Introduction

Cardiac morbidity and mortality are closely related to cardiac volumes, myocardial mass and global left ventricular (LV) function, expressed as the LV ejection fraction (LVEF). Accurate assessment of these parameters is required for the prediction of prognosis in individual patients as well as in entire cohorts [13].

Assessment of LVEF can be performed by a number of different imaging modalities, ranging from echocardiography to direct contrast ventriculography. The current standard of reference for global LV function analysis is short-axis magnetic resonance imaging (MRI), using steady-state free precession (SSFP) sequences [46].

The main advantage of MRI is its excellent temporal resolution without exposing the patient to ionizing radiation or iodine-based contrast agents. In addition, MRI does not rely on geometrical assumptions for assessment of LV function parameters. However, MRI is contra-indicated in a substantial number of patients for various reasons, e.g., non-MR-compatible implants or claustrophobia [7].

Furthermore, assessment of coronary artery anatomy using MRI is currently held to be inferior to assessment by CT [8].

For this indication, CT is gaining ground due to several major extensive technological improvements. The most marked development has been the introduction of the multiple-detector computed tomography (MDCT). With ever decreasing gantry rotation times and the introduction of a second X-ray source in dual-source CT (DSCT) technology, the temporal resolution of CT is slowly approaching the temporal resolution obtained with MRI. On the other hand, spatial resolution obtained with contemporary CT techniques already outperforms the spatial resolution of contemporary MRI acquisitions.

Different CT techniques offer different advantages and disadvantages. Sixty-four-slice MDCT offers a high spatial resolution with a relatively low temporal resolution. DSCT combines both a high temporal resolution with a high spatial resolution. Moreover, the image quality of DSCT at higher heart rates is superior to that of conventional MDCT systems.

Since DSCT is used ever more frequently in clinical practice, it is increasingly relevant to assess its diagnostic accuracy for analysis of global LV function parameters, since they can easily be calculated from the raw data of a gated DSCT examination of the heart. A recently published meta-analysis showed that the diagnostic accuracy of CT-derived global LV function parameters has increased with the introduction of more advanced CT systems [9].

Therefore, we sought to investigate the diagnostic accuracy of the new DSCT system.

Materials and methods

Patients

The present study was conducted at the University Medical Center Groningen. The protocol was reviewed and approved by the medical ethics committee. Thirty-four consecutive clinical patients with diverse cardiac disease, who were scheduled for DSCT by their treating physician in routine clinical practice were approached for the study. After the DSCT assessment, the patient underwent an MRI examination of the heart. If LV function modifying interventions, such as coronary interventions or medication changes, were carried out in the time between both acquisitions the patient was excluded.

All patient characteristics are displayed in Table 1. There were no limitations for inclusion with regard to the underlying cardiac disease. All patients in the present study had sinus rhythm, allowing adequate ECG triggering.

Table 1 Baseline clinical characteristics (n = 29). Data are displayed as (n) percentage, unless otherwise indicated (CABG coronary artery bypass grafting, MI myocardial infarction, PCI percutaneous coronary intervention, NYHA class New York Heart Association functional class)

DSCT data acquisition

All cardiac CT angiographies were performed on a DSCT system (Somatom Definition, Siemens Medical Solutions, Forcheim, Germany). Retrospective ECG-triggered images were obtained, from the carina to the apex, during one breath-hold. The following scan protocol was used: tube voltage was 120 kV and tube current 410 mAs/rot for both tubes. The gantry rotation time was 0.33 s and the pitch ranged from 0.2 to 0.5 s, adapted automatically to the heart rate. The ECG pulsing window was 20–70% of the RR-interval for all patients. Scanning was performed with a 64 × 0.6-mm collimation. Coronary enhancement was achieved by intravenous injection of 80 ml of non-ionic contrast agent (Iomeron 400 mg/ml, Bracco, Milan, Italy) followed by a saline bolus flush. Bolus timing was performed automatically with a region of interest set in the descending aorta with a threshold of 100 Hounsfield units. For LV function analysis, multiphase reconstructions were made at every 10% of the RR-interval. A mono-segmental reconstruction algorithm was used for all image-reconstructions.

MRI data acquisition

All patients were examined on a 1.5-Tesla MR system (Siemens Magnetom Sonata, Erlangen, Germany) using a 2 × 6 channel body-coil. After single-shot localizer images, a retrospectively gated cine SSFP sequence (TR/TE 57.46/1.10 ms, α 59°, FOV 284 × 350 mm, matrix 125 × 192 mm, voxel size 1.7 × 1.7 × 6 mm, interslice gap 4 mm.) was implemented to acquire short-axis cine-loops, covering the entire left ventricular cavity from base to apex. No contrast agents were administered. An example of a short axis image of both DSCT and MRI is provided in Fig. 1.

Fig. 1
figure 1

Example of a short-axis image from MRI (right) and DSCT (left)

Visual quality assessment

Before post-processing, all DSCT datasets were visually scored for image-quality. Contrary to the quality-assessment of the DSCT dataset for coronary artery analysis, where the opacification of the coronary arteries and the ability to follow them from base to distal is essential, the quality assessment for LV functional analysis focuses more on opacifictaion of the LV cavity and the presence of “stairstep” artifacts caused by ventricular extra-systoles, breathing artifacts or segmentation errors [10].

