Abstract
To compare magnetic resonance imaging (MRI), 64-slice multi-detector computed tomography (MDCT) and dual-source computed tomography (DSCT) in assessing global function parameters using a moving heart phantom. A moving heart phantom with known volumes (215–258 ml) moving at 50–100 beats per minute was examined by three different imaging modalities using clinically implemented scanning protocols. End-diastolic and end-systolic volumes were calculated by two experienced observers using dedicated post-processing tools. Ejection fraction (EF) and cardiac output (CO) were calculated and mutually compared using Bland-Altman plots. MRI underestimated the ejection EF by 16.1% with a Bland-Altman interval (B-A) of [-4.35 (-2.48) -0.60]. Sixty-four-slice MDCT overestimated the EF by 2.6% with a relatively wide B-A interval of [-3.40 (0.40) 4.20]. DSCT deviated the least from the known phantom volumes, underestimating the volumes by 0.8% with a B-A interval of [-1.17 (-0.13) 0.91]. CO analysis showed similar results. Furthermore, a good correlation was found between DSCT and MRI for EF and CO results. MRI systematically underestimates functional cardiac parameters, ejection fraction and cardiac output of a moving heart phantom. Sixty-four-slice MDCT underestimates or overestimates these functional parameters depending on the heart rate because of limited spatial resolution. DSCT deviates the least from these functional parameters compared to MRI, EBT and 64-slice MDCT.
Similar content being viewed by others
Introduction
Functional parameters of the left ventricle, such as the ejection fraction and the cardiac output, are closely related to cardiac morbidity and mortality. These parameters can predict the prognosis of coronary artery disease in individual patients and in entire cohorts if assessed accurately [1–4]. The current standard of reference to assess functional parameters is short-axis magnetic resonance imaging (MRI) using steady-state free precession (SSFP) sequences [5–10]. The main advantage of MRI is its excellent temporal resolution without the need to expose the patient to ionising radiation or nephrotoxic contrast agents. However, MRI is contra-indicated in a substantial number of patients for various reasons, e.g., non-MR-compatible implants or claustrophobia [11].
As an alternative to MRI, multi-detector computed tomography (MDCT) can be used to assess functional parameters of the left ventricle. Although MDCT has a lower temporal resolution (165 ms), its spatial resolution (voxel size 0.6 × 0.6 × 3.0 mm) is higher than MRI (voxel size 1.7 × 1.7 × 6.0 mm). Another CT technique, dual-source CT (DSCT), has a temporal resolution more similar to the temporal resolution of MRI (83 ms). DSCT combines this relatively high temporal resolution with a relatively high spatial resolution (voxel size 0.6 × 0.6 × 3.0 mm) [12]. These given resolutions are based upon clinically implemented scan protocols. Although these resolutions can be further optimised, this could influence negatively other aspects of the scan, such as scan duration, patient dose or noise levels.
Since CT is now used more frequently in clinical practice than MRI, it is relevant to assess its diagnostic accuracy for analysis of functional parameters of the left ventricle, since these parameters can easily be calculated from the raw data of a gated CT examination of the heart. Whereas CT is the preferred modality for assessing morphology of the heart and the coronary arteries, CT might also play an important role in functional assessment as well.
Therefore, the purpose of this study was to compare MRI, 64-slice MDCT and DSCT in assessing the ejection fraction and cardiac output of an anthropomorphic moving heart phantom at a range of different heart rates.
Materials and methods
An anthropomorphic moving heart phantom (Limbs & Things, Bristol, UK) with known volumes was connected to a gas pump with variable flow settings (Fig. 1). The gas pump also generated a trigger signal that enabled synchronised ECG gating of the imaging modalities with the movement of the phantom. The phantom was designed and adjusted for CT and MRI applicability [13]. Heart rates of 50–100 beats per minute (bpm) with an interval of 10 bpm were simulated. The known phantom volumes were used as the reference. The reference value for the end-diastolic volume (EDV) was determined by filling the phantom with water. This value was considered to be constant at varying heart rates. The reference end-systolic volume (ESV) was calculated by determination of the stroke volume (SV) using a gas flow analyser (VT Plus HF, Fluke, Everett, WA) and using the relation SV = EDV – ESV.
