes before the scintigram, patients were instructed to ingest 500 mL of water. Then, patients lay in a supine position on the table of a single-headed gamma camera (Argus, ADAC/Philips). The camera detector was located below the table 30 cm from the patient. A cold vial of commercial DTPA (TechneScan® DTPA, Mallinckrodt) was reconstituted with Tc-99m pertechnetate (chemical impurity <10 μg Al/mL Tc-99m, radiochemical purity >95 %) that was eluted out of the Mo-99/Tc-99m generator (Technetium-99m Generator, Samyoung Unitech), yielding Tc-99m DTPA. A butterfly needle was first inserted into an antecubital vein in an upper extremity, and then Tc-99m DTPA (185 MBq) was injected. Without delay, 10 mL of normal saline was rapidly flushed into the vein. Dynamic serial images were acquired for 20 min thereafter: twenty 3-s images for flow and eighty 15-s images for function. The 20 % energy window was set at an energy peak of 140 keV. The image matrix was 128 × 128. Before the injection, pre-injection counts were obtained for 60 s; immediately after patient image acquisition, post-injection counts were obtained for another 60 s. Injected radioactivity was defined as the difference between pre- and post-injection counts, measured in counts per second.

GFR calculation from the Tc-99m DTPA GFR scintigraphy

GFR measurement was performed using quantitative software (PMOD, version 3.13, PMOD Technologies). The contours of individual kidneys and the corresponding backgrounds were drawn as shown in Fig. 1a. The area under the curve (AUC), representing the radioactivity of individual kidneys, was obtained 2–3 min post-injection of Tc-99m DTPA (Fig. 1b), which had been the optimal time point for GFR estimation in our previous study [28]. The renal radioactivity was attenuation-corrected using the previously determined formula for renal depth [23]. The resultant equation used for %renal DTPA uptake in the current study was as follows:

$$ \% renal\;\mathrm{DTPA}\;\mathrm{uptake}=\left[\left(\mathrm{RK}-\mathrm{RB}\right)/{\mathrm{e}}^{-\upmu \mathrm{RD}}+\left(\mathrm{LK}-\mathrm{LB}\right)/{\mathrm{e}}^{-\upmu \mathrm{LD}}\right]\times 100/\mathrm{injected}\;\mathrm{counts} $$

where RK is the number of counts from the right kidney, RB is the number of counts in the right kidney background, RD is the renal depth of the right kidney (in centimetres), LK is the number of counts from the left kidney, LB is the number of counts from the left kidney background, LD is the renal depth of the left kidney (in centimetres), and μ is the attenuation coefficient of Tc-99m in soft tissue (0.153/cm).

Fig. 1
figure 1

Method used to delineate regions of interest (ROIs) for %renal uptake. a ROIs were drawn along the renal contour for renal radioactivity. Curvilinear ROIs were drawn at the lower lateral portion of the individual kidneys for background activity correction. b At 2–3 min post-injection of Tc-99m DTPA, the radioactivity counts for the area under curve (AUC) were obtained for %renal uptake of Tc-99m DTPA. Please note that the right kidney shows a smaller contour than the left kidney resulting from a previous partial nephrectomy

Cr-51 EDTA GFR test

The gold standard in the current study was the Cr-51 EDTA GFR test with two plasma samples [29, 30]. Cr-51 EDTA stock solution (37 MBq/10 mL, GE Healthcare) was diluted 1:10 with normal saline. Whole blood (5 mL) was aspirated from the patients 3 h and 5 h post-injection of Cr-51 EDTA [dose = 1.85–2.96 MBq (50–80 μCi)]. After centrifugation (3,000 rpm for 8 min), 2 mL of plasma was obtained from each blood sample, the plasma was separated into two aliquots for duplication (1 mL per aliquot) and stored at room temperature until gamma counting. Two or 3 days later, radioactivity in each aliquot was counted for 20 min using an automatic gamma counter (1480 WIZARD 3”, Perkin-Elmer). Decay and background corrections were performed. The average of the duplicated plasma aliquots was used for data analysis. The clearance of Cr-51 EDTA was calculated using the following equations.

