Introduction

Inguinal hernia repair is one of the most commonly performed operations in the world. Every year almost 60,000 inguinal hernia repairs (IHR) using transabdominal preperitoneal (TAPP) and extraperitoneal (TEP) techniques are performed in Germany [1]. In the majority of cases postoperative recovery is without complications and return to normal activities is possible within 1 week [2]. However, some patients complain about postoperative pain for several weeks and some suffer from chronic groin pain for months or even for years. Chronic pain was defined as “lasting for 3 months or more” using the definition of the International Association for the Study of Pain [3]. In their meta-analysis of randomized controlled trials, McCormack et al. found the incidence of chronic pain after inguinal hernia repair using the TAPP technique to be at 13.8% [4]. In the literature, several risk factors for chronic inguinal pain are mentioned. These are young age, female gender, medial hernia, open repair technique, postoperative complications, need for operation for recurrence [5], presence of preoperative pain, neurotoxic chemotherapy, depression, neuroticism, anxiety [6, 7], and an observation period of fewer than 3 years after surgery [8]. However, the staples used for mesh fixation [9] and the alloplastic mesh itself may also be major causes for chronic postoperative pain [10].

Animal studies proved that there is a correlation between the weight, biocompatibility, and shrinkage of polypropylene meshes [10]. There have been numerous clinical reports pointing out the tendency of implanted mesh to shrink extensively and migrate and penetrate neighboring organs, have an oversized load of material, and cause an associated chronic inflammatory reaction and even carcinogenic effects [1012]. A great effort was put into the development of meshes that have improved biocompatibility.

Since 2001 titanium-covered polypropylene meshes, referring to a new type of compound material, are available. The titanium is covalently bonded to the alloplastic surface of the mesh. There is evidence from animal experiments that the titanium-coated polypropylene meshes induce a less pronounced foreign body reaction (i.e., less shrinkage) compared with identical meshes lacking a titanium coating [13].

To improve the outcome for their patients some surgeons have begun performing inguinal hernia repair using these titanized meshes. However, it was not clear which mesh to use for which kind of hernia, and it was also not clear how the 35-g/m2 (Light) and the 16-g/m2 (Extralight) versions differed concerning the rate of recurrence or chronic pain [14].

The aim of this prospective, randomized, single-blinded clinical trial was to compare the incidence of chronic inguinal pain and the rate of recurrence after laparoscopic transabdominal preperitoneal (TAPP) hernia repair using 35-g/m2 (Light) and 16-g/m2 (Extralight) titanized polypropylene mesh.

Methods

Patients who were at least 18 years old and presented at the General Surgery Unit of Agatharied Academic Teaching Hospital of the Ludwig-Maximilians-University of Munich with a diagnosis of an inguinal or a femoral hernia were eligible to enter the study. The hernia repair was to be a laparoscopic tension-free TAPP technique with mesh fixation using titanium staples. Patients were randomized intraoperatively to receive either a 35-g/m2 (TiMESH-Light™, pfm medical ag, Cologne, Germany) or a 16-g/m2 (TiMESH-Extralight™, pfm medical ag) mesh for the repair of the inguinal hernia. The protocol was approved by the Institutional Review Board of the Ludwig-Maximilians-University of Munich (Ethical Committee) and informed consent was obtained from all participants.

All patients who were class IV or class V as defined by the American Society of Anesthesiologists (ASA) were excluded, as were those who had contraindications for general anesthesia. Further exclusion criteria were acute intra-abdominal infection such as peritonitis, previous major abdominal procedures, or a history of ileus. Patients who were participating in another trial, younger than 18 years, and mentally immature were also excluded.