The anthropomorphic moving heart phantom was positioned along the z-axis of the scanner and scanned with three different imaging modalities using clinically implemented scanning protocols: MRI, 64-slice MDCT and DSCT. The parameters for 1.5-T MR (Siemens Magnetom Sonata, Erlangen, Germany) were: TR/TE 57.46/1.10 ms, α 59o, FOV 284 × 350 mm, matrix 125 × 192 mm, voxel size 1.7 × 1.7 × 6 mm and interslice gap 4 mm with a retrospectively gated cine steady-state free-precession sequence. The parameters for 64-slice MDCT (Siemens Somatom Sensation 64, Erlangen, Germany) were: rotation time 330 ms, slice thickness 3.0 mm increment 3.0 mm, tube voltage 120 kV and tube current 250 ms. The parameters for DSCT (Siemens Somatom Definition, Forcheim, Germany) were: rotation time 330 ms, slice thickness 3.0 mm, increment 3.0 mm, tube voltage 120 kV and tube current 120 mAs/rot, a protocol similar to that of the 64-slice MDCT.
After data acquisition the 64-slice MDCT and DSCT images were reconstructed with a medium smooth kernel at every 10% of the RR interval. A bi-segmental reconstruction was used for the 64-slice MDCT data for heart rates over 65 bpm as suggested by the manufacturer, and a single-segmental reconstruction was used for the DSCT data.
Then, the reconstructed axial images were analysed by two independent experienced observers. Both observers had more than 3 years of experience in analysing and post-processing clinical cardiac MRI and CT scans. Two dedicated post-processing tools, using the same algorithm and the same visual interface, were used for this analysis. The volume analysis of the MRI and DSCT data was performed using QMass (Medis, Leiden, The Netherlands). The volume analysis of the 64-slice MDCT data was performed using CardIQ (GE Medical Systems, Milwaukee, WI). In the volume analysis, the end-diastolic volume (EDV) and the end-systolic volume (ESV) were calculated. Since the phantom was scanned along the z-axis, short-axis images were obtained automatically. Apical and basal slices were defined as the first and last slice in which the ventricular lumen was present. The diastolic phase for the EDV was defined as the phase with the greatest luminal cavity. The end-systolic phase for the ESV was defined as the phase with the smallest luminal cavity. The observers manually traced the area of the cavity in each slice in the appropriate phases, defining the total volume. The obtained volumes of the analysis were used to calculate the stroke volume (SV), ejection fraction (EF) and cardiac output (CO) at every heart rate. Stroke volume is calculated as SV = EDV - ESV. The ejection fraction is calculated as EF = (SV/EDV) * 100%. The cardiac output is calculated as CO = HR * SV.
Data analysis
The functional parameters of the imaging modalities were compared to the functional parameters of the reference and assessed using an over- or underestimation and a relative difference. The over- or underestimation is defined as the difference in functional parameter between one of the four modalities and the reference. The relative difference is defined as the over- or underestimation in functional parameter by a modality as a percentage of the functional parameter of the reference.
The method of Bland and Altman was used to display the average over- or underestimation and limits of agreement between the imaging modalities and the reference [14]. The results are shown in the form of [average - 2x standard deviation (average) average + 2x standard deviation]. A small over- or underestimation indicates a high accuracy, and small limits of agreement indicate a high precision.
A statistical test was used to investigate the significance of the obtained volumes and estimations. A paired t-test was used to asses the differences in the measured volumes by the two observers. The same statistical test was used to asses the over- and underestimations in EF and CO as a function of HR by the different modalities. A p-value < 0.05 was considered to be significant for the test.
Results
Volumetric data
The reference maximum volume, end-diastolic volume (EDV), of the heart phantom is 258 ± 1 ml. The reference minimum volume, end-systolic volume (ESV), of the heart phantom varies at different heart rates. The ESV increases with increasing heart rates, ranging from 215.0 ml at 50 bpm to 223.7 ml at 100 bpm.
Averaged EDV values were between 263.2 and 269.6 ml for MRI, between 274.6 and 283.2 ml for 64-slice MDCT and between 256.9 and 261.7 ml for DSCT, showing exaggerated EDVs compared to the reference for all modalities except for DSCT. Averaged ESV values were between 229.0 and 236.9 ml for MRI, between 230.2 and 248.4 ml for 64-slice MDCT and between 219.8 and 220.6 ml for DSCT, showing exaggerated ESVs compared to the reference for all modalities except for DSCT again. The measured volumes are given in detail in Tables 1 and 2, and axial slices of the heart phantom obtained with the three imaging modalities are shown in Fig. 2.
The volumes were used to calculate the different functional parameters. The ejection fraction (EF) of the reference is smaller at higher heart rates, a pattern observed with the three modalities as well (Fig. 3). The EF by MRI was underestimated compared to the reference at all heart rates. The average relative difference of -16.1% to the reference for MRI was significant (p < 0.01). The EF by 64-slice MDCT was overestimated compared to the reference except at heart rates of 60 and 100 bpm. However, the average relative difference of 2.6% to the reference was not significant (p = 0.62). The EF by DSCT showed a small underestimation at lower heart rates and a small overestimation at higher heart rates. The average relative difference of 0.8% to the reference was not significant as well (p = 0.57).