An example of a “stairstep” artifact is shown in Fig. 2. Four categories were used: good quality (without any artifacts), minor artifacts (however, LV delineation sufficiently possible), major artifact (LV delineation not reliable) and poor image quality (no full coverage of LV or overall distortion). Image quality was scored separately by two observers, blinded for each others results. Only the DSCT datasets rated to be either of good quality and those scored to have minor artifacts were analyzed.

Fig. 2
figure 2

Example of a “stairstep” artifact on a short axis reformation of a DSCT dataset

Post-processing

Post-processing was performed using a dedicated software package, Mass (Medis, Leiden, the Netherlands), for both DSCT and MRI data. The two dedicated post-processing tools make use of the same algorithm and visual interface. The DSCT data were reformatted to short-axis cine-loops of the LV cavity with a slice thickness of 6 mm and an interslice gap of 4 mm, matching the MRI specifications. The end-diastolic phase was defined as the phase with the greatest visually estimated luminal cavity at mid-ventricular level, and the systolic phase as the phase with the smallest luminal cavity. Readers manually traced the area of the cavity in each slice in the appropriate phases. The obtained parameters of the analysis consisted of the end-diastolic volumes (EDV) and end-systolic volumes (ESV). From these parameters stroke volume (SV) and left ventricular ejection fraction (LVEF) were calculated. Stroke volume was calculated by SV = EDV - ESV. The ejection fraction was calculated by EF = (SV/EDV) × 100%. Myocardial mass was calculated by delineating the epicardial contour as well as the endocardial contour, thereby isolating the myocardial muscle volume. By multiplying this volume by a volume-weight constant, myocardial mass was calculated. The papillary muscles were regarded as part of the ventricular cavity in all analyses. All DSCT analyses were performed by a single blinded investigator. All MRI analyses were performed by one single investigator, blinded for the DSCT data.

Data analysis

The method of Bland and Altman was used to display the average difference and limits of agreement between the reference values of MRI and the functional parameters of DSCT [11].

The results are given in the form of mean ± 2SD. A small difference to the reference is associated with high accuracy and small standard deviations are associated with high precision. Pearson’s correlation coefficient was calculated to assess the correlation between MRI and DSCT.

Results

Thirty-four consecutive patients underwent both DSCT and MRI assessment of global LV function parameters. The calculated mean effective radiation dose based on the used scan protocol is 7.3 mSv [12].

Indications for DSCT, delay to MRI and image quality of DSCT are displayed in Table 2. Interobserver-agreement for image quality of DSCT datasets was excellent and results are shown in Table 3. Twenty-nine DSCT datasets were considered to be of sufficient quality to be analyzed [good quality (n = 16) and minor artifacts (n = 13)].

Table 2 Indications for DSCT, delay to MRI and image quality of DSCT datasets. Data are displayed as (n) percentage, unless otherwise indicated (CAD coronary artery disease)
Table 3 Interobserver quality scoring

Ventricular volumes

Average EDV measured by DSCT was 214.8 ml (±87.3 ml). Average EDV measured by MRI was 203.8 ml (±98.4 ml). Average ESV measured by DSCT was 94.9 ml. (±56.8 ml). Average ESV measured by MRI was 91.4 ml (±64.1 ml). The average overestimation of DSCT for EDV was 10.96 ml. (±27.4 ml.) and 3.54 ml. (±15.9 ml.) for ESV. Bland-Altman intervals for EDV and ESV were [−43.83 (10.96) 65.76] and [−28.34 (3.54) 35.42], respectively. The Bland-Altman plots for EDV and ESV are displayed in Figs. 3 and 4. The Pearson’s correlation coefficient for EDV and ESV were 0.96 and 0.97, respectively (p < 0.001 for both).

Fig. 3
figure 3

Bland-Altman plot comparing EDV measured by DSCT with EDV measured by MRI

Fig. 4
figure 4

Bland-Altman plot comparing ESV measured by DSCT with ESV measured by MRI

Myocardial mass

Average myocardial mass measured by DSCT was 122.1 g (±46.2 g). Compared with myocardial mass measured by MRI, there was an average overestimation of 0.9 g. Average myocardial mass measured by MRI was 121.2 g. (±46.5 g) The Bland-Altman interval for myocardial mass was [−31.0 (0.9) 32.8]. The Bland-Altman plot for myocardial mass is shown in Fig. 5. The Pearson’s correlation coefficient for myocardial mass was 0.94 (p < 0.001).

Fig. 5
figure 5

Bland-Altman plot comparing myocardial mass measured by DSCT with myocardial mass measured by MRI

LVEF

The average LVEF measured by DSCT datasets was 57.3%-point (±8.1%-point). Compared with the average LVEF measurement by MRI of 56.9%-point (±9.0%-point), there was an average overestimation of 0.40%-point (±4.54%-point). The Bland-Altman interval for LVEF was [−8.67 (0.40) 9.48] (Fig. 6). The Pearson’s correlation coefficient for LVEF was 0.90 (p = 0.001).