$$ \mathrm{E}=\mathrm{Uncorrected}\;\mathrm{GFR}\left(\frac{\mathrm{mL}}{ \min}\right)=\mathrm{Injected}\;\mathrm{activity}\times \frac{\mathrm{Ln}\left(\frac{\mathrm{C}1}{\mathrm{C}2}\right)}{\delta \mathrm{t}}\times \frac{\mathrm{C}{2}^{\left(\frac{\mathrm{t}1}{\delta 2}\right)}}{\mathrm{C}{1}^{\left(\frac{\mathrm{t}2}{\delta \mathrm{t}}\right)}} $$
$$ \mathrm{Corrected}\;\mathrm{GFR}\left(\frac{\mathrm{mL}}{ \min}\right)=\left\{\left(0.990778\times \mathrm{E}-0.001218\times {\mathrm{E}}^2\right)-3.7\right\}\times 1.1 $$
$$ \mathrm{Gold}\;\mathrm{standard}\;\mathrm{GFR}\left(\frac{\mathrm{m}\mathrm{L}}{ \min}\mathrm{per}\;1.73\;{\mathrm{m}}^2\right)=\mathrm{corrected}\;\mathrm{GFR}\left(\frac{\mathrm{m}\mathrm{L}}{ \min}\right)\times \frac{1.73}{\mathrm{BSA}} $$

where t1 = 3 h, t2 = 5 h, δt = 2 h, C1 = first plasma sample counts and C2 = second plasma sample counts. BSA indicates body surface area calculated using the DuBois equation: BSA(m2) = 0.007184 × weight(kg)0.425 × height(cm)0.725.

One day after the first session of gamma counting as described above, the whole-plasma samples were gamma counted once more. The average GFR from the two sessions of gamma counting was used for the final data analysis.

Serum creatinine-based estimation of GFR

The abbreviated MDRD equation was used for GFR estimation from the serum creatinine (sCr) level [3, 8]. The sCr level was measured using a chemistry analyser (Toshiba 200FR, Toshiba Medical Systems) with Roche kinetic compensated Jaffe reagent. The assay is calibrated with a lyophilised human serum calibrator (C.f.a.s., Roche Diagnostics) in which the sCr level is traceable to an isotope-dilution mass spectrometry determination. The equation used for GFR estimation was compatible with the abbreviated MDRD equation as follows:

$$ \mathrm{eGFR}\left(\frac{\mathrm{m}\mathrm{L}}{ \min}\mathrm{per}\;1.73\;{\mathrm{m}}^2\right)=175\times {\mathrm{sCr}}^{-1.154}\times {\mathrm{age}}^{-0.203}\times \left(0.742\;\mathrm{if}\;\mathrm{female}\right) $$

Statistical analysis

Data are represented as the mean ± standard deviation if not otherwise indicated. Reproducibility was analysed using coefficients of variation (CV) and intra-class correlations (ICC) between the first and second estimates. Paired t tests, Pearson’s correlation test and Bland–Altman analysis were used for comparisons between the two estimates. Non-parametric analyses, such as the Wilcoxon signed-rank sum test, or Mann–Whitney U test were used for comparisons when a normal distribution could not be assumed. Serial change of GFRs was tested using generalised estimating equation. STATA (version 12.0.0) was used for the statistical analysis. P values less than 0.05 were considered statistically significant.

Results

Reproducibility of Tc-99m DTPA GFR scintigraphy

First, we investigated the reproducibility of %renal uptake and gamma camera-based GFR measurements in comparison with measurements obtained from the sCr level and the eGFR. Here, we used Gates’ original GFR equation in order to represent the gamma camera-based GFR.

The 12 volunteers (male/female = 7:5; age, 25.0 ± 6.65 years) underwent Tc-99m DTPA GFR scintigram twice within a 1-week interval. The sCr levels and eGFR were also checked twice on the same day as the Tc-99m DTPA GFR scintigram.

The %renal uptakes were 10.24 % ± 1.60 % and 9.95 % ± 1.62 % for the first and second measurements, respectively; a significant difference was not found between the two measurements (P = 0.1763, Wilcoxon signed-rank sum test). The sCr levels were 0.88 ± 0.18 mg/dL and 0.88 ± 0.28 mg/dL for the first and second measurements, respectively, and were not significantly different, either (P = 0.9658). However, the measurement of %renal uptake was more reproducible than that of sCr level as demonstrated in Fig. 2a and b, which was quantitatively supported by the lower %CV (5.41 %) and the greater ICC (0.94) for the %renal uptake compared with the %CV (9.50 %) and the ICC (0.93) for the sCr level (Table 2).