Randomization was carried out using a lottery system. Letters were prepared prior to starting the trial and stored in a box. Each letter was drawn from the box and opened by a nurse just after dissection of the inguinal region and hemorrhage control were finished. The only influence that the surgeon had was the choice of the size of the mesh to be 10 × 15 cm or larger. In patients with bilateral hernias, both sides were repaired with the same type of mesh. Patients were not aware of which kind of mesh was used for the repair. During the consent process (Fig. 1) prior to entering the trial, patients were informed that two new types of titanized polypropylene meshes are available and that we wanted to find out which one is the better with respect to postoperative pain and recurrence.

Fig. 1
figure 1

Design of the study

All patients underwent standardized TAPP repair in the presence of a surgeon experienced in TAPP repair who either performed the procedure himself or supervised it in a teaching situation. There were 13 surgeons who performed the operations. General anesthesia was used in all cases. All patients received standardized postoperative oral pain medication consisting of diclofenac 2 × 50 mg, novaminsulfone 4 × 500 mg, and omeprazole 1 × 20 mg. Four times per day pain levels were routinely controlled using the VAS ranging from 1 to 10. If at any time levels above VAS 3 were detected, additional medication (oxycodone/naloxone [Targin®] 10 mg or piritramide [Dipidolor®] 15 mg) was administered until the patient reported levels below VAS 3. All patients were given standardized postoperative instructions that did not limit their activities. They could participate in sports as usual within the first 10 days except heavy-weight body building and anything that caused high-frequency movements of the psoas muscle, e.g., mountain biking.

Surgical procedure

The herniotomy was carried out with the patient under general anesthesia with a single shot of prophylactic antibiotic (cefuroxime 750 mg). The operation was performed with the patient in the Trendelenburg position. The bladder was drained using a catheter. After pneumoperitoneum was induced, the 10-mm optic trocar was introduced at the umbilicus. Following inspection of the abdomen and the hernia, a 5-mm trocar was placed at the left lateral midclavicle. On the right lateral midclavicular line, a 12-mm trocar was placed to insert the stapler.

From the umbilical artery medially to 4–6 cm past the inguinal ring laterally, the abdominal wall (peritoneum) was incised describing an arc. The epigastric vessels were identified. The medial compartment was completely dissected to the midline, exposing the rectus muscle, the horizontal pubic ramus, Hesselbach’s triangle, Cooper’s ligament, and the iliopubic tract. The femoral hernia orifice was exposed. The dissection went deep into the prevesical space. The peritoneum was dissected from the spermatic cord cranially to about midpsoas. In lateral hernias, the sac was completely reduced and separated from the deferent duct and testicular vessels. In direct hernias, the pseudosac was completely dissected and the hernia orifice was freed from tissue. Perihernial lipomas were excised. In all cases, titanized meshes, either TiMESH-Light or TiMESH-Extralight (15 cm × 10 cm or 15 cm × 15 cm mostly cut down to 15 × 12 cm) were used. Meshes were not incised.

Staple fixation of the mesh was done in the vast majority of repairs. In special cases, e.g., extremely large hernias, suture fixation was used, and in very small lateral hernias no fixation was used. Mesh was fixed with two to six titanium staples by using either an Endopath Multifire stapler (EMS, Ethicon Endo-Surgery Inc., Cincinnati, OH, USA) or an Endo-Universal stapler (Auto-Suture, Covidien, Mansfield, MA, USA). One staple each was placed at the transverse fascia at the ventral border of the mesh medially and laterally to the epigastric vessels. The other staples were optional and were placed as follows: one staple to Cooper’s ligament, one to the suprapubic connective tissue in the midline, and the last staple could be variably placed, always keeping at least a 3-cm safety distance to the inguinal ligament.

The peritoneum was closed using either staples or a running suture. In patients with bilateral hernias, the same procedure was performed on both sides. The placement of one Redon drain (Suru International, Mumbai, India) at the end of the procedure was optional. After removal of the trocars, the umbilical port site was closed at the level of the fascia. The skin was sutured. Umbilical hernias, when detected, were closed simultaneously by suture.