In contrast to the EF, the cardiac output (CO) of the reference is larger at higher heart rates, a pattern observed for all three investigations (Fig. 4). The CO by MRI was underestimated compared to the reference at all heart rates. The average relative difference of -12.7% to the reference for MRI was significant (p < 0.02). The CO by 64-slice MDCT was overestimated compared to the reference except at heart rates of 60 and 100 bpm. However, the average relative difference of 10.7% to the reference was not significant (p = 0.15). The CO by DSCT was underestimated except at a heart rate of 100 bpm. The average relative difference of 0.3% to the reference was not significant as well (p = 0.84). The relative differences to the reference for each modality are given in Tables 1 and 2.
The accuracies (size of over- or underestimation) and precisions (range of limits of agreement) of the different imaging modalities in assessing the EF are shown using Bland-Altman plots (Fig. 5). The accuracy and precision of MRI were both moderate in assessing the EF. The accuracy of 64-slice MDCT was much better, but the precision was relatively low. The accuracy and precision of DSCT were both high. The precise Bland-Altman limits for the different modalities in assessing the EF were [-4.35 (-2.48) -0.60] for MRI, [-3.40 (0.40) 4.20] for MDCT and [-1.17 (-0.13) 0.91] for DSCT.
The accuracies and precisions found in the analysis of the CO results were different from the previous EF results (Fig. 6). MRI and 64-slice MDCT both showed a moderate accuracy. However, the precisions were different. The precision of MRI was high, but the precision of 64-slice MDCT was low. DSCT showed again a high accuracy and a high precision. The precise Bland-Altman limits for the different modalities in assessing the CO were [-0.72 (-0.37) -0.02] for MRI, [-0.58 (0.31) 1.20] for MDCT and [-0.18 (0.01) 0.20] for DSCT. The accuracy of each modality is given in Table 3.
Inter-observer data
Finally, the inter-observer analysis showed that, overall, the difference in volumes as determined by the two observers was not significantly different (p > 0.08). However, when analysing the difference in EDV and the difference in ESV separately, the difference in EDV was not significant (p > 0.90), whereas the difference in ESV was significant (p < 0.01).
Discussion
Our results show that MRI underestimates functional parameters, ejection fraction and cardiac output of a moving heart phantom. Sixty-four-slice MDCT underestimates or overestimates these parameters depending on the heart rate. DSCT deviates least from these parameters compared to MRI and 64-slice MDCT.
The different imaging modalities inevitably used slightly different scan protocols in this study. Although it would have been possible to use more similar parameters, such as comparable slice thicknesses and increments, it was considered to be more clinically relevant to use typical protocols used in daily clinical practice.
MRI overestimated the end-diastolic volume (EDV) and end-systolic volume (ESV) structurally, as Mao et al. found as well [15]. This resulted in a structural underestimation of the ejection fraction (EF) and cardiac output (CO). The structural overestimation of the physical volumes might be the result of the limited spatial resolution of MRI. For MRI, the section thickness was 6 mm, a reduced spatial resolution compared to the other modalities, explaining this overestimation. Sequences with higher spatial resolution are available for MRI, and it is expected that using these optimised sequences will improve the results. However, the duration of the scan will be increased.
Sixty-four-slice MDCT is not hampered by a limited spatial resolution; however, its temporal resolution is low. This low temporal resolution causes increased blurring in the CT images, making a proper delineation of the luminal cavity more difficult. Therefore, an observer might easily over- or underestimates a volume at an individual heart rate (e.g., the rather large deviation at 80 bpm), whereas the average measurement over all heart rates has a high accuracy. This explains the variable results with large limits of agreement for 64-slice MDCT. The average measurement is accurate, but it shows large outliners. Similar results were reported previously; data acquired with a temporal resolution of 165 ms showed large limits of agreement, whereas data acquired with a temporal resolution of 83 ms showed very small limits of agreement [16]. Although the temporal resolution can be improved with multi-segment reconstruction, leading to better reproducibility in a phantom study [17], it was shown that this technique did not improve results for studies in human subjects [18].
DSCT with a high temporal and a high spatial resolution approximated the physical volumes the best. Another phantom study by Mahnken et al. showed also good approximations to the physical volumes with DSCT [16]. In vivo studies have also reported on the use of DSCT for functional assessment. However, these studies only result in a comparison between DSCT and usually MRI, because physical values are not known. The reported correlations between DSCT and MRI are very good [19–21].