Fig. 2
figure 2

Reproducibility of the gamma camera-based glomerular filtration rate (GFR) measurements in comparison with serum creatinine (sCr)-based GFR estimation (eGFR) in 12 healthy volunteers. The measurement of %renal uptake (a) was more reproducible than that of sCr levels (b). Furthermore, the measurement of Gates’ GFR (c) was also more reproducible than that of eGFR (d). Those results were quantitatively supported in the analyses of the %CV and ICC (Table 2)

Table 2 Reproducibility of the %renal uptake, gamma camera-based GFR (Gates’ GFR), serum creatinine (sCr) level and eGFR values

The gamma camera-based GFRs derived from the %renal uptakes were 86.4 ± 17.3 mL/min/1.73 m2 and 83.4 ± 15.8 mL/min/1.73 m2 for the first and second measurements, respectively, and were not significantly different (P = 0.1763). The eGFRs calculated from the sCr levels were 97.05 ± 13.7 mL/min/1.73 m2 and 100.5 ± 20.6 mL/min/1.73 m2 for the first and second measurements, respectively; the eGFRs of the two measurements were not significantly different either (P = 0.7646). However, data plotting (Fig. 2c, d) clearly showed that the gamma camera-based GFR was more reproducibly measured than the eGFR, which was also reflected by the lower %CV (6.60 %) and the higher ICC (0.94) for the gamma camera-based GFR compared with the %CV (10.8 %) and the ICC (0.76) for the eGFR (Table 2).

A revised equation for improved GFR measurement

The Cr-51 EDTA GFR (97.0 ± 31.9 mL/min/1.73 m2; range, 13.6–153.2 mL/min/1.73 m2) was compared with Gates’ original GFR in 45 subjects (14 healthy volunteers and 31 unilaterally nephrectomised patients). The Cr-51 EDTA GFRs correlated significantly with Gates’ original GFRs (Pearson’s r = 0.70, P < 0.0001, 95 % confidence interval = 0.54–0.81), but GFRs from Gates’ original equation (70.1 ± 20.5 mL/min/1.73 m2) were significantly lower than those from the Cr-51 EDTA GFR test (97.0 ± 31.9 mL/min/1.73 m2) (P < 0.0001, paired t test), which indicated that Gates’ original GFR systematically underestimated the gold standard GFR.

In this regard, we performed a linear regression analysis using the Cr-51 EDTA GFR as a dependent variable and %renal uptake of Tc-99m DTPA as an explanatory variable in order to derive a revised equation for improved GFR measurement, and generated the following equation:

$$ \mathrm{Improved}\;\mathrm{GFR}\left(\mathrm{mL}/ \min \right)=\left(\% renal\;\mathrm{uptake}\times 11.7773\right)-0.7354 $$

The revised equation has a greater slope than Gates’ original equation. As a result, the new, improved GFRs were always greater than Gates’ original GFRs for any given value of %renal uptake, and the difference became greater with higher values of %renal uptake (Fig. 3).

Fig. 3
figure 3

Comparison of the equations for Gates’ original GFR and the new improved GFR. The revised equation [GFR(mL/min) = (%renal uptake × 11.7773) − 0.7354] always yielded higher GFRs than Gates’ original equation [GFR(mL/min) = (%renal uptake × 9.8127) − 6.82519] in the clinically relevant range of %renal uptake. Furthermore, greater %renal uptake led to larger GFR increments with the revised GFR equation

Using Gates’ original equation [GFR(mL/min) = (%renal uptake × 9.8127) − 6.82519] [22], the %renal uptake had significantly underestimated the Cr-51 EDTA GFRs, especially in patients with high GFR levels, which is clearly demonstrated by the Bland–Altman and correlation analyses (Fig. 4a, c). However, the revised equation resulted in improved GFRs (98.0 ± 26.3 mL/min/1.73 m2) that were not different from (P = 0.7360) and maintained a significant correlation with Cr-51 EDTA GFRs (r = 0.73, P < 0.0001, 95 % CI = 0.57–0.83) (Fig. 4b, d).