Primary outcome

The primary outcome of the trial was chronic inguinal pain more than 3 years after inguinal hernia repair with TAPP technique. Chronic pain was defined as pain that persisted for more than 3 months after surgery and was present at the time of follow-up examination. It included patients who regularly consumed painkillers to control inguinal pain after hernia repair at the time of follow-up examination. The postoperative follow-up period for every hernia was at least 3 years. All patients were personally questioned and examined in detail at least 3 years after surgery by the same examiner.

Measurement of pain for all patients was also performed by only one examiner who used a standardized pain protocol (Table 1). The intensity was measured with a 10-cm VAS from 0 (no pain) to 10 (agonizing). Patients describing persistent pain were referred for further neurological, orthopedic, urological, gynecological, and psychiatric investigation (Fig. 2). When no extrainguinal cause for the pain could be found, the pain was considered to be a result of the procedure.

Table 1 Hernia follow-up: standardized pain protocol accounted for every single hernia repaired
Fig. 2
figure 2

Approach to patients with pain

Patients who described postoperative pain at the time of the follow-up examination which, according to memory, subsided within 3 months after surgery did not limit their activities and returned to their prehernia lifestyles. They were documented to have suffered from acute pain. This historic perspective of the patient’s postoperative period was soft data that was not considered to be a primary outcome measure.

Secondary outcomes

Secondary outcomes were early and late recurrences rates. Recurrences were confirmed by clinical examination, ultrasound, and, in some cases, laparoscopy.

Statistics

The study was designed to detect a 3% difference in chronic inguinal pain between the groups, with a sample of at least 200 hernias per group, a power of 70%, and an α error of 5%. The results were analyzed using multiple logistic regression (SPSS 15.0 for Windows and Microsoft Excel 2000). In the primary analysis, the rates of chronic inguinal pain in patients who received TiMESH-Extralight and TiMESH-Light were compared. The values are expressed as median first and third quartile, mean and range, or relative frequencies. A one-sided P < 0.05 was considered significant.

Results

Baseline characteristics of the patients

Between October 2002 and January 2006, all 380 patients with 466 inguinal hernias underwent inguinal hernia repair with the TAPP technique. The 3-year follow-up period for the last patient ended in January 2009. Of the 380 patients operated on, 344 came personally for a physical examination and questioning in detail, for a follow-up rate of 90%. The response rate was the same in both groups: 90.0% (225 of 250 hernias) for Group A (Light) and 89.8% (194 of 216 hernias) for Group B (Extralight).

Table 2 shows the baseline characteristics of the patients according to the two treatment groups. Demographic characteristics, the characteristics of hernia, and ASA classification were similar in both groups. There was no statistical difference (P > 0.05, Table 2).

Table 2 Baseline characteristics of the patients according to treatment group

Chronic pain

Postoperative pain was reported by 75 of 419 (17.9%) investigated inguinal hernias. In 60 patients, the pain fell into the category of acute and was limited to the immediate postoperative period of 3 months or less. Acute pain was more common in Group B (Extralight), with 22.4%, than in Group A (Light), with only 13.2%. The difference, however, was not significant (P > 0.05).

In 15 inguinal hernia cases (3.6%), patients reported chronic inguinal pain. Of these 15 inguinal hernias, 3 (1.5%) in two patients were found in Group B (Extralight) and 12 (5.3%) in Group A (Light). Chronic inguinal pain was significantly more common after implantation of TiMESH-Light mesh than after implantation of TiMESH-Extralight mesh (P = 0.037, Fig. 3). There were no intraoperative complications in any of these 15 patients with chronic pain during the postoperative course. There were no documented lesions to the spermatic cord, the nerves, or the blood vessels. One of the 15 patients with chronic inguinal pain suffered from a recurrent hernia. One patient in the TiMESH-Light group who had bilateral hernia repair had both meshes removed because the pain had been so severe. In the TiMESH-Light group, one case with chronic pain occurred after the repair of a recurrent hernia. We found no effect of the operating surgeon on the occurrence of chronic inguinal pain in the data collected.