There were no clinically significant differences in image quality for both CT systems with respect to delineation of the ventricular cavity (Fig. 2). It should be noted that the smooth surface of the ventricular cavity in the heart phantom is not necessarily representative of the situation in vivo. The human left and right ventricular cavities are lined with varying degrees of inhomogeneous porous myocardial tissue, often referred to as trabecularisation [22]. Furthermore, the contrast between blood and myocardial tissue in vivo is different than the contrast between artificial material and gas in the heart phantom. Therefore, the aspect of contrast and image quality could not be fully taken into account in the present study.
The measured EDVs and ESVs resulted in very low EFs and COs. Such low EFs and COs are uncommon in clinical practice. Besides these uncommon EFs and COs, there are two more aspects that make this phantom study deviate from clinical practice. First, the moving heart phantom has a fixed ejection fraction and cardiac output at a certain heart rate, whereas in patients the functional parameters may vary. Second is the phenomenon of trabecularisation of the left ventricle, which makes exact delineation of the endocardial border in patients more dependent on visible contrast between the blood pool and the ventricular wall as stated previously. Despite these shortcomings, the use of a moving heart phantom offered the advantage of ensuring highly reproducible conditions at each of the three imaging modalities, whereas studies in human subjects intrinsically suffer from lack of consistency due to different variables, such as physiological non-constant heart rates caused by breathing or fluctuations in fluid status.
Although the moving heart phantom has some shortcomings and deviates from clinical practice, the results might be extrapolated to EFs and COs more common in clinical practice. The small over- and underestimations of the functional parameters of the left ventricle might argue for interchangeable results for different imaging modalities. However, these over- and underestimations are due to the fact that the simulated functional parameters were relatively small as well. If these results are extrapolated to values associated with normal EFs and COs [23, 24], the over- and underestimations might be around 10% for the EF and around 0.9 l/min for the CO.
In addition to the use of different imaging modalities resulting in different values of functional parameters, human error might increase this difference. Surprisingly, there was a significant difference in the ESV between the two observers, whereas the EDV showed no difference. These differences cannot be explained by visual selection of different phases or systematic differences in delineation. These differences are most likely explained by larger differences in terms of percentage between the intrinsically lower quantitative values for ESV compared to EDV.
Limitations
Reconstructions were made at every 10% of the RR interval, although smaller intervals are possible. Smaller intervals, for example intervals of 5%, have been used in previous studies [25, 26]. However, reconstructions at intervals of 10% are thought to be adequate for clinical use according to Suzuki et al. [27]. In addition, it was reported that DSCT and MRI show similar differences in functional parameters when 10 or 20 phases are used for reconstruction of the DSCT data [28].
Different post-processing tools were used for the volume analysis, because the individual post-processing tools could only process CT or MR data and not both. Furthermore, not all post-processing tools were available throughout the entire study, and since the DSCT data were added in a later stadium, it could not be analysed using CardIQ. However, all post-processing tools were developed by the same company (Medis, Leiden, The Netherlands) and use the same algorithm and visual interface. Therefore, the differences in end-diastolic and end-systolic volumes are not considered to be influenced by using the different post-processing tools.
Conclusion
Our results show that a clinically implemented MRI protocol structurally underestimates functional parameters, ejection fraction and cardiac output of a moving heart phantom. A clinical implemented protocol using 64-slice MDCT underestimates or overestimates these functional parameters depending on the heart rate. A clinical protocol using DSCT deviates the least from these functional parameters compared to MRI and 64-slice MDCT.