Fig. 4
figure 4

Improvement in gamma camera-based GFR measurements using the revised equation. The GFRs from Gates’ original equation of [GFR(mL/min) = (%renal uptake × 9.8127) − 6.82519] were significantly lower than those from the Cr-51 EDTA test (P < 0.0001), particularly in patients with a high GFR. The Bland–Altman (a) and correlation analyses (c) readily demonstrate that the Gates’ GFRs systematically underestimate the Cr-51 EDTA GFRs, especially within the range over 60 mL/min/1.73 m2 of Cr-51 EDTA GFR. On the other hand, the improved GFRs from the revised equation of [GFR(mL/min) = (%renal uptake × 11.7773) − 0.7354] were not significantly different from those of the Cr-51 EDTA test (P = 0.7360). It is of note that the revised equation minimises the GFR differences between the improved GFRs and Cr-51 EDTA GFRs even at high GFR levels as shown in the Bland–Altman analysis (b) and that the improved GFRs and Cr-51 EDTA GFRs are consistent with the line of equality as demonstrated in the correlation analysis (d)

Application of the improved GFR to partial nephrectomy

Next, we retrospectively applied the improved GFR from the revised equation to patients following partial nephrectomy (n = 25).

The total GFRs significantly decreased 3 months (P < 0.001) and 6 months (P < 0.001) post-surgery, which were clearly demonstrated by the new improved GFRs (Fig. 5a). The eGFRs also demonstrated the same tendency in GFR decline over a 3- to 6-month period post-surgery, but the preoperative eGFRs of 2 patients could not be provided because the formal report for the eGFR was just greater than 60 mL/min/1.73 m2 without specific value and some aberrant data were noticed (e.g. higher eGFRs at 6 months postoperatively than at baseline or 3 months postoperatively; data now shown).

Fig. 5
figure 5

Assessment of perioperative GFR changes using the improved GFR equation. a The preoperative total GFRs had significantly decreased 3 months (P < 0.001) and 6 months post-surgery (P < 0.001), which was readily demonstrated by the improved GFRs. The kidneys that were operated on had significant GFR declines 3 months (P < 0.001) and 6 months (P < 0.001) post-surgery (b), whereas the contralateral kidneys did not experience significant GFR changes over the 3- to 6-month period post-surgery (P > 0.05) (c). *P < 0.001

The individual GFR data from the revised GFR equation clearly demonstrated the perioperative GFR changes in both the kidneys that were operated on and the contralateral kidneys (Fig. 5b, c). The kidneys that were operated on suffered from significant GFR declines 3 months post-surgery (P < 0.001), and the GFR declines continued until 6 months post-surgery (P < 0.001; Fig. 5b), whereas the contralateral kidneys did not experience significant GFR changes over the 3- to 6-month period post-surgery (P > 0.05; Fig. 5c).

Discussion

Our current approach allowed reliable and accurate measurement of moderate-to-high GFRs i.e. the GFR range of patients undergoing nephron-sparing surgery or partial nephrectomy. The revised GFR equation [GFR(mL/min) = (%renal uptake × 11.7773) − 0.7354] derived from the linear regression analysis between the gamma camera imaging-based parameter (%renal uptake of Tc-99m DTPA) and Cr-51 EDTA GFR could be successfully applied to the measurement of individual and total GFRs post-partial nephrectomy.

In our institute, the gamma camera imaging-based measurement of GFR was more reproducible than the sCr-based eGFR (Fig. 2). Furthermore, the %CVs for %renal uptake (5.41 %) and gamma camera-based GFR (6.60 %) in the current study were not inferior to reported %CVs of inulin clearance (7.1–11.3 %) [31], iothalamate clearance (6.3–18.7 %) [32] or iohexol clearance (5.6–7.6 %) [33], suggesting that the gamma camera imaging-based GFR from the %renal uptake of Tc-99m DTPA is reliable enough for repeated measurements of GFR (Figs. 2 and 5).

Not only the reproducibility but also the accuracy is important for any GFR study to be used in clinical practice. Gates’ original GFR method has been seriously criticised in this regard [23, 25]. In the current study, Gates’ method of GFR measurement could be significantly improved, particularly for patients with normal-to-high GFRs (Figs. 3 and 4). In fact, the clinical need for an accurate GFR measurement technique was not as prevalent for high GFR levels compared with low GFR levels until nephron-sparing surgery and partial nephrectomy became urological topics over the last decade [34, 35]. Only indirect GFR surrogates or the unreliable eGFR have been used so far instead of directly measured GFR [12, 13]. Thus, the clinical implication of the current study is that a revised and improved Gates’ equation allows Tc-99m DTPA GFR scintigraphy to play a critical role in the evaluation of perioperative GFR changes in individual kidneys, especially in partial nephrectomy patients.

