Abstract
Purpose: Magnetic resonance (MR)-guided percutaneous vascular interventions have evolved to a practical possibility with the advent of open-configuration MR systems and real-time tracking techniques. The purpose of this study was to assess an MR-tracking percutaneous transluminal angioplasty (PTA) catheter with regard to its safety profile and functionality.
Methods: Real-time, biplanar tracking of the PTA catheter was made possible by incorporating a small radiofrequency (RF) coil in the catheter tip and connecting it to a coaxial cable embedded in the catheter wall. To evaluate potentially hazardous thermal effects due to the incorporation of the coil, temperature measurements were performed within and around the coil under various canning and tracking conditions at 1.5 Tesla (T). Catheter force transmission and balloon-burst pressure of the MR-tracking PTA catheter were compared with those of a standard PTA catheter. The dilatative capability of the angioplasty balloon was assessed in vitro as well as in vivo, in an isolated femoral artery segment in a swine.
Results: The degree of heating at the RF coil was directly proportional to the power of the RF pulses. Heating was negligible with MR tracking, conventional spin-echo and low-flip gradient-echo sequences. Sequences with higher duty cycles, such as fast spin echo, produced harmful heating effects. Force transmission of the MR-tracking PTA catheter was slightly inferior to that of the standard PTA catheter, while balloon-burst pressures were similar to those of conventional catheters. The MR-tracking PTA catheter functioned well both in vitro and in vivo.
Conclusion: The in vivo use of an MR-tracking PTA catheter is safe under most scanning conditions.
Similar content being viewed by others
References
Edelmann RR (1993) MR angiography: Present and future. AJR 161:1–11
Pelc NJ, Herfkens RJ, Shimakawa A, Enzmann DR (1991) Phase contrast cine magnetic resonance imaging. Magn Reson Q 7:229–254
Schenck JF, Jolesz FA, Roemer PB, Cline HE, Lorensen WE, Kikinis R, Silverman SG, Hardy CJ, Barber WD, Laskaris ET (1995) Superconducting open-configuration MR imaging system for image-guided therapy. Radiology 195:805–814
Dumoulin CL, Souza SP, Darrow RD (1993) Real-time position monitoring of invasive devices using magnetic resonance. Magn Reson Med 29:411–415
Ackerman JL, Offut MC, Buxton RB, Brady TJ (1986) Rapid 3D tracking of small RF coils. In: Book of Abstracts: Society of Magnetic Resonance in Medicine. Society of Magnetic Resonance in Medicine, Berkeley, CA, p 1131
Dumoulin CL, Souza SP, Darrow RD, Pelc NJ, Adams WJ, Ash SA (1991) Simultaneous acquisition of phase contrast angiograms and stationary tissue images with Hadamard encoding of flow-induced phase shifts. J Magn Reson Imaging 1:399–404
Leung DA, Debatin JF, Wildermuth S, Heske N, Dumoulin CL, Darrow RD, Hauser M, Davis CP, von Schulthess GK (1995) Real-time biplanar needle tracking for interventional MR imaging procedures. Radiology 197:485–488
Leung DA, Debatin JF, Wildermuth S, McKinnon GC, Holtz D, Dumoulin CL, Darrow RD, Hofmann E, von Schulthess GK (1995) Intravascular MR tracking catheter: Preliminary experimental evaluation. AJR 164:1265–1270
Koechli VD, McKinnon GC, Hofmann E, von Schulthess GK (1994) Vascular interventions guided by ultrafast MR imaging: Evaluation of different materials. Magn Reson Med 31:309–314
Lufkin RB, Teresi L, Hanafee WN (1987) New needle for MR-guided aspiration cytology of the head and neck. AJR 149:380–382
Wildermuth S, Debatin JF, Leung DA, Hofmann E, Dumoulin CL, Darrow RD, Schöpke WD, Uhlschmid G, McKinnon GC, von Schulthess GK (1995) MR-guided percutaneous intravascular interventions: In vivo assessment of potential applications. In: Book of Abstracts: Society of Magnetic Resonance in Medicine. Society of Magnetic Resonance in Medicine, Berkeley, CA, p 428
Buchli R, Boesiger P, Meier D (1988) Heating effects of metallic implants by MRI examinations. Magn Reson Med 7:255–261
Knapp WH, Debatin JF, Helus F, Sinn HJ, Ostertag H (1989) Increased thermal response to ultrasound in the Walker carcinosarcoma treated with vasoactive drugs. Cancer Res 49:1768–1772
Mansfield P (1977) Multiplanar image formation using NMR spinechoes. J Phys C 10:L55-L58
Wetter DR, McKinnon GC, Bebatin JF, von Schulthess GK (1995) Cardiac echo-planar MR imaging: Comparison of single- and multipleshot techniques. Radiology 194:765–770
Pelc NJ, Bernstein MA, Shimakawa A, Glover GH (1991) Encoding strategies for three-direction phase-contrast MR imaging of flow. J Magn Reson Imaging 1:405–413
Hausmann R, Lewin JS, Laub G (1991) Phase-contrast MR angiography with reduced acquisition time: New concepts in sequence design. J Magn Reson Imaging 1:415–422
Ladd ME, Erhart P, Debatin JF, Hofmann E, Boesiger P, von Schulthess GK, McKinnon GC (1997) Guidewire antennas for MR fluoroscopy. Magn Reson Med 37:891–897
Hurst GC, Hua J, Duerk JL, Cohen AM (1992) Intravascular (catheter) NMR receiver probe: Preliminary design analysis and application to canine iliofemoral imaging. Magn Reson Med 24:343–357
Kandarpa K, Jakab P, Patz S, Schoen FJ, Jolesz FA (1993) Prototype miniature endoluminal MR imaging catheter. J Vasc Interv Radiol 4:419–427
Zimmermann GG, Erhart P, Schneider J, von Schulthess GK, Schmidt M, Debatin JF (1997) Intravascular MR imaging of atherosclerotic plaque: Ex vivo analysis of human femoral arteries with histologic correlation. Radiology 204:769–774
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Wildermuth, S., Dumoulin, C.L., Pfammatter, T. et al. MR-guided percutaneous angioplasty: Assessment of tracking safety, catheter handling and functionality. Cardiovasc Intervent Radiol 21, 404–410 (1998). https://doi.org/10.1007/s002709900288
Issue Date:
DOI: https://doi.org/10.1007/s002709900288