References
Emond M, Mock MB, Davis KB et al (1994) Long-term survival of medically treated patients in the Coronary Artery Surgery Study (CASS) Registry. Circulation 90:2645–2657
Ghali JK, Liao Y, Simmons B et al (1992) The prognostic role of left ventricular hypertrophy in patients with or without coronary artery disease. Ann Intern Med 117:831–836
Levy D, Garrison RJ, Savage DD et al (1990) Prognostic implications of echocardiographically determined left ventricular mass in the Framingham Heart Study. N Engl J Med 322:1561–1566
Moise A, Bourassa MG, Theroux P et al (1985) Prognostic significance of progression of coronary artery disease. Am J Cardiol 55:941–946
Thiele H, Paetsch I, Schnackenburg B et al (2002) Improved accuracy of quantitative assessment of left ventricular volume and ejection fraction by geometric models with steady-state free precession. J Cardiovasc Magn Reson 4:327–339
Sechtem U, Pflugfelder PW, Gould RG et al (1987) Measurement of right and left ventricular volumes in healthy individuals with cine MR imaging. Radiology 163:697–702
Rominger MB, Bachmann GF, Pabst W et al (1999) Right ventricular volumes and ejection fraction with fast cine MR imaging in breath-hold technique: applicability, normal values from 52 volunteers, and evaluation of 325 adult cardiac patients. J Magn Reson Imaging 10:908–918
Rerkpattanapipat P, Hundley WG (2007) Dobutamine stress magnetic resonance imaging. Echocardiography 24:309–315
Futamatsu H, Wilke N, Klassen C et al (2007) Usefulness of cardiac magnetic resonance imaging for coronary artery disease detection. Minerva Cardioangiol 55:105–114
Bellenger NG, Grothues F, Smith GC et al (2000) Quantification of right and left ventricular function by cardiovascular magnetic resonance. Herz 25:392–399
Tornqvist E, Mansson A, Larsson EM et al (2006) It's like being in another world–patients’ lived experience of magnetic resonance imaging. J Clin Nurs 15:954–961
Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268
Greuter MJ, Dorgelo J, Tukker WG et al (2005) Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit. Eur Radiol 15:995–1007
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310
Mao S, Takasu J, Child J et al (2003) Comparison of LV mass and volume measurements derived from electron beam tomography using cine imaging and angiographic imaging. Int J Cardiovasc Imaging 19:439–445
Mahnken AH, Bruder H, Suess C et al (2007) Dual-source computed tomography for assessing cardiac function: a phantom study. Invest Radiol 42:491–498
Mahnken AH, Hohl C, Suess C et al (2006) Influence of heart rate and temporal resolution on left-ventricular volumes in cardiac multislice spiral computed tomography: a phantom study. Invest Radiol 41:429–435
Juergens KU, Maintz D, Grude M et al (2005) Multi-detector row computed tomography of the heart: does a multi-segment reconstruction algorithm improve left ventricular volume measurements? Eur Radiol 15:111–117
Bastarrika G, Arraiza M, De Cecco CN et al (2008) Quantification of left ventricular function and mass in heart transplant recipients using dual-source CT and MRI: initial clinical experience. Eur Radiol 18(9):1784–1790, Epub 2008 May 20
Busch S, Johnson TR, Wintersperger BJ et al (2008) Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings. Eur Radiol 18:570–575
Brodoefel H, Kramer U, Reimann A et al (2007) Dual-source CT with improved temporal resolution in assessment of left ventricular function: a pilot study. AJR Am J Roentgenol 189:1064–1070
Steen H, Nasir K, Flynn E et al (2007) Is magnetic resonance imaging the ‘reference standard’ for cardiac functional assessment? Factors influencing measurement of left ventricular mass and volumes. Clin Res Cardiol 96:743–751
Ren X, Ristow B, Na B et al (2007) Prevalence and prognosis of asymptomatic left ventricular diastolic dysfunction in ambulatory patients with coronary heart disease. Am J Cardiol 99:1643–1647
Maurer MS, Burkhoff D, Fried LP et al (2007) Ventricular structure and function in hypertensive participants with heart failure and a normal ejection fraction: the cardiovascular health study. J Am Coll Cardiol 49:972–981
Juergens KU, Grude M, Maintz D et al (2004) Multi-detector row CT of left ventricular function with dedicated analysis software versus MR imaging: initial experience. Radiology 230:403–410
Schlosser T, Pagonidis K, Herborn CU et al (2005) Assessment of left ventricular parameters using 16-MDCT and new software for endocardial and epicardial border delineation. AJR Am J Roentgenol 184:765–773
Suzuki S, Furui S, Kaminaga T et al (2006) Accuracy and efficiency of left ventricular ejection fraction analysis, using multidetector row computed tomography: effect of image reconstruction window within cardiac phase, slice thickness, and interval of short-axis sections. Circ J 70:289–296
Puesken M, Fischbach R, Wenker M et al (2008) Global left-ventricular function assessment using dual-source multidetector CT: effect of improved temporal resolution on ventricular volume measurement. Eur Radiol 18(10):2087–2094, Epub 2008 May 1
Acknowledgements
The authors would like to thank J. Overbosch for performing the volume assessment and E.J.K Noach for reviewing this manuscript.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://creativecommons.org/licenses/by-nc/2.0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
About this article
Cite this article
Groen, J.M., van der Vleuten, P.A., Greuter, M.J.W. et al. Comparison of MRI, 64-slice MDCT and DSCT in assessing functional cardiac parameters of a moving heart phantom. Eur Radiol 19, 577–583 (2009). https://doi.org/10.1007/s00330-008-1197-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-008-1197-1