Skip to main content
Log in

Influence of Surgical Staples on Radiofrequency Ablation Using Multitined Expandable Electrodes

  • Laboratory Investigation
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose

During radiofrequency ablation (RFA), there is a risk that the multitined expandable electrode will come into contact with one of the surgical staples used to treat local recurrence after surgical operations. Our objective was to evaluate whether a surgical staple would influence the RFA of egg white using a multitined expandable electrode.

Methods

Multitined expandable electrodes, LeVeen needles (expandable diameter 3.0 cm), were sunk into an egg white bath with (a) no surgical staple, (b) a surgical staple touching one of the tines, or (c) a surgical staple touching two of the tines simultaneously. By connecting the LeVeen needle and copper plate at the bottom of the bath, RFA was then performed on the egg whites as a substitute for human tissue. Ten egg white baths were ablated under each of conditions (a), (b), and (c), for a total of 30 sets of coagulated egg white.

Results

There was no significant difference in the time from the power-on to the roll-off (i.e., the completion and shutting off of the electric circuit) or in the maximum diameter of the thermal lesion between conditions (a) and (b) or (a) and (c). However, the minimum diameter of the thermal lesion was significantly smaller in (c) compared with (a) (p < 0.01).

Conclusions

Surgical staples have the capacity to interfere with the electromagnetic field and decrease the minimum diameter of the thermal lesion in the event that a staple touches two of the tines of a multitined expandable electrode during RFA. Although the difference might be small enough to be neglected under many clinical circumstances, we recommend that, if possible, the tines not be expanded near metallic material.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Lencioni R, Cioni D, Crocetti L, et al. (2005) Early-stage hepatocellular carcinoma in patients with cirrhosis: Long-term results of percutaneous image-guided radiofrequency ablation. Radiology 234:961–967

    Article  PubMed  Google Scholar 

  2. Simon CJ, Dupuy DE, DiPetrillo TA, et al. (2007) Pulmonary radiofrequency ablation: Long-term safety and efficacy in 153 patients. Radiology 243:268–275

    PubMed  Google Scholar 

  3. Gervais DA, McGovern FJ, Arellano RS, et al. (2003) Renal cell carcinoma: Clinical experience and technical success with radio-frequency ablation of 42 tumors. Radiology 226:417–424

    Article  PubMed  Google Scholar 

  4. Shirato K, Morimoto M, Tomita N, et al. (2002) Small hepatocellular carcinoma: Therapeutic effectiveness of percutaneous radio frequency ablation therapy with a LeVeen needle electrode. J Ultrasound Med 21:67–76

    PubMed  Google Scholar 

  5. Solmi L, Nigro G, Roda E (2006) Therapeutic effectiveness of echo-guided percutaneous radiofrequency ablation therapy with a LeVeen needle electrode in hepatocellular carcinoma. World J Gastroenterol 12:1098–1104

    PubMed  Google Scholar 

  6. Hiraki T, Sakurai J, Tsuda T, et al. (2006) Risk factors for local progression after percutaneous radiofrequency ablation of lung tumors: Evaluation based on a preliminary review of 342 tumors. Cancer 107:2873–2880

    Article  PubMed  Google Scholar 

  7. Cheng YC, Brown RW, Chung YC, et al. (1998) Calculated RF electric field and temperature distributions in RF thermal ablation: Comparison with gel experiments and liver imaging. J Magn Reson Imaging 8:70–76

    Article  PubMed  CAS  Google Scholar 

  8. Goldberg SN, Ryan TP, Hahn PF, et al. (1997) Transluminal radiofrequency tissue ablation with use of metallic stents. J Vasc Interv Radiol 8:835–843

    Article  PubMed  CAS  Google Scholar 

  9. Oshima F, Yamakado K, Akeboshi M, et al. (2004) Lung radiofrequency ablation with and without bronchial occlusion: Experimental study in porcine lungs. J Vasc Interv Radiol 15:1451–1456

    PubMed  Google Scholar 

  10. Boll DT, Lewin JS, Duerk JL, MerkLe EM (2003) Do surgical clips interfere with radiofrequency thermal ablation? AJR Am J Roentgenol 180:1557–1560

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yusuke Sakuhara.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Sakuhara, Y., Shimizu, T., Abo, D. et al. Influence of Surgical Staples on Radiofrequency Ablation Using Multitined Expandable Electrodes. Cardiovasc Intervent Radiol 30, 1201–1205 (2007). https://doi.org/10.1007/s00270-007-9174-7

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-007-9174-7

Keywords

Navigation