Fig. 3
figure 3

Distribution of chronic pain

In Group A, one inguinal hernia repair was an emergency intervention. All other 14 hernia repairs were performed as elective surgery. Figure 4 shows the intensity of chronic pain in the Group A (Light) and in Group B (Extralight).

Fig. 4
figure 4

Intensity of chronic pain on a visual analog scale

Recurrences

In the 3-year observation period, 12 recurrent hernias (2.86%) were found by postoperative investigation. Hernia recurrence was more common in Group A (Light), with seven hernias (3.1%), than in Group B (Extralight), with five hernias (2.6%). The difference, however, was not significant (P = 0.724, Fig. 5). All recurrent hernias were repaired in a second operation. There were no more complications in any of the patients.

Fig. 5
figure 5

Recurrence rate in both groups

Discussion

In the past, the recurrence rate was a major outcome measure in studies dealing with inguinal hernia repair (IHR). Meanwhile, the attention has turned to other measures of success. Chronic pain has become a major focus in the last few years and there have been numerous studies dealing with pain after IHR, its type and frequency of occurrence, its risk factors, and profiles of patients predisposed to the condition. The roots of the problem are nerve injury, tissue injury, the material used for the repair, and the method of mesh fixation. Anatomical changes also might be a cause [15].

We found in our prospective, randomized, single-blinded trial that 3 years after IHR with the TAPP technique 5.3% of the patients suffered from chronic pain after the implantation of the 35-g/m2 titanized polypropylene mesh (TiMESH-Light). After hernia repair with the 16-g/m2 titanized polypropylene mesh (TiMESH-Extralight), which made up the second arm in the trial, chronic pain was detected in only 1.5% of the patients. This difference was significant (P = 0.037). The only difference between the two meshes was the material load. There was no difference in the baseline characteristics between the groups, so the reduction in chronic pain in Group B can be attributed only to the lighter-weight mesh. This is in line with the experimental results of Junge et al. [16] who very clearly demonstrated significantly improved functional consequences and morphologic tissue response with material-reduced mesh compared to a common heavy-weight polypropylene mesh. Klosterhalfen et al. [11] found in histological studies of long-term implanted meshes that the reduction of mesh material reduced the acute inflammatory reaction, caused less formation of connective tissue, and hardly maintained a chronic inflammatory process. There is much experimental evidence in the literature that material composition and mesh structure may significantly affect foreign body reaction [17]. A recent extensive review by Weyhe et al. of current randomized controlled trials and retrospective studies led to the conclusion that light meshes seem to have some advantages with respect to postoperative pain and foreign body sensation [17]. However, Light meshes offer no advantage with respect to alleviating “severe chronic groin pain.” We confirmed this in our study. The only two patients in the TiMESH-Extralight group with chronic pain reported it to be quite strong. Weyhe et al. also found that the use of Light meshes was associated with an increased recurrence rate [17]. In the present study we did not find a difference between the 35- and the 16-g/m2 versions with respect to recurrence. Nonetheless, light-weight meshes have replaced their heavy-weight counterparts almost completely. This is clinical reality today, and it is independent of the fact that it may be difficult to prove why these Light meshes are favored. In our study, the lower rate of chronic pain detected in the arm with the lighter mesh confirmed expectations.

In a recent review of the literature, the incidence of chronic pain after IHR was reported to range between 25 and 30% in the majority of studies, with 2-9% reported as severe pain [18]. McCormack et al. found that the incidence of chronic pain after IHR with TAPP technique was 13.8% in a meta-analysis of randomized controlled trials [4], which was more than 5% lower than for the examined open repairs. Aasvang and Kehlet found in their review of the present literature that inguinal herniorrhaphy may be followed by a risk of chronic pain in about 12% of patients (18% after open and 6% after laparoscopic repair), with a clinically significant effect on daily activities [19]. Grant et al. [20] found pain and numbness to be less of a problem after laparoscopic repair than after open repair. A well-conducted, large, randomized controlled multicenter trial found a 10% incidence of chronic pain after laparoscopy and 14% after open surgery in 1983 patients (P < 0.05) without significant differences in VAS scores between the two groups at 6, 12, and 24 months. Kroninger [21] reported chronic pain at a rate of 15% after TAPP repair in a long-term follow up (>52 months).