Indeed, it is not difficult to examine the function of the remnant kidney after total nephrectomy, because the functioning renal tissues come from only one remnant kidney. Therefore, perioperative GFR changes after total nephrectomy have been well described in both animal and clinical studies [36]; total nephrectomy would elicit compensatory hyperfunction or hypertrophy of the remaining kidney up to several years post-surgery, but would accelerate functional impairment of the remnant kidney in the long run [37]. On the other hand, only a few imaging studies have examined the functional consequences of kidneys that were operated on and contralateral kidneys after partial nephrectomy, and only indirect estimates of GFR such as tubular function or parenchymal volume could be assessed [20]. One of the key issues in partial nephrectomy is how to efficiently preserve the function of the remnant parenchyma of the kidney that was operated on, which is the major goal of modern nephron-sparing surgery [16, 17]. Imaging studies like Tc-99m DTPA GFR scintigraphy reflecting the renal function of individual kidneys may have a clinical impact in this regard. Figure 5 demonstrates that the kidneys that were operated on suffered from GFR declines 3 to 6 months post-partial nephrectomy (Fig. 5b), whereas the contralateral kidneys did not show drastic changes in their individual GFRs, albeit with a little fluctuation (Fig. 5c). These findings would be explained as the loss of functioning renal tissue from the surgery itself, the occurrence of perioperative renal damage or a mixture of both in the kidney that was operated on to the extent that compensatory hyperfunction of the contralateral kidney was not required at 3–6 months post-surgery.

The revised equation for improved GFR [GFR(mL/min) = (%renal uptake × 11.7773) − 0.7354] was derived from the imaging parameter (%renal uptake of Tc-99m DTPA) which had been originally adopted by Gates [GFR (mL/min) = (%renal uptake × 9.8127) – 6.82519] [21, 22]. Linear regression analysis was used in both cases, generating another linear relationship between the two equations (Fig. 3), which means that Gates’ GFR data obtained from the gamma cameras containing commercial software can be converted to new, improved GFR data with ease by using a simple conversion equation like the following:

$$ \mathrm{New}\;\mathrm{improved}\;\mathrm{GFR}=\left(\mathrm{Original}\;\mathrm{Gates}'\mathrm{GFR}\times 1.20021\right)+7.45626 $$

The employment of Cr-51 EDTA plasma clearance as a gold standard for true GFR has some advantages over Gates’ original method in which 24-h urinary creatinine clearance was used [21]. Because of the technical difficulties of urine collection over a 24-h period, the urinary clearance of creatinine has been reported to be less reproducible, even in comparison with eGFR [1, 32]. In fact, urine collection was the main reason for the poor reproducibility of the inulin clearance test [31]. On the other hand, Cr-51 EDTA clearance using multiple plasma samples has been reported to be highly reproducible (%CV = 2.6–3.5 %) [38]. A small underestimation of the inulin clearance GFR by Cr-51 EDTA GFR was reported about 40 years ago [39]. Nonetheless, Cr-51 EDTA GFR has been widely used as a reference standard because of its reliability and convenience in a variety of trials investigating new GFR agents [38]. Moreover, the kinetics of Cr-51 EDTA and Tc-99m DTPA have been reported to be similar, justifying the use of the Cr-51 EDTA test in the current study [40].

Inclusion of patients representing the entire range of impaired renal function is usually recommended in this kind of establishment study for GFR. The patients recruited in the current study were either healthy volunteers or almost nephrologically disease-free oncology patients (Table 1). Taking into account this limitation, the current study results should be applied mainly to nephrectomy patients. However, the new Tc-99m DTPA GFR scintigraphy may be applicable to low-GFR patients with medical illnesses, which warrants further investigation.

In conclusion, this new Tc-99m DTPA GFR scintigraphic method is very promising for the accurate evaluation of perioperative GFR changes in patients undergoing nephron-sparing partial nephrectomy.