Fig. 6
figure 6

Bland-Altman plot comparing LVEF measured by DSCT on 29 DSCT datasets with LVEF measured by MRI

Discussion

The present study shows that global LV functional parameters, calculated from DSCT datasets acquired in daily clinical practice show acceptable limits of agreement with global LV parameters obtained with MRI.

The use of MDCT for the assessment of LV functions has been investigated extensively. Studies using older generation CT showed that function assessment could be assessed, but the temporal resolution of the four- and 16-slice MDCT systems hampered the sensitivity for detection and accurate classification of regional LV wall motion abnormalities [13].

Although the temporal resolution can be improved with multi-segment reconstruction, leading to better reproducibility in a phantom study [14], it was shown that this technique did not improve results for studies in human subjects [15].

More recently introduced MDCT devices capable of achieving better temporal resolution were studied previously. Butler et al. [16] concluded that 64-slice MDCT could serve as an alternative for suboptimal echocardiography results, and Salem et al. [17] found that 64-slice MDCT enabled LV function analysis in patients with regular sinus rhythm.

A phantom study by Mahnken et al. [14] showed that reliable assessment of LV volumes was also possible at increased heart rates.

Since DSCT has only recently been introduced, there are few data on the diagnostic accuracy of this technique. A recent study using a moving heart-phantom by Mahnken et al. [18] showed a good correlation between DSCT and MRI, with no significant differences, which is in accordance to the present in-vivo study.

Only two small studies in patients are available at the present time [19, 20]. A striking difference between the two studies is the wide B-A interval in the study by Busch et al. [20] in which two different post-processing software-packages have been implemented for the MRI and DSCT on the one hand, and the narrow B-A interval in the study by Brodoefel et al. [19] in which the same post-processing tool was used for both investigations on the other hand, suggesting a possible effect of the post-processing on the results.

A number of different post-processing software packages are currently commercially available for both DSCT and MRI. Little is known regarding the accuracy and interchangeability of these post-processing tools and this needs to be elucidated before these other software packages are used in clinical practice. Since in the present study both the DSCT and the MRI data were analyzed using the identical short-axis parameters and post-processing software (Mass; Medis, Leiden, The Netherlands) the differences in volume assessment are not considered to be influenced by differences in post-processing. Since DSCT datasets cover the entire left ventricle, it is tempting to speculate that reconstruction of short-axis slices without interslice gap or even partially overlapping slices may further enhance diagnostic accuracy. However, data regarding those protocols are currently unavailable.

One of the major findings of the present study is the fact that clinical reliability of LV functional assessment can be improved dramatically if the raw DSCT dataset is checked for artifacts, influencing the short-axis cine-loops. In comparison with coronary artery analyses, when only single frames from a manually selected phase in the cardiac cycle are used, the quality analyses for LV functional analyses focus more on LV wall motion delineation. It is, therefore, important to note the fact that if coronary artery analysis is possible on a particular DSCT dataset, this does not directly imply that accurate LV functional analysis is possible as well.

The average delay between DSCT and MRI of 6.3 days in the present study can be considered a limitation. Some previous studies using other CT systems have examined patients with less delay. It is conceivable that changes in either volume status or heart rate may have influenced the intrinsic volumes of the individual patients over time, thereby influencing the direct comparison between the two imaging modalities. However, LVEF is known to be notoriously stable over time and care was taken to confirm that no significant clinical interventions or changes in medication had taken place between the two investigations in all patients.

Although MRI and DSCT both provide two- and three-dimensional anatomical information, each technique has its own specific advantages and drawbacks. DSCT is particularly useful in patients with intracorporal devices such as pacemakers and pacemakerleads. These devices cause large image artifacts on the MRI and their intra-cardiac leads may warm up and possibly even cause perforation at the tip. Although there is increasing evidence that MRI examination may be performed safely in some specific cases, pacemakers and defibrillators still remain a contraindication [21].

DSCT is a very fast technique, which allows examination of less stable patients, patients with intra-corporal devices, severely claustrophobic patients or patients who cannot lie flat on their back for the time necessary for MRI assessment.

However, DSCT requires the use of ionizing radiation. Radiation exposure is currently widely debated and many efforts have been made to reduce the effective dose in DSCT. In the present study, a retrospective triggered protocol was implemented, covering the entire RR-interval. However ECG pulsing was used to vary the tube current in order to reduce radiation exposure in the part of the cardiac cycle which is usually ignored in coronary artery patency analysis. Recently, a number of reports have been published on full prospective ECG triggering, significantly reducing the effective radiation dose, at equal image quality [22, 23].

However, it should be noted that LV function analysis is not possible using these protocols, as the maximal end-diastolic and end-systolic phase cannot be determined and may even not be covered by the scanning window.

Conclusion

Volumetric analysis renders interesting adjunctive data in DSCT performed for coronary artery evaluation. Global LV functional parameters calculated from DSCT datasets acquired in daily clinical practice correlate well with MRI and may be considered interchangeable. However, visual assessment of the image quality of the short-axis cine slices should be performed to detect any artifacts in the DSCT data which could influence accuracy.