Several trials reported the incidence of chronic pain due to IHR after a follow-up of 1 year [4]. Our patients were followed up for 3 years after surgery, so the number of patients feeling pain may be reduced as a result of the follow-up time interval used. Courtney et al. [22] used the Wisconsin Brief Pain Questionnaire, which measures pain severity and the impact of pain on function using 0-10 rating scale, to assess the long-term outcomes in 126 patients who reported severe or very severe pain within 3 months after IHR. Of the 86 who responded after 2.5 years, 71% still reported pain, which was severe or very severe in 26% of the cases and mild or very mild in 45% of the cases. This study demonstrates that the number of patients complaining and the severity of the pain both diminish over time. However, despite some available data, there is an overall lack of well-designed prospective studies quantifying the time course of post-herniorrhaphy pain. Even in view of the evidence suggesting a “burn-out” effect of the pain complaints, we find that the very low rate detected in both of our TiMESH groups cannot be attributed only to the time factor.

The incidence of short- or long-term chronic pain after laparoscopic IHR with TAPP technique as reported in the literature is generally higher than the results of the present trial. Chronic pain in 5.3% of the patients after 3 years after IHR with TAPP technique using 35-g/m2 titanized polypropylene mesh (Light) offers a good result in comparison with the literature. A rate of 1.5% when 16-g/m2 titanized polypropylene mesh (Extralight) is used can be classified as a rare event. The data are rather solid since it is derived from a prospective randomized study with a large sample size of 466 hernias repairs and a high-quality follow-up, i.e., physical examination and personal questioning of 90% of the patients operated on. de Jonge et al. [23] stated after an extensive review of randomized trials, considered to be the “gold standard” of evidence, that the majority were using the VAS to assess pain, suggesting that there is a certain amount of standardization that can better enable interstudy comparison. Even though this particular quality standard was met by the present study, current instruments used for the evaluation of chronic pain need to be improved and prospectively validated.

The mesh used is the most probable reason for the low rate of chronic pain in this study. There are several modern concepts integrated into these prosthetic meshes, such as the light weight, pore size ≥ 1 mm [12], monofile fiber, and titanium surface, that may have made a clinical difference, at least in this study, when compared to other meshes used in hernia repair. The low rate of chronic pain detected in both groups in the present study may have to do not only with the reduced weight but also with the unique titanium surface. With respect to several histological and histochemical parameters of biocompatibility, the titanized polypropylene mesh proved advantageous compared with the other modern meshes in animal experiments. Heavy-weight titanium-coated polypropylene meshes induce a less pronounced foreign body reaction than identical meshes with no titanium coating. Since the amount of material implanted is identical, this must be attributed solely to the titanium coating [13]. The favorable biocompatibility of the titanized meshes that supposedly cause less foreign body reaction and less shrinkage than other modern meshes as demonstrated in animal experiments [13, 24] may also hold true in humans.

What causes the pain in those patients still suffering after 3 years? In addition to the mesh there is surgical trauma, direct nerve injury, and the use of staples that may have also played a role here. There are numerous published reports of nerve entrapment by tacks placed during laparoscopic repair. Laparoscopic hernia repair [25] without the use of tacks has been described and it eliminates the risk of misplaced tacks [26]. In addition, it is more cost effective because staplers are expensive. Others do advise to use staples because the rate of recurrence may be reduced [27]. In the present study, staples were used in both groups to an equal extent. The rate of recurrence was the same in both groups but not the rate of chronic pain. The role of staples as a major cause of chronic inguinal pain cannot be supported